IntelePACS - InteleConnect / TechPortal Class II

Reason: Software application that receives digital images and data to be communicated, processed, manipulated, enhanced, stored, displayed has a bug that could cause data loss in Relevant Clinical Info field when modifying studies through Case Editor in either Technologist Portal or Referring Physician Portal, which could result in loss of clinical information, which could impact clinical decision-making.

INTELERAD MEDICAL SYSTEMS INCORPORATED 20251017 Ongoing
Distributed: US Nationwide distribution in the states of AZ, CT, AL, IL, WA, MD, NJ,CA, TN, IN, FL, NC.
Cold Form Wrap: Back, REF: DP163CT03-BLK-L,XL, and Shoulder, REF: DP163CT02 Class II

Reason: The IFU (Information for Use) on the wrap instructs the user to heat the hot and cold compression wrap in the microwave for 3 minutes, which may cause the wrap to leak or break at the seams and/or reach temperatures capable of causing a second degree burn.

DJO, LLC 20250925 Ongoing
Distributed: US Nationwide distribution in the states of WI, TN, TX, MD, GA, UT, WA, MI, NC, NE, CA, MA, CO, SC, OH, IN, MN, NM, PA, NY, IL, FL, NJ, VA, AZ, VT, CT, OK.
Isoflurane (ISO) V90 Electronic Vaporizer with the A9 Anesthesia System. Labeled as the following: 1. Iso E_Vap. (Safety Filling adapter), Part Number 115-066758-00. 2. Exchange Iso E_Vap. (Safety Filling adapter), Part Number E115-066758-00. Product Description: The Anesthesia System is a device used to administer to a patient, continuously or intermittently, a general inhalation anesthetic and to maintain a patient's ventilation. The A9 Anesthesia System utilizes V90 electronic vaporizers for anesthetic agent delivery. Class II

Reason: Potential for anesthesia leakage.

Mindray DS USA, Inc. dba Mindray North America 20251020 Ongoing
Distributed: Worldwide - US Nationwide distribution in the states of California, Florida, Georgia, Kentucky, Maryland, Mississippi, Missouri, New Jersey, Puerto Rico, Texas, Washington, West Virginia and the country of Canada.
Sevoflurane (SEVO) V90 Electronic Vaporizer with the A9 Anesthesia System. Labeled as the following: 1. Sev E_Vap. (Safety Filling adapter), Part Number: 115-066756-00. 2. Exchange Sev E_Vap. (Safety Filling adapter), Part Number: E115-066756-00. Product Description: The Anesthesia System is a device used to administer to a patient, continuously or intermittently, a general inhalation anesthetic and to maintain a patient's ventilation. The A9 Anesthesia System utilizes V90 electronic vaporizers for anesthetic agent delivery. Class II

Reason: Potential for anesthesia leakage.

Mindray DS USA, Inc. dba Mindray North America 20251020 Ongoing
Distributed: Worldwide - US Nationwide distribution in the states of California, Florida, Georgia, Kentucky, Maryland, Mississippi, Missouri, New Jersey, Puerto Rico, Texas, Washington, West Virginia and the country of Canada.
Philips Azurion 7M20 systems with FlexArm ceiling-mounted system. Model Number: 722234; Class II

Reason: The motorized longitudinal movement of the FlexArm stand may be inconsistent (not smooth) and eventually may become unavailable. This issue is due to grease leakage from the bearings and/or anti-corrosion oil applied on the bearings, because of which the lubrication on the ceiling rail may be excessive, reducing the friction between the rails and the friction wheel. Expansion to add additional model.

PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. 20251002 Ongoing
Distributed: Domestic: AL, AR, AZ, CA, CO, FL, GA, HI, IL, IN, KS, KY, MD, MN, MO, NM, NY, OH, OK, PA, RI, TX, UT, VA, WA, WI, WV, WY; International: Australia, Austria, Belgium, Bulgaria, Canada, Czech Republic, Denmark, Finland, France, Germany, Hong Kong, Iraq, Italy, Japan, Netherlands, Norway, Philippines, Poland, Romania, South Africa, South Korea, Spain, Sweden, Switzerland, Thailand, United Kingdom, Vietnam;
IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. IV ADMIN SET, 110 IN.; Catalog Number: 490566. Class II

Reason: Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion).

B Braun Medical Inc 20251029 Ongoing
Distributed: Worldwide distribution - US distribution and international distribution to Canada, Germany, Guatemala, and Singapore.
IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. IV ADMIN SET W/ 15DROPS/ML; Catalog Number: 490508. Class II

Reason: Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion).

B Braun Medical Inc 20251029 Ongoing
Distributed: Worldwide distribution - US distribution and international distribution to Canada, Germany, Guatemala, and Singapore.
Outlook set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. OUTLOOK IV SET 15DR W/2 CARESITE FILTER; Catalog Number: 354214. Class II

Reason: Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion).

B Braun Medical Inc 20251029 Ongoing
Distributed: Worldwide distribution - US distribution and international distribution to Canada, Germany, Guatemala, and Singapore.
Vista set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. VISTA BASIC PUMP IV SET W/ 15 DROP/ML; Catalog Number: US9923F. Class II

Reason: Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion).

B Braun Medical Inc 20251029 Ongoing
Distributed: Worldwide distribution - US distribution and international distribution to Canada, Germany, Guatemala, and Singapore.
ULTRASITE set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. IV ADMIN SET W/ 2 ULTRASITE VLV, 104 IN; Catalog Number: 352049-C. 2. IV ADMIN SET W/ 3 ULTRASITE VLV, 112 IN; Catalog Number: 375100-C. Class II

Reason: Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion).

B Braun Medical Inc 20251029 Ongoing
Distributed: Worldwide distribution - US distribution and international distribution to Canada, Germany, Guatemala, and Singapore.
Page 270 of 4,030 | Showing 10 recalls