I.T.S. Straight Plate with Angular Stability with the below descriptions and corresponding article numbers. 1. Forearm Plate, Curved; Article Numbers: 21604-6, 21604-9. 2. Forearm Plate, Straight; Article Numbers: 21605-11, 21605-6, 21605-9. 3. Straight Compression Plate; Article Numbers: 21105-6, 21105-8. 4. Straight Plate, 1.5mm; Article Numbers: 21104-4, 21104-5, 21104-6, 21104-7, 21104-8, 21104-9. 5. Straight Plate, 2.0mm; Article Numbers: 21101-5, 21101-6, 21101-7, 21101-8, 21101-9, 21101-11, 21101-13. 6. Straight Plate, 3.5mm; Article Number: 21102-6. 7. Straight Plate, 4.5mm; Article Numbers: 21103-8, 21103-11, 21103-10, 21103-11, 21103-12, 21103-13, 21103-8, 21103-9. Class II

Reason: Updated MRI safety testing demonstrates higher RF-induced temperature increases under certain MRI conditions than previously reflected in the Instructions for Use (IFU).

I.T.S. GmbH 20260123 Ongoing
Distributed: US Nationwide distribution in the states of AL, AZ, BVI, CA, CO, FL, GA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NE, NJ, NM, NV, OH, OK, PA, SC, TX, USVI, WV, W.
Olecranonplate with Angular Stability with the below descriptions and corresponding article numbers. 1. Olecranon Hook Plate; Article Numbers: 21112-7, 21114-5, 21113-9. 2. Olecranon Plate; Article Numbers: 21111-6, 21111-8, 21111-12. Class II

Reason: Updated MRI safety testing demonstrates higher RF-induced temperature increases under certain MRI conditions than previously reflected in the Instructions for Use (IFU).

I.T.S. GmbH 20260123 Ongoing
Distributed: US Nationwide distribution in the states of AL, AZ, BVI, CA, CO, FL, GA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NE, NJ, NM, NV, OH, OK, PA, SC, TX, USVI, WV, W.
Pilonplate with Angular Stability, with the below descriptions and corresponding article numbers. 1. Anterolateral Tibia Plate; Article Numbers: 21311-11, 21311-15, 21311-7, 21312-11, 21312-15, 21312-7. 2. Pilon Plate 4mm; Article Numbers: 21094-10, 21094-4, 21094-6, 21094-8. 3. Pilon Plate, Small; Article Numbers: 21093-4, 21093-6. Class II

Reason: Updated MRI safety testing demonstrates higher RF-induced temperature increases under certain MRI conditions than previously reflected in the Instructions for Use (IFU).

I.T.S. GmbH 20260123 Ongoing
Distributed: US Nationwide distribution in the states of AL, AZ, BVI, CA, CO, FL, GA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NE, NJ, NM, NV, OH, OK, PA, SC, TX, USVI, WV, W.
FR.O.H. Calcaneus Repair System with the following description: Anterolateral Calcaneus Plate. Article Number: 21064-60. Class II

Reason: Updated MRI safety testing demonstrates higher RF-induced temperature increases under certain MRI conditions than previously reflected in the Instructions for Use (IFU).

I.T.S. GmbH 20260123 Ongoing
Distributed: US Nationwide distribution in the states of AL, AZ, BVI, CA, CO, FL, GA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NE, NJ, NM, NV, OH, OK, PA, SC, TX, USVI, WV, W.
Brand Name: Humeral Head Plate with Angular Stability, with the following description: Proximal Humeral Plate. Article Numbers: 211341-10, 211341-4, 211341-7, 21134-4, 21134-5, 21135-4, 21136-4, 21136-5, 21136-7, 21137-4, 21137-7, 21138-4, 21139-4. Class II

Reason: Updated MRI safety testing demonstrates higher RF-induced temperature increases under certain MRI conditions than previously reflected in the Instructions for Use (IFU).

I.T.S. GmbH 20260123 Ongoing
Distributed: US Nationwide distribution in the states of AL, AZ, BVI, CA, CO, FL, GA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NE, NJ, NM, NV, OH, OK, PA, SC, TX, USVI, WV, W.
Claviculaplate with Angular Stability, with the below descriptions and associated article numbers. 1. Anterior Clavicle Plate, Medial; Article Numbers: 21116-8, 21117-10, 21118-10. 2. Clavicle Plate; Article Numbers: 21121-6, 21121-8. 3. Clavicle Plate, 3.5mm; Article Numbers: 21122-10, 21122-6, 21122-8. 4. Clavicle Plate, Lateral; Article Numbers: 21123-4, 21124-4, 21127-6, 21128-6. Class II

Reason: Updated MRI safety testing demonstrates higher RF-induced temperature increases under certain MRI conditions than previously reflected in the Instructions for Use (IFU).

I.T.S. GmbH 20260123 Ongoing
Distributed: US Nationwide distribution in the states of AL, AZ, BVI, CA, CO, FL, GA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NE, NJ, NM, NV, OH, OK, PA, SC, TX, USVI, WV, W.
Volar Radius Plate with Angular Stability with the following description: Radius Plate PROlock II. Article Numbers: 21031-3, 21031-5, 21032-11, 21032-3, 21032-5, 21033-3, 21033-5, 21034-3, 21034-5, 21035-3, 21036-3. Class II

Reason: Updated MRI safety testing demonstrates higher RF-induced temperature increases under certain MRI conditions than previously reflected in the Instructions for Use (IFU).

I.T.S. GmbH 20260123 Ongoing
Distributed: US Nationwide distribution in the states of AL, AZ, BVI, CA, CO, FL, GA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NE, NJ, NM, NV, OH, OK, PA, SC, TX, USVI, WV, W.
Imactis CT-Navigation System, stereotaxic accessory for Computer Tomography (CT), systems, Model Numbers E85101LA and E85101LP Class II

Reason: There is a potential issue that can lead to discordance between the simulated needle trajectory/tip position and the actual needle trajectory/tip position.

GE Medical Systems, LLC 20260203 Ongoing
Distributed: Worldwide distribution - US Nationwide.
Handy Solutions Neck & Shoulders Heating Pad, Number: 25607, Manufactured by Ninbo Unico Products Co., Ltd Class I

Reason: If heating pad is folded while in use, the power density increases in the folded area causing excessive temperatures, which can also be caused by placing pad under the body, then the high limit device is unable to sense the higher temperature so power is not terminated. May cause a burning smell prior to melting/burning potentially causing burns to persons and property.

Navajo Manufacturing Company 20260218 Ongoing
Distributed: US Nationwide distribution in the states of CA, MS, FL, IN, LA, VA, TX.
BioFlo DuraMax Catheter, REF: H965103028011/A, H965103028021/A, H965103028021/EU, H965103028031/A, H965103028031/EU, H965103028041/A, H965103028041/EU, H965103028051/A, H965103028051/EU, H965103028061/A, H965103028061/EU, H965103028070/A, H965103028070/EU, H965103028080/A, H965103028080/EU, H965103028090/A, H965103028090/EU, H965103028161/A, H965103028181/A, H965103028191/A, H965103028191/EU, H965103028201/A, H965103028211/A, H965103028220/A, H965103028230/A, H965103028240/A, H965103038181/A Class I

Reason: 16F dual-valved splittable sheath introducer due to a design defect, may not split as intended, which may result in hemorrhage, foreign bodies, procedure delay, embolization/thrombosis, impaired catheter function, loss of vessel for future vascular access.

Merit Medical Systems, Inc. 20260213 Ongoing
Distributed: Worldwide distribution: US (nationwide) to states of: TX, LA, OR, NM, MI, FL, CO, CA, GA, NC, AK, NY, PA, WI, NJ, VA, AZ, IN, KY, MO, DE, SC, MT, NE, WV, OH, MA, MN, CT, MD, KS, WA, TN, ID, UT, DC, MS, IL, OK, AL, RI, AR, NV and OUS (International) to countries of : Canada, Spain, United Arab Emirates, Brazil, Jordan, Iraq, France, Indonesia, Malaysia, Thailand, Slovakia, Colombia, Saudi Arabia, Belize, Hong Kong, Mexico, Austria, Belgium, Chile, Germany, Sweden, Ireland, Poland, Latvia, Italy, South Africa, Switzerland, Bulgaria, Peru, Argentina, Qatar, Greece, Georgia, United Kingdom, Brunei Darussalam, Ecuador, Panama, Netherlands, Venezuela, Bolivarian Republic of, Costa Rica, Viet Nam, Cayman Islands, Iran, Islamic Republic of, Australia, Nicaragua, Dominican Republic, Bolivia, Plurinational State of, El Salvador, Guatemala, Honduras, Japan