ClariTEE Miniaturized TEE probe. Model Number: CLT-010. Ultrasound miniaturized TEE probe for use with Mobile Ultrasound, ImaCor model ZHH. Class II

Reason: Production assembly error resulting in the potential for reversed articulation, where moving the lever forward caused the probe tip to anteflex rather than retroflex, and vice versa.

ImaCor Inc. 20250829 Ongoing
Distributed: US Nationwide distribution in the states of South Carolina, New Jersey, Pennsylvania, Florida, and Georgia.
This device is a digital radiography/fluoroscopy system used in a diagnostic and interventional angiography configuration. The system is indicated for use in diagnostic and angiographic procedures for blood vessels in the heart, brain, abdomen and lower extremities. Class II

Reason: It has been found that when SPOT Fluoro is first turned on, the default ROI is selected and the collimator blades move to the boundary of the ROI, and the static image is overlayed around the ROI, covering the collimated area. If the customer changes the ROI, the collimator blades and static images are adjusted accordingly. However, when the acquisition program is changed, the collimator returns to the default ROI position which may be larger than the specified one, leaving the X-ray area partially covered by the static image. This is a software problem with the SPOT Fluoro program and will not occur if the function is not used.

Canon Medical System, USA, INC. 20250902 Ongoing
Distributed: U.S.
BD Alaris Pump Infusion Sets for the following reference numbers: C24101E, Amber Tubing, 15 Micron Filter, 0.2 Micron Filter, SmartSite Y-Site 10015414, 180 Micron Filter 2447-0007, Ball Valve SmartSite Port (Burette) 2 SmartSites Y-Sites 24201-0007, 2 SmartSite Y-sites 10012645, Back Check Valve, Non-vented, SmartSite Y-site 2426-0500, Back Check Valve, 3 SmartSite Y-sites 2450-0500, Check Valve 2 Ganged 3-Way Stopcocks 3 SmartSite Y-sites 11522558, Ball Valve, Back Check Valve, 3 SmartSite Y-sites 2410-0500, 2 SmartSite Y-Sites 2232-0007, 0.2 Micron Filter 11613191, 0.2 Micron Filter, Ball Valve, SmartSite" Port (Burette), 2 SmartSite Y-sites 10012144, 3 SmartSite Y-Sites, Check Valve 10013034, 3 SmartSite Y-Sites, Check Valve, 2-Ganged 4-way Stopcock 10013037, Non-Vented, 180 Micron Filter, Low Sorbing Tubing Segment, 1 SmartSite, Y-Site 11582773, 3 SmartSite Y-Sites, Check Valve, 2-Ganged 3-way Stopcock Class I

Reason: Infusion pump module used with compatible pump infusion sets may perform outside the established performance ranges for flow rate and bolus accuracy, downstream and upstream occlusion time to alarm, and post-occlusion bolos volume.

CareFusion 303, Inc. 20250911 Ongoing
Distributed: US: Nationwide OUS: United Arab Emirates Australia Belgium Bahrain Botswana Canada Switzerland China France United Kingdom Gibraltar Israel India Kuwait Mexico Malaysia Netherlands New Zealand Philippines Pakistan Qatar Saudi Arabia Singapore Taiwan, Province of China South Africa
NOxBOXi Nitric Oxide Delivery System; Model Numbers: NOXBOX-I Class I

Reason: The unexpected internal fault / system diagnostic error to result when a rapid succession of button selections is made to the user interface without waiting for the device to respond to user prompts.

NOXBOX LTD 20250909 Ongoing
Distributed: Domestic: TN;
NOxBOXi Nitric Oxide Delivery System; Model Numbers: NOXBOX-I Class I

Reason: Fluctuations may be more likely to occur if the total flow through the device sensor (used by the device to determine flow) is less than the minimum device requirement of 0.5 LPM. In addition, dose fluctuations have been observed when Bunnel LifePulse HFJV system or conventional ventilators have the following device settings: low service pressures generally less than 3.0 PSI, a pressure difference greater than 5.0cm H2O; or at breath per minute rates of 300 (5 Hz).

NOXBOX LTD 20250909 Ongoing
Distributed: Domestic: TN
Ashvins variant HEYEX 2 / HEYEX PACS. Software Version: Version 2.6.0 (Build 2088) to 2.6.8 (Build 2220). Device Type: Software (Medical Image Management, Picture Archiving and Communication System) Class II

Reason: Potential that the measured value may be smaller than the actual area.

MedicalCommunications GmbH 20250909 Ongoing
Distributed: US Nationwide distribution.
Argus PB-3000, Reference Numbers: 1A.701307,3.921002, 3.921030,3.921031, 3.921006 Class II

Reason: Vital sign monitoring instrument may trigger an error message during blood pressures measurements, Specifically, when the initial inflation pressure is set in the high-pressure range.

Schiller, Ag 20250808 Ongoing
Distributed: US Nationwide distribution in the states of MT, WI, IL.
VITEK¿ 2 Gram-negative AST Test Kits (Cards) containing Colistin formulation cs02n. VITEK¿ 2 is an automated system consisting of instruments, software and reagent cards designed for the identification and antimicrobial susceptibility testing of bacteria and yeast. Export only Class II

Reason: Potential for false resistant results for Colistin (cs02n) with multi-drug-resistant Pseudomonas aeruginosa and Acinetobacter baumannii complex used in specific cards

bioMerieux, Inc. 20250626 Ongoing
Distributed: International distribution in the countries of Albania, Armenia, Australia, Austria, Belgium, Bosnia, Bulgaria, China, Croatia, Cyprus, Czech Republic/Slovakia, France, Germany, Greece, Iberia, India, Israel, Italy, Japan, Lithuania, Macedonia, Malaysia/Brunei, Moldova, Netherlands, Pakistan, Poland, Portugal, Romania, Russia, Serbia, South Korea, Spain, Switzerland, Taiwan, Thailand, and Turkey.
Medline Kits: 1) VALVE PACK-LF, Model Number: DYNJ0415366P; 2) VALVE PACK-LF, Model Number: DYNJ0415366Q Class I

Reason: The kits contain certain lots of cannula products where the catheter may not retain its shape.

Medline Industries, LP 20250821 Ongoing
Distributed: US Nationwide distribution in the state of CA.
MAGNETOM Vida Fit. Model Number: 11410481. Class I

Reason: There is a potential for an ice blockage to form or currently exist within the magnet venting system. In the event of a quench, helium gas may be unable to escape through the designed venting paths, leading to a pressure build-up within the helium containment system. This pressure build-up could ultimately rupture the helium containment system, potentially resulting in a helium leak into the scanning room.

Siemens Medical Solutions USA, Inc 20250828 Ongoing
Distributed: Worldwide distribution - US Nationwide and the countries of United Arab Emirates, Armenia, Angola, Argentina, Austria, Australia, Azerbaijan, Bosnia and Herzegovina, Bangladesh, Belgium, Bulgaria, Bahrain, Brunei Darussalam, Brazil, Botswana, Belarus, Canada, Switzerland, Chile, China, Colombia, Cyprus, Czechia, Germany, Denmark, Algeria, Ecuador, Estonia, Egypt, Spain, Ethiopia, Finland, Faroe Islands, France, United Kingdom of Great Britian and Northern Ireland, Georgia, Greece, Hong Kong, Croatia, Hungary, Canary Islands, Indonesia, Ireland, Israel, India, Iran, Iceland, Italy, Jordan, Japan, Kenya, Korea, Kuwait, Kazakhstan, Lebanon, Sri Lanka, Lithuania, Luxembourg, Latvia, Morocco, Monaco, Moldova, Montenegro, Republic of North Macedonia, Myanmar, Mongolia, Macao, Malta, Maldives, Mexico, Malaysia, Mozambique, Nigeria, Netherlands, Norway, Nepal, New Zealand, Oman, Panama, Peru, Philippines, Pakistan, Poland, Portugal, Paraguay, Qatar, Reunion, Romania, Serbia, Russian Federation, Saudi Arabia, Sweden, Singapore, Slovenia, Slovakia, Senegal, Syrian Arab Republic, Thailand, Turkmenistan, Tunisia, Turkey, Trinidad and Tobago, Taiwan, Tanzania, Ukraine, Uganda, Uzbekistan, Viet Nam, Kosovo, South Africa.