Reason: Potential risk of Quality Control failures and/or false resistant antibiotic results when testing isolates of Enterobacterales/Enterobacteriaceae species and/or Pseudomonas aeruginosa with the identified antibiotic formulations. Issue may lead to QC failure or false resistant results.
Distributed: Worldwide - US Nationwide distribution.Reason: Sample preparation system designed to automate staining, lysing, cell washing has a software error that causes samples to be dispensed close to or touching tube bottoms, which could dislodge fluid line, result in sample loss, spillage, make instrument inoperable and will lead to unintentional removal of 60-100 microliters from final sample, may lead to incorrect diagnosis or patient management.
Distributed: Worldwide - US Nationwide distribution in the states of MO, WA, MA, IL, AL, OR, IL and the countries of AB, BE, ES, FR, NL, ON, DE, AT, VIC, AU, GB, SK, MX, CH.Reason: 942-01-40G acetabular system package was found to contain 942-01-36H acetabular system; and several other acetabular system and knee insert devices may also have a packaging discrepancy, which may cause surgical delay to exchange devices or revision surgery.
Distributed: US: TX, IA, MI, MS, MA, RI, NJ, PA, NY, MO, OK, PR, AL, CAReason: 942-01-40G acetabular system package was found to contain 942-01-36H acetabular system; and several other acetabular system and knee insert devices may also have a packaging discrepancy, which may cause surgical delay to exchange devices or revision surgery.
Distributed: US: TX, IA, MI, MS, MA, RI, NJ, PA, NY, MO, OK, PR, AL, CAReason: Due to a software implementation error that allowed instruments arms failing an important diagnostic test to remain in clinical use. This test was supposed to be used to detect Universal Surgical Manipulator insertion axis ball screws that exhibit the symptoms of being bent and misaligned which is a precursor to potential ball screw fatigue failure/fraction
Distributed: U.S.: AL, AR, CA, CO, CT, FL, ID, IL, IN, KS, KY, MD, MI, MN, MS, MO, NC, NJ, NV, NY, OH, OK, OR, PA, TN, TX, UT, VA, WA, and WI. O.U.S.: Austria, Brazil, China, France, Germany, Greece, India, Italy, Japan, Lebanon, Netherlands, South Korea, Spain, Switzerland, Taiwan and United KingdomReason: There is a potential dislodgement of the Prismaflex Set deaeration chamber from the Prismaflex Control Unit.
Distributed: US Nationwide distribution.Reason: There is a potential dislodgement of the Prismaflex Set deaeration chamber from the Prismaflex Control Unit.
Distributed: US Nationwide distribution.Reason: There is a potential dislodgement of the Prismaflex Set deaeration chamber from the Prismaflex Control Unit.
Distributed: US Nationwide distribution.Reason: There is a potential dislodgement of the Prismaflex Set deaeration chamber from the Prismaflex Control Unit.
Distributed: US Nationwide distribution.Reason: There is a potential dislodgement of the Prismaflex Set deaeration chamber from the Prismaflex Control Unit.
Distributed: US Nationwide distribution.