Halyard ORGAN RECOVERY OR PACK, Kit Code: LLOG1000-15. Class I

Reason: Devices are not suitable for organ transplant.

AVID Medical, Inc. 20251111 Ongoing
Distributed: US Nationwide distribution in the states of FL, GA and Puerto Rico.
LUMINOS Lotus Max, Model 11574100 and Luminos dRF Max, Model 10762471 Class II

Reason: A potential issue associated with LUMINOS Lotus Max and Luminos dRF Max systems with software version VF11 with bucky wall stand for mobile detectors and without ceiling stand .When switching quickly between organ programs (OGPs) the collimation may be incorrect.

Siemens Medical Solutions USA, Inc 20251205 Ongoing
Distributed: U.S. and OUS
The AlignRT InBore system contains six Class 1 lasers which enables the system s cameras to track the patient s position and movement. Class II

Reason: Vision RT is informing customers of an omission identified in the Instructions for Use for AlignRT InBore systems. Information identifying the lasers and their characteristics was omitted from the Instructions for Use for the subject system as required by 21 CFR 1040,10(h), via conformance with recognized consensus standard IEC 60825-1 Ed. 3 (FDA Laser Notice 56).

Vision RT Ltd 20250827 Ongoing
Distributed: U.S. and OUS
BVI Leos Laser and Endoscopy System. Model Number: OME6000U. Class II

Reason: Manufacturing deviation in which the ferrites specified for the USB cable assembly linking the Nurse Monitor touchscreen to the System Controller were not installed as intended. The omission of the ferrites on the USB communication cable results in non-conformance to the design specifications, which require two ferrites (one at each end of the cable).

Beaver-Visitec International, Inc. 20251205 Ongoing
Distributed: US Nationwide distribution in the states of FL, MI, OK, TX, WI.
Philips EPIQ Ultrasound Systems (various models) used in conjunction with the X5-1c transducer REF: , , 795200, 795201, 795231, 795232, 795234 REF: 795117, 795122 Class II

Reason: Retroactive: Ultrasound may experience an unexpected automatic reboot, resulting in damage to transducer.

Philips Ultrasound, LLC 20220714 Completed
Distributed: Worldwide - US Nationwide distribution in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY and the countries of Argentina, Australia, Belgium, Canada, Chile, Colombia, Czech Republic, Denmark, Ecuador, Estonia, Finland, France, Germany, Greece, Hong Kong, Hungary, India, Iran, Iraq, Ireland, Israel, Italy, Japan, Korea, Republic of Latvia, Lithuania, Morocco, Netherlands , New Zealand, Oman, Panama, Philippines, Poland, Romania, Russian Federation, Saudi Arabia, Serbia, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Thailand, Turkey, United Arab Emirates, United Kingdom.
Sheathes3D Seamless Viral Barrier, Latex-Free, 7/8"D Tapered to 2.9"W x 11.8"L (2.2cm Tapered to 7.4cm x 30cm), Non-Sterile, Ultrasound Probe Cover, Vaginal/Rectal, Tapered, Rolled, Extra-Long, Individually Wrapped, Box of 50 Item Number 76339, Class II

Reason: Vaginal/rectal ultrasound probe covers with a viral barrier may have product defects, which may result in contamination during use, posing an infection risk.

Sheathing Technologies Inc 20251117 Ongoing
Distributed: US Nationwide distribution in the state of WA.
Brand Name: Mastisol Liquid Adhesive Product Name: Liquid Adhesive Model/Catalog Number: 0523-48 Product Description: Liquid adhesive, 2/3 mL vial, 48 vials per carton Component: No Class II

Reason: Butyrate tube cracks during actuation, rendering product unusable.

Ferndale Laboratories, Inc. 20251125 Ongoing
Distributed: Worldwide distribution - US Nationwide and the country of UK.
Vortex Surgical 25ga Illuminated Flex-Tip Laser Probe, Catalog Number VS0135.25; Class II

Reason: Reason for the voluntary recall is some products in this lot are not passing through a 25ga cannula smoothly.

Vortex Surgical Inc. 20251119 Ongoing
Distributed: US and Japan
Material Processing Laser Products Class II

Reason: Lack of Remote Interlock Connector - required per 21 CFR 1040.10(f)(3)

SHENZHEN ATOMSTACK TECHNOLOGIES 20250930 Ongoing
Distributed: U.S.
BECKMAN COULTER BICARBONATE REF: OSR6137 OSR6237 OSR6637 OSR6x37 Bicarbonate is a system reagent for the quantitative determination of Bicarbonate in human serum and plasma on Beckman Coulter AU analyzers. Class II

Reason: Bicarbonate reagent may generate falsely high Bicarbonate results due to interference with Lactate Dehydrogenase (LDH) in the test sample.

Beckman Coulter Ireland, Inc. 20251014 Ongoing
Distributed: Worldwide - U.S. Nationwide distribution in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, Guan, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV, and WY. The countries of Argentina, Brazil, Canada, Ecuador, Ireland, Mexico, Netherlands, Panama, Taiwan, and Province of China.
Page 244 of 4,030 | Showing 10 recalls
Data provided by openFDA