Detachable 10in Pendant Upgrade component for the AIRO Computed Tomography (CT) X-ray System. Class II

Reason: AIRO pendant replacement kits and upgrade do not possess the "Safe Hand Position Warning" label.

Mobius Imaging, LLC 20251204 Ongoing
Distributed: US Nationwide distribution in the states of AK, AL, AZ, CA, CO, CT, DC, FL, IL, KY, MA, MD, MI, MN, MO, NC, NY, OH, PA, RI, SC, TN, TX, UT, VA, WI.
Detachable 10in Pend FRU-1 Pend component for the AIRO Computed Tomography (CT) X-ray System. Class II

Reason: AIRO pendant replacement kits and upgrade do not possess the "Safe Hand Position Warning" label.

Mobius Imaging, LLC 20251204 Ongoing
Distributed: US Nationwide distribution in the states of AK, AL, AZ, CA, CO, CT, DC, FL, IL, KY, MA, MD, MI, MN, MO, NC, NY, OH, PA, RI, SC, TN, TX, UT, VA, WI.
Medicrea, IB3D PL Instruments Set, Rx Only REF: SPS03174 Class II

Reason: Due to complaints, orthopedic rotation tool handle may untread from the shaft resulting in the inability to rotate implant.

Medicrea International 20251029 Ongoing
Distributed: US Nationwide distribution in the states of TN, CA, MN, NC, TX, FL, OH, RI, VA.
Medicrea, IB3D Universal Implant Inserter, Rx Only REF: A24000000 Class II

Reason: Due to complaints, orthopedic rotation tool handle may untread from the shaft resulting in the inability to rotate implant.

Medicrea International 20251029 Ongoing
Distributed: US Nationwide distribution in the states of TN, CA, MN, NC, TX, FL, OH, RI, VA.
MAVERICK¿ Measurement Module Spectrometer PN 820-01367 (originally part of a kit PN 820-01519) Finished Goods PN 820-01496 Class II

Reason: Potential Elevated Laser Output A disconnected cable or a poorly crimped cable under mechanical bending stress can cause loss of electrical contact. At one connection on the board, the failure can drive the laser output to approximately 1,600 mW. Potential Unintended Laser Firing A disconnected, or poorly crimped cable can falsely signal a fire command during measurement cycles. At a secondary location on the board, this fault can lead to unexpected laser activation during operation. Safety logic prevents firing when the system is idle.

REPLIGEN CORPORATION 20251210 Ongoing
Distributed: U.S.
FreeStyle Libre 3 Plus Sensor REF 72080-01 UDI-DI code: 00357599819002 The FreeStyle Libre 3 Plus sensor is a component of the FreeStyle Libre 3 Continuous Glucose Monitoring System. Class I

Reason: Due to a manufacturing process issue that may cause carbon build up in the sensor that may result in incorrect low glucose results

Abbott Diabetes Care, Inc. 20251124 Ongoing
Distributed: U.S.: AK, AL, AZ, CA, CO, FL, GA, IA, ID, IL, IN, KS, LA, MA, MD, MI, MN, MO, MS, MT, NC, NE, NJ, NV, NY, OH, OK, PA, SC, SD, TX, UT, VA, WI, WV, and WY. O.U.S.: Andorra, Austria, Belgium, Canada, Denmark, Finland, France, Germany, Italy, Luxembourg, Netherlands, New Zealand, Norway, Qatar, San Marino, Spain, Sweden, Switzerland, and United Kingdom.
FreeStyle Libre 3 Plus Sensor REF 78769-01 UDI-DI code: 00357599843014 The FreeStyle Libre 3 Plus sensor is a component of the FreeStyle Libre 3 Continuous Glucose Monitoring System. Class I

Reason: Due to a manufacturing process issue that may cause carbon build up in the sensor that may result in incorrect low glucose results

Abbott Diabetes Care, Inc. 20251124 Ongoing
Distributed: U.S.: AK, AL, AZ, CA, CO, FL, GA, IA, ID, IL, IN, KS, LA, MA, MD, MI, MN, MO, MS, MT, NC, NE, NJ, NV, NY, OH, OK, PA, SC, SD, TX, UT, VA, WI, WV, and WY. O.U.S.: Andorra, Austria, Belgium, Canada, Denmark, Finland, France, Germany, Italy, Luxembourg, Netherlands, New Zealand, Norway, Qatar, San Marino, Spain, Sweden, Switzerland, and United Kingdom.
FreeStyle Libre 3 Plus Sensor REF 78768-01 UDI-DI code: 00357599844011 The FreeStyle Libre 3 Plus sensor is a component of the FreeStyle Libre 3 Continuous Glucose Monitoring System. Class I

Reason: Due to a manufacturing process issue that may cause carbon build up in the sensor that may result in incorrect low glucose results

Abbott Diabetes Care, Inc. 20251124 Ongoing
Distributed: U.S.: AK, AL, AZ, CA, CO, FL, GA, IA, ID, IL, IN, KS, LA, MA, MD, MI, MN, MO, MS, MT, NC, NE, NJ, NV, NY, OH, OK, PA, SC, SD, TX, UT, VA, WI, WV, and WY. O.U.S.: Andorra, Austria, Belgium, Canada, Denmark, Finland, France, Germany, Italy, Luxembourg, Netherlands, New Zealand, Norway, Qatar, San Marino, Spain, Sweden, Switzerland, and United Kingdom.
FreeStyle Libre 3 Sensor REF 72081-01 UDI-DI code: 00357599818005 The FreeStyle Libre 3 sensor is a component of the FreeStyle Libre 3 Continuous Glucose Monitoring System. Class I

Reason: Due to a manufacturing process issue that may cause carbon build up in the sensor that may result in incorrect low glucose results

Abbott Diabetes Care, Inc. 20251124 Ongoing
Distributed: U.S.: AK, AL, AZ, CA, CO, FL, GA, IA, ID, IL, IN, KS, LA, MA, MD, MI, MN, MO, MS, MT, NC, NE, NJ, NV, NY, OH, OK, PA, SC, SD, TX, UT, VA, WI, WV, and WY. O.U.S.: Andorra, Austria, Belgium, Canada, Denmark, Finland, France, Germany, Italy, Luxembourg, Netherlands, New Zealand, Norway, Qatar, San Marino, Spain, Sweden, Switzerland, and United Kingdom.
VANTAGE TITAN 3T Model MRT-3010/MEXL-3010 Class II

Reason: There is a potential for formation of ice occurring in the venting system of the superconducting magnet of the MRI systems. if a quench occurs in this condition, helium gas may not be able to escape via the designed vent paths, and the pressure in the helium vessel may rise. This pressure build-up could eventually rupture the helium vessel, causing helium gas to be released into the MR scanning room.

Canon Medical System, USA, INC. 20251219 Ongoing
Distributed: U.S. Nationwide distribution in the states of CA, CO, FL, IA, IL, IN, KS, MD, MO, MT, NV, NY, OH, OR, PA, PR, TX, VA, and WI.
Page 241 of 4,030 | Showing 10 recalls
Data provided by openFDA