MEDLINE beds labeled as: 1) Semi Electric Basic Homecare Bed, REF MDR107002E; 2) Semi Electric Basic Homecare Bed, 4-pack,REF MDR107002E-4 Class I

Reason: Medline Industries has identified a potential hazard involving the hand control pendant and associated wires. In rare and atypical scenarios such as operating the bed above its specified weight limit, encountering an obstruction to bed movement, or experiencing jammed components/motor the hand pendant and pendant cords may overheat and, in rare instances, pose a risk of fire. Medline Industries, LP has received a total of (58) complaints associated with reports of pendant sparking, burning, melting, smoking or fire.

Medline Industries, LP 20251126 Ongoing
Distributed: US, CA, PA, GUAM, VI
Portex Spinal Tray, Item No. A3729-24 A3595 Class II

Reason: Affected lots of syringes may have a crack on the syringe collar, which may lead to delayed or repeated blood draw/fluid delivery. This issue may also lead to blood leakage during sampling of arterial blood or leakage during fluid delivery.

ICU Medical Inc. 20251219 Ongoing
Distributed: US Nationwide distribution in the states of CA, FL, IL, IN, MA, MI, MN, MO, NE, NJ, NY, OH, OK, TX, WA.
Pulsator Arterial Blood Sampling Kit, Item Numbers 4044EG-1 4080P-2 4084P-1 4080PE 4080PG Class II

Reason: Affected lots of syringes may have a crack on the syringe collar, which may lead to delayed or repeated blood draw/fluid delivery. This issue may also lead to blood leakage during sampling of arterial blood or leakage during fluid delivery.

ICU Medical Inc. 20251219 Ongoing
Distributed: US Nationwide distribution in the states of CA, FL, IL, IN, MA, MI, MN, MO, NE, NJ, NY, OH, OK, TX, WA.
CentriMag Blood Pump, REF: 102953, 201-20003, 201-90010, 201-90016; CentriMag RVAD HDE KIT, REF: 104318; CentriMag Pre-connected Pack, REF: CMAEK00; CentriMag VAD KIT, REF: 201-51006, 201-51007, 201-90001 201-90001T; Blood Pump with CentriMag Acute Circulatory Support System for ECMO, REF: CMAEK01. CentriMag Blood Pump is a component to be used with CentriMag Acute Circulatory Support System Class II

Reason: Acute circulatory support system pump may not fit into the motor and reports received of the pump not being securely locked into the motor using the screw-in locking feature. Failure to properly lock the pump into motor my not cause pump not inserted alarm, and could lead to particulate, increased heat, anxiety, inconvenience, delay in hemodynamic/ECMO support, stroke, embolism, death.

Thoratec LLC 20251010 Ongoing
Distributed: Worldwide - US Nationwide distribution including in the states of MN, PA, OK, FL, CA, OR, IL, NY, DE, GA, NJ, NC, WI, OH, AZ, AR, TX, TN, MO, MA, MT, KY, NE, WV, VA, ME, ND, AL, CO, DC, MS, IN, MI, CT, SC, WA, NM, MD, HI, ID, IA, LA, UT, RI, NV and the countries of Algeria, Australia, Austria, Bahrain, CANADA, Cayman Islands, Colombia, Costa Rica, Cyprus, Czech Republic, Denmark, Egypt, Estonia, Finland, France, Germany, Hong Kong, Italy, Jordan, Kuwait, Lithuania, Macedonia, Malaysia, Mexico, Netherlands, Oman, Peru, Poland, Portugal, Qatar, Romania, Russian Fed., Saudi Arabia, Serbia, Singapore, Slovenia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, United Arab Emirates, United Kingdom.
Abre Venous Self-Expanding Stent System, Model numbers AB9G14120090 AB9G12100090 AB9G14080090 AB9G12080090 AB9U12060090 AB9U14100090 AB9U14080090 AB9U12080090 AB9U14060090 AB9G14100090 AB9G12060090 Class II

Reason: Stents may be impacted by a Nitinol anomaly which causes a reduced stent diameter when deployed and corresponding reduced radial force. Issue may result in stent migration.

Medtronic Inc 20251204 Ongoing
Distributed: Worldwide - US Nationwide and the countries of Argentina, Australia, Austria, Belgium, Chile, Colombia, Costa Rica, Denmark, Dominican Republic, Finland, Germany, Jordan, Lebanon, Malaysia, Netherlands, New Zealand, Peru, Poland, Russia, Singapore, Spain, Switzerland, Taiwan, Thailand, United Arab Emirates, United Kingdom, Uzbekistan, Vietnam.
da Vinci SP, SP Access Port Kit, Large Incision Reference Numbers:430073 (2.7-4cm,SP1098, Box) & 430075 (2.7-7cm,SP1098,Box) Class II

Reason: Access Port System tray may develop cracks potentially resulting in a sterility breach.

Intuitive Surgical, Inc. 20251215 Ongoing
Distributed: Worldwide - US Nationwide distribution including in the states of Alabama, Arizona, California, Colorado, Connecticut, Florida, Georgia, Illinois, Indiana, Iowa, Kansas, Kentucky, Louisiana, Maine, Maryland, Massachusetts, Michigan, Minnesota, Mississippi, Missouri, Nevada, New Hampshire, New Jersey, New York, North Carolina, Ohio, Oklahoma, Oregon, Pennsylvania, South Carolina, South Dakota, Tennessee, Texas, Utah, Virginia, Washington, West Virginia, Wisconsin and the countries of Australia, Austria, Belgium, China, Czech Republic, Denmark, France, Germany, Greece, Hong Kong, Ireland, Italy, Japan, Poland, Romania, Saudi Arabia, South Korea, Spain, Sweden, Switzerland, Taiwan, United Arab Emirates, United Kingdom.
The product is used for the evaluation of the intrinsic coagulation pathway, APTT substitution test and the monitoring of heparin therapy. Class II

Reason: Potential for microbial contamination.

Instrumentation Laboratory 20251211 Ongoing
Distributed: Worldwide - US Nationwide distribution in the states of CA, FL, IL, MN, MS, NE, OH, TX, WI and the countries of Belgium, Netherlands, Luxembourg, Switzerland, China, Hong Kong, Israel, Italy, South Korea, Lithuania, Macao, Malaysia, Philippines, Qatar, Romania, Turkey, Kosovo.
Plum Duo Infusion System, List Number 40002-04-01 Class II

Reason: ICU Medical received one lot of defective primary speakers from our supplier that may fail to produce an audible sound. If the primary speaker in your Plum Duo Infusion Pump fails, the pump may not produce audible alarms, alerts, or provide audible feedback for touchscreen inputs.

ICU Medical, Inc. 20251215 Ongoing
Distributed: Worldwide - US Nationwide distribution.
ThinPrep¿ CytoLyt¿ Solution. Buffered Cell Wash Solution. Class II

Reason: Fungal contamination of affected lot with Parengyodontium album.

Hologic, Inc 20251212 Ongoing
Distributed: US Nationwide distribution.
Brand Name: Catalyst+ Product Name: Catalyst, Catalyst+ Model/Catalog Number: SP001-0026, SP003-0002 Software Version: cSRS module in combination with cMotion module Product Description: GMDN: 40893 Patient positioning device, diagnostic imaging/radiotherapy, laser Component: Not applicable Class II

Reason: Due to issues with the system's stereotactic radiosurgery (SRS) treatment cannot be guaranteed for all couch angles and system setups. n certain cases, the system may indicate the patient as correctly positioned at the isocenter, even when the isocenter is positioned outside of the indicated tolerance.

C-RAD POSITIONING AB 20251009 Ongoing
Distributed: The software module cSRS has been installed in 37 countries Worldwide - U.S. Nationwide distribution in the states of AK, AL, AR, CA, CT, FL, MO, NC, NM, NY, OH, OK, OR, PA, TN, TX, VA, and WA. The countries of Argentina, Australia, Austria, Belarus, Belgium, Benin, Canada, China, Colombia, Egypt, El Salvador, Finland, France, Germany, Greece, Hungary, India, Italy, Japan, Jordan, Mexico, Morocco, Netherlands, New Zealand, Norway, Portugal, Romania, Russia, Singapore, South Korea, Spain, Sweden, Switzerland, Taiwan, Ukraine, and Vietnam.