VANTAGE GALAN 3T Model MRT-3020/MEXL-3020 Class II

Reason: There is a potential for formation of ice occurring in the venting system of the superconducting magnet of the MRI systems. if a quench occurs in this condition, helium gas may not be able to escape via the designed vent paths, and the pressure in the helium vessel may rise. This pressure build-up could eventually rupture the helium vessel, causing helium gas to be released into the MR scanning room.

Canon Medical System, USA, INC. 20251219 Ongoing
Distributed: U.S. Nationwide distribution in the states of CA, CO, FL, IA, IL, IN, KS, MD, MO, MT, NV, NY, OH, OR, PA, PR, TX, VA, and WI.
Codman CereLink ICP Sensor Basic Kit. Catalog Number: 826850. The Codman CereLink ICP Sensor Basic Kit consists of the CereLink ICP Sensor (ICP Sensor) and a 14-gauge Tuohy needle with stylet. The ICP Sensor is a nylon tube with a microminiature strain gauge pressure transducer (sensing element) mounted at one end and an electrical connector at the other end. It is designed for use with a Codman intracranial pressure monitoring device. The ICP Sensor Basic Kit is designed for use with the Codman Cranial Hand Drill. The drill facilitates access to the intraparenchymal area. The drill is also available as a component of the Codman Cranial Access Kit. Use of the ICP Sensor Basic Kit is indicated when direct ICP monitoring is required. The kit is indicated for use in both subdural and intraparenchymal pressure monitoring applications only. Class II

Reason: Potential corrosion stains on the surface of the 14-gauge Tuohy Needle that is included in the Microsensor and Cerelink ICP Kits.

Integra LifeSciences Corp. (NeuroSciences) 20251212 Ongoing
Distributed: Worldwide - US Nationwide distribution in the states of AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NJ, NV, NY, OH, OK, OR, PA, SC, SD, TN, TX UT, VA WA and the countries of Argentina, Australia, Austria, Bangladesh, Belgium, Bolivia (Plurinational State of), Brazil, Canada, Chile, China, Colombia, Cyprus, Czechia, France, Georgia, Germany, Greece, Hong Kong, Iceland, India, Ireland, Israel, Italy, Japan, Kenya, Korea (the Republic of), Lithuania, Nepal, New Zealand, Poland, Portugal, Romania, Saudi Arabia, Singapore, Slovenia, South Africa, Spain, Sweden, Taiwan (Province of China), Thailand, Turkey, United Arab Emirates (the), United Kingdom of Great Britain and Northern Ireland (the).
Codman Microsensor Basic Kit. Catalog Number: 626631US. Use of the CODMAN MICROSENSOR Basic Kit is indicated when direct ICP monitoring is required. The kit is indicated for use in both subdural and intraparenchymal pressure monitoring applications only. Class II

Reason: Potential corrosion stains on the surface of the 14-gauge Tuohy Needle that is included in the Microsensor and Cerelink ICP Kits.

Integra LifeSciences Corp. (NeuroSciences) 20251212 Ongoing
Distributed: Worldwide - US Nationwide distribution in the states of AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NJ, NV, NY, OH, OK, OR, PA, SC, SD, TN, TX UT, VA WA and the countries of Argentina, Australia, Austria, Bangladesh, Belgium, Bolivia (Plurinational State of), Brazil, Canada, Chile, China, Colombia, Cyprus, Czechia, France, Georgia, Germany, Greece, Hong Kong, Iceland, India, Ireland, Israel, Italy, Japan, Kenya, Korea (the Republic of), Lithuania, Nepal, New Zealand, Poland, Portugal, Romania, Saudi Arabia, Singapore, Slovenia, South Africa, Spain, Sweden, Taiwan (Province of China), Thailand, Turkey, United Arab Emirates (the), United Kingdom of Great Britain and Northern Ireland (the).
Medline Hudson RCI Medium Concentration Oxygen Masks: 1041CE MASK,OXYGEN,UND CHIN,MED CO OUS ONLY; HUD1035 O2 MASK,MED CONC,SHORT,PEDI,7' TUB SC; HUD1042 O2 MASK,MED CONC,PEDI,7' TUB SC; HUD1930 O2 MASK,MED CONC,ADULT,7' TUB UC; HUDRHO41U MASK,OXYGEN,MEDIUM-CONCENTRATION,7,SC Class II

Reason: Firm received multiple complaints regarding tubing disconnecting from oxygen mask. If issue occurs during use, may lead to delay in patient care and patient may experience shortness of breath leading to hypoxia, which may require medical intervention to prevent further respiratory deterioration.

Medline Industries, LP 20251121 Ongoing
Distributed: Worldwide distribution - US Nationwide and the countries of Australia, Bahamas, Canada, China, Dominican Republic, Guam, South Korea, Mexico, Panama, and Singapore.
Brand Name: HFD100 Product Name: Head Fixation Device HFD100 Model/Catalog Number: 119695-000, 119629-000, 119630-000, 113803-000, 113803-600, 113802-000, 113802-600 Software Version: Not Applicable Product Description: The HFD100 Head Fixation Device provides rigid skeletal fixation within the optimal imaging envelope while introducing minimal artifacts in the acquired images. There are two different assemblies for the HFD100. The standard HFD100 is equipped with a table adapter for use with either the operating room table or the angiography table. For sites with a Hill-Rom TruSystem 7500 Hybrid Neuro MR Operating Table (ORT400), the HFD100 has a split block clamp assembly (Trumpf Mount Clamp Assembly) which connects directly to the table back section. The HFD100 system is also comprised of the linkage components to mount the skull clamp to the table adapter, and the 3-pin skull clamp. The linkage system is used to position the skull clamp in the necessary position, without colliding with the magnet bore. Component: Not a component. Class II

Reason: Torque screw may crack or separate. If used, an affected screw could lead to inadequate head stabilization, compromised surgical precision, procedural delays, requirement for alternate head fixation methods, and risk of patient injury.

IMRIS Imaging Inc 20251208 Ongoing
Distributed: Worldwide distribution - US Nationwide and the countries of Canada, Sweden, Switzerland, Japan, Australia, Qatar, and China.
Philips Smart-hopping 2.0 AP 1.4 GHz. Patient Monitor. Class II

Reason: It was found that the MX40 device could not reconnect to the PIC iX when moving between Radiohead and Trident 1.4 GHz access points if the signal strength changed quickly.

Philips North America Llc 20251212 Ongoing
Distributed: US Nationwide distribution.
Arial 900 MHz Call Station, Model Number CSK200-900MR, with Firmware version: 1.6.3 Class II

Reason: Securitas Healthcare has become aware of a firmware issue in the Call Stations in the Affected Lots that may cause Call Stations in the Affected Lots to (1) not transmit a low battery alert to the Arial Wireless Emergency Call System prior to battery failure, or (2) not transmit a low battery alert to the Arial Wireless Emergency Call system at least seven (7) days prior to the battery failing.

Securitas Healthcare LLC 20251208 Ongoing
Distributed: Worldwide distribution -US Nationwide and the country of Canada.
Arial 900 MHz Call Station, Model Number CSK200-900, with Firmware version: 1.6.3 Class II

Reason: Securitas Healthcare has become aware of a firmware issue in the Call Stations in the Affected Lots that may cause Call Stations in the Affected Lots to (1) not transmit a low battery alert to the Arial Wireless Emergency Call System prior to battery failure, or (2) not transmit a low battery alert to the Arial Wireless Emergency Call system at least seven (7) days prior to the battery failing.

Securitas Healthcare LLC 20251208 Ongoing
Distributed: Worldwide distribution -US Nationwide and the country of Canada.
Impella 5.5 with SmartAssist¿ S2, Impella 5.5 with SmartAssist S2 Set, AU. Class II

Reason: Device packaged in incorrect outer box carton.

Abiomed, Inc. 20251204 Ongoing
Distributed: International distribution to the country of Australia and Taiwan.
The Pressio 2 ICP Monitoring System, Model Number PSO-4000; Intracranial Pressure Monitor Class II

Reason: Customer complaints of Pressio monitor rebooting.

Sophysa 20251117 Ongoing
Distributed: US Nationwide distribution in the states of AZ, WI, TX, OR, CA, OK, ID, FL, NJ.