Hemochron Jr. Citrate APTT Cuvette 510(k) K014008 A unitized microcoagulation test intended to be used in performing a quantitative, one -stage APTT. The citrate APTT test is used for evaluation of low doses of heparin anticoagulant (up to 1.5 units/mL), depending upon individual patient heparin sensitivity. Class II

Reason: ITC has determined that some Citrate APTT cuvettes may recover higher than expected results in normal individuals.

ITC-Nexus Dx 20131219 Terminated
Distributed: Worldwide Distribution: US (nationwide) and countries of: Austria, Belgium, Canada, Serbia, Germany, Estonia, Italy, Spain and Saudia Arabia.
The Symbionic Leg is an integrated prosthetic leg for transfemoral and knee disarticulation amputees. The prosthesis is composed of the RHEO KNEE and the PROPRIO FOOT, both already existing products from ¿ssur. Both are a battery powered micro-processor controlled prosthetic device and are connected through a shared battery. The knee and the foot have different sensor systems that provide real-time information used to control knee joint resistance and the flexion/extension of the ankle during the gait cycle. Default control parameters are provided with the device but can be manually adjusted by a certified prosthetist through software (knee), or by a certified prosthetist or the user through controls on the device itself (foot). SYMBIONIC LEG is intended for continuous use by amputees of low to moderate impact levels that weigh less than 125 kg for foot categories 25  30 and less than 100kg for foot Categories 22, 23 and 24. The system is not intended for athletic use. Class II

Reason: OSSUR initiated this recall of the Symbionic Leg because it may have a possible defect in a small batch of electronics.

Ossur Americas, Inc. 20120212 Terminated
Distributed: US Distribution including the states of AL, DC, GA, NY, NJ, OK and PA.
LOCATOR¿ Implant Attachment System. The LOCATOR¿ Implant Attachment System is a universal hinge, resilient attachment for endosseous implants. The LOCATOR¿ Implant Attachment System is designed for use with overdentures or partial dentures, retained in whole or in part, by endosseous implants in the mandible or maxilla. FDA 510(k): K994257, K072878 Class II

Reason: Zest Anchors, LLC is voluntarily recalling specific lots of LOCATOR¿ Implant Attachment Systems because it is currently provided with a label that indicates that the final torque applied to the abutment should be 30 N-cm. It was recently noted by Zest that in order to provide an optimal performance by increasing the factor of safety between the failure load of the abutment screw and the recommend

Zest Anchors LLC 20140116 Terminated
Distributed: Distributed in Ohio and Switzerland.
3.5mm Locking Cortical Bone Screw, T15 Hexalobe, Self Tapping, 20mm Part No. 00-0903-2620 OrthoPediatrics Pediloc" Tibia Plates are indicated for fractures, osteotomies, and non-unions of the pediatric and small stature adult tibia. The subject recalled device is a Bone Screw designed for use with this system. Class II

Reason: OrthoPediatrics has initiated a Voluntary RECALL of Part # 00-0903-2620, 3.5MM Locking Cortical Screw, Lot # 009SV because the device within the package may be incorrect. These packages incorrectly contain a 3.5mm Non Locking Cortical Screw # 00-1050-3524.

OrthoPediatrics Corp 20140108 Terminated
Distributed: Worldwide Distribution - USA including OK, KY, FL, LA, OH, CO, MO, MN and Internationally to Ireland, Saudi Arabia, Italy, Turkey, South Africa, and Germany.
The Gaymar Medi-Therm device provides a means of regulating patient temperature by supplying temperature-controlled water through a connector hose to accessory Gaymar Hyper/Hypothermia blanket(s)/body wrap(s). The blanket/body wrap provides an interface for heating or cooling the patient. Accessory YSI 400 series patient probe interfaces between the Medi-Therm and patient to sense patient temperature, which is displayed on the device's control panel. The Medi-Therm device controls output water temperature by mixing hot and cold water using hot and cold solenoid valves under microprocessor control. The device includes a circulating pump, heater and refrigeration system. The Medi-Therm controller, connector hose, blanket/body wrap and patient probe comprise the Medi-Therm Hyper/Hypothermia System. Class II

Reason: Stryker Medical has identified that on some units of the MTA7900, there is a lack of a water temperature display while the unit is in Auto mode. While in Auto mode, only the patient temperature is displayed along with the set point. The water display will be blank during Auto mode. If the user wants to see the water temperature displayed and switches out of Auto mode to Manual mode, then back to

Stryker Medical Division of Stryker Corporation 20131227 Terminated
Distributed: nationwide and worldwide
Diagnostic imaging systems (GEMINI LXL Model #882412, TruFlight Select Model #882438, GEMINI TF PET/CT16 Model #882470, GEMINI TF PET/CT 64 Model #882471, GEMINI TF Ready Model #882473 & GEMINI TF Big Bore Model #882476) Class II

Reason: The PET Reconstruction Server (PRS) database may randomly lock up after the low dose CT acquisition is complete and will not allow the PET acquisition to begin, resulting in an unexpected incomplete study.

Philips Medical Systems (Cleveland) Inc 20131204 Terminated
Distributed: Product was shipped to the following states: AL, AR, AZ, CA, CO, DC, FL, GA, HI, IA, IL, IN, KY, LA, MA, MD, MI, MN, MO, MS, NC, ND, NE, NJ, NV, NY, NY , OH, OR, PA, TN, TX, VA, VT, WA & WI. Product was also shipped to the following countries: Australia, Belgium, Brazil, Bulgaria, Canada, China, Croatia, Cuba, Denmark, Egypt, France, Germany, Hong Kong, Hungary, India, Indonesia, Italy, Japan, Kuwait, Lebanon, Malta, Mexico, Netherlands, Pakistan, Philippines, Poland, Portugal, Republic of Korea, Russia, Russian Federation, Saudi Arabia, Singapore, Slovakia, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, Ukraine, Ukraine, United Kingdom, Venezuela & Vietnam.
Trifurcated Extension Set, model no. ME 1224. Trifurcated Extension Set, model no. MP 9230. Class II

Reason: CareFusion is recalling the Trifurcated Extension Set, models ME 1224 and MP 9230, because they may have a defect in the trifurcated component that could result in a component leak.

CareFusion 303, Inc. 20140116 Terminated
Distributed: Nationwide in US and Canada
ADVIA Centaur¿ Systems Phenytoin Calibrator N ( PHTN) 6-pack SMN 10314977 For in vitro diagnostic use in the quantitative determination of phenytoin in serum or plasma using the ADVIA Centaur and ADVIA Centaur XP systems Class II

Reason: ADVIA Centaur¿ Systems Phenytoin Calibrator N, lots CN58 and higher, change in recovery observed with the ADVIA Centaur Systems Phenytoin assay compared to recoveries using previous Calibrator N lots.

Siemens Healthcare Diagnostics, Inc 20131218 Terminated
Distributed: Worldwide Distribution - USA (nationwide) and Internationally to Argentina, Australia,Brazil,Canada, Columbia, China, Ecuador, Egypt, India, Japan, Korea, Mexico, Saudi Arabia,Singapore, Thailand, and Uruguay.
ADVIA Centaur¿ Systems Phenytoin Calibrator N ( PHTN) 2-pack SMN 10311391 For in vitro diagnostic use in the quantitative determination of phenytoin in serum or plasma using the ADVIA Centaur and ADVIA Centaur XP systems Class II

Reason: ADVIA Centaur¿ Systems Phenytoin Calibrator N, lots CN58 and higher, change in recovery observed with the ADVIA Centaur Systems Phenytoin assay compared to recoveries using previous Calibrator N lots.

Siemens Healthcare Diagnostics, Inc 20131218 Terminated
Distributed: Worldwide Distribution - USA (nationwide) and Internationally to Argentina, Australia,Brazil,Canada, Columbia, China, Ecuador, Egypt, India, Japan, Korea, Mexico, Saudi Arabia,Singapore, Thailand, and Uruguay.
XiO RTP System. Used to create treatment plans for any cancer patient for whom external beam radiation therapy or brachytherapy has been prescribed. Class II

Reason: When bolus is present, the Effective Depth (with bolus) to the weight point returned on the Source Data Report is incorrect.

Elekta, Inc. 20131125 Terminated
Distributed: Worldwide Distribution-USA (nationwide) including DC and Puerto Rico and the states of AL, AK, AZ, CA, CO, CT, DE, FL, GA, IL, IN, KS, KY, LA, ME, MD, MI, MN, MS, MO, MT, NE, NV, NJ, NM, NY, NC, ND, OH, OK, OR, PA, RI, SC, SD, TN, TX, QA, WA, WV, WI, and WY, and the countries of Albania, Algeria, Argentina, Austria, Australia, Bosnia, Bulgaria, Belgium, Bahrain, Brazil, Belarus, Botswana, Bangladesh, Canada, China, Columbia, Costa Rica, Croatia, Cuba, Cyprus, Chile, Czech Republic, Ecuador, Egypt, Estonia, France, Finland, Gabon, Germany, Great Britain, Greece, Guatemala, Hungary, India, Indonesia, Italy, Ireland, Israel, Iraq, Jordan, Japan, Kosovo, Kuwait, Lithuania, Libya, Latvia, Morocco, Malaysia, Malta, Mexico, Myanmar, Netherland, New Zealand, Pakistan, Peru, Philippines, Poland, Portugal, Romania, Russia, Saudi Arabia, Serbia, Singapore, Slovenia, Slovakia, South Africa, South Korea, Spain, Sudan, Syria, Taiwan, Thailand, Turkey, Turkmenistan, Ukraine, Venezuela, and Vietnam.