Reprocessed ACUSON AcuNav Ultrasound Catheter for intracardiac and intra-luminal visualization of cardiac and great vessel anatomy and physiology as well as visualization of other devices in the heart. For imaging guidance only, not treatment delivery, during cardiac interventional percutaneous procedures. Medline Item Numbers 10135910RH 10135936RH Class I

Reason: Specific lots of reprocessed electrophysiology and ultrasound catheters may contain small residual particulates. If devices with residual particulates on patient-contacting surfaces are used, there is a potential risk of an inflammatory response or systemic infection, and if particulate were to dislodge during use and contact circulating blood and/or the heart, it may produce a granulomatous reaction or thrombus formation, which may lead to cerebral or pulmonary embolism, or deep vein thrombosis.

Medline Industries, LP 20250731 Ongoing
Distributed: Domestic US distribution nationwide. International distribution pending. No international distribution.
Reprocessed ViewFlex Xtra ICE Catheter to visualize cardiac structures; blood flow and other devices within the heart. Medline Item Numbers D087031RH Class I

Reason: Specific lots of reprocessed electrophysiology and ultrasound catheters may contain small residual particulates. If devices with residual particulates on patient-contacting surfaces are used, there is a potential risk of an inflammatory response or systemic infection, and if particulate were to dislodge during use and contact circulating blood and/or the heart, it may produce a granulomatous reaction or thrombus formation, which may lead to cerebral or pulmonary embolism, or deep vein thrombosis.

Medline Industries, LP 20250731 Ongoing
Distributed: Domestic US distribution nationwide. International distribution pending. No international distribution.
Reprocessed Inquiry Steerable Diagnostics Catheter for electrogram recording and cardiac stimulation during diagnostic electrophysiologic studies; Medline Item Numbers 81102RH 81104RH 81107RH 81174RH 81402RH 81404RH 81405RH 81472RH 81473RH 81474RH 81531RH 81532RH Class I

Reason: Specific lots of reprocessed electrophysiology and ultrasound catheters may contain small residual particulates. If devices with residual particulates on patient-contacting surfaces are used, there is a potential risk of an inflammatory response or systemic infection, and if particulate were to dislodge during use and contact circulating blood and/or the heart, it may produce a granulomatous reaction or thrombus formation, which may lead to cerebral or pulmonary embolism, or deep vein thrombosis.

Medline Industries, LP 20250731 Ongoing
Distributed: Domestic US distribution nationwide. International distribution pending. No international distribution.
Reprocessed Webster CS Catheter 10 Electrodes for electrophysiological mapping of cardiac structures. Medline Item Numbers BD710DF282CRH BD710DF282RRH BD710FJ282CRH BD710FJ282RRH Class I

Reason: Specific lots of reprocessed electrophysiology and ultrasound catheters may contain small residual particulates. If devices with residual particulates on patient-contacting surfaces are used, there is a potential risk of an inflammatory response or systemic infection, and if particulate were to dislodge during use and contact circulating blood and/or the heart, it may produce a granulomatous reaction or thrombus formation, which may lead to cerebral or pulmonary embolism, or deep vein thrombosis.

Medline Industries, LP 20250731 Ongoing
Distributed: Domestic US distribution nationwide. International distribution pending. No international distribution.
Reprocessed Livewire Steerable Electrophysiology Catheter Item Numbers 401575RH 401582RH 401600RH 401603RH 401606RH 401652RH 401653RH 401654RH 401904RH 401905RH 401908RH 401914RH 401915RH 401918RH 401932RH 401933RH 401934RH 401938RH 401939RH 401940RH 401941RH 401991RH Class I

Reason: Specific lots of reprocessed electrophysiology and ultrasound catheters may contain small residual particulates. If devices with residual particulates on patient-contacting surfaces are used, there is a potential risk of an inflammatory response or systemic infection, and if particulate were to dislodge during use and contact circulating blood and/or the heart, it may produce a granulomatous reaction or thrombus formation, which may lead to cerebral or pulmonary embolism, or deep vein thrombosis.

Medline Industries, LP 20250731 Ongoing
Distributed: Domestic US distribution nationwide. International distribution pending. No international distribution.
Sterile Medline Convenience Kits: 1) EXTREMITY PACK, Model Number: DYNJ45701B Class I

Reason: Convenience kits labeled as sterile have not gone through the sterilization process.

Medline Industries, LP 20250808 Ongoing
Distributed: US Distribution to States of: FL, MD, NJ, TN
Sterile Medline Convenience Kits: 1) MAJOR PACK L-F, Model Number: DYNJ0382730O; 2) LB BASIC CUSTOM PACK, Model Number: DYNJ61038B Class I

Reason: Convenience kits labeled as sterile have not gone through the sterilization process.

Medline Industries, LP 20250808 Ongoing
Distributed: US Distribution to States of: FL, MD, NJ, TN
GORE ACUSEAL Vascular Graft, REF: ECH050020J, ECH050020W, ECH050050J, ECH050050W, ECH060010A, ECH060020A, ECH060020J, ECH060020W, ECH060040, ECH060040A, ECH060040W, ECH060050A, ECH060050J, ECH460045A, ECH460045J, ECH470045, ECH470045A; used with Instructions for Use. Class II

Reason: Reports of vascular graft delamination, which may lead to reduced access flow; cannulation difficulties; bleeding or bruising; and harms related to reintervention procedures, so Instructions for Use will be updated to modify existing warning to 1) further clarify techniques that may contribute to the risk of delamination, 2) Addition of delamination to device-related adverse event section.

W L Gore & Associates, Inc. 20250826 Ongoing
Distributed: US: OR, AL, OH, MD, VA, CA, FL, NC, OK, MA, MO, AZ, TX, IL, GA, PA, NH, HI, MN, MS, KY, ID, LA, NJ, TN, DE, IN, MI, WI, NY, DC, SC, CT, NE, IA, WA, AR, NV, WV, AK, UT, ND, KS, CO, ME, MT, NM, SD, PR, WY, RI. OUS: PANAMA, CANADA, CHINA, TAIWAN, GERMANY, SLOVENIA, NORWAY, UNITED KINGDOM, ITALY, SPAIN, FINLAND, NETHERLANDS, SWEDEN, FRANCE, UNITED ARAB EMIRATES, DENMARK, PORTUGAL, AUSTRIA, SWITZERLAND, BELGIUM, TURKIYE, SLOVAK REPUBLIC, CYPRUS, SAUDI ARABIA, SOUTH AFRICA, LUXEMBOURG, FRENCH GUIANA, ANDORRA, IRELAND, NEW CALEDONIA, POLAND, ISRAEL, GREECE, GUADELOUPE, CROATIA, REUNION, BAHRAIN, HUNGARY, JAPAN
Artegraft Collagen Vascular Grafts Model Numbers: (1) AG 616, (2) AG 630, (3) AG 636, (4) AG 640, (5) AG 645, (6) AG 715, (7) AG 730, (8) AG 735, (9) AG 740, (10) AG 745, (11) AG 750, (12) AG 830, (13) AG 840, (14) AG 1015, (15) AG 1030; Class II

Reason: The bovine carotid arteries used to produce the impacted grafts from affected lots were sourced using raw material from a supplier that had not been reviewed and approved by the appropriate regulatory authority.

LeMaitre Vascular, Inc. 20250902 Ongoing
Distributed: US Nationwide distribution in the states of CA, CT, FL, GA, IL, MA, MS, NC, ND, NJ, NY, OK, PA, TN, TX, VA, WA.
Medtronic, Simplera Sensor, REF: MMT-5100JD1 Class II

Reason: The device cap label may not adhere to the sensor, compromising the sterility of the sensor, and increasing the risk of infection.

Medtronic MiniMed, Inc. 20250507 Ongoing
Distributed: US: No distribution OUS: Austria, Belgium, Italy, Spain, Switzerland, United Kingdom
Page 340 of 4,030 | Showing 10 recalls
Data provided by openFDA