Shanghai United Imaging Healthcare Co., Ltd. Computed Tomography X-ray System, Model:uCT 780 MD, Rx Only Class II

Reason: Third-party outer covers can become trapped between patient table and gantry, causing mechanical interference during table movement.

UIH Technologies LLC 20250912 Ongoing
Distributed: US: OUS:
Shanghai United Imaging Healthcare Co., Ltd. Computed Tomography X-ray System, Model:uCT 760 MD, Rx Only Class II

Reason: Third-party outer covers can become trapped between patient table and gantry, causing mechanical interference during table movement.

UIH Technologies LLC 20250912 Ongoing
Distributed: US: OUS:
Shanghai United Imaging Healthcare Co., Ltd. Computed Tomography X-ray System, Model:uCT 550 MD, Rx Only Class II

Reason: Third-party outer covers can become trapped between patient table and gantry, causing mechanical interference during table movement.

UIH Technologies LLC 20250912 Ongoing
Distributed: US: OUS:
Shanghai United Imaging Healthcare Co., Ltd. Computed Tomography X-ray System, Model:uCT 530 MD, Rx Only Class II

Reason: Third-party outer covers can become trapped between patient table and gantry, causing mechanical interference during table movement.

UIH Technologies LLC 20250912 Ongoing
Distributed: US: OUS:
DxI 9000 Access Immunoassay Analyzer, Part Number C11137 Class II

Reason: Beckman Coulter has confirmed a software anomaly on DxI 9000 Access Immunoassay Analyzer and the DxC 500i Clinical Analyzer that may prevent correct application of user-specified manual dilution factors when ordering additional tests for samples already "In Progress" using the same "Sample ID number". This can lead to false low or high patient results and potential delays in reporting, risking erroneous diagnoses if undetected.

Beckman Coulter, Inc. 20250919 Ongoing
Distributed: Worldwide - US Nationwide distribution.
DxC 500i Clinical Analyzer Modules, DxC 500 AU Module w/ISE, DxC 500i (Part Number C63522), DxC 500 AU Module, DxC 500i (Part Number C63521), and Access 2 Module, DxC 500i (Part Number C13252) Class II

Reason: Beckman Coulter has confirmed a software anomaly on DxI 9000 Access Immunoassay Analyzer and the DxC 500i Clinical Analyzer that may prevent correct application of user-specified manual dilution factors when ordering additional tests for samples already "In Progress" using the same "Sample ID number". This can lead to false low or high patient results and potential delays in reporting, risking erroneous diagnoses if undetected.

Beckman Coulter, Inc. 20250919 Ongoing
Distributed: Worldwide - US Nationwide distribution.
GE Healthcare Carescape Central Station (CSCS) V3 MAS800 Desktop with software version V3.0.5, Part Number 5867474-03; Medical Display, cathode ray tube Class II

Reason: Carescape Central Station (CSCS), software version V3.0.5, can cause the system to enter a continuous reboot cycle. If this occurs, a loss of central monitoring of connected patients could result. This issue occurs only when the number of central stations connected to the network is 118, 119 or 120.

Wipro GE Healthcare Private Ltd. 20250905 Ongoing
Distributed: Worldwide distribution.
GE Healthcare Carescape Central Station (CSCS) V3 MAS800 Desktop with software version V3.0.5; Part Number 5867474-04; Medical Display, cathode ray tube Class II

Reason: Carescape Central Station (CSCS), software version V3.0.5, can cause the system to enter a continuous reboot cycle. If this occurs, a loss of central monitoring of connected patients could result. This issue occurs only when the number of central stations connected to the network is 118, 119 or 120.

Wipro GE Healthcare Private Ltd. 20250905 Ongoing
Distributed: Worldwide distribution.
BIOMET ZipTight, Acute AC Joint Implant, Single Ziploop , Model/Catalog Number: 904834; Fixation System for AC Joint Repair Class II

Reason: There is a potential for a missing slotted button assembly in one lot of ZipTight, Acute AC Joint Implant, Single Ziploop

Biomet, Inc. 20250923 Ongoing
Distributed: Refer to the attached ZFA 2025-00166_Initial Consignee List and ZFA 2025-00166_Initial Distribution History containing the distribution and the identification of the domestic consignees.
Salem Sump PVC Tubes: Product Code/Product Name: 1180264408 Salem Sump" PVC Gastroduodenal Tube, 10 Fr/Ch (3.3 mm) x 90 cm 1180264416 Salem Sump" PVC Gastroduodenal Tube, 12 Fr/Ch (4.0 mm) x 114 cm 1180264424 Salem Sump" PVC Gastroduodenal Tube, 14 Fr/Ch (4.7 mm) x 114 cm 1180264432 Salem Sump" PVC Gastroduodenal Tube, 16 Fr/Ch (5.3 mm) x 114 cm 1180264440 Salem Sump" PVC Gastroduodenal Tube, 18 Fr/Ch (6.0 mm) x 114 cm 1180264457 Salem Sump" PVC Gastroduodenal Tube, 20 Fr/Ch (6.7 mm) x 114 cm 1180264523 Salem Sump" Polyurethane Gastroduodenal Tube, 14 Fr/Ch (4.7 mm) x 114 cm 1180264531 Salem Sump" Polyurethane Gastroduodenal Tube, 16 Fr/Ch (5.3 mm) x 114 cm 1180264549 Salem Sump" Polyurethane Gastroduodenal Tube, 18 Fr/Ch (6.0 mm) x 114 cm 1180264556 Salem Sump" Polyurethane Gastroduodenal Tube, 20 Fr/Ch (6.7 mm) x 114 cm 1180264895 Salem Sump PVC Gastroduodenal Tube, 20 Fr/Ch (3.3mm), 108 cm 8888264804 Salem Sump PVC Gastroduodenal Tube, 10Fr/Ch (3.3 mm) x 85 cm 8888264820 Salem Sump PVC Gastroduodenal Tube, 12 Fr/Ch (4.0 mm) x 108 cm 8888264846 Salem Sump PVC Gastroduodenal Tube, 14 Fr/Ch (4.7mm), 108 cm 8888264861 Salem Sump PVC Gastroduodenal Tube, 16 Fr/Ch (5.3mm), 108 cm 8888264887 Salem Sump PVC Gastroduodenal Tube, 18 Fr/Ch (6.0mm), 108 cm SKUs not distributed in US Class II

Reason: Firm has received complaints regarding Anti-Reflux Valve (ARV) breakage due to excessive force during use. Firm is providing additional guidance for use. Expansion of RES 95214.

Cardinal Health 200, LLC 20250918 Ongoing
Distributed: Worldwide - US Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY, PR, GU and the countries of Australia, Austria, Bahrain, Belgium, Bulgaria, Canada, Chile, Colombia, Croatia, Czech Republic, Denmark, Ecuador, Estonia, Finland, France, French Polynesia, Germany, Greece, Hong Kong, Hungary, Ireland, Israel, Italy, Jordan, Africa, Kuwait, Luxembourg, Mexico Monaco, Netherlands, New Zealand, Northern Ireland, Norway, Pakistan, Panama, Philippines, Poland, Portugal, Qatar, Romania, Russia, Saudi Arabia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Taiwan, UAE, United Kingdom, Vietnam.