LIFEPAK CR2 Part / Catalog Numbers / UDI-DI code: CR2-2-001451 99512-001434 883873785911 LIFEPAK CR2: The LIFEPAK CR2 defibrillator is indicated for use on patients 1 year of age or older in cardiopulmonary arrest. The patient must be unresponsive (unconscious), not breathing normally, and showing no signs of circulation (for example, no pulse, no coughing, or no movement). The cprCOACHTM Feedback Technology in the LIFEPAK CR2 defibrillator is indicated for use on cardiopulmonary arrest patients and provides CPR guidance in accordance with the AHA Guidelines for patients 1 year of age or older. The LIFEPAK CR2 defibrillator is intended for use by personnel who have been trained in its operation. Users should have received training in basic life support/AED, advanced life support, or a physician-authorized emergency medical response training program. Class II

Reason: Due to required inspections not being performed on products/units that have gone through servicing.

Physio-Control, Inc. 20250722 Ongoing
Distributed: U.S. Nationwide distribution in the states of CO and MT.
LIFEPAK 1000 Part / Catalog Numbers / UDI-DI code: 320371500023 / 99425-000025 / 885074698953 320371500229 / 99425-000025 / 885074698953 331706400001 / 99425-000025 / 885074698953 LIFEPAK 1000: The defibrillator is to be used in AED mode only on patients who are in cardiopulmonary arrest. The patient must be unresponsive, not breathing normally, and showing no signs of circulation. ECG monitoring is for use on conscious and unconscious patients of all ages for the purpose of ECG rhythm recognition and heart rate monitoring. Class II

Reason: Due to required inspections not being performed on products/units that have gone through servicing.

Physio-Control, Inc. 20250722 Ongoing
Distributed: U.S. Nationwide distribution in the states of CO and MT.
LIFEPAK 15 Part / Catalog Numbers / UDI-DI code: V15-7-000022 / 99577-001957 / 883873871690 V15-2-000051 / 99577-000046 / 883873988022 V15-2-000099 / 99577-000093 / 883873979020 V15-2-001603 / 99577-001256 / 883873911631 LIFEPAK 15: The LP15 is intended for use by trained medical personnel in out-of-doors and indoor emergency care settings within the environmental conditions specified in the Operating Instructions. The LP15 is designed to be used during ground transportation except when specified otherwise. Manual mode monitoring and therapy functions are intended for use on adult and pediatric patients. Automated external defibrillation mode is intended for use on patients eight years of age and older. Class II

Reason: Due to required inspections not being performed on products/units that have gone through servicing.

Physio-Control, Inc. 20250722 Ongoing
Distributed: U.S. Nationwide distribution in the states of CO and MT.
InPen App, CFN: MMT-8060 (iOS Users) and MMT-8061 (Android Users) Class II

Reason: Medtronic MiniMed, Inc. is recalling InPen App for iOS and Android users due to software design errors that could lead to a missed short-acting insulin dose reminder and a recommendation to correct a high glucose value. It does not impact insulin delivery, long-acting insulin dose reminders, or CGM alerts, and users can still use the pen itself to calculate a dose, deliver insulin, record the dose date/time, and view CGM data. This issue was identified during internal testing before release in the US but after release to OUS customers, no complaints or MDRs related to this recall have been reported. Use of the affected device may result hyperglycemia by failing to alert the user and delay treatment of diabetes.

Medtronic MiniMed, Inc. 20250616 Ongoing
Distributed: International distribution to the countries of Argentina, Australia, Austria, Belgium, Bermuda, Chile, Czech Republic, Denmark, Egypt, Finland, France, Georgia, Germany, Greece, Hong Kong, Hungary, Iceland, India, Ireland, Israel, Italy, Luxembourg, Netherlands, New Zealand, Norway, Poland, Portugal, Romania, Saudi Arabia, Serbia, Singapore, South Africa, Spain, Sweden, Switzerland, United Arab Emirates, United Kingdom.
RayStation with the following product descriptions: 1. RayStation 2023B, Software Version: 14.0.0.3338. 2. RayStation 2024A, Software Version: 15.0.0.430. 3. RayStation 2024A SP3, Software Version: 15.1.3.10. Product Description: Radiation Therapy Treatment Planning System. Class II

Reason: DICOM SOP Instance UID and Series Instance UID from RayGateway, i.e. the interface to Accuray's iDMS, are not guaranteed to be unique.

RAYSEARCH LABORATORIES AB 20250808 Ongoing
Distributed: Worldwide - US Nationwide distribution in the states of AL, CA, FL, GA, IL, IA, KY, LA, MD, MN, MS, NY, NC, OH, OK, SC, WA, WI and the countries of Canada, China, Colombia, France, Germany, Hong Kong S.A.R., China, Italy, Japan, Netherlands, Poland, Spain, Sweden, Switzerland, ENG, Thailand, Turkey, UK, Ukraine.
RayStation with the following product descriptions: 1. RayStation 6 Service Pack 1, Software Version: 6.1.1.2. 2. RayStation 6 Service Pack 2, Software Version: 6.2.0.7. 3. RayStation 6 Service Pack 3, Software Version: 6.3.0.6. Product Description: Radiation Therapy Treatment Planning System. Class II

Reason: DICOM SOP Instance UID and Series Instance UID from RayGateway, i.e. the interface to Accuray's iDMS, are not guaranteed to be unique.

RAYSEARCH LABORATORIES AB 20250808 Ongoing
Distributed: Worldwide - US Nationwide distribution in the states of AL, CA, FL, GA, IL, IA, KY, LA, MD, MN, MS, NY, NC, OH, OK, SC, WA, WI and the countries of Canada, China, Colombia, France, Germany, Hong Kong S.A.R., China, Italy, Japan, Netherlands, Poland, Spain, Sweden, Switzerland, ENG, Thailand, Turkey, UK, Ukraine.
RayStation with the following product descriptions: 1. RayStation 12A, Software Version: 13.0.0.1547. 2. RayStation 12A Service Pack 1, Software Version: 13.1.0.144. 3. RayStation 12A Service Pack 2, Software Version: 13.1.1.89. Product Description: Radiation Therapy Treatment Planning System. Class II

Reason: DICOM SOP Instance UID and Series Instance UID from RayGateway, i.e. the interface to Accuray's iDMS, are not guaranteed to be unique.

RAYSEARCH LABORATORIES AB 20250808 Ongoing
Distributed: Worldwide - US Nationwide distribution in the states of AL, CA, FL, GA, IL, IA, KY, LA, MD, MN, MS, NY, NC, OH, OK, SC, WA, WI and the countries of Canada, China, Colombia, France, Germany, Hong Kong S.A.R., China, Italy, Japan, Netherlands, Poland, Spain, Sweden, Switzerland, ENG, Thailand, Turkey, UK, Ukraine.
RayStation with the following product descriptions: 1. RayStation 11B, Software Version: 12.0.0.932. 2. RayStation 11B Service Pack 1, Software Version: 12.1.0.1221. 3. RayStation 11B Service Pack 2, Software Version: 12.0.3.68. 4. RayStation 11B Service Pack 3, Software Version: 12.0.4.12. 5. RayStation 11B Service Pack Toshiba 1, Software Version: 12.3.0.119. Product Description: Radiation Therapy Treatment Planning System. Class II

Reason: DICOM SOP Instance UID and Series Instance UID from RayGateway, i.e. the interface to Accuray's iDMS, are not guaranteed to be unique.

RAYSEARCH LABORATORIES AB 20250808 Ongoing
Distributed: Worldwide - US Nationwide distribution in the states of AL, CA, FL, GA, IL, IA, KY, LA, MD, MN, MS, NY, NC, OH, OK, SC, WA, WI and the countries of Canada, China, Colombia, France, Germany, Hong Kong S.A.R., China, Italy, Japan, Netherlands, Poland, Spain, Sweden, Switzerland, ENG, Thailand, Turkey, UK, Ukraine.
RayStation with the following product descriptions: 1. RayStation 11A, Software Version: 11.0.0.951. 2. RayStation 11A Service Pack 1, Software Version: 11.0.1.29. 3. RayStation 11A Service Pack 2, Software Version: 11.0.3.116. 4. RayStation 11A Service Pack 3, Software Version: 11.0.4.15. Product Description: Radiation Therapy Treatment Planning System. Class II

Reason: DICOM SOP Instance UID and Series Instance UID from RayGateway, i.e. the interface to Accuray's iDMS, are not guaranteed to be unique.

RAYSEARCH LABORATORIES AB 20250808 Ongoing
Distributed: Worldwide - US Nationwide distribution in the states of AL, CA, FL, GA, IL, IA, KY, LA, MD, MN, MS, NY, NC, OH, OK, SC, WA, WI and the countries of Canada, China, Colombia, France, Germany, Hong Kong S.A.R., China, Italy, Japan, Netherlands, Poland, Spain, Sweden, Switzerland, ENG, Thailand, Turkey, UK, Ukraine.
RayStation with the following product descriptions: 1. RayStation 10B, Software Version: 10.1.0.613. 2. RayStation 10B Service Pack 1, Software Version: 10.1.1.54. Product Description: Radiation Therapy Treatment Planning System. Class II

Reason: DICOM SOP Instance UID and Series Instance UID from RayGateway, i.e. the interface to Accuray's iDMS, are not guaranteed to be unique.

RAYSEARCH LABORATORIES AB 20250808 Ongoing
Distributed: Worldwide - US Nationwide distribution in the states of AL, CA, FL, GA, IL, IA, KY, LA, MD, MN, MS, NY, NC, OH, OK, SC, WA, WI and the countries of Canada, China, Colombia, France, Germany, Hong Kong S.A.R., China, Italy, Japan, Netherlands, Poland, Spain, Sweden, Switzerland, ENG, Thailand, Turkey, UK, Ukraine.
Page 357 of 4,030 | Showing 10 recalls
Data provided by openFDA