Atrium Ocean Water Seal Chest Drain ATS Blood Recovery System with stopcock and in-line connectors Sterile fluid path packaging Product Code: 2050-070. To evacuate air and/or fluid from the chest cavity or mediastinum. Class II

Reason: Chest Drain tubing of of the ATS Blood recovery may leak or disconnect

Atrium Medical Corporation 20131119 Terminated
Distributed: Worldwide Distribution-USA (nationwide) and the countries of Australia, Canada, Austria, Ireland, Romania, South Africa, Colombia, Lebanon Netherlands, Jordan, Russia, and Taiwan.
Atrium Ocean Water Seal Chest Drain ATS (Autotransfusion) Blood Recovery System with stopcock and in-line connectors; Product Code: 2050-000. To evacuate air and/or fluid from the chest cavity or mediastinum. Class II

Reason: Chest Drain tubing of of the ATS Blood recovery may leak or disconnect

Atrium Medical Corporation 20131119 Terminated
Distributed: Worldwide Distribution-USA (nationwide) and the countries of Australia, Canada, Austria, Ireland, Romania, South Africa, Colombia, Lebanon Netherlands, Jordan, Russia, and Taiwan.
Oxoid Antimicrobial Susceptibility Testing Disc, MEM10, REF: CT0774B, 10 mcg meropenem, Packed in packs containing 5 cartridges per pack, 50 discs per cartridge. Firm name on the label is Oxoid, Ltd., Basingstoke, United Kingdom. Used in the semi-quantitative agar diffusion test method for in vitro susceptibility testing to assist in the determination of a suitable antibiotic agent for treatment purposes. Class II

Reason: Individual discs in the lot may not be sufficiently impregnated with antibiotic, resulting in smaller than expected zone sizes.

Remel Inc 20131121 Terminated
Distributed: Nationwide Distribution including the states of AL, AZ, CA, CO, FL, HI, IA, IL, KY, LA, MD, MI, MN, MO, MS, NC, NY, OH, SC, TN, TX, VA, WI, and WV
Consult Diagnostics¿ Eon Calcium Reagent For In-Vitro Diagnostic Use PN: 1429 For the quantitative measurement of calcium in plasma and serum Class II

Reason: High positive calcium bias on plasma sample versus results on serum

Vital Diagnostics, Inc. 20130829 Terminated
Distributed: Nationwide Distribution
Vital Diagnostics Envoy Calcium Reagent For In-Vitro Diagnostic Use. PN 55285 For the quantitative measurement of calcium in plasma and serum Class II

Reason: High positive calcium bias on plasma sample versus results on serum

Vital Diagnostics, Inc. 20130829 Terminated
Distributed: Nationwide Distribution
Centricity PACS System versions 3.X and higher and 4.X and higher Product Usage: The Centricity PACS Workstation is intended for use as a primary diagnostic and analysis tool for diagnostic images by trained healthcare professionals, including radiologists, physicians, technologists, clinicians and nurses. It is also intended for use as a clinical review workstation throughout the healthcare facility. the workstation interface provides the user with a means to display, manipulate, archive, print, and export images when connected with the Centricity PACS infrastructure. Class II

Reason: There is a potential loss of image(s) associated with the Centricity to Centricity (C2C) exam transfer module of Centricity PACS System versions 3.X and higher and 4.X and higher when transferring from the source to a destination server. When another process in the destination server attempts to access the same object or table, the transfer process of a particular image may be terminated. Once ter

Ge Healthcare It 20120915 Terminated
Distributed: Worldwide Distribution - US: Nationwide (AK, AL, AR, AZ, CA, CO, CT, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WV, WY); FOREIGN: Australia, Austria, Belgium, Bermuda, China, Denmark, Egypt, France, Germany, Hong Kong, India, Ireland, Israel, Italy, Japan, Korea, Kuwait, Malaysia, Malta, Netherlands, Portugal, Saudi Arabia, Singapore, South Africa, Spain, Sweden, Switzerland, Taiwan, Turkey, United Kingdom.
Palacos¿ R+G Radioplaque Bone Cement 1 x 40 g Single (with Gentamicin), 40.8 g Methyl Acrylate Copolymer, 20 ml Methyl Methacrylate Monomer 0.5 g Gentamicin, Distributed by Zimmer Dover, OH 44522 The cement is indicated for use in the second stage of a two stage revision for total joint arthroplasty after the initial infection has been cleared. Class II

Reason: The patient label associated with the product may be incorrect in that the lot number on the patient label may be different than the lot number on the outer carton label.

Zimmer, Inc. 20131104 Terminated
Distributed: Nationwide Distribution including AZ, CA, CO, CT, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, ME, MI, MN, MO, NC, NJ, NV, NY, OH, OK, OR, PA, TN, TX, UT, VA, VT, WA, WI, WV & WY.
Anspach Carbide Cutting Burrs & Drums: 2MM Carbide Fluted Ball, Extended part # S-2B-C and 7MM Carbide Fluted Ball part # S-7B-C Anspach cutting burrs are intended for cutting and shaping bone, including the spine and cranium Class II

Reason: Some carbide cutting burr fractured during use.

The Anspach Effort, Inc. 20131217 Terminated
Distributed: Worldwide distribution: US (Nationwide) including state of Hawaii and Internationally to: Australia, Poland, Canada, South Africa, Israel, and Switzerland.
Anspach Carbide Cutting Burrs & Drums: 1.5MM x 7.5MM Cylindrical Carbide Drum part # S-15DRMC Anspach cutting burrs are intended for cutting and shaping bone, including the spine and cranium Class II

Reason: Some carbide cutting burr fractured during use.

The Anspach Effort, Inc. 20131217 Terminated
Distributed: Worldwide distribution: US (Nationwide) including state of Hawaii and Internationally to: Australia, Poland, Canada, South Africa, Israel, and Switzerland.
Anspach Carbide Cutting Burrs: 1.5MMx7.5MM Cylindrical Carbide Drum part # S-15DRMC-L, 1MM Carbide Fluted ball, Extended part # S-1B-C, 1MM Carbide Fluted ball part # S-SB-C, 2MM Carbide Fluted ball part # S-2SB-C, 3MM Carbide Fluted ball Extended part # S-3B-C, 3MM Carbide Fluted ball part # S-3SB-C, 4MM Carbide Fluted ball part # S-4B-C, 4MM Carbide Fluted ball Extended 27MM part # S-4BL-C, 5MM Carbide Fluted ball part # S-5B-C, and 6MM Carbide Fluted ball part # S-6B-C. Anspach cutting burrs are intended for cutting and shaping bone, including the spine and cranium Class II

Reason: Some carbide cutting burr fractured during use.

The Anspach Effort, Inc. 20131217 Terminated
Distributed: Worldwide distribution: US (Nationwide) including state of Hawaii and Internationally to: Australia, Poland, Canada, South Africa, Israel, and Switzerland.
Page 3,656 of 4,030 | Showing 10 recalls