Brand Name: Dexcom G6 Continuous Glucose Monitoring System Product Name: Dexcom G6 and G6 Pro Android US CGM App Model/Catalog Number: SW11678 Software Version: version 1.15.0 Product Description: Dexcom G6 and G6 Pro Android US CGM App SW11678 version 1.15.0 Component: Dexcom G6 Continuous Glucose Monitoring System Class I

Reason: A software defect in version v1.15.0 of the G6 Android app can cause the app to terminate unexpectedly, which may result in the user not receiving estimated glucose values, alarms, alerts or notifications. This could result in the missed detection of a hyperglycemic or hypoglycemic event, protentional resulting in severe hyperglycemia, diabetic ketoacidosis (DKA), or hyperosmolar hyperglycemic state (HHS).

Dexcom, Inc. 20250828 Ongoing
Distributed: This product was distributed throughout the US. G6 Android US v1.15.0 SW11678 G6 and G6 Pro Android app release has been halted <1% rollout after this issue was identified. Please see the attached consignee data for additional details.
Magseed Pro 12 cm soft tissue marker. Model Number: MS2-17-1-12. Product Description: The Endomag Magseed Pro Magnetic Marker System is a sterile, single use device composed of a marker preloaded in a 17ga Needle Delivery System. The Endomag Magseed Pro Marker is indicated for use to radiographically mark soft tissue during a surgical procedure or for future surgical procedures. Class II

Reason: Potential for contamination with cotton fibers.

ENDOMAGNETICS LTD 20251001 Ongoing
Distributed: Worldwide - US Nationwide distribution in the states of CA, CO, CT, DC, FL, GA, IA, ID, IN, KY, MD, ME, MI, MN, MO, NC, ND, NE, NH, NJ, NY, OH, TX, WA and the country of Canada.
Magseed Pro 7 cm soft tissue marker. Model Number: MS2-17-1-07. Product Description: The Endomag Magseed Pro Magnetic Marker System is a sterile, single use device composed of a marker preloaded in a 17ga Needle Delivery System. The Endomag Magseed Pro Marker is indicated for use to radiographically mark soft tissue during a surgical procedure or for future surgical procedures. Class II

Reason: Potential for contamination with cotton fibers.

ENDOMAGNETICS LTD 20251001 Ongoing
Distributed: Worldwide - US Nationwide distribution in the states of CA, CO, CT, DC, FL, GA, IA, ID, IN, KY, MD, ME, MI, MN, MO, NC, ND, NE, NH, NJ, NY, OH, TX, WA and the country of Canada.
Brand Name: AESCULAP Product Name: REDUCING CONVERTER 10/12MM TO 5MM Model/Catalog Number: EK087P Software Version: N/A Product Description: REDUCING CONVERTER 10/12MM TO 5MM Component: N/A Class II

Reason: It was determined there is the potential of silicone fragments detaching from the yellow sealing component.

Aesculap Inc 20250924 Ongoing
Distributed: Worldwide - US Nationwide distribution in the states of AR, AZ, CA, CT, MD, NE, NM, NY, OR, PA, WA, WI and the country of Canada.
Brand Name: AESCULAP Product Name: SEALING UNIT FOR 10/12MM TROCARS Model/Catalog Number: EK086P Software Version: N/A Product Description: SEALING UNIT FOR 10/12MM TROCARS Component: N/A Class II

Reason: It was determined there is the potential of silicone fragments detaching from the yellow sealing component.

Aesculap Inc 20250924 Ongoing
Distributed: Worldwide - US Nationwide distribution in the states of AR, AZ, CA, CT, MD, NE, NM, NY, OR, PA, WA, WI and the country of Canada.
Brand Name: Aesculap Product Name: SEALING CAP FOR 10/12MM TROCARS Model/Catalog Number: EK085P Software Version: N/A Product Description: SEALING CAP FOR 10/12MM TROCARS Component: N/A Class II

Reason: It was determined there is the potential of silicone fragments detaching from the yellow sealing component.

Aesculap Inc 20250924 Ongoing
Distributed: Worldwide - US Nationwide distribution in the states of AR, AZ, CA, CT, MD, NE, NM, NY, OR, PA, WA, WI and the country of Canada.
Brand Name: Aesculap Product Name: SEALING UNIT F/10/12MM TROCARS W.REDUCER Model/Catalog Number: EK083P Software Version: N/A Product Description: SEALING UNIT F/10/12MM TROCARS W.REDUCER Component: N/A Class II

Reason: It was determined there is the potential of silicone fragments detaching from the yellow sealing component.

Aesculap Inc 20250924 Ongoing
Distributed: Worldwide - US Nationwide distribution in the states of AR, AZ, CA, CT, MD, NE, NM, NY, OR, PA, WA, WI and the country of Canada.
Brand Name: Alcon Laboratories, Inc. Product Name: Accessory devices for Unity and Constellation surgical consoles Model/Catalog Number: Many - See 'Attachment 1 - Affected Product List' 8065000356 27GA HP ENTRY SYSTEM, 4 MM 1 COUNT 8065000358 25GA HP ENTRY SYSTEM, 6 MM 1 COUNT 8065000359 27GA HP ENTRY SYSTEM, 6 MM 1 COUNT 8065000467 25GA INFUSION CANNULA(4&6MM ES)+TRAY+TMD 8065000468 27GA INFUSION CANNULA(4&6MM ES)+TRAY+TMD Software Version: N/A Product Description: These products are standalone accessory devices meant for vitrectomy surgical procedures. They interface with a corresponding Alcon surgical console. Component: N/A Class II

Reason: Due to incomplete seals in the pouch which provide the sterile barrier.

Alcon Research LLC 20250805 Ongoing
Distributed: Worldwide distribution: US (nationwide): AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NY, OH, OK, OR, PA, RI, SC,SD,TN, TX, VA, WA, WI, and WV; and OUS (foreign) to countries of: Argentina, Australia, Austria, Azerbaijan, Bahrain, Bangladesh, Belgium, Brazil, Brunei, Canada, Chile, China, Colombia, Croatia, Czechia, Denmark, Finland, France, Germany, Greece, Guadeloupe, Hong Kong, Hungary, Iceland, India, Ireland, Israel, Italy, Japan, Korea, Kosovo, Luxembourg, Malaysia, Mexico, Moldova, Montenegro, Morocco, Netherlands, New Zealand, Pakistan, Peru, Philippines, Poland, Portugal, Qatar, Reunion, Russian Federation, Saudi Arabia, Serbia, Slovakia, South Africa, Spain, Sri Lanka, Sweden, Switzerland, Taiwan, Turkey, United Arab Emirates, and United Kingdom,
Brand Name: Alcon Laboratories, Inc. Product Name: Accessory devices for Unity and Constellation surgical consoles Model/Catalog Number: Many - See 'Attachment 1 - Affected Product List' 8065000252 UNITY 27 GA CHANDELIER 8065000253 UNITY 27+DS WIDE ANGLE LIGHT PROBE Software Version: N/A Product Description: These products are standalone accessory devices meant for vitrectomy surgical procedures. They interface with a corresponding Alcon surgical console. Component: N/A Class II

Reason: Due to incomplete seals in the pouch which provide the sterile barrier.

Alcon Research LLC 20250805 Ongoing
Distributed: Worldwide distribution: US (nationwide): AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NY, OH, OK, OR, PA, RI, SC,SD,TN, TX, VA, WA, WI, and WV; and OUS (foreign) to countries of: Argentina, Australia, Austria, Azerbaijan, Bahrain, Bangladesh, Belgium, Brazil, Brunei, Canada, Chile, China, Colombia, Croatia, Czechia, Denmark, Finland, France, Germany, Greece, Guadeloupe, Hong Kong, Hungary, Iceland, India, Ireland, Israel, Italy, Japan, Korea, Kosovo, Luxembourg, Malaysia, Mexico, Moldova, Montenegro, Morocco, Netherlands, New Zealand, Pakistan, Peru, Philippines, Poland, Portugal, Qatar, Reunion, Russian Federation, Saudi Arabia, Serbia, Slovakia, South Africa, Spain, Sri Lanka, Sweden, Switzerland, Taiwan, Turkey, United Arab Emirates, and United Kingdom,
Brand Name: Alcon Laboratories, Inc. Product Name: Accessory devices for Unity and Constellation surgical consoles Model/Catalog Number: Many - See 'Attachment 1 - Affected Product List' 8065000393 UNITY HP VISCOUS FLUID CONTROL VFC PACK Software Version: N/A Product Description: These products are standalone accessory devices meant for vitrectomy surgical procedures. They interface with a corresponding Alcon surgical console. Component: N/A Class II

Reason: Due to incomplete seals in the pouch which provide the sterile barrier.

Alcon Research LLC 20250805 Ongoing
Distributed: Worldwide distribution: US (nationwide): AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NY, OH, OK, OR, PA, RI, SC,SD,TN, TX, VA, WA, WI, and WV; and OUS (foreign) to countries of: Argentina, Australia, Austria, Azerbaijan, Bahrain, Bangladesh, Belgium, Brazil, Brunei, Canada, Chile, China, Colombia, Croatia, Czechia, Denmark, Finland, France, Germany, Greece, Guadeloupe, Hong Kong, Hungary, Iceland, India, Ireland, Israel, Italy, Japan, Korea, Kosovo, Luxembourg, Malaysia, Mexico, Moldova, Montenegro, Morocco, Netherlands, New Zealand, Pakistan, Peru, Philippines, Poland, Portugal, Qatar, Reunion, Russian Federation, Saudi Arabia, Serbia, Slovakia, South Africa, Spain, Sri Lanka, Sweden, Switzerland, Taiwan, Turkey, United Arab Emirates, and United Kingdom,
Page 310 of 4,030 | Showing 10 recalls
Data provided by openFDA