MADSEN ACCUSCREEN TEOAE/DPOAE/ABR Probe (P/N 8-69-41100) Class II

Reason: Following a component change in the DPOAE probe (speaker replacement), technical distortions may occur at high stimulus levels combined with unusually low OAE amplitude acceptance thresholds. These distortions can mimic true DPOAE responses and may result in a false PASS outcome.

PATH 20251106 Ongoing
Distributed: Domestic: WI. International: Ireland.
HemosIL SynthAFax. Partial Thromboplastin Time Tests. Class III

Reason: Recalled lots were manufactured with double the amount of preservative concentration.

Instrumentation Laboratory 20251118 Ongoing
Distributed: Domestic: CA, CO, FL, NY; Foreign: AR, AT, CA, CZ, DE, DK, ES, FR, GB, IE, IL, IN, IT, JP, LK, NL, PL, PT, RO, RS, TH, VN.
Orthofix NorthStar OCT Navigation Surgical Technique Guide, and instruments: Navigation Driver, REF: PC2-400012; Navigation Straight Bone Probe, REF: PC2-400013; Navigation, 3.5mm Drill, REF: PC2-400135; Navigation, 4.0mm Drill, REF: PC2-400140; Navigation, 4.5mm Drill, PC2-400145; Navigation, 5.5mm Drill, REF: PC2-400155; Navigation, 3.5mm Tap, REF: PC2-400235; Navigation, 4.0mm Tap, REF: PC2-400240; Navigation, 4.5mm Tap, REF: PC2-400245; Navigation, 4.5mm Dual Lead Tap, REF: PC2-400345; Navigation, 5.5mm Dual Lead Tap, REF: PC2-400355 Class II

Reason: The NorthStar OCT Navigation Surgical Technique Guide (STG) incorrectly identified both Medtronic s Solera and Medtronic s Infinity system toolcards as compatible with the NorthStar OCT Navigation Instruments. However, the NorthStar OCT Navigation Instruments are compatible for use with Medtronic s Solera toolcards only.

SEASPINE ORTHOPEDICS CORPORATION 20251117 Ongoing
Distributed: US: CT, TX, NV, CO, CA, WA, NJ, OR, IN, MI, FL, PA, IL, VA, AL, RI, MO, AZ, MS, SC, UT, NY, OK, SD. OUS: Australia, Mexico, Chile
5008X CAREsystem +CLiC +CDX; Class II

Reason: Several software anomalies with the potential to impact patient treatment, of which one includes the possibility to result in treatment stop if certain conditions are fulfilled.

Fresenius Medical Care Holdings, Inc. 20251025 Ongoing
Distributed: Domestic: LA, MA, MN;
GE HealthCare ViewPoint 6, Catalog Number H47601AA; Radiological image processing system Class II

Reason: GE HealthCare has become aware that the Patient History tab in ViewPoint 6 versions 6.15.3 and 6.15.4 ultrasound software may not include complete patient medical history information.

GE Healthcare GmbH 20251103 Ongoing
Distributed: Worldwide
Cardiosave Rescue Class III

Reason: The IFU addendum updates the Vibration and Shock Table to reference the correct standards.

Datascope Corp. 20251023 Ongoing
Distributed: Domestic: Nationwide distribution. International: Albania, Algeria, Angola, Argentina, Armenia, Republic of, Australia, Austria, Azerbaijan, Bahrain, Bangladesh, Belarus, Belgium, Bosnia And Herzegovina, Botswana, Republic of, Brazil, Brunei, Bulgaria, Canada, Chile, China, Colombia, Costa Rica, C¿te d Ivoire, (Ivory Coast), Croatia, Cyprus, Czech Republic, Denmark, Ecuador, Egypt, El Salvador, Estonia, Finland, France, Germany, Ghana, Guatemala, Hong Kong, Hungary, Iceland, India, Indonesia, Iran, Islamic Republic of, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kenya, Korea, Republic of, Kosovo, Kuwait, Latvia, Lebanon, Libya, Lithuania, Luxembourg, Malaysia, Mauritius, Mexico, Montenegro, Morocco, Myanmar, (Burma), Netherlands, New Zealand, Nicaragua, Nigeria, North Macedonia, Norway, Oman, Pakistan, Palestine Territory, Peru, Philippines, Poland, Portugal, Qatar, Russian Federation, Rwanda, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Sweden, Switzerland, Syrian Arab Republic, Taiwan, Republic Of China, Tanzania, United Republic Of, Thailand, Trinidad and Tobago, Tunisia, Turkey, Turkmenistan, Ukraine, United Arab Emirates, United Kingdom, Uruguay, Uzbekistan, Vietnam, Yemen, Zambia;
Cardiosave Hybrid Class III

Reason: The IFU addendum updates the Vibration and Shock Table to reference the correct standards.

Datascope Corp. 20251023 Ongoing
Distributed: Domestic: Nationwide distribution. International: Albania, Algeria, Angola, Argentina, Armenia, Republic of, Australia, Austria, Azerbaijan, Bahrain, Bangladesh, Belarus, Belgium, Bosnia And Herzegovina, Botswana, Republic of, Brazil, Brunei, Bulgaria, Canada, Chile, China, Colombia, Costa Rica, C¿te d Ivoire, (Ivory Coast), Croatia, Cyprus, Czech Republic, Denmark, Ecuador, Egypt, El Salvador, Estonia, Finland, France, Germany, Ghana, Guatemala, Hong Kong, Hungary, Iceland, India, Indonesia, Iran, Islamic Republic of, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kenya, Korea, Republic of, Kosovo, Kuwait, Latvia, Lebanon, Libya, Lithuania, Luxembourg, Malaysia, Mauritius, Mexico, Montenegro, Morocco, Myanmar, (Burma), Netherlands, New Zealand, Nicaragua, Nigeria, North Macedonia, Norway, Oman, Pakistan, Palestine Territory, Peru, Philippines, Poland, Portugal, Qatar, Russian Federation, Rwanda, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Sweden, Switzerland, Syrian Arab Republic, Taiwan, Republic Of China, Tanzania, United Republic Of, Thailand, Trinidad and Tobago, Tunisia, Turkey, Turkmenistan, Ukraine, United Arab Emirates, United Kingdom, Uruguay, Uzbekistan, Vietnam, Yemen, Zambia;
Halyard, HEAD AND NECK TRAY. Catalog Number: EURO016-09. Class II

Reason: Potential for incomplete seal on header bag.

AVID Medical, Inc. 20251114 Ongoing
Distributed: US distribution to GA, IL, NE, MD.
Halyard, MAJOR PACK. Catalog Number: JEMJ10-01. Class II

Reason: Potential for incomplete seal on header bag.

AVID Medical, Inc. 20251114 Ongoing
Distributed: US distribution to GA, IL, NE, MD.
Halyard, MIDTOWN GENERAL ARTERIOGRAM. Catalog Number: EUHM009-01. Class II

Reason: Potential for incomplete seal on header bag.

AVID Medical, Inc. 20251114 Ongoing
Distributed: US distribution to GA, IL, NE, MD.