Philips HeartStart MRx Monitor/Defibrillator; Models: M3535A, M3536A, M3536J, M3536M, M3536M5, and M3536MC. Used for the termination of ventricular tachycardia and ventricular fibrillation. Class II

Reason: ECG trunk cable and connector block of the MRx could be susceptible to accelerated wear, which could result in an interrupted ECG signal

Philips Medical Systems, Inc. 20140218 Terminated
Distributed: Worldwide Distribution-USA (nationwide) and the countries of AFGHANISTAN ALBANIA ARGENTINA ARUBA AUSTRALIA AUSTRIA AZERBAIJAN BAHRAIN BANGLADESH BELGIUM BOLIVIA BOSNIA AND HERZEGOWINA BRAZIL BRUNEI DARUSSALAM BULGARIA CANADA CHILE CHINA COLOMBIA COSTA RICA CROATIA (local name: Hrvatska) CYPRUS CZECH REPUBLIC DENMARK ECUADOR EGYPT ESTONIA ETHIOPIA FINLAND FRANCE GERMANY GREECE GUADELOUPE HONG KONG ICELAND INDIA INDONESIA IRELAND ISRAEL ITALY JAPAN JORDAN KENYA KOREA, REPUBLIC OF KUWAIT LAOS LATVIA LEBANON LIBYAN ARAB JAMAHIRIYA LITHUANIA LUXEMBOURG MACAU MACEDONIA MALAYSIA MALTA MEXICO MOROCCO NAMIBIA NETHERLANDS NETHERLANDS ANTILLES NEW CALEDONIA NEW ZEALAND NICARAGUA NIGERIA NORWAY OMAN PAKISTAN PANAMA PAPUA NEW GUINEA PERU PHILIPPINES POLAND PORTUGAL PUERTO RICO QATAR REUNION ROMANIA RUSSIAN FEDERATION SAUDI ARABIA SINGAPORE SLOVENIA SOUTH AFRICA SPAIN SRI LANKA SWEDEN SWITZERLAND TAIWAN, PROVINCE OF CHINA TANZANIA, UNITED REPUBLIC OF THAILAND TRINIDAD AND TOBAGO TUNISIA TURKEY UGANDA UKRAINE UNITED ARAB EMIRATES UNITED KINGDOM UZBEKISTAN VENEZUELA VIET NAM YEMEN.
XL CONVALESCENT RECLINER 5291/529S with Padded armrests with guard, Dual fold down side-tables, Removable side panels, Built-in head-rest. The products are a multi position recliners with a steel frame used primarily in long term care or convalescent facilities. The recliners are suitable as general purpose seating for relaxation, reclining, eating, patient transport within the facility, and as a patient room chair for visitors. Class II

Reason: A supplier manufacturing assembly defect with the right hand recline mechanism. Use of this product may result in the loss of functionality of the three (3) position lock bar when in a reclined position.

Winco Mfg., LLC 20130906 Terminated
Distributed: Worldwide Distribution - USA nationwide including Alabama, Arkansas, Arizona, California, Colorado, Connecticut, Florida, Georgia, Idaho, Illinois, Indiana, Kanas, Kentucky, Louisiana, Massachusetts, Maryland, Maine, Michigan, Minnesota, Missouri, Mississippi, North Carolina, North Dakota, New Jersey, New Mexico, New York, Ohio, Oklahoma, Pennsylvania, South Carolina, South Dakota, Tennessee, Texas, Virginia, Vermont, Washington, Wisconsin, West Virginia, Puerto Rico and Internationally to the country of Canada.
DROP-ARM CONVALESCENT RECLINER 5271/5281 with One-touch release for drop-arms, Built-in headrest, Padded push-handle, Three-position lock, Over-lap tray. The products are a multi position recliners with a steel frame used primarily in long term care or convalescent facilities. The recliners are suitable as general purpose seating for relaxation, reclining, eating, patient transport within the facility, and as a patient room chair for visitors. Class II

Reason: A supplier manufacturing assembly defect with the right hand recline mechanism. Use of this product may result in the loss of functionality of the three (3) position lock bar when in a reclined position.

Winco Mfg., LLC 20130906 Terminated
Distributed: Worldwide Distribution - USA nationwide including Alabama, Arkansas, Arizona, California, Colorado, Connecticut, Florida, Georgia, Idaho, Illinois, Indiana, Kanas, Kentucky, Louisiana, Massachusetts, Maryland, Maine, Michigan, Minnesota, Missouri, Mississippi, North Carolina, North Dakota, New Jersey, New Mexico, New York, Ohio, Oklahoma, Pennsylvania, South Carolina, South Dakota, Tennessee, Texas, Virginia, Vermont, Washington, Wisconsin, West Virginia, Puerto Rico and Internationally to the country of Canada.
CONVALESCENT RECLINER 5251/5261, 525S/526S with Side Storage Pocket, Upholstered side-panels, Recessed hand-grips, Webbed suspension seating, Non-porous tray. The products are a multi position recliners with a steel frame used primarily in long term care or convalescent facilities. The recliners are suitable as general purpose seating for relaxation, reclining, eating, patient transport within the facility, and as a patient room chair for visitors. Class II

Reason: A supplier manufacturing assembly defect with the right hand recline mechanism. Use of this product may result in the loss of functionality of the three (3) position lock bar when in a reclined position.

Winco Mfg., LLC 20130906 Terminated
Distributed: Worldwide Distribution - USA nationwide including Alabama, Arkansas, Arizona, California, Colorado, Connecticut, Florida, Georgia, Idaho, Illinois, Indiana, Kanas, Kentucky, Louisiana, Massachusetts, Maryland, Maine, Michigan, Minnesota, Missouri, Mississippi, North Carolina, North Dakota, New Jersey, New Mexico, New York, Ohio, Oklahoma, Pennsylvania, South Carolina, South Dakota, Tennessee, Texas, Virginia, Vermont, Washington, Wisconsin, West Virginia, Puerto Rico and Internationally to the country of Canada.
PREMIER CARE RECLINER 5570 with Urethane armrests, Multiple side-table options, 3 position lock-bar, Fully upholstered, OPTIONAL Heat or Heat & Massage, OPTIONAL LiquiCell. The products are a multi position recliners with a steel frame used primarily in long term care or convalescent facilities. The recliners are suitable as general purpose seating for relaxation, reclining, eating, patient transport within the facility, and as a patient room chair for visitors. Class II

Reason: A supplier manufacturing assembly defect with the right hand recline mechanism. Use of this product may result in the loss of functionality of the three (3) position lock bar when in a reclined position.

Winco Mfg., LLC 20130906 Terminated
Distributed: Worldwide Distribution - USA nationwide including Alabama, Arkansas, Arizona, California, Colorado, Connecticut, Florida, Georgia, Idaho, Illinois, Indiana, Kanas, Kentucky, Louisiana, Massachusetts, Maryland, Maine, Michigan, Minnesota, Missouri, Mississippi, North Carolina, North Dakota, New Jersey, New Mexico, New York, Ohio, Oklahoma, Pennsylvania, South Carolina, South Dakota, Tennessee, Texas, Virginia, Vermont, Washington, Wisconsin, West Virginia, Puerto Rico and Internationally to the country of Canada.
Oxoid Legionella BCYE Growth Supplement, For 500 ml of medium, SR0110C, packaged 10 vials/box, lyophilized. The firm name on the label is Oxoid Ltd., Basingstoke, Hants, England. The product is used for the isolation of Legionella species from clinical and environmental samples. When the product is added to CYE Agar base, it stabilizes pH and provides essential growth factors for the isolation of the Legionella spp. Class II

Reason: The product may contain high levels of microbial contamination

Remel Inc 20140124 Terminated
Distributed: US Distribution including Puerto Rico and the states of FL, IL, MA, ME, MI, and PA.
cobas¿ PCR Urine Kit 100 PKT IVD cobas¿ PCR Female Swab Kit 100 PKT IVD cobas¿ PCR media 100T IVD cobas¿ PCR Female Swab Kit 100 PKT JPN-IVD Roche Molecular Systems, Inc. Branchburg, NJ 08876 USA Roche Diagnostics 9115 Hague Road Indianapolis, IN 46250-0457 The cobas¿¿ PCR media serves as a nucleic acid stabilizing transport and storage medium for gynecological specimens and is contained in the cobas¿¿ PCR Female Swab Kit, cobas¿¿ PCR Urine kit, and cobas¿¿ PCR media 100 T kit. -The cobas¿¿ PCR Urine Sample Kit is used to collect and transport urine specimens. Use this collection kit only with the cobas¿¿ CT/NG Test. - The cobas¿¿ PCR Female Swab Sample Kit is used to collect and transport endocervical and vaginal swab specimens. Use this collection kit only with the cobas¿¿ CT/NG Test NOTE: This collection kit should not be used for collection of female urine specimens. o For endocervical swab specimens, collections are to be performed by a clinician. o For vaginal swab specimens, collections are to be performed by a clinician or by the patient (self-collection) -In US labeling, it indicates Self collection in a clinical setting. - In ex-US labeling, it indicates Clinician Instructed Self- Collection and Ensure the patient has read and understood the following self-collection instructions before providing a collection kit. - In some countries (e.g., UK), the sample collection kit is provided as self-collection for home / personal use, which may not be used in a controlled clinical setting and it is unknown if any instructions are provided. - The cobas¿¿ PCR media kit is used to stabilize and transport urine specimens. Use this kit only with the cobas¿¿ CT/NG Test. Class II

Reason: Numerous complaints have been filed that leaking cobas ¿ PCR media 4.3 mL IVD have been received from different lots of the cobas ¿ PCR Urine kits, cobas ¿ PCR Female Swab kits, and cobas ¿ PCR media kits.

Roche Molecular Systems, Inc. 20140106 Terminated
Distributed: Nationwide Distribution
Philips HeartStart MRx Monitor/Defibrillator used Model numbers: M3535A, M3536A, M3536J, M3536M, M3536MC Class II

Reason: HeartStart MRx Defibrillator Monitor therapy connection could experience accelerated wear causing risk for patients and/or caregivers

Philips Medical Systems, Inc. 20140204 Terminated
Distributed: Worldwide Distribution: USA (nationwide) and to countries of: AFGHANISTAN ALBANIA ARGENTINA ARUBA AUSTRALIA AUSTRIA AZERBAIJAN BAHRAIN BANGLADESH BELGIUM BOLIVIA BOSNIA AND HERZEGOWINA BRAZIL BRUNEI DARUSSALAM BULGARIA CAMBODIA CHILE CHINA COLOMBIA COSTA RICA CROATIA (local name: Hrvatska) CYPRUS CZECH REPUBLIC DENMARK ECUADOR EGYPT EL SALVADOR ESTONIA ETHIOPIA FINLAND France GERMANY GHANA GREECE GUADELOUPE HONG KONG ICELAND INDIA INDONESIA IRAQ IRELAND ISRAEL ITALY JAPAN JORDAN KAZAKHSTAN KENYA KOREA, REPUBLIC OF KUWAIT LAOS¿ LATVIA LEBANON LIBYAN ARAB JAMAHIRIYA LITHUANIA LUXEMBOURG MACEDONIA MALAYSIA MEXICO MONGOLIA MOROCCO MOZAMBIQUE MYANMAR (Burma)¿ NAMIBIA NEPAL NETHERLANDS NETHERLANDS ANTILLES NEW CALEDONIA NEW ZEALAND NICARAGUA NORWAY OMAN PAKISTAN PALESTINE PANAMA PAPUA NEW GUINEA PARAGUAY PERU PHILIPPINES POLAND PORTUGAL PUERTO RICO QATAR REUNION ROMANIA RUSSIAN FEDERATION SAUDI ARABIA SERBIA SINGAPORE SLOVENIA SOUTH AFRICA SPAIN SRI LANKA SWEDEN SWITZERLAND TAIWAN, PROVINCE OF CHINA TANZANIA, UNITED REPUBLIC OF THAILAND TRINIDAD AND TOBAGO TUNISIA TURKEY TURKMENISTAN UGANDA UKRAINE UNITED ARAB EMIRATES UNITED KINGDOM UZBEKISTAN VENEZUELA VIET NAM YEMEN
Philips HeartStart XL, Defibrillator/Monitor Model number M4735A Product Usage: The M4735A HeartStart XL Defibrillator/Monitor is for use in the hospital by qualified medical personnel trained in the operation of the device and qualified by training in basic life support, advanced cardiac life support, or defibrillation. It must be used by or on the order of a physician. When operating as a semi-automatic external defibrillator in AED Mode, the M4735A HeartStart XL Defibrillator/Monitor is suitable for use by medical personnel trained in basic life support that includes the use of an AED Class II

Reason: Philips HeartStart XL may experience a power board failure causing failure to defibrillate

Philips Medical Systems, Inc. 20140214 Terminated
Distributed: Worldwide Distribution and USA Nationwide and the countries of AUSTRALIA, BANGLADESH, BELGIUM, BRAZIL, CANADA, CHINA, COLOMBIA, ETHIOPIA, FINLAND, FRANCE, GEORGIA, GERMANY, HUNGARY, INDIA, INDONESIA, IRELAND, ITALY, JAPAN, KAZAKHSTAN, KOREA, REPUBLIC OF, MALAYSIA, MEXICO, MONACO, NEPAL, NETHERLANDS, PORTUGAL, SINGAPORE, SOUTH AFRICA, SPAIN, SRI LANKA, SWITZERLAND, TAIWAN, PROVINCE OF CHINA, THAILAND, UNITED KINGDOM, UZBEKISTAN, VENEZUELA, and VIET NAM,
M3536A HeartStart MRx with Wireless Link Upgrade (861289) The HeartStart MRx is for use for the termination of ventricular tachycardia and ventricular fibrillation. The device is for use by qualified medical personnel trained in the operation of the device and qualified by training in basic life support, advanced cardiac support, or defibrillation. It must be used by or on the order of a physician. Class II

Reason: Wireless Link may be unable to transmit data when configured for a specific data flow.

Philips Medical Systems, Inc. 20140218 Terminated
Distributed: Worldwide Distribution - USA including AZ, CA, FL, IL, IN, KS, KY, MA, MD, MI, MN, MO, NC, NY, OH, OK, OR, PA, SC, TN, TX,and VA and Internationally to Australia, Norway, and Spain.
Page 3,619 of 4,030 | Showing 10 recalls