10mL Non-Sterile BD Luer-Lok Tip Syringe Only. Catalog Number: 301029. 850 bulk non-sterile units per case. Class II

Reason: During the bulk packaging process, some Luer Slip syringes were incorrectly included in a batch designated for Luer Lok syringes.

Becton Dickinson & Company 20260127 Ongoing
Distributed: US Nationwide distribution in the states of California and Texas.
MEDLINE UNNA-Z STRETCH ZINC PASTE BANDAGE, REF NONUNNAS40 KOB article #40307010 Class II

Reason: Due to labeling error. Incomplete ingredients listed. Labeling did not include castor oil or calamine.

KOB GmbH 20260115 Ongoing
Distributed: U.S.: IL O.U.S.: Not provided as product is manufactured O.U.S.
Brand Name: Olympus ShockPulse-SE Lithotripsy System with Generator Product Name: ShockPulse-SE Lithotripsy System - Reuseable Probes Model/Catalog Number: SPL-SR, contains generator SPL-G Product Description: An assembly of devices that uses a combination of, or individually applied, pneumatically-created ballistic shock waves and ultrasonic waves for the intracorporeal fragmentation of stones (calculi) in the urinary tract (i.e., kidney, ureter, and bladder) for their removal. It typically includes a pressure and power regulator/control unit, handpieces, and two types of probes of various sizes. The probes are attached to the two handpieces which are connected to the control unit that supplies the energy for both the pneumatic and ultrasonic functions. Energy is applied to a calculus which is fragmented in situ and removed via a suction system facilitated by the lithotripsy system. Class II

Reason: Mis-wired component-the improperly wired component can result in additional noise on the power supply output. Noise on the ultrasonic input circuit is not consistent with electromagnetic compatibility (EMC) standards and may decrease the overall reliability of the system and result in delay of treatment. It was discovered that additional serial numbers of ShockPulse generators may also have the mis-wired component. As a result, this is expanding the action to the additional impacted devices.

Olympus Corporation of the Americas 20260108 Ongoing
Distributed: International distribution in the country of Canada, Germany, Singapore, Australia, and India.
Brand Name: HiResolution Bionic Ear System Product Name: Sky CI M90 Sound Processor Model/Catalog Number: CI-5295-150 Class II

Reason: Behind-the-ear sound processer packaging label is different then included product.

Advanced Bionics, LLC 20251127 Ongoing
Distributed: Unknown
Gelsoft Plus, REF: 635024PE, 631514PE, 631222PE, 633008PE, 633010PE, 633007PE, 636010PE, 632522PE, 633012PE, 635018PE, 631516PE, 636016PE, 631220PE, 631608PE, 631206PE, 631407P50E, 631508PE, 636008PE, 636006PE, 632518PE, 634008PE, 634006PE, 631218PE, 632211PE, 634007PE, 633006PE, 631506PE, 632524PE, 636007PE, 631512PE, 636012PE, 636014PE, 632211P50E, 633016PE, 631407PE, 631809PE, 632010PE, 631608P50E, 632520PE, 632412PE, 631510PE, 635022PE, 631224PE, 632412P50E, 631507PE, 633014PE, 632010P50E, 635020PE, 631809P50E. Gelweave, REF: 736038E, 736036E, 736034E, 736032ST/10DE, 736032E, 736030ST/10DE, 736030E, 736028ST/10DE, 736028E, 736026ST/10DE, 736026E, 736024ST/10DE, 736024E, 736022ST/10DE, 736022E, 736020E, 736018E, 736016E, 736014E, 736012E, 736010E, 736008E, 736006E, 735034ST/8AE, 735034/8RME, 735034/10SEE, 735034/10RME, 735032STE, 735032ST/8AE, 735032/8SE, 735032/8RME, 735032/10SEE, 735032/10RME, 735030STE, 735030ST/8AE, 735030/8SE, 735030/8RME, 735030/10SEE, 735030/10RME, 735028STE, 735028ST/8AE, 735028E, 735028/8SE, 735028/8RME, 735028/10RME, 735026STE, 735026ST/8AE, 735026E, 735026/8SE, 735026/8RME, 735026/10RME, 735024STE, 735024ST/8AE, 735024E, 735024/8SE, 735024/8RME, 735024/10SEE, 735024/10RME, 735022STE, 735022E, 735022/8SE, 735022/8RME, 735020E, 735020/8RME, 735018E, 734038E, 734036E, 734034E, 734034CX4RMSE, 734034/8E, 734034/10E, 734032E, 734032CX4RMSE, 734032/8E, 734032/10E, 734030E, 734030CX4RMSE, 734030/8E, 734030/10E, 734028CX4RMSE, 734028/8E, 734028/10E, 734026CX4RMSE, 734026/8E, 734026/10E, 734024CX4RMSE, 734024/8E, 734024/10E, 734022CX4RMSE, 734022/8E, 734022/10E, 734020CX4RMSE, 734020/8E, 734020/10E, 734018/8E, 73341088E, 73341088/8RME, 73341088/8E, 73341088/10RME, 73341088/10E, 73321488/8ARME, 7332148/10RMFE, 7332148/10RMEE, 733212810/10STDE, 7332128/10RMFE, 7332128/10RMEE, 73321088E, 73321088/8SDE, 73321088/8RME, 73321088/8E, 73321088/10RME, 73321088/10E, 7330888/8E, 7330668/6E, 733038E, 733036E, 733034E, 733032E, 733030E, 733028E, 733026E, 733024E, 733024/6AE, 733022E, 733022/6AE, 733020E, 733020/6AE, 733018E, 733018/6AE, 733016E, 733014E, 73301488/8ARME, 7330148/10RMFE, 7330148/10RMEE, 73301412/10ETE, 733012E, 733012810/10STDE, 7330128/10RMFE, 7330128/10RMEE, 733010E, 73301088E, 73301088/8SDE, 73301088/8RME, 73301088/8E, 73301088/10RME, 73301088/10E, 733010108/8S4E, 733010108/8RME, 733010/10RME, 733008E, 733006E, 7328888/8E, 7328668/6E, 73281488/8ARME, 7328148/10RMEE, 73281412/10ETE, 73281288/8E, 732812810/10STDE, 7328128/10RMFE, 7328128/10RMEE, 73281088E, 73281088/8SDE, 73281088/8RME, 73281088/8E, 73281088/10RME, 73281088/10E, 732810108/8S4E, 7326888/8E, 7326668/6E, 73261488/8ARME, 7326148/10RMEE, 732612810/10STDE, 7326128/10RMFE, 7326128/10RMEE, 73261088E, 73261088/8SDE, 73261088/8RME, 73261088/8E, 73261088/10RME, 73261088/10E, 732610108/8S4E, 732610/10RME, 732528E, 732526E, 732524E, 732522E, 732520E, 732518E, 7324888/8E, 7324668/6E, 73241488/8ARME, 7324148/10RMEE, 732412E, 732412810/10STDE, 7324128/10RMEE, 732412/6E, 73241088E, 73241088/8SDE, 73241088/8RME, 73241088/8E, 73241088/10RME, 73241088/10E, 732410108/8S4E, 73221488/8ARME, 732214/10/10CRME, 732212810/10STDE, 732212/10/8CRME, 732212/10/8ARME, 732211E, 73221088/8RME, 73221088/8E, 73221088/10E, 732210108/8S4E, 7320888/8E, 7320668/6E, 732038E, 732034E, 732034CX4RME, 732034/8E, 732034/10E, 732032E, 732032CX4RME, 732032/8E, 732032/10E, 732030E, 732030CX4RME, 732030/8E, 732030/10E, 732028CX4RME, 732028/8E, 732028/10E, 732026CX4RME, 732026/8E, 732026/10E, 732024CX4RME, 732024/8E, 732024/10E, 732022/10E, 732014/8X2E, 732014/8X2AE, 732014/10X2E, 732014/10X2AE, 732012810/10STDE, 732012/8X2E, 732012/8X2AE, 732012/10X2E, 732010E, 73201088/8RME, 73201088/8E, 732010108/8S4E, 732010/6E, 731809E, 731809/6E, 731608XL55E, 731608E, 731608/8/10RME, 731608/6E, 731608/10/10RME, 731516E, 731514E, 731512E, 731510E, 731508E, 731506E, 7314108/10RME, 731407XL55E, Class II

Reason: Gelatin sealed woven polyester prostheses devices, designed for vascular repair, were incorrectly labeled with an additional month of shelf life. Device expiry should have been calculated from gel impregnation (date of manufacture) but it was incorrectly calculated from device packaging.

Vascutek, Ltd. 20251218 Ongoing
Distributed: US Nationwide distribution including in the states of FL, PA, TX, NC, NY, CA, NJ, OH, MI, DC, CO, VA, GA, IL, OR, ND, VT, MS, MA, IN, WA, AZ, MN, MD, IA, OK, MO, AL, WI, LA, AR, TN, CT, ME, NE, SC, KS, NV, KY, RI, WV, UT, MT, ID, AK, NH, DE, NM, SD, HI, WY.
Brand Name: STA Liatest Free Protein S Product Name: STA Liatest Free Protein S Model/Catalog Number (REF): 00516 Product Description: The STA¿ - Liatest¿ Free Protein S kits are intended for use on STA-R¿ and STA Compact¿ for the quantitative antigenic assay of free protein S in human citrated plasma by the immuno-turbidimetric method. Component: Not Applicable Class II

Reason: The potential of out-of-range results and an underestimation of the free protein S level in normal patient plasmas. The defect only affects normal-range values.

Diagnostica Stago, Inc. 20260204 Ongoing
Distributed: Worldwide - US Nationwide distribution in the states of AZ, CA, CO, FL, GA, HI, IL, KY, MI, MN, MO, MT, NC, ND, NJ, NY, OH, PA, TN, TX, UT, VA, WA and the countries of ALBANIA, ALGERIA, ARGENTINA, AUSTRALIA, AUSTRIA, BAHRAIN, BANGLADESH, BELGIUM, BIELORUSSIE, BRAZIL, BULGARIA, CANADA, CHILE, CHINA, COSTA RICA, DENMARK, FINLAND, FRANCE, GERMANY, GREECE, GUADELOUPE, GUATEMALA, INDIA, IRAQ, IRELAND, ITALY, JORDAN, KENYA, KOREA REPUBLIC OF, LITUANIA, LUXEMBOURG, MALAYSIA, MEXICO, MOROCCO, NETHERLANDS, NEW CALEDONIA, NEW ZEALAND, PAKISTAN, PALESTINIAN TERRITORY, PARAGUAY, PORTUGAL, ROMANIA, RUSSIE, SAUDI ARABIA, SERBIA, SOUTH AFRICA, SPAIN, SWEDEN, SWITZERLAND, TAIWAN, THAILAND, TUNISIE, TURKEY, UNITED ARAB EMIRATES, UNITED KINGDOM, URUGUAY;
Esaote Endocavity ultrasonic probe, Model E 3-12, REF 120000038 Class II

Reason: A potential weakness has been identified in the probe body, so that, in a specific production batch, liquid may leak from the terminal part of the casing near the cable.

ESAOTE S.P.A. 20250314 Completed
Distributed: US: Nationwide distribution in the states of NC and TX.
Olympus Single Use Biopsy Valve. Model/Catalog Number: MAJ-1218. 20 units per box. The MAJ-1218 has been designed to be attached to the instrument channel port of compatible endoscopes to prevent leakage of body fluids. Class II

Reason: Potential for rubber fragment detachment during use.

Olympus Corporation of the Americas 20260129 Ongoing
Distributed: US Nationwide distribution.
Olympus Single use Biopsy Valve. Model/Catalog Number: MAJ-210. 20 units per box. The MAJ-210 has been designed to be attached to the instrument channel port of compatible endoscopes to prevent reflux of body fluids. Class II

Reason: Potential for rubber fragment detachment during use.

Olympus Corporation of the Americas 20260129 Ongoing
Distributed: US Nationwide distribution.
CS300 IABP. Software Version CS300 IABP C.01. Class II

Reason: The firm has identified that the battery runtime and cycle specifications contained in the devices Instructions for Use (IFU) need to be updated.

Datascope Corp. 20260123 Ongoing
Distributed: Worldwide distribution - US Nationwide and the countries of Afghanistan, Albania, Algeria, Argentina, Australia, Austria, Azerbaijan, Bahrain, Bangladesh, Barbados, Belarus, Belgium, Bolivia, Bosnia And Herzegovina, Brazil, Bulgaria, Cambodia, Canada, Chile, China, Colombia, Costa Rica, Cyprus, Czech Republic, Dominican Republic, Ecuador, Egypt, El Salvador , Estonia, France, Georgia, Germany, Ghana, Greece, Guatemala, Guyana, Honduras, Hong Kong, Hungary, India, Indonesia, Iran, Islamic Republic of, Iraq, Ireland, Israel, Italy, Jamaica, Japan, Jordan, Kenya, Kosovo, Kuwait, Latvia, Lebanon, Libya, Lithuania, Luxembourg, Macau, Macedonia, Malaysia, Maldives, Malta, Mauritius, Mexico, Morocco, Myanmar, Namibia, Nepal, Netherlands, New Zealand, Nigeria, North Macedonia, Norway, Oman, Pakistan, Palestine, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, Republic of Korea, Romania, Russian Federation, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Sudan, Sweden, Switzerland, Syria, Taiwan, Tanzania, Thailand, Trinidad, Tunisia, Turkey, Turkmenistan, Uganda, Ukraine, United Arab Emirates, United Kingdom, Uruguay, Venezuela, Vietnam, Yemen.
Page 195 of 4,030 | Showing 10 recalls