ADDitIV utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog numbers: 1. Primary Gravity IV Set, 2 Non-Needle-Free Injection Sites; Catalog Number: V1443. 2. ADDITIVE SET W/ 3 INJ SITES -DR*; Catalog Number: V1447. 3. ADD PEDIATRIC 3 INJ LUER LOCK; Catalog Number: V1448. 4. ADULT ADD 2 INJ CKV LUER LOCK; Catalog Number: V1484. Class II

Reason: Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion).

B Braun Medical Inc 20251029 Ongoing
Distributed: Worldwide distribution - US distribution and international distribution to Canada, Germany, Guatemala, and Singapore.
BioFire Respiratory Panel 2.1, IVD, REF: 423742 (30 test kit) Class II

Reason: Due to unintended movement of pouch chemistry, nucleic acid test may result in elevated risk of false negative results and risk of control failures.

BioFire Diagnostics, LLC 20251022 Ongoing
Distributed: US:AZ, FL,GA,ID NC, NH, NJ, NY, OR, PA, SC OUS: None
NICO BrainPath; Product Number/Product Name NN-8024 BrainPath Shephard's Hook - Greenberg; NN-8025 BrainPath Shephard's Hook - Sugita; NN-8026 BrainPath Shephard's Hook - Budde; NN-8042 BrainPath Shephard's Hook - Greenberg LK; Class II

Reason: Tape used to secure components within packaging contains latex, despite products being labeled as latex-free. Latex could potentially cause an allergic reaction.

Stryker Corporation 20251031 Ongoing
Distributed: Worldwide distribution - US Nationwide and the countries of Australia, Belgium, Canada, England, Ireland, Italy, Kuwait, Singapore, Slovenia, Spain, and the Netherlands.
NICO Myriad Illumination Pack Product No. NN-1000 11g Illumination Sleeve/Fiber; Product No. NN-1001 13g Illumination Sleeve/Fiber Class II

Reason: Tape used to secure components within packaging contains latex, despite products being labeled as latex-free. Latex could potentially cause an allergic reaction.

Stryker Corporation 20251031 Ongoing
Distributed: Worldwide distribution - US Nationwide and the countries of Australia, Belgium, Canada, England, Ireland, Italy, Kuwait, Singapore, Slovenia, Spain, and the Netherlands.
NICO Myriad Handpiece Product Number/Product Name: NN-8000 15x10 MHP; NN-8001 15x 13 MHP; NN-8002 15x 25 MHP - MINOP; NN-8003 13x10 MHP; NN-8004 13x13 MHP; NN-8005 15x25 MHP GAAB; NN-8006 11X10 MHP; NN-8007 11x13 MHP; NN-8008 19x28 MHP OI; NN-8010 17x31.5 MHP DECQ; NN-8015 NICO Fluid System (NFS); NN-8017 11x13 MHP/PB; NN-8018 13X13 MHP/PB; NN-8019 19x28 MHP Little LOTTA; NN-8020 19x21.5 MHP PaediScope; NN-8021 15x26.5 MHP LOTTA; NN-8500 15x10 MHP; NN-8501 15x13 MHP; NN-8502 15x25 MINOP MHP; NN-8503 13x10 MHP; NN-8504 13x13 MHP; NN-8505 15x25 GAAB MHP; NN-8506 11X10 MHP; NN-8507 11x13 MHP; NN-8508 19x28 OI MHP; NN-8510 17x31.5 DECQ MHP; NN-8517 13X13 MHP/PB; NN-8518 11x13 MHP/PB; NN-8519 19x28 Little LOTTA MHP; NN-8520 19x21.5 PaediScope MHP; NN-8521 15X26.5 LOTTA MHP; Class II

Reason: Tape used to secure components within packaging contains latex, despite products being labeled as latex-free. Latex could potentially cause an allergic reaction.

Stryker Corporation 20251031 Ongoing
Distributed: Worldwide distribution - US Nationwide and the countries of Australia, Belgium, Canada, England, Ireland, Italy, Kuwait, Singapore, Slovenia, Spain, and the Netherlands.
Brand Name: Galaxy System Product Name: Galaxy System Model/Catalog Number: GAL-001 Product Description: The Galaxy SystemTM and its accessories are intended to provide bronchoscopic visualization of and access to patient airways for diagnostic and therapeutic procedure. Class II

Reason: Due to Users not having access to the Instructions for Use/User Manual due to it being password protected.

Noah Medical Corporation 20250912 Ongoing
Distributed: US distribution to states of: NY, MA, WV, WY, TN, OH, NJ, MN, CA, MO, ID, IN, PA, MD, VA, WI, CO, SC, IL, NC and OUS (Foreign) to countries of: Hong Kong and Australia
FlexLab (FLX); Version: FLX-217-40; Class II

Reason: The Aliquoter Module (mALX) may generate secondary sample tubes without labels or with mismatched labels which may result in delayed results or erroneous results, this error occurs if the operators open the printer head of mALX to perform error recoveries or maintenance procedures and do not empty the printer tubes buffer to remove all the secondary sample tubes

Inpeco S.A. 20251030 Ongoing
Distributed: US Domestic: MI. OUS International: Australia, Finland, France, Germany, Italy, Portugal, Spain, and UK.
FlexLab (FLX); Version: FLX-217-10; Class II

Reason: The Aliquoter Module (mALX) may generate secondary sample tubes without labels or with mismatched labels which may result in delayed results or erroneous results. This error occurs if the operators open the printer head of mALX to perform error recoveries or maintenance procedures and do not empty the printer tubes buffer to remove all the secondary sample tubes.

Inpeco S.A. 20251030 Ongoing
Distributed: US Domestic: MI. OUS International: Australia, Finland, France, Germany, Italy, Portugal, Spain, and UK.
CareLink Clinic, REF: MMT-7350 Class II

Reason: Software error causing incorrect data to be displayed on the 24-hour Sensor Glucose Overview Graph, any potential therapy decisions were made based on the incorrect data displayed on the 24-hour sensor glucose overview graph may lead to hypoglycemia or hyperglycemia.

Medtronic MiniMed, Inc. 20251021 Ongoing
Distributed: Worldwide Distribution: US (nationwide) including states of: TX, NC, MN, CA, VA, ID, WA, GA, MI, MA, OH, IL, IA, CT, FL, PA, NY, TN, AZ, LA, AL, NJ, CO, SC, MS, OR, MT, MD, KS, AK, NH, NE, WY, ND, IN, MO, WI, KY, UT, NV, SD, WV, NM, HI, AR, OK, VT, RI; and OUS (foreign) to countries of: Australia (AU), Sweden (SE), Belgium (BE), Great Britain (GB), Netherlands (NL), Finland (FI), Canada (CA), Germany (DE), Spain (ES), Switzerland (CH), Italy (IT), South Africa (ZA), South Korea (KR), Singapore (SG), New Zealand (NZ), Poland (PL), Israel (IL), Denmark (DK), Norway (NO), Luxembourg (LU), Lithuania (LT), Czech Republic ( CZ), Mexico (MX), Brazil (BR), Colombia (CO), Slovakia (SK), Ireland (IE), Argentina (AR), Greece (GR), Kuwait (KW), Portugal (PT), Iceland (IS), Japan (JP), Turkey (TR), Austria (AT), Qatar (QA), Hong Kong (HK), Chile (CL), Ukraine (UA), Hungary (HU), Estonia (EE), RS, Latvia (LV), Taiwan (TW), Saudia Arabia (SA), Bahrain (BH), Oman (OM), Romania (RO), India (IN), Uruguay (UY), Malaysia (MY), United Arab Emirates (AE), Algeria (DZ), Croatia (HR), France (FR), Russia (RU), Cyprus (CY), Thailand (TH), Egypt (EG), Bosnia and Herzegovina (BA), Libya (LY), and Iraq (IQ)
Option"ELITE Vena Cava Filter System REF 352506070E UDI-DI code: 00886333217151 The Option"ELITE Filter is indicated for the prevention of recurrent pulmonary embolism (PE) via percutaneous placement in the inferior vena cava (IVC) in the following conditions: " Pulmonary thromboembolism when anticoagulants are contraindicated " Failure of anticoagulant therapy in thromboembolic disease " Emergency treatment following massive pulmonary embolism where anticipated benefits of conventional therapy are reduced " Chronic, recurrent pulmonary embolism where anticoagulant therapy has failed or is contraindicated Class II

Reason: Due to complaints of increased resistance when advancing the dilator within the introducer sheath. This issue could potential result in procedural delays and/or risk of venous intimal injuries if excessive force is applied

Argon Medical Devices, Inc 20250912 Ongoing
Distributed: U.S. (nationwide) distribution to states of: AL, AR, AZ, CA, CO, CT, DC, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, TN, TX, UT, VA, WA, WI, and WV. O.U.S. (international) to countries of: China, Panama and Trinidad and Tobago