Allura CV20; Product Code: 722031; Associated 510(k) numbers: K102005, K130842, K130638, K133292, K141979, K161563, K162859, K033737, K041949; Class II

Reason: Systems may not perform as intended due to the deterioration of certain internal components (the CMOS battery, hard disk drive, and/or power supply unit). If this deterioration occurs, motorized movements of the system will become unavailable. Manual movements of the stand (longitudinal, transversal and rotational for ceiling mounted systems and rotational for floor mounted systems) and table longitudinal movements remain available. Imaging (X-ray) functionality remains available.

PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. 20250903 Ongoing
Distributed: Worldwide distribution - US Nationwide and the countries of Albania, Algeria, Argentina, Australia, Austria, Azerbaijan, Bangladesh, Belarus, Belgium, Brazil, Bulgaria, Canada, Chile, China, Colombia, Costa Rica, Cuba, Cyprus, Czech Republic, Denmark, Dominican Rep, Ecuador, Egypt, Estonia, Fiji, Finland, France, FrenchPolynesia, Georgia, Germany, Greece, Honduras, Hong Kong, Hungary, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Kosovo, Lao, Lebanon, Libya, Lithuania, Macedonia, Malaysia, Malta, Martinique, Mauritius, Mexico, Mongolia, Morocco, Myanmar, Nepal, Netherlands, New Caledonia, New Zealand, Norway, Oman, Pakistan, Palestine, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, R¿union, Romania, Russian Fed., Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sri Lanka, Sudan, Sweden, Switzerland, Syria, Taiwan, Tanzania, Thailand, Tunisia, T¿rkiye, Ukraine, United Kingdom, Uruguay, Utd.Arab.Emir., Uzbekistan, Venezuela, Vietnam, Yemen.
Allura Xper FD20/15 OR Table; Product Code: 722059; Associated 510(k) numbers: K102005, K130842, K130638, K133292, K141979, K161563, K162859, K033737, K041949; Class II

Reason: Systems may not perform as intended due to the deterioration of certain internal components (the CMOS battery, hard disk drive, and/or power supply unit). If this deterioration occurs, motorized movements of the system will become unavailable. Manual movements of the stand (longitudinal, transversal and rotational for ceiling mounted systems and rotational for floor mounted systems) and table longitudinal movements remain available. Imaging (X-ray) functionality remains available.

PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. 20250903 Ongoing
Distributed: Worldwide distribution - US Nationwide and the countries of Albania, Algeria, Argentina, Australia, Austria, Azerbaijan, Bangladesh, Belarus, Belgium, Brazil, Bulgaria, Canada, Chile, China, Colombia, Costa Rica, Cuba, Cyprus, Czech Republic, Denmark, Dominican Rep, Ecuador, Egypt, Estonia, Fiji, Finland, France, FrenchPolynesia, Georgia, Germany, Greece, Honduras, Hong Kong, Hungary, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Kosovo, Lao, Lebanon, Libya, Lithuania, Macedonia, Malaysia, Malta, Martinique, Mauritius, Mexico, Mongolia, Morocco, Myanmar, Nepal, Netherlands, New Caledonia, New Zealand, Norway, Oman, Pakistan, Palestine, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, R¿union, Romania, Russian Fed., Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sri Lanka, Sudan, Sweden, Switzerland, Syria, Taiwan, Tanzania, Thailand, Tunisia, T¿rkiye, Ukraine, United Kingdom, Uruguay, Utd.Arab.Emir., Uzbekistan, Venezuela, Vietnam, Yemen.
Allura Xper FD20/15; Product Code: 722058; Associated 510(k) numbers: K102005, K130842, K130638, K133292, K141979, K161563, K162859, K033737, K041949; Class II

Reason: Systems may not perform as intended due to the deterioration of certain internal components (the CMOS battery, hard disk drive, and/or power supply unit). If this deterioration occurs, motorized movements of the system will become unavailable. Manual movements of the stand (longitudinal, transversal and rotational for ceiling mounted systems and rotational for floor mounted systems) and table longitudinal movements remain available. Imaging (X-ray) functionality remains available.

PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. 20250903 Ongoing
Distributed: Worldwide distribution - US Nationwide and the countries of Albania, Algeria, Argentina, Australia, Austria, Azerbaijan, Bangladesh, Belarus, Belgium, Brazil, Bulgaria, Canada, Chile, China, Colombia, Costa Rica, Cuba, Cyprus, Czech Republic, Denmark, Dominican Rep, Ecuador, Egypt, Estonia, Fiji, Finland, France, FrenchPolynesia, Georgia, Germany, Greece, Honduras, Hong Kong, Hungary, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Kosovo, Lao, Lebanon, Libya, Lithuania, Macedonia, Malaysia, Malta, Martinique, Mauritius, Mexico, Mongolia, Morocco, Myanmar, Nepal, Netherlands, New Caledonia, New Zealand, Norway, Oman, Pakistan, Palestine, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, R¿union, Romania, Russian Fed., Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sri Lanka, Sudan, Sweden, Switzerland, Syria, Taiwan, Tanzania, Thailand, Tunisia, T¿rkiye, Ukraine, United Kingdom, Uruguay, Utd.Arab.Emir., Uzbekistan, Venezuela, Vietnam, Yemen.
Allura Xper FD20/10; Product Code: 722029; Associated 510(k) numbers: K102005, K130842, K130638, K133292, K141979, K161563, K162859, K033737, K041949; Class II

Reason: Systems may not perform as intended due to the deterioration of certain internal components (the CMOS battery, hard disk drive, and/or power supply unit). If this deterioration occurs, motorized movements of the system will become unavailable. Manual movements of the stand (longitudinal, transversal and rotational for ceiling mounted systems and rotational for floor mounted systems) and table longitudinal movements remain available. Imaging (X-ray) functionality remains available.

PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. 20250903 Ongoing
Distributed: Worldwide distribution - US Nationwide and the countries of Albania, Algeria, Argentina, Australia, Austria, Azerbaijan, Bangladesh, Belarus, Belgium, Brazil, Bulgaria, Canada, Chile, China, Colombia, Costa Rica, Cuba, Cyprus, Czech Republic, Denmark, Dominican Rep, Ecuador, Egypt, Estonia, Fiji, Finland, France, FrenchPolynesia, Georgia, Germany, Greece, Honduras, Hong Kong, Hungary, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Kosovo, Lao, Lebanon, Libya, Lithuania, Macedonia, Malaysia, Malta, Martinique, Mauritius, Mexico, Mongolia, Morocco, Myanmar, Nepal, Netherlands, New Caledonia, New Zealand, Norway, Oman, Pakistan, Palestine, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, R¿union, Romania, Russian Fed., Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sri Lanka, Sudan, Sweden, Switzerland, Syria, Taiwan, Tanzania, Thailand, Tunisia, T¿rkiye, Ukraine, United Kingdom, Uruguay, Utd.Arab.Emir., Uzbekistan, Venezuela, Vietnam, Yemen.
Allura Xper FD20/20 OR Table; Product Code: 722039; Associated 510(k) numbers: K102005, K130842, K130638, K133292, K141979, K161563, K162859, K033737, K041949; Class II

Reason: Systems may not perform as intended due to the deterioration of certain internal components (the CMOS battery, hard disk drive, and/or power supply unit). If this deterioration occurs, motorized movements of the system will become unavailable. Manual movements of the stand (longitudinal, transversal and rotational for ceiling mounted systems and rotational for floor mounted systems) and table longitudinal movements remain available. Imaging (X-ray) functionality remains available.

PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. 20250903 Ongoing
Distributed: Worldwide distribution - US Nationwide and the countries of Albania, Algeria, Argentina, Australia, Austria, Azerbaijan, Bangladesh, Belarus, Belgium, Brazil, Bulgaria, Canada, Chile, China, Colombia, Costa Rica, Cuba, Cyprus, Czech Republic, Denmark, Dominican Rep, Ecuador, Egypt, Estonia, Fiji, Finland, France, FrenchPolynesia, Georgia, Germany, Greece, Honduras, Hong Kong, Hungary, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Kosovo, Lao, Lebanon, Libya, Lithuania, Macedonia, Malaysia, Malta, Martinique, Mauritius, Mexico, Mongolia, Morocco, Myanmar, Nepal, Netherlands, New Caledonia, New Zealand, Norway, Oman, Pakistan, Palestine, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, R¿union, Romania, Russian Fed., Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sri Lanka, Sudan, Sweden, Switzerland, Syria, Taiwan, Tanzania, Thailand, Tunisia, T¿rkiye, Ukraine, United Kingdom, Uruguay, Utd.Arab.Emir., Uzbekistan, Venezuela, Vietnam, Yemen.
Allura Xper FD20/20; Product Code: 722038; Associated 510(k) numbers: K102005, K130842, K130638, K133292, K141979, K161563, K162859, K033737, K041949; Class II

Reason: Systems may not perform as intended due to the deterioration of certain internal components (the CMOS battery, hard disk drive, and/or power supply unit). If this deterioration occurs, motorized movements of the system will become unavailable. Manual movements of the stand (longitudinal, transversal and rotational for ceiling mounted systems and rotational for floor mounted systems) and table longitudinal movements remain available. Imaging (X-ray) functionality remains available.

PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. 20250903 Ongoing
Distributed: Worldwide distribution - US Nationwide and the countries of Albania, Algeria, Argentina, Australia, Austria, Azerbaijan, Bangladesh, Belarus, Belgium, Brazil, Bulgaria, Canada, Chile, China, Colombia, Costa Rica, Cuba, Cyprus, Czech Republic, Denmark, Dominican Rep, Ecuador, Egypt, Estonia, Fiji, Finland, France, FrenchPolynesia, Georgia, Germany, Greece, Honduras, Hong Kong, Hungary, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Kosovo, Lao, Lebanon, Libya, Lithuania, Macedonia, Malaysia, Malta, Martinique, Mauritius, Mexico, Mongolia, Morocco, Myanmar, Nepal, Netherlands, New Caledonia, New Zealand, Norway, Oman, Pakistan, Palestine, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, R¿union, Romania, Russian Fed., Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sri Lanka, Sudan, Sweden, Switzerland, Syria, Taiwan, Tanzania, Thailand, Tunisia, T¿rkiye, Ukraine, United Kingdom, Uruguay, Utd.Arab.Emir., Uzbekistan, Venezuela, Vietnam, Yemen.
Allura Xper FD20 Biplane OR Table Product Codes: (1)722020, (2) 722025; Associated 510(k) numbers: K102005, K130842, K130638, K133292, K141979, K161563, K162859, K033737, K041949; Class II

Reason: Systems may not perform as intended due to the deterioration of certain internal components (the CMOS battery, hard disk drive, and/or power supply unit). If this deterioration occurs, motorized movements of the system will become unavailable. Manual movements of the stand (longitudinal, transversal and rotational for ceiling mounted systems and rotational for floor mounted systems) and table longitudinal movements remain available. Imaging (X-ray) functionality remains available.

PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. 20250903 Ongoing
Distributed: Worldwide distribution - US Nationwide and the countries of Albania, Algeria, Argentina, Australia, Austria, Azerbaijan, Bangladesh, Belarus, Belgium, Brazil, Bulgaria, Canada, Chile, China, Colombia, Costa Rica, Cuba, Cyprus, Czech Republic, Denmark, Dominican Rep, Ecuador, Egypt, Estonia, Fiji, Finland, France, FrenchPolynesia, Georgia, Germany, Greece, Honduras, Hong Kong, Hungary, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Kosovo, Lao, Lebanon, Libya, Lithuania, Macedonia, Malaysia, Malta, Martinique, Mauritius, Mexico, Mongolia, Morocco, Myanmar, Nepal, Netherlands, New Caledonia, New Zealand, Norway, Oman, Pakistan, Palestine, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, R¿union, Romania, Russian Fed., Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sri Lanka, Sudan, Sweden, Switzerland, Syria, Taiwan, Tanzania, Thailand, Tunisia, T¿rkiye, Ukraine, United Kingdom, Uruguay, Utd.Arab.Emir., Uzbekistan, Venezuela, Vietnam, Yemen.
Allura Xper FD20 OR Table Product Codes: (1)722023, (2) 722035, (3) 722015; Associated 510(k) numbers: K102005, K130842, K130638, K133292, K141979, K161563, K162859, K033737, K041949; Class II

Reason: Systems may not perform as intended due to the deterioration of certain internal components (the CMOS battery, hard disk drive, and/or power supply unit). If this deterioration occurs, motorized movements of the system will become unavailable. Manual movements of the stand (longitudinal, transversal and rotational for ceiling mounted systems and rotational for floor mounted systems) and table longitudinal movements remain available. Imaging (X-ray) functionality remains available.

PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. 20250903 Ongoing
Distributed: Worldwide distribution - US Nationwide and the countries of Albania, Algeria, Argentina, Australia, Austria, Azerbaijan, Bangladesh, Belarus, Belgium, Brazil, Bulgaria, Canada, Chile, China, Colombia, Costa Rica, Cuba, Cyprus, Czech Republic, Denmark, Dominican Rep, Ecuador, Egypt, Estonia, Fiji, Finland, France, FrenchPolynesia, Georgia, Germany, Greece, Honduras, Hong Kong, Hungary, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Kosovo, Lao, Lebanon, Libya, Lithuania, Macedonia, Malaysia, Malta, Martinique, Mauritius, Mexico, Mongolia, Morocco, Myanmar, Nepal, Netherlands, New Caledonia, New Zealand, Norway, Oman, Pakistan, Palestine, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, R¿union, Romania, Russian Fed., Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sri Lanka, Sudan, Sweden, Switzerland, Syria, Taiwan, Tanzania, Thailand, Tunisia, T¿rkiye, Ukraine, United Kingdom, Uruguay, Utd.Arab.Emir., Uzbekistan, Venezuela, Vietnam, Yemen.
Allura Xper FD20 Biplane Product Codes: (1)722013, (2) 722008; Associated 510(k) numbers: K102005, K130842, K130638, K133292, K141979, K161563, K162859, K033737, K041949; Class II

Reason: Systems may not perform as intended due to the deterioration of certain internal components (the CMOS battery, hard disk drive, and/or power supply unit). If this deterioration occurs, motorized movements of the system will become unavailable. Manual movements of the stand (longitudinal, transversal and rotational for ceiling mounted systems and rotational for floor mounted systems) and table longitudinal movements remain available. Imaging (X-ray) functionality remains available.

PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. 20250903 Ongoing
Distributed: Worldwide distribution - US Nationwide and the countries of Albania, Algeria, Argentina, Australia, Austria, Azerbaijan, Bangladesh, Belarus, Belgium, Brazil, Bulgaria, Canada, Chile, China, Colombia, Costa Rica, Cuba, Cyprus, Czech Republic, Denmark, Dominican Rep, Ecuador, Egypt, Estonia, Fiji, Finland, France, FrenchPolynesia, Georgia, Germany, Greece, Honduras, Hong Kong, Hungary, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Kosovo, Lao, Lebanon, Libya, Lithuania, Macedonia, Malaysia, Malta, Martinique, Mauritius, Mexico, Mongolia, Morocco, Myanmar, Nepal, Netherlands, New Caledonia, New Zealand, Norway, Oman, Pakistan, Palestine, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, R¿union, Romania, Russian Fed., Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sri Lanka, Sudan, Sweden, Switzerland, Syria, Taiwan, Tanzania, Thailand, Tunisia, T¿rkiye, Ukraine, United Kingdom, Uruguay, Utd.Arab.Emir., Uzbekistan, Venezuela, Vietnam, Yemen.
Allura Xper FD20 Product Codes: (1)722012, (2) 722028, (3) 722006; Associated 510(k) numbers: K102005, K130842, K130638, K133292, K141979, K161563, K162859, K033737, K041949; Class II

Reason: Systems may not perform as intended due to the deterioration of certain internal components (the CMOS battery, hard disk drive, and/or power supply unit). If this deterioration occurs, motorized movements of the system will become unavailable. Manual movements of the stand (longitudinal, transversal and rotational for ceiling mounted systems and rotational for floor mounted systems) and table longitudinal movements remain available. Imaging (X-ray) functionality remains available.

PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. 20250903 Ongoing
Distributed: Worldwide distribution - US Nationwide and the countries of Albania, Algeria, Argentina, Australia, Austria, Azerbaijan, Bangladesh, Belarus, Belgium, Brazil, Bulgaria, Canada, Chile, China, Colombia, Costa Rica, Cuba, Cyprus, Czech Republic, Denmark, Dominican Rep, Ecuador, Egypt, Estonia, Fiji, Finland, France, FrenchPolynesia, Georgia, Germany, Greece, Honduras, Hong Kong, Hungary, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Kosovo, Lao, Lebanon, Libya, Lithuania, Macedonia, Malaysia, Malta, Martinique, Mauritius, Mexico, Mongolia, Morocco, Myanmar, Nepal, Netherlands, New Caledonia, New Zealand, Norway, Oman, Pakistan, Palestine, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, R¿union, Romania, Russian Fed., Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sri Lanka, Sudan, Sweden, Switzerland, Syria, Taiwan, Tanzania, Thailand, Tunisia, T¿rkiye, Ukraine, United Kingdom, Uruguay, Utd.Arab.Emir., Uzbekistan, Venezuela, Vietnam, Yemen.