ChemoLock Bag Spike, 01 units REF: CL-10-10 Class II

Reason: Port weld of drug transfer device may separate or break during use and potentially result in a leak.

ICU Medical, Inc. 20250908 Ongoing
Distributed: Worldwide distribution: US (Nationwide) including states of: AK AL AR AZ CA CO CT DE FL GA IA ID IL KS KY LA MA MD MI MN MS MT NC NE NJ NM NV NY OH OK OR PA SD TN TX VA WA WY and OUS (foreign) to countries of: Canada, France, and Japan.
VANTAGE GALAN 3T. CMSC Model: MRT-3020/U7. Model Number: MEXL-3020/B7. Class II

Reason: Affected system contains components that do not meet the required distance between electrodes for compliance with the IEC 60601-1 standard for medical electrical equipment.

Canon Medical System, USA, INC. 20251016 Ongoing
Distributed: US distribution to Ohio and Nevada
Alcon PRECISION1 One-Day Contact Lenses, -2.25D. Class II

Reason: Potential for a weak seal in some units resulting in compromise in sterility.

Alcon Research LLC 20250930 Ongoing
Distributed: Worldwide - US distribution in the states of Alabama, Arizona, Arkansas, Connecticut, Florida, Idaho, Iowa, Louisiana, Maryland, Massachusetts, Michigan, Minnesota, New Jersey, New York, Oklahoma, Oregon, Tennessee, Texas, Washington and the country of Canada.
Surgical Planner, BrainSTORM, software version BST.9.3.0, orthopedic augmented reality Class II

Reason: A software anomaly, under specific conditions when large rotational alignment values are applied in 2D Image Fusion, may lead to misalignment of any secondary image layer(s) to the primary image layer.

Surgical Theater Inc 20250916 Terminated
Distributed: Worldwide - US Nationwide distribution in the states of MI, FL, DC, MN, CA, TX, NJ, PA, AZ, NY, KY, IN, WV and the countries of Italy, Germany, Israel, UK.
SpineAR Surgical Navigation Advanced Platform Patient Engagement (SyncAR SNAP), software version SPR.1.2.0, orthopedic augmented reality Class II

Reason: A software anomaly, under specific conditions when large rotational alignment values are applied in 2D Image Fusion, may lead to misalignment of any secondary image layer(s) to the primary image layer.

Surgical Theater Inc 20250916 Terminated
Distributed: Worldwide - US Nationwide distribution in the states of MI, FL, DC, MN, CA, TX, NJ, PA, AZ, NY, KY, IN, WV and the countries of Italy, Germany, Israel, UK.
Endo Surgical Navigation Advanced Platform Patient Engagement (EndoSNAP), software version SNP.1.12.0, imaging processing radiological system Class II

Reason: A software anomaly, under specific conditions when large rotational alignment values are applied in 2D Image Fusion, may lead to misalignment of any secondary image layer(s) to the primary image layer.

Surgical Theater Inc 20250916 Terminated
Distributed: Worldwide - US Nationwide distribution in the states of MI, FL, DC, MN, CA, TX, NJ, PA, AZ, NY, KY, IN, WV and the countries of Italy, Germany, Israel, UK.
SyncAR Surgical Navigation Advanced Platform Patient Engagement (SyncAR SNAP), software version SNP.1.12.0, imaging processing radiological system Class II

Reason: A software anomaly, under specific conditions when large rotational alignment values are applied in 2D Image Fusion, may lead to misalignment of any secondary image layer(s) to the primary image layer.

Surgical Theater Inc 20250916 Terminated
Distributed: Worldwide - US Nationwide distribution in the states of MI, FL, DC, MN, CA, TX, NJ, PA, AZ, NY, KY, IN, WV and the countries of Italy, Germany, Israel, UK.
Surgical Navigation Advanced Platform Patient Engagement (PE SNAP), software version SNP.3.12.0, imaging processing radiological system Class II

Reason: A software anomaly, under specific conditions when large rotational alignment values are applied in 2D Image Fusion, may lead to misalignment of any secondary image layer(s) to the primary image layer.

Surgical Theater Inc 20250916 Terminated
Distributed: Worldwide - US Nationwide distribution in the states of MI, FL, DC, MN, CA, TX, NJ, PA, AZ, NY, KY, IN, WV and the countries of Italy, Germany, Israel, UK.
Surgical Navigation Advanced Platform (SNAP), software version SNP.3.12.0, imaging processing radiological system Class II

Reason: A software anomaly, under specific conditions when large rotational alignment values are applied in 2D Image Fusion, may lead to misalignment of any secondary image layer(s) to the primary image layer.

Surgical Theater Inc 20250916 Terminated
Distributed: Worldwide - US Nationwide distribution in the states of MI, FL, DC, MN, CA, TX, NJ, PA, AZ, NY, KY, IN, WV and the countries of Italy, Germany, Israel, UK.
SuRgical Planner (SRP), software version SRP.7.12.0, imaging processing radiological system Class II

Reason: A software anomaly, under specific conditions when large rotational alignment values are applied in 2D Image Fusion, may lead to misalignment of any secondary image layer(s) to the primary image layer.

Surgical Theater Inc 20250916 Terminated
Distributed: Worldwide - US Nationwide distribution in the states of MI, FL, DC, MN, CA, TX, NJ, PA, AZ, NY, KY, IN, WV and the countries of Italy, Germany, Israel, UK.
Page 309 of 4,030 | Showing 10 recalls