C-Rex UNO Microinterventional Goniotomy Instrument. Model Number: CRX-120. Product Description: The C-Rex UNO Microinterventional Goniotomy Instrument is a sterile single use disposable manual surgical instrument used in ophthalmic procedures for ab interno goniotomy and inner wall trabeculotomy. Class II

Reason: Potential for compromised integrity of the sterile packaging.

Iantrek Inc. 20250906 Ongoing
Distributed: US Nationwide distribution in the states of AL, AZ, CA, CT, DE, FL, GA, ID, IL, MD, MI, MN, MO, MS, NE, NJ, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA, WA, WI.
Cyclopen Microinterventional Cyclodialysis System. Model Number: CP-300. Product Description: The CycloPen Microinterventional Cyclodialysis device is a manual surgical instrument used for the ab-interno construction or modification of a cyclodialysis, followed by delivery of viscous materials, bio-tissue/allograft, or other materials to reinforce and maintain the cleft. Class II

Reason: Potential for compromised integrity of the sterile packaging.

Iantrek Inc. 20250906 Ongoing
Distributed: US Nationwide distribution in the states of AL, AZ, CA, CT, DE, FL, GA, ID, IL, MD, MI, MN, MO, MS, NE, NJ, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA, WA, WI.
FERNO POWER X2 POWERED AMBULANCE COT, Model/Catalog Number: 0015816; FOR USE IN GROUND TRANSPORTATION AMBULANCES. Class II

Reason: The fastening post hardware could loosen allowing the fastening post to detach from the cot

Ferno-Washington Inc 20250911 Ongoing
Distributed: US Nationwide distribution in the states of AL, AR, AZ, CA, CO, FL, IS, IL, KY, LA, MN, MO, NC, ND, NE, NM, NY, OH, OK, PA, TN, TX, UT, VA, WI, WV.
The failure occurs if a user on the Omnipod 5 iOS application selects and holds down on a manual entry field, to magnify the visual field using the magnifying feature of the iPhone, while simultaneously selecting a single number on the keyboard. Class II

Reason: The failure occurs if a user on the Omnipod 5 iOS application selects and holds down on a manual entry field, to magnify the visual field using the magnifying feature of the iPhone, while simultaneously selecting a single number on the keyboard. This causes the single digit number to duplicate. This could result in errors in manual entry of digits - examples: bolus entry, I:C ratio, bolus calculator, prior to starting the bolus insulin delivery. Therefore, this may result in over delivery of insulin which may cause hypoglycemia.

Insulet Corporation 20250911 Ongoing
Distributed: US Nationwide distribution.
Neuroguard IEP 3 in 1 Carotid Stent and Post-Dilation Balloon System with Integrated Embolic Protection (Neuroguard IEP System), REF: NG-0740-140-2, NG-NV-7-40, NG-0640-140-2, NG-NV-6-40 Class II

Reason: Carotid stent and post-dilation balloon system with integrated embolic protection system has an integrated filter that may not open or may not fully open during sheath retraction and operation of the filter knob. The stent is used with a primary distal embolic protection device.

Contego Medical LLC 20250904 Ongoing
Distributed: US Nationwide distribution in the states of FL, CO, CA, IL, NC, KY, TX, PA, DE, NJ, IN, NY, OH, LA, OK, SC, CT, WI, MA, KS, UT, MD, AL.
TactiFlex Sensor Enabled Ablation Catheter, 8F, 115 cm, F, (UniD Curve F), REF A-TFSE-F; cardiac catheter Class I

Reason: Abbott has become aware of a limited number of tip detachment events during catheter removal from the packaging.

Abbott 20250910 Ongoing
Distributed: Worldwide
TactiFlex Sensor Enabled Ablation Catheter, 8F, 115 cm, FJ, (BiD Curve F-J), REF A-TFSE-FJ; cardiac catheter Class I

Reason: Abbott has become aware of a limited number of tip detachment events during catheter removal from the packaging.

Abbott 20250910 Ongoing
Distributed: Worldwide
TactiFlex Sensor Enabled Ablation Catheter, 8F, 115 cm, DF, (BiD Curve D-F), REF A-TFSE-DF; cardiac catheter Class I

Reason: Abbott has become aware of a limited number of tip detachment events during catheter removal from the packaging.

Abbott 20250910 Ongoing
Distributed: Worldwide
TactiFlex Sensor Enabled Ablation Catheter, 8F, 115 cm, DD, (BiD Curve D-D), REF A-TFSE-DD; cardiac catheter Class I

Reason: Abbott has become aware of a limited number of tip detachment events during catheter removal from the packaging.

Abbott 20250910 Ongoing
Distributed: Worldwide
TactiFlex Sensor Enabled Ablation Catheter, 8F, 115 cm, D, (UniD Curve D), REF A-TFSE-D; cardiac catheter Class I

Reason: Abbott has become aware of a limited number of tip detachment events during catheter removal from the packaging.

Abbott 20250910 Ongoing
Distributed: Worldwide
Page 328 of 4,030 | Showing 10 recalls