Oertli 1. 23G Laser Probe OS4 Flex-Tip VK401113 2. 23G Laser Probe OS4 Illuminated Flex-Tip VK401213 3. 23G Laser Probe OS4 Intuitive Exendable VK401323 4. 25G Laser Probe OS4 Flex-Tip VK401115 5. 25G Laser Probe OS4 Illuminated Flex-Tip VK401215 6. 25G Laser Probe OS4 Intuitive Exendable VK401325 7. 27G Laser Probe OS4 Intuitive Exendable VK401327 Class II

Reason: There may be voids located in the seal of Tyvek pouches associated with identified catalog numbers and lots. A compromised sterile barrier has potential of bioburden contamination which could lead to infection.

Vortex Surgical Inc. 20251216 Ongoing
Distributed: Worldwide Distribution: US (nationwide) to states of: AL, AR, AZ, CA, CT, FL, HI, IA, IL, IN, KY, MA, MI, MN, MS, NC, NE, NJ, NM, NY, OH, OR, PA, RI, SC, SD, TN, TX, WA, WI, VA; and OUS (foreign) to countries of: Australia, Chile, China, France, Germany, India, Ireland, Italy, Japan, Korea, Mexico, Netherlands, Philippines, Poland, Puerto Rico, Spain, Switzerland, Taiwan, United Kingdom, Vietnam.
Vortex Surgical 1. 23GA Laser Probe Curved Cat No. VS0120.23 2. 25GA Laser Probe Curved Cat No. VS0120.25 3. 25GA Laser Probe Flex-Tip Cat No. VS0130.25 4. 25GA Laser Probe Illuminated Curved Cat No. VS0125.25B 5. 25GA Laser Probe MaxReach Cat No. VS0140.25 6. 25GA Laser Probe Straight Cat No. VS0110.25 Class II

Reason: There may be voids located in the seal of Tyvek pouches associated with identified catalog numbers and lots. A compromised sterile barrier has potential of bioburden contamination which could lead to infection.

Vortex Surgical Inc. 20251216 Ongoing
Distributed: Worldwide Distribution: US (nationwide) to states of: AL, AR, AZ, CA, CT, FL, HI, IA, IL, IN, KY, MA, MI, MN, MS, NC, NE, NJ, NM, NY, OH, OR, PA, RI, SC, SD, TN, TX, WA, WI, VA; and OUS (foreign) to countries of: Australia, Chile, China, France, Germany, India, Ireland, Italy, Japan, Korea, Mexico, Netherlands, Philippines, Poland, Puerto Rico, Spain, Switzerland, Taiwan, United Kingdom, Vietnam.
Volk Single Use Vitrectomy Lenses 1. Flat Vitrectomy Lens, Catalog VFD 2. Volk Magnifying Lens, Catalog VMD Class II

Reason: There may be voids located in the seal of Tyvek pouches associated with identified catalog numbers and lots. A compromised sterile barrier has potential of bioburden contamination which could lead to infection.

Vortex Surgical Inc. 20251216 Ongoing
Distributed: Worldwide Distribution: US (nationwide) to states of: AL, AR, AZ, CA, CT, FL, HI, IA, IL, IN, KY, MA, MI, MN, MS, NC, NE, NJ, NM, NY, OH, OR, PA, RI, SC, SD, TN, TX, WA, WI, VA; and OUS (foreign) to countries of: Australia, Chile, China, France, Germany, India, Ireland, Italy, Japan, Korea, Mexico, Netherlands, Philippines, Poland, Puerto Rico, Spain, Switzerland, Taiwan, United Kingdom, Vietnam.
Focalyx Fusion Class II

Reason: Software device that is an accessory for image-guided interventional and diagnostic procedures involving prostate has accessories that may not function properly with Windows 10, which may cause patient harm during biopsy or cause treatment/therapy delays, so firm is recommending stop use until device is verified/validated with Windows 11.

Focalyx Technologies, LLC. 20251223 Ongoing
Distributed: Worldwide - US Nationwide distribution in the states of NY, MA, FL and the countries of ES, VE, DO.
NAEOTOM Alpha.Pro Software applications: syngo.CT Brain Quantification syngo.CT Vessel Hyperdensities syngo.CT ASPECTS Class II

Reason: To remove the software applications from certain CT systems as the applications have not received FDA 510(k) clearance.

Siemens Medical Solutions USA, Inc 20251219 Ongoing
Distributed: Worldwide distribution - US Nationwide and the countries of Canada, Costa Rica, India, Israel.
NAEOTOM Alpha.Prime Software applications: syngo.CT Brain Quantification syngo.CT Vessel Hyperdensities syngo.CT ASPECTS Class II

Reason: To remove the software applications from certain CT systems as the applications have not received FDA 510(k) clearance.

Siemens Medical Solutions USA, Inc 20251219 Ongoing
Distributed: Worldwide distribution - US Nationwide and the countries of Canada, Costa Rica, India, Israel.
NAEOTOM Alpha Software applications: syngo.CT Brain Quantification syngo.CT Vessel Hyperdensities syngo.CT ASPECTS Class II

Reason: To remove the software applications from certain CT systems as the applications have not received FDA 510(k) clearance.

Siemens Medical Solutions USA, Inc 20251219 Ongoing
Distributed: Worldwide distribution - US Nationwide and the countries of Canada, Costa Rica, India, Israel.
SOMATOM X.ceed Software applications: syngo.CT Brain Quantification syngo.CT Vessel Hyperdensities syngo.CT ASPECTS Class II

Reason: To remove the software applications from certain CT systems as the applications have not received FDA 510(k) clearance.

Siemens Medical Solutions USA, Inc 20251219 Ongoing
Distributed: Worldwide distribution - US Nationwide and the countries of Canada, Costa Rica, India, Israel.
MiniMed 780G Insulin Pump Catalog Numbers: MMT-1884 MMT-1886 With Software Version 6.60 The MiniMed 780G insulin pump is indicated for use by patients age 7-80 years with Type 1 diabetes, whose total daily dose of insulin is 8 units per day or more. The MiniMed 780G system is intended for continuous delivery of basal insulin at selectable rates, and the administration of insulin boluses at selectable amounts for the management of type 1 diabetes mellitus in persons seven years of age and older requiring insulin. The system is also intended to continuously monitor glucose values in the fluid under the skin. The MiniMed 780G system includes SmartGuard technology, which can be programmed to automatically adjust insulin delivery based on continuous glucose monitoring (CGM) sensor glucose values and can suspend delivery of insulin when the SG value falls below or is predicted to fall below predefined threshold values. Class II

Reason: A software timing and processor communication issue in MiniMed" 780G software version 6.60 may trigger Pump Error 43 and/or Pump Error 41, resulting in suspension of insulin delivery.

Medtronic MiniMed, Inc. 20251102 Ongoing
Distributed: Worldwide - U.S. National distribution including in the states of AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, VT, WA, WI, and WV. The country of Israel.
TORNIER PERFORM REVERSED PERIP SCREW 5.0MMX18MM NON STERILE, catalog number DWJ318 Class II

Reason: A specific lot of screws is incorrectly labeled. Screws are labeled as 5.0mmx18mm but actual screws are 5.00mmx14mm. If mismatch is detected during surgery, the issue may cause a minor delay in surgery. If not detected and incorrect screw is used, issue may increase risk of earlier mechanical failure, loosening, and need for revision surgery.

Tornier, Inc 20251219 Ongoing
Distributed: US Nationwide distribution.
Page 220 of 4,030 | Showing 10 recalls