Smith & Nephew RENASYS-F XL with Soft Port REF #: 66800797 (The RENASYS Soft Port Dressing Kits are used in conjunction with negative pressure wound therapy pumps) Class II

Reason: Reports of wound fluid and /or blood that were not evacuated from beneath the wound dressing and the pump blockage alarm did not activate has been associated with inadequate Soft Port sealing at the wound site or improper positioning of the Soft Port orifice over the wound contact dressing drape. Inadequate sealing can result in air leaks and contribute to inefficient fluid evacuation from the wo

Smith & Nephew Inc. 20140109 Terminated
Distributed: Worldwide Distribution-USA (nationwide) including Puerto Rico and the states of AK, AL, AR, AZ, CA, CO, CT, FL, GA, ID, IL, KY, LA, MA, MS, MT, NC, NJ, NV, NY, OH, OR, PA, TN, TX, VA, WA, WI, WV, HI and the countries Dubai, Curacao, Argentina, Australia, Chile, Germany, France, England, Hong Kong, Kuwait, Lebanon, Libya, Sri Lanka, Malaysia, New Zealand, Manila, Saudi Arabia, South Africa, Singapore, Turkey, Taipei and Vietnam.
Smith & Nephew RENASYS-F Foam Dressing Kit with Soft Port REF #'s: 66800794, 66800795, and 66800796 (The RENASYS Soft Port Dressing Kits are used in conjunction with negative pressure wound therapy pumps) Class II

Reason: Reports of wound fluid and /or blood that were not evacuated from beneath the wound dressing and the pump blockage alarm did not activate has been associated with inadequate Soft Port sealing at the wound site or improper positioning of the Soft Port orifice over the wound contact dressing drape. Inadequate sealing can result in air leaks and contribute to inefficient fluid evacuation from the wo

Smith & Nephew Inc. 20140109 Terminated
Distributed: Worldwide Distribution-USA (nationwide) including Puerto Rico and the states of AK, AL, AR, AZ, CA, CO, CT, FL, GA, ID, IL, KY, LA, MA, MS, MT, NC, NJ, NV, NY, OH, OR, PA, TN, TX, VA, WA, WI, WV, HI and the countries Dubai, Curacao, Argentina, Australia, Chile, Germany, France, England, Hong Kong, Kuwait, Lebanon, Libya, Sri Lanka, Malaysia, New Zealand, Manila, Saudi Arabia, South Africa, Singapore, Turkey, Taipei and Vietnam.
Smith & Nephew RENASYS Soft Port REF: 66800799 (The RENASYS Soft Port Dressing Kits are used in conjunction with negative pressure wound therapy pumps) Class II

Reason: Reports of wound fluid and /or blood that were not evacuated from beneath the wound dressing and the pump blockage alarm did not activate has been associated with inadequate Soft Port sealing at the wound site or improper positioning of the Soft Port orifice over the wound contact dressing drape. Inadequate sealing can result in air leaks and contribute to inefficient fluid evacuation from the wo

Smith & Nephew Inc. 20140109 Terminated
Distributed: Worldwide Distribution-USA (nationwide) including Puerto Rico and the states of AK, AL, AR, AZ, CA, CO, CT, FL, GA, ID, IL, KY, LA, MA, MS, MT, NC, NJ, NV, NY, OH, OR, PA, TN, TX, VA, WA, WI, WV, HI and the countries Dubai, Curacao, Argentina, Australia, Chile, Germany, France, England, Hong Kong, Kuwait, Lebanon, Libya, Sri Lanka, Malaysia, New Zealand, Manila, Saudi Arabia, South Africa, Singapore, Turkey, Taipei and Vietnam.
Panaroma Central Station Product Usage: This device can view real time, store, print, graph, and trend patient clinical and demographic data. The device can also set independent alarm limits for data send by the bedside monitor. Class II

Reason: Mindray has identified two issues with the Panorama Central Station. Full Disclosure data may lag behind the Panorama system time clock when the Panorama Central Station is in use with Mindray¿s V-Series Monitor. Additionally, event indicators lag behind the Panorama system time clock when the Panorama Central Station is in use with a Panorama Telepack.

Mindray DS USA, Inc. dba Mindray North America 20140108 Terminated
Distributed: Worldwide Distribution - USA Nationwide Distribution
Philips HeartStart MRx Monitor/Defibrillator Model numbers: M3535A, M3536A, M3536M,M3536M4,M3536M6,861288,861289,861464,861483,and 861491 Class II

Reason: MRx may display a Red X in the Ready For Use (RFU) indicator during automated tests informing the user that a failure has been detected that may prevent the delivery of a shock and pacing

Philips Medical Systems, Inc. 20140218 Terminated
Distributed: Nationwide Foreign: AUSTRIA BAHRAIN BELGIUM BRAZIL CANADA CHINA EGYPT ESTONIA FINLAND FRANCE GERMANY GHANA ICELAND INDIA INDONESIA IRELAND ISRAEL ITALY JAPAN JORDAN KENYA KOREA, REPUBLIC OF LATVIA LEBANON LIBYAN ARAB JAMAHIRIYA LITHUANIA MALAYSIA MEXICO MOZAMBIQUE MYANMAR (Burma) NETHERLANDS NETHERLANDS ANTILLES NEW CALEDONIA NEW ZEALAND NORWAY PAKISTAN PERU POLAND PORTUGAL QATAR RUSSIAN FEDERATION SAUDI ARABIA SINGAPORE SOUTH AFRICA SPAIN SWEDEN SWITZERLAND TAIWAN, PROVINCE OF CHINA THAILAND TURKMENISTAN UNITED ARAB EMIRATES UNITED KINGDOM VIET NAM
RetCam 3, RetCam Portable and RetCam Shuttle; manufactured by Clarity Medical Systems, Inc.; Pleasanton, CA 94588. General ophthalmic imaging including retinal, corneal and external imaging. Class II

Reason: Software anomaly for RetCam 3, RetCam Shuttle and RetCam Portable with software versions 6.0.x, 6.1. x and 6.2x where the patients age is calculated incorrectly.

Clarity Medical Systems Inc 20140124 Terminated
Distributed: Worldwide Distribution-USA (nationwide) including DC and the states of NM, LA, CA, MI, PR, GA, MD, MT, AL, MA, NY, IL, OH, TX, VA, MI, FL, MO, KY, TN, CT, NJ, PA, UT, SD, OK, MS, and NC and the countries of Algeria, Australia, Austria, Bangladesh, Belarus, The Netherlands, Brazil, Bulgaria, Chile, China, Columbia, Czech Republic, Denmark, Ecuador, Egypt, Finland, France,India, Indonesia, Iran, United Kingdom, Israel, Italy, Kuwait, Lebanon, Malaysia, Mexico, Sweden, Pakistan, Peru, Philippines, Poland, Portugal, Saudi Arabia, Taiwan, Thailand, Turkey, UAE, Argentina, Uzbekistan, Oman, Canada, Germany, and Switzerland.
Treatment packs used on the VisuMax Laser Keratome. Size M Manufactured by Carl Zeiss Meditec AG, Jena, Germany. The VisuMax Laser Keratome is indicated for the following: In the creation of a corneal flap in patients undergoing LASIK surgery or other treatment requiring lamellar resection of the cornea. Patients undergoing surgery or other treatment requiring initial lamellar resection of the cornea. In the creation of a lamellar cut/resection of the cornea for lamellar keratoplasty. In the creation of a cut/incision for penetrating keratoplasty and corneal harvesting. Class II

Reason: Carl Zeiss Treatment pack, lot number M 130010, size M, may actually contain size S.

Carl Zeiss Meditec AG 20140203 Terminated
Distributed: US Distribution: NV only.
HER2 Image Analysis Manual Read of HER2 Digital Slides ER/PR Image Analysis IHC PR Breast Tissue Manual Read of Digital Slides The ScanScope¿¿ System is an automated digital slide creation, management, viewing and analysis system. It is intended for in vitro diagnostic use as an aid to the pathologist in the display, detection, counting and classification of tissues and cells of clinical interest based on particular color, intensity, size, pattern and shape. The IHC HER2 Image Analysis application is intended for use as an aid to the pathologist in the detection and semi-quantitative measurement of HER2/neu (c-erbB-2) in formalin-fixed, paraffin-embedded normal and neoplastic tissue. The IHC HER2 Image Analysis application is intended for use as an accessory to the Dako HercepTest" to aid in the detection and semi-quantitative measurement of Her2/neu (c-erbB-2) in formalin-fixed, paraffin-embedded normal and neoplastic tissue. When used with the Dako HercepTest", it is indicated for use as an aid in the assessment of breast cancer patients for whom HERCEPTIN¿¿(Trastuzumab) treatment is being considered. Note: The IHC HER2 Image Analysis application is an adjunctive computer-assisted methodology to assist the reproducibility of a qualified pathologist in the acquisition and measurement of images from microscope slides of breast cancer specimens stained for the presence of HER-2 receptor protein. The accuracy of the test result depends upon the quality of the immunohistochemical staining. It is the responsibility of a qualified pathologist to employ appropriate morphological studies and controls as specified in the instructions for the Dako HercepTest" to assure the validity of the IHC HER2 Image Analysis application assisted HER-2/neu score. The actual correlation of the Dako HercepTest" to Herceptin¿¿ clinical outcome has not been established. Manual Read of HER2 Digital Slides: The ScanScope¿¿ System is an automated digital slide creation, management, viewing and analysis system. It is intended for in vitro diagnostic use as an aid to the pathologist in the display, detection, counting and classification of tissues and cells of clinical interest based on particular color, intensity, size, pattern and shape. The IHC HER2 Manual Read of Digital Slides application is intended for use as an aid to the pathologist in the detection and semi-quantitative measurement of HER-2/neu (c-erbB-2) by manual examination of the digital slide of formalin-fixed, paraffin-embedded normal and neoplastic tissue immunohistochemically stained for HER-2 receptors on a computer monitor. HER-2 results are indicated for use as an aid in the management, prognosis and prediction of therapy outcomes of breast cancer. The IHC HER2 Manual Read of Digital Slides application is intended for use as an accessory to the Dako HercepTestT to aid the pathologist in the detection and semi-quantitative measurement of HER-2/neu (c-erbB-2) by manual examination of the digital slide of formalin-fixed, paraffin-embedded normal and neoplastic tissue immunohistochemically stained for HER-2 receptors on a computer monitor. When used with the Dako HercepTestT, it is indicated for use as an aid in the assessment of breast cancer patients for whom HERCEPTIN¿¿ (Trastuzumab) treatment is being considered. Note: The actual correlation of the Dako HercepTestT to Herceptin¿¿ clinical outcome has not been established. ER/PR Image Analysis: The ScanScope¿¿ XT System is an automated digital slide creation, management, viewing and analysis system. It is intended for in vitro diagnostic use as an aid to the pathologist in the display, detection, counting and classification of tissues and cells of clinical interest based on particular color, intensity, size, pattern and shape. The IHC ER Image Analysis application is intended for use as an aid to the pathologist in the detection and quantitative measurement of ER (Estrogen Receptor) in formalin-fixed paraffin-embedded normal and neo Class II

Reason: In an abundance of caution, Aperio (now known as Lecia Biosystems) is providing additional information to clarify that the use of Digital IHC Breast/ER/Dako Image Analysis, Digital IHC Breast/HER2 Digital Read, Digital IHC Breast/HER2/Dako Image Analysis, Digital IHC Breast/PR Digital Read and Digital IHC Breast/PR/Dako Image Analysis with the RUO products identified in that earlier notification i

Leica Biosystems Imaging, Inc. 20140207 Terminated
Distributed: US Distribution including the states of FL, WA, PA, MS, TX, MN CA, NY, NJ, NC, CO and OH.
Hand Control 12FR Suction Coagulator with 10 ft cord and holster, Sterile, Disposable, Model Number 0043-25, packaged Tyvek¿ polyester peel pouch, 25 pouches per box. This device is intended to conduct monopolar electrosurgical energy from an electrosurgical generator (ESU) to target tissue during ENT and general surgical procedures. It is also a suction apparatus used to remove surgical smoke and fluids from the site. The device is intended for single use; it is not intended to be cleaned or reused. Class II

Reason: Megadyne Medical Products, Inc. is voluntarily recalling Hand Control 12Fr Suction Coagulator with 10ft cord and 132405 Holster, Sterile, Disposable, Lot 132405 due to the potential for a small number of these suction coagulators to self-activate immediately on connection to the electrosurgical generator and to remain activated until it is disconnected.

Megadyne Medical Products, Inc. 20140227 Terminated
Distributed: USA including FL, MN, LA, NJ, NM, PA, TX and Internationally to Canada.
Mobile X-ray System, Model: MUX-100, MUX-100H Product Usage: The device is a mobile X-ray system in order to take general radiography of patients in the hospital who cannnot move and/or of outpatients in emergency Class II

Reason: Shimadzu Corporation is recalling the Shimadzu Mobile X-ray Sytems (MUX-100 and MUX-100H) because the connecting metal plate which connects the main chassis and the column of MUX-100 may crack.

Shimadzu Medical Systems 20131115 Terminated
Distributed: USA Nationwide Distribution