MEDLINE convenience kits labeled as: 1) BREAST BIOPSY PACK-LF, REF DYNJ0160598C; 2) BIOPSY BREAST PACK, REF DYNJ65361A; 3) BREAST BIOPSY/MASTECTOMY, REF DYNJ908839A. Class II

Reason: Medline Industries, LP is issuing a recall for specific item(s) and lot(s) of Medline Kits containing SafeAir Smoke Evacuation Pencils. Stryker Instruments has expanded the scope of their recall of SafeAir Smoke Evacuation Pencils previously initiated in October 2024, adding an additional 47 lots to the scope of the recall. The initial recall was due to the potential for product to activate without manual input when plugged into a power source or remain active after buttons are released.

Medline Industries, LP 20250930 Ongoing
Distributed: US Nationwide distribution.
MEDLINE convenience kits labeled as: 1) CENTRAL LINE PACK, REF DYNJ00281P; 2) FREE FLAP PACK, REF DYNJ68213C. Class II

Reason: Medline Industries, LP is issuing a recall for specific item(s) and lot(s) of Medline Kits containing SafeAir Smoke Evacuation Pencils. Stryker Instruments has expanded the scope of their recall of SafeAir Smoke Evacuation Pencils previously initiated in October 2024, adding an additional 47 lots to the scope of the recall. The initial recall was due to the potential for product to activate without manual input when plugged into a power source or remain active after buttons are released.

Medline Industries, LP 20250930 Ongoing
Distributed: US Nationwide distribution.
MEDLINE convenience kits labeled as: 1) GENERAL AAA #11-RF, REF CDS840261AB; 2) CAROTID ENDARTERECTOMY CDS-LF, REF CDS983637I; 3) PERIPHERAL VASCULAR CDS, REF CDS983723J; 4) ACH OPEN HEART, REF CDS983773J; 5) VASCULAR, REF CDS984255K; 6) OPEN HEART CHILDRENS PACK-LF, REF DYNJ0161262C; 7) VASCULAR PACK-LF, REF DYNJ0421380P; 8) GEN FEM POP #14-RF, REF DYNJ21877W; 9) VASCULAR PACK, REF DYNJ22350L; 10) GENERAL CAROTID/AV TRAY #15-RF, REF DYNJ27432AC; 11) OPEN HEART PACK, REF DYNJ37284K; 12) ACH CABG ADD-ON, REF DYNJ38175L; 13) CARDIOVASCULAR PACK, REF DYNJ42921G; 14) CABG PACK-LF, REF DYNJ43207K; 15) HEART DRAPING PACK-LF, REF DYNJ43214F; 16) AAA PACK, REF DYNJ44847I; 17) CASTRO CABG-COMPONENT PACK, REF DYNJ44857U; 18) FISTULA PACK, REF DYNJ44869K; 19) MAJOR VASCULAR LEG-TEAM, REF DYNJ44879K; 20) PERMANENT PACING, REF DYNJ44886M; 21) CAROTID VASCULAR ENDARTERECTMY, REF DYNJ50528M; 22) DR DANIELSON AC PACK, REF DYNJ51038R; 23) VASCULAR CAROTID PACK, REF DYNJ56838F; 24) VASCULAR FEM PACK, REF DYNJ56839D; 25) PK CUST OPEN HEART A AND B, REF DYNJ60548B; 26) PK CUST CAROTID ST MICHAE, REF DYNJ60783B; 27) CAROTID PACK, REF DYNJ61659A; 28) VASCULAR MINOR PACK, REF DYNJ62097C; 29) VASCULAR PACK-LF, REF DYNJ65039C; 30) PACEMAKER PACK, REF DYNJ66033B; 31) DR MELLINGER AC PACK, REF DYNJ80609D; 32) HEART PACK II, REF DYNJ81581B; 33) AV FISTULA PACK, REF DYNJ81605; 34) MAJOR VASCULAR PACK, REF DYNJ81610B; 35) CABG ACCESSORY PACK, REF DYNJ82166; 36) VALVE ACCESSORY PACK, REF DYNJ82217; 37) PERIVASCULAR CUSTOM PACK, REF DYNJ84017B; 38) CAROTID ENDARTERECTOMY PACK-LF, REF DYNJ85109; 39) ACH OPEN HEART A&B PK, REF DYNJ85110; 40) VASCULAR ABDOMINAL PACK, REF DYNJ85118; 41) AV FISTULA PACK-LF, REF DYNJ85121; 42) CARDIAC - DEEIK PACK, REF DYNJ85637; 43) VASCULAR, REF DYNJ900293K; 44) VASCULAR-LF, REF DYNJ900714C; 45) HEART/MAJOR VASCULAR, REF DYNJ901066K; 46) MODULE OPEN HEART, REF DYNJ903042P; 47) VASCULAR/OR ANGIO, REF DYNJ905079F; 48) OPEN HEART, REF DYNJ905497B; 49) AAA, REF DYNJ906175B; 50) VASCULAR ABDOMINAL, REF DYNJ907105D; 51) CV COMPLETE, REF DYNJ907904A; 52) KIT CV NURSE CENTRAL DUPAGE, REF DYNJ908820; 53) KIT CV NURSE CENTRAL DUPAGE, REF DYNJ908820A. Class II

Reason: Medline Industries, LP is issuing a recall for specific item(s) and lot(s) of Medline Kits containing SafeAir Smoke Evacuation Pencils. Stryker Instruments has expanded the scope of their recall of SafeAir Smoke Evacuation Pencils previously initiated in October 2024, adding an additional 47 lots to the scope of the recall. The initial recall was due to the potential for product to activate without manual input when plugged into a power source or remain active after buttons are released.

Medline Industries, LP 20250930 Ongoing
Distributed: US Nationwide distribution.
MEDLINE convenience kits labeled as: CATH LAB PACEMAKER PACK, REF DYNJ59818A Class II

Reason: Medline Industries, LP is issuing a recall for specific item(s) and lot(s) of Medline Kits containing SafeAir Smoke Evacuation Pencils. Stryker Instruments has expanded the scope of their recall of SafeAir Smoke Evacuation Pencils previously initiated in October 2024, adding an additional 47 lots to the scope of the recall. The initial recall was due to the potential for product to activate without manual input when plugged into a power source or remain active after buttons are released.

Medline Industries, LP 20250930 Ongoing
Distributed: US Nationwide distribution.
MEDLINE convenience kits labeled as LOCAL BASIC, REF DYNJ34418M Class II

Reason: Medline Industries, LP is issuing a recall for specific item(s) and lot(s) of Medline Kits containing SafeAir Smoke Evacuation Pencils. Stryker Instruments has expanded the scope of their recall of SafeAir Smoke Evacuation Pencils previously initiated in October 2024, adding an additional 47 lots to the scope of the recall. The initial recall was due to the potential for product to activate without manual input when plugged into a power source or remain active after buttons are released.

Medline Industries, LP 20250930 Ongoing
Distributed: US Nationwide distribution.
MEDLINE STERILE BASIC EYE PACK 16: 1) REF: DYNJ37743D; 2) REF: DYNJ37743F. Class II

Reason: Medline has identified specific sterile eye procedure kits with pouch packaging that may exhibit open seals.

Medline Industries, LP 20251003 Ongoing
Distributed: US: MO
MC3 VitalFlow Console, REF 58100; Blood pump of ecmo Class II

Reason: As of August 6, 2025, Medtronic has received eleven reports of VitalFlow Consoles displaying an E70 error code during normal operation. When this occurs, the touch screen may become temporarily unresponsive and go blank for up to two minutes before recovering to full functionality. Importantly, the console continues to maintain set pump speed and function throughout.

Medtronic Perfusion Systems 20251003 Ongoing
Distributed: US Nationwide distribution.
Brand Name: HeartMate Product Name: HeartMate II¿ Controller (standalone) Model/Catalog Number: 106017, HeartMate II¿ Pocket Controller, US; 106762, HeartMate II¿ Pocket Controller, US; 107801, HeartMate II¿ Pocket Controller Software Version: N/A Product Description: The HeartMate II System Controller is a small computer that controls and monitors system operation. It sends power and operating signals to the Left Ventricular Assist Device and adjusts device operation to maintain programmed levels of cardiac support. The System Controller identifies alarm conditions and initiates Hazard and Advisory alarms. It records device performance and alarm data, and transfers the information to the HeartMate Touch Communication System. Component: Yes, HeartMate II¿ LVAS Implant Kit Class II

Reason: Due to potential corrosion issues associated with the Backup Battery cable connector that may cause a Backup Battery Fault Alarm.

Thoratec LLC 20251009 Ongoing
Distributed: Worldwide - US Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, NC, NE, NJ, NM, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA, WA, WI, WV and the countries of Algeria, Australia, Austria, Azerbaijan, Bahrain, Belgium, Brazil, Bulgaria, Canada, Cayman Islands, Chile, China, Colombia, Costa Rica, Cyprus, Czech Republic, Denmark, Egypt, Estonia, Finland, France, Georgia, Germany, Greece, India, Iran, Iraq, Ireland, Israel, Italy, Japan, Kuwait, Lebanon, Lithuania, Macedonia, Malaysia, Martinique, Montenegro, Morocco, Netherlands, New Zealand, Norway, Oman, Palestine, Peru, Poland, Portugal, Qatar, Reunion, Romania, Russia, Saudi Arabia, Serbia, Singapore, Slovenia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, Tunisia, Turkey, UK, United Arab Emirates, Uzbekistan.
Brand Name: HeartMate Product Name: HeartMate II¿ LVAS Implant Kit Model/Catalog Number: 106015, HeartMate II¿ LVAS Kit, with Pocket Controller, Sealed, US; 106016, HeartMate II¿ LVAS Kit, with Pocket Controller, Sealed, EU Software Version: N/A Product Description: The HeartMate II LVAS is an axial-flow, rotary ventricular assist system that generates flows up to 10 liters per minute (lpm). One end of the Left Ventricular Assist Device is attached to the apex of the left ventricle; the other end of the device connects to the ascending aorta. The Left Ventricular Assist Device diverts blood from the weakened left ventricle and propels it to the aorta. A small external computer, the System Controller, monitors system operation. A driveline, which passes through the patient s abdomen, connects the implanted pump to the System Controller. The system is powered by a Power Module or Mobile Power Unit (MPU) that is connected to an AC electrical outlet, or by two HeartMate 14 Volt Lithium-Ion batteries. Component: No Class II

Reason: Due to potential corrosion issues associated with the Backup Battery cable connector that may cause a Backup Battery Fault Alarm.

Thoratec LLC 20251009 Ongoing
Distributed: Worldwide - US Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, NC, NE, NJ, NM, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA, WA, WI, WV and the countries of Algeria, Australia, Austria, Azerbaijan, Bahrain, Belgium, Brazil, Bulgaria, Canada, Cayman Islands, Chile, China, Colombia, Costa Rica, Cyprus, Czech Republic, Denmark, Egypt, Estonia, Finland, France, Georgia, Germany, Greece, India, Iran, Iraq, Ireland, Israel, Italy, Japan, Kuwait, Lebanon, Lithuania, Macedonia, Malaysia, Martinique, Montenegro, Morocco, Netherlands, New Zealand, Norway, Oman, Palestine, Peru, Poland, Portugal, Qatar, Reunion, Romania, Russia, Saudi Arabia, Serbia, Singapore, Slovenia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, Tunisia, Turkey, UK, United Arab Emirates, Uzbekistan.
Brand Name: HeartMate Product Name: HeartMate 3 Controller (Standalone) Model/Catalog Number: 106531US, HeartMate 3 Controller, US; 106531INT, HeartMate 3 Controller, OUS; 106531LF2, HeartMate 3 Low Flow 2.0 Controller Software Version: N/A Product Description: The HeartMate 3" System Controller acts as the central power and communication hub for the HeartMate 3 LVAS. It passes power from the Power Module, the Mobile Power Unit, Lithium-Ion Batteries, or its own integrated emergency backup supply, down to the LVAD via the Driveline. The HeartMate 3 System Controller constantly monitors system performance through communication with the implanted LVAD and Controller internal measurements and alerts the user to any alarm conditions by activating membrane panel LEDs and integrated audio annunciators. Further information on alarm conditions as well as system status can be attained by the user from the front panel LCD on the System Controller. When connected to a HeartMate Touch" Communication System, the System Controller sends information regarding the System Controller and Pump Status once per second to provide additional information to the user. This link also allows the clinician to set new patient operating parameters (e.g. pump speed) and provides a link for downloading trend and/or event recorder data. Component: Yes, HeartMate 3 LVAS Class II

Reason: Due to potential corrosion issues associated with the Backup Battery cable connector that may cause a Backup Battery Fault Alarm.

Thoratec LLC 20251009 Ongoing
Distributed: Worldwide - US Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, NC, NE, NJ, NM, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA, WA, WI, WV and the countries of Algeria, Australia, Austria, Azerbaijan, Bahrain, Belgium, Brazil, Bulgaria, Canada, Cayman Islands, Chile, China, Colombia, Costa Rica, Cyprus, Czech Republic, Denmark, Egypt, Estonia, Finland, France, Georgia, Germany, Greece, India, Iran, Iraq, Ireland, Israel, Italy, Japan, Kuwait, Lebanon, Lithuania, Macedonia, Malaysia, Martinique, Montenegro, Morocco, Netherlands, New Zealand, Norway, Oman, Palestine, Peru, Poland, Portugal, Qatar, Reunion, Romania, Russia, Saudi Arabia, Serbia, Singapore, Slovenia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, Tunisia, Turkey, UK, United Arab Emirates, Uzbekistan.
Page 290 of 4,030 | Showing 10 recalls
Data provided by openFDA