Brand Name: Xpert¿ BCR-ABL Ultra Model/Catalog Number: GXBCRABL-10 Class II

Reason: Invitro diagnostic test kits with specimens that have elevated white blood cell count (>30,000 cell) may cause inaccurate results.

Cepheid 20251030 Ongoing
Distributed: US: AL AR AZ CA CO FL GA IA IL KS KY LA MD ME MI MN MO NC NE NH NJ NY OH PA RI TX VA WA WI OUS: Algeria Armenia Australia Austria Bahrain Bangladesh Belgium Brazil Burkina Faso Cambodia Canada Canary islands Chile Colombia Cyprus Czech Republic Dem. Rep. Congo Djibouti Dominican Rep. Ecuador El Salvador Eswatini Finland France Frenc.Polynesia Gambia Georgia Germany Ghana Greece Guadeloupe Guatemala Honduras Hong Kong India Indonesia Iraq Israel Italy Jamaica Jordan Kazakhstan Kenya Kuwait Latvia Lebanon Lithuania Luxembourg Malawi Malaysia Mauritania Mauritius Mexico Morocco Mozambique Myanmar Netherlands New Caledonia New Zealand Nicaragua Norway Oman Pap. New Guinea Peru Philippines Poland Portugal Qatar Rep.of Congo Reunion Rwanda Saudi Arabia Senegal Seychelles Singapore South Africa Spain Sri Lanka Sweden Switzerland Taiwan Tajikistan Tanzania Thailand Togo Tunisia Uganda United Kingdom Uruguay Utd.Arab Emir. Venezuela Zambia Zimbabwe
Brand Name: Xpert¿ BCR-ABL Ultra Model/Catalog Number: GXBCRABL-US-10 Class II

Reason: Invitro diagnostic test kits with specimens that have elevated white blood cell count (>30,000 cell) may cause inaccurate results.

Cepheid 20251030 Ongoing
Distributed: US: AL AR AZ CA CO FL GA IA IL KS KY LA MD ME MI MN MO NC NE NH NJ NY OH PA RI TX VA WA WI OUS: Algeria Armenia Australia Austria Bahrain Bangladesh Belgium Brazil Burkina Faso Cambodia Canada Canary islands Chile Colombia Cyprus Czech Republic Dem. Rep. Congo Djibouti Dominican Rep. Ecuador El Salvador Eswatini Finland France Frenc.Polynesia Gambia Georgia Germany Ghana Greece Guadeloupe Guatemala Honduras Hong Kong India Indonesia Iraq Israel Italy Jamaica Jordan Kazakhstan Kenya Kuwait Latvia Lebanon Lithuania Luxembourg Malawi Malaysia Mauritania Mauritius Mexico Morocco Mozambique Myanmar Netherlands New Caledonia New Zealand Nicaragua Norway Oman Pap. New Guinea Peru Philippines Poland Portugal Qatar Rep.of Congo Reunion Rwanda Saudi Arabia Senegal Seychelles Singapore South Africa Spain Sri Lanka Sweden Switzerland Taiwan Tajikistan Tanzania Thailand Togo Tunisia Uganda United Kingdom Uruguay Utd.Arab Emir. Venezuela Zambia Zimbabwe
InPen App, Model/CFN Number: MMT-8061 (Android Users) Class II

Reason: When app is uninstalled and reinstalled, insulin pen software issue causes Choose Notification Style Screen to not show during setup so users can't choose to allow notifications to override phone settings when on mute/Do Not Disturb, and previously set up override permission is deleted, so audible and vibratory notifications not received, which may lead to delayed insulin therapy, hyperglycemia.

Medtronic MiniMed, Inc. 20251113 Ongoing
Distributed: US: IL, AZ, MN, NE, KS, TN, CA, FL, TX, WA, MO, NC, NY, WI, IN, ME, CT, RI, MD, PA, OK, KY, MA, DE, UT, ID, VA, MI, OH, LA, NJ, IA, MT, HI, AR, OR, SC, AL, CO, WY, VT, NV, GA, MS, SD, AK, WV, NH, ND, NM. OUS: Austria, Belgium, Chile, Czech Republic, Denmark, Egypt, Estonia, Finland, France, Georgia, Germany, Greece, Hungary, Iceland, Ireland, Israel, Italy, Luxembourg, Netherlands, Norway, Poland, Portugal, Romania, South Africa, Spain, Sweden, Switzerland, United Kingdom
BIOSURE HA SCREW, 6MM x 20 MM. Threaded metallic bone fixation fastener. Class II

Reason: Removal of affected lot of screws due to labeling error.

Smith & Nephew, Inc. 20251124 Ongoing
Distributed: US (nationwide) distribution to states of : AL, CA, CO, FL, IA, IL, KY, MN, NC, NE, NY, OK, TN, TX, UT; OUS (Foreign) to countries of: Australia, Czech Republic, Finland, France, Germany, Indonesia, Italy, Malaysia, Martinique, Mexico, Netherlands, Norway, Poland, Singapore, Switzerland.
Myosa for Kids KS1 REF MyOSAKS1_Sb/Sp/Mb/Mp/Lb/Lc KS2 REF MyOSAKS2_Sb/Sc/Mb/Mc/Mp/Lb/Lc Myosa for Kids is a mouthguard in 3 different sizes (sm, med, lg) that can be used in children ages 3-15 to address breathing and myofunctional disorders. Catalog Numbers: Class II

Reason: Due to firm mistakenly distributing product within the U.S. without FDA premarket clearance/approval.

Myofunctional Research Company USA 20251009 Ongoing
Distributed: U.S. (nationwide) distribution to states of: Nationwide: AK, AR, CA, CO, FL, GA, HI, IL, KS, KY, LA, MA, MD, MI, MN, NC, NJ, NM, NY, OH, OR, PA, TX, UT, VA, and WA; O.U.S. (foreign) to country of: Australia
Brand Name: Rocket Product Name: Rocket 20Fg Seldinger Catheter Procedure Tray Model/Catalog Number: R54549-20-PK Software Version: N/A Product Description: Chest tube manufactured from PVC, with fenestrations and a barium stripe to the indicated French gauge size. Supplied with a procedure pack to allow percutaneous insertion, supplied with connectors to allow attachment to chest drain bottle devices. Component: N/A Class II

Reason: IFU and device kit labeling contains misleading statements leading to improper securing of the device resulting in the potential to fall out during use.

Rocket Medical Plc 20251107 Ongoing
Distributed: Domestic: CO, CT, IL, MA, ME, NH, NJ, NM, OH, RI, SC, TX; International: UAE, Austria, Australia, Canada, Germany, Denmark, Finland, France, United Kingdom, Greece, Hong Kong, Ireland, Israel, India, Italy, Jersey, Kenya, Sri Lanka, Malta, Netherlands, Norway, New Zealand, Portugal, Saudi Arabia, Singapore, Slovenia, South Africa
Brand Name: Rocket Product Name: Rocket 16Fg Seldinger Catheter Procedure Tray Model/Catalog Number: R54549-16-PK Software Version: N/A Product Description: Chest tube manufactured from PVC, with fenestrations and a barium stripe to the indicated French gauge size. Supplied with a procedure pack to allow percutaneous insertion, supplied with connectors to allow attachment to chest drain bottle devices. Component: N/A Class II

Reason: IFU and device kit labeling contains misleading statements leading to improper securing of the device resulting in the potential to fall out during use.

Rocket Medical Plc 20251107 Ongoing
Distributed: Domestic: CO, CT, IL, MA, ME, NH, NJ, NM, OH, RI, SC, TX; International: UAE, Austria, Australia, Canada, Germany, Denmark, Finland, France, United Kingdom, Greece, Hong Kong, Ireland, Israel, India, Italy, Jersey, Kenya, Sri Lanka, Malta, Netherlands, Norway, New Zealand, Portugal, Saudi Arabia, Singapore, Slovenia, South Africa
Brand Name: Rocket Product Name: Rocket 18Fg Seldinger Catheter with Safety Guard Model/Catalog Number: R54544-18-SG Software Version: N/A Product Description: Chest tube manufactured from PVC, with fenestrations and a barium stripe to the indicated French gauge size. Supplied with a procedure pack to allow percutaneous insertion, supplied with connectors to allow attachment to chest drain bottle devices. Class II

Reason: IFU and device kit labeling contains misleading statements leading to improper securing of the device resulting in the potential to fall out during use.

Rocket Medical Plc 20251107 Ongoing
Distributed: Domestic: CO, CT, IL, MA, ME, NH, NJ, NM, OH, RI, SC, TX; International: UAE, Austria, Australia, Canada, Germany, Denmark, Finland, France, United Kingdom, Greece, Hong Kong, Ireland, Israel, India, Italy, Jersey, Kenya, Sri Lanka, Malta, Netherlands, Norway, New Zealand, Portugal, Saudi Arabia, Singapore, Slovenia, South Africa
Brand Name: Rocket Product Name: Rocket 18Fg Seldinger Catheter Procedure Tray Model/Catalog Number: R54544-18-PK Software Version: N/A Product Description: Chest tube manufactured from PVC, with fenestrations and a barium stripe to the indicated French gauge size. Supplied with a procedure pack to allow percutaneous insertion, supplied with connectors to allow attachment to chest drain bottle devices. Component: N/A Class II

Reason: IFU and device kit labeling contains misleading statements leading to improper securing of the device resulting in the potential to fall out during use.

Rocket Medical Plc 20251107 Ongoing
Distributed: Domestic: CO, CT, IL, MA, ME, NH, NJ, NM, OH, RI, SC, TX; International: UAE, Austria, Australia, Canada, Germany, Denmark, Finland, France, United Kingdom, Greece, Hong Kong, Ireland, Israel, India, Italy, Jersey, Kenya, Sri Lanka, Malta, Netherlands, Norway, New Zealand, Portugal, Saudi Arabia, Singapore, Slovenia, South Africa
Brand Name: Rocket Product Name: Rocket 12Fg Seldinger Catheter with Safety Guard Model/Catalog Number: R54544-12-SG Software Version: N/A Product Description: Chest tube manufactured from PVC, with fenestrations and a barium stripe to the indicated French gauge size. Supplied with a procedure pack to allow percutaneous insertion, supplied with connectors to allow attachment to chest drain bottle devices. Component: N/A Class II

Reason: IFU and device kit labeling contains misleading statements leading to improper securing of the device resulting in the potential to fall out during use.

Rocket Medical Plc 20251107 Ongoing
Distributed: Domestic: CO, CT, IL, MA, ME, NH, NJ, NM, OH, RI, SC, TX; International: UAE, Austria, Australia, Canada, Germany, Denmark, Finland, France, United Kingdom, Greece, Hong Kong, Ireland, Israel, India, Italy, Jersey, Kenya, Sri Lanka, Malta, Netherlands, Norway, New Zealand, Portugal, Saudi Arabia, Singapore, Slovenia, South Africa
Page 249 of 4,030 | Showing 10 recalls
Data provided by openFDA