ISIS HVT Tracheal Tube Cuffed with Subglottic Secretion Suction Port, 7.0, Rx only, Product Usage: A tracheal tube is a device inserted into a patients trachea via the nose or mouth and used to maintain an open airway. Class I

Reason: The tracheal tube can kink during patient use.

Teleflex Medical 20140106 Terminated
Distributed: Worldwide Distribution - USA (nationwide) Puerto Rico and Internationally to Canada, Barbados, Buenos Aires, Bolivia, El Salvador, Mexico, Columbia, Chile, Costa Rica, Dominican Republic, Ecuador, Barbados, Belgium, South Korea, Thailand, New Zealand, Netherlands.
ISIS HVT Tracheal Tube Cuffed with Subglottic Secretion Suction Port, 6.5, Rx only, Product Usage: A tracheal tube is a device inserted into a patients trachea via the nose or mouth and used to maintain an open airway. Class I

Reason: The tracheal tube can kink during patient use.

Teleflex Medical 20140106 Terminated
Distributed: Worldwide Distribution - USA (nationwide) Puerto Rico and Internationally to Canada, Barbados, Buenos Aires, Bolivia, El Salvador, Mexico, Columbia, Chile, Costa Rica, Dominican Republic, Ecuador, Barbados, Belgium, South Korea, Thailand, New Zealand, Netherlands.
ISIS HVT Tracheal Tube Cuffed with Subglottic Secretion Suction Port, 6.0, Rx only, Product Usage: A tracheal tube is a device inserted into a patients trachea via the nose or mouth and used to maintain an open airway. Class I

Reason: The tracheal tube can kink during patient use.

Teleflex Medical 20140106 Terminated
Distributed: Worldwide Distribution - USA (nationwide) Puerto Rico and Internationally to Canada, Barbados, Buenos Aires, Bolivia, El Salvador, Mexico, Columbia, Chile, Costa Rica, Dominican Republic, Ecuador, Barbados, Belgium, South Korea, Thailand, New Zealand, Netherlands.
Gelzone Shoulder Sleeve, Catalog Number SLV634, Lot Number 852766. Product Usage: To be used where limb support and/or elevation is considered necessary. Class II

Reason: Implantech initiated the voluntary recall of Gelzone Shoulder Sleever, SLV-634, Lot Number 852766 because it may not satisfy customer requirements for the robustness of the Velcro closure.

Implantech Associates Inc 20140129 Terminated
Distributed: Worldwide Distribution - USA (nationwide) in the states of: CA, AZ, IL, NC, CO and the country of: South Africa.
Bausch & Lomb AKREOS AO Micro Incision Lens The Akreos IOL intended for primary implantation in the capsular bag of the eye for the visual correction of aphakia secondary to the removal of a cataractous lens in adult patients. Class II

Reason: Lens was manufactured with incorrect raw material.

Bausch & Lomb Surgical, Inc. 20131016 Terminated
Distributed: Worldwide Distribution: US (nationwide) and Internationally to: Great Britain, France, Spain, Portugal, Sweden, Russia and Guadeloupe.
Synergy XVI The Elekta Synergy S is intended to be used for radiation therapy treatment of malignant neoplastic diseases, as determined by a licensed medical practitioner. Class II

Reason: Potential collision risk when using XVI and external beam shaping devices.

Elekta, Inc. 20131125 Terminated
Distributed: Worldwide Distribution: US (nationwide) including Puerto Rico and the states of: AL, AK, AZ, CA, CO, CT, FL, GA, IL, IA, KY, MD, MA, MI, MN, MS, MT, NE, NV, NJ, NM, NY, NC, OH, OK, OR, PA, SC, TN, TX, VT, VA, WA, WV, WI, DC, and Internationally to: Algeria, Australia, Austria, Belarus, Belgium, Brazil, Canada, Chile, China, Cyprus, Czech Republic, France, Germany, Greece, Hong Kong, India, Israel, Italy, Japan, Malaysia, Mexico, Pakistan, Portugal, Russia, Singapore, Slovakia, South Korea, Spain, Taiwan, Thailand, and United Kingdom.
Synergy XVI XVI can incorrectly calculate the target position of the treatment table, because of an error that does not occur very frequently. This is the sequence of events when this error occurs - (1) You do an online VolumeView acquisition and registration. (2) You accept the registration and the Table Move Assistant dialog box appears. (3) The Relative Actual column of the Table Move Assistant dialog box incorrectly appears blank. (4) When this occurs, XVI does not read the Absolute Actual position of the treatment table. This causes XVI to calculate the treatment table moves from the zero position of the treatment table (+0.0, + 0.0, + 0.0). (5) In XVI R5.0, if the Relative Set values are less than or equal to + or -2.5cm (for earlier releases of XVI, the value is + or -2cm), you can do remote automatic table movement (RATM) from the function keypad (FKP). (6) When you press the Table ASU (automatic setup) buttons, the treatment table moves to a position that was incorrectly calculated from XVI. Class II

Reason: XVI can incorrectly calculate the target position of the treatment table.

Elekta, Inc. 20131125 Terminated
Distributed: Worldwide Distribution - USA (nationwide) including Puerto Rico and the states of AL, AK, AZ, AR, CA, CT, FL, GA, ID, IL, KY, LA, MD, MA, MI, MN, MO, NH, NJ, NM, NY, NC, ND, OH, OK, OR, PA, RI, SD, TN, TX, VT, VA, WA, WV, WI, WY and DC., and the countries of Algeria, Australia, Austria, Bahrain, Belarus, Belgium, Brazil, Bulgaria, Canada, China, Columbia, Croatia, Cuba, Cyprus, Czech Republic, Denmark, Ecuador, Estonia, Finland, France, Germany, Greece, Hong Kong, Hungary, India, Iraq, Ireland, Israel, Italy, Japan, Jordan, Malaysia, Mexico, Morocco, Netherlands, New Zealand, Norway, Pakistan, Poland, Portugal, Romania, Russia, Saudi Arabia, Singapore, Slovenia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, United Kingdom and Venezuela
External Implant RBT, D3 Threadform 4.0 X 10.5mm, REF 40105D3, STERILE R, Rx Only, BioHorizons(R), 2300 Riverchase Center, Birmingham, AL 35244 Usage: dental Class II

Reason: The body length of the External Implant 40105D3 Lot 1302762 labeled as 10.5mm in length is in fact 12mm in length.

BioHorizons Implant Systems Inc 20131210 Terminated
Distributed: Worldwide distribution: US (nationwide) to states of: LA, CA, NC, IL, PA, and NY; and Internationally to: Seoul, Korea; New Delhi, India; Taipei City, Taiwan; and Madrid, Spain.
Synthes Driving Cap/Threaded Both parts are intended for use in the Tibia system, LFN and RAFN systems. Their function is to insert a nail. The Synthes Driving Cap/Threaded is threaded into the insertion handle and secured with a wrench. If desired, the nail can then be inserted using light hammer blows. Class II

Reason: The Synthes Driving Cap/Threaded can bind in the IM Nail Radiolucent Insertion Handles when assembled becoming difficult to separate.

Synthes, Inc. 20140107 Terminated
Distributed: USA (nationwide) Distribution.
CartoUnivu Module, Model: Carto 3 Software versions 3.2.2 and 3.2.3; Catalog No. KT-5400-124 WITH THE CARTOUNIVU MODULE, FLUOROSCOPIC CAPTURES (IMAGES AND CINE CLIPS) CAN BE IMPORTED TO AND DISPLAYED IN THE CART0 3 SYSTEM. THESE IMAGES ARE NOT FOR DIAGNOSTIC PURPOSES. WHEN FLUOROSCOPIC CAPTURES ARE INTEGRATED INTO THE CART0 3 SYSTEM MAP VIEWERS (MAP VIEWER AND ADDITIONAL VIEW), CART0 3 MAPS, MRl OR CT IMAGES, AND CATHETER VISUALIZATION ARE SUPERIMPOSED ON THE FLUOROSCOPIC CAPTURE. Class II

Reason: Biosense Webster has initiated a recall of the CartoUnivu Module within the Carto 3 System (V3.2.2 and V3.2.3) when used with Siemens Axiom Artis VB Fluoroscopy systems in particular procedural configurations only. If the table rotation is not returned to zero point, there is a misalignment between the Carto 3 System map display and the fluoroscopic capture.

Biosense Webster, Inc. 20140212 Terminated
Distributed: Worldwide Distribution - USA including CA, KY, NJ, MA, and UT and Internationally to Belgium, Austria, and Germany.