Locking Cortical Screw 3.5mmX38mm REF 8161-35-038 Class II

Reason: 3.5mmx38mm Locking Cortical Screw, implants are being recalled since an investigation identified that a 3.5mmx28mm Cortical Locking Screw is labeled as a 3.5mmx38mm Cortical Locking Screw.

Biomet, Inc. 20131219 Terminated
Distributed: US Distribution including states of: AZ, CA, GA, IL, KY, MI, NC, NH, NY, OH, PA, SD, TX, UT, and VA.
Anspach MCA Cutting Burr, DFU Rx Only Designed for cutting and shaping delicate bone, primarily in otology procedures Class II

Reason: Current directions for use for the MCA cutting burs (item 18-0001) do not provide enough details regarding the intended use specifically for the MCA burrs.

The Anspach Effort, Inc. 20120329 Terminated
Distributed: Worldwide Distribution - USA (nationwide) and Internationally to Great Britain, Australia, Switzerland, Sweden, France, Japan, Spain, Italy, Canada, and Netherlands.
Draeger Infinity Acute Care System Monitoring Solution Class II

Reason: After a test in which a low SpO2 (yellow) medium grade alarm initiated in neo-natal mode and was subsequently paused, the high grade life threatening SpO2 alarm (red) did not activate when the SpO2 values were decreased to reach the red alarm limit.

Draeger Medical Systems, Inc. 20131210 Terminated
Distributed: US Distribution including the states of: MA, NH, IN, CA, TX, NY, OK, and IL.
pO2 D999 membrane units. Distributed by Radiometer America, Inc., Westlake, OH 44145 pO2 membrane units, ordering number 942-042 Intended use is in vitro testing of samples of whole blood for pO2 (partial pressure of oxygen). Class II

Reason: RADIOMETER became aware that the regarded lot pO2 units 942-042R0971 were produced with inner solution for pCO2 membrane units. This solution is a salt solution similar to the pO2 inner solution, but also contains glycerol which can be reduced at the cathode causing a suddenly rising zero current, causing the membrane unit to fail the following Cal2 calibration.

Radiometer America Inc 20130610 Terminated
Distributed: Worldwide Distribution - US (nationwide) including states of: CO, CT, GA, MA, MO & MS and country of: Canada.
Boston Scientific Extractor Pro XL Retrieval Balloon Catheter 12-15 mm Injects Below UPN :M00547140 Class II

Reason: Extractor Pro XL Retrieval Balloon Catheters may have the incorrect skive hole location or balloon size

Boston Scientific Corporation 20131218 Terminated
Distributed: Worldwide Distribution - USA (Nationwide Distribution and Puerto Rico) and Internationally.
Boston Scientific Extractor Pro XL Retrieval Balloon Catheter 12-15 mm Injects Above UPN: M00547110 Class II

Reason: Extractor Pro XL Retrieval Balloon Catheters may have the incorrect skive hole location or balloon size

Boston Scientific Corporation 20131218 Terminated
Distributed: Worldwide Distribution - USA (Nationwide Distribution and Puerto Rico) and Internationally.
Boston Scientific Extractor Pro XL Retrieval Balloon Catheter 9-12 mm Injects Above UPN: M00547100 The Extractor Pro XL Retrieval Balloon Catheter is used endoscopically to 1) remove stones from the biliary system; or 2) to facilitate injection of contrast medium while occluding the duct with the balloon. Class II

Reason: Extractor Pro XL Retrieval Balloon Catheters may have the incorrect skive hole location or balloon size

Boston Scientific Corporation 20131218 Terminated
Distributed: Worldwide Distribution - USA (Nationwide Distribution and Puerto Rico) and Internationally.
King Systems KLTSD414 Quantity, 5 PER CASE KING LTSD KIT SIZE 4 King LT(S)-DTM supraglottic airway Identification Part Number(s): KLTSD414 KING LTSD KIT/KIT/TROUSSE/KIT SIZE 4 Contents: 1 - KLTSD Size 4......(155 - 180cm) .. 1 - 60cc Syringe 1 - Sterile Lubricant 32-5332 05/10 Intended for use in adult patients (in excess of 25 kg) for controlled ventilation during anesthesia for procedures that are short in duration. Class II

Reason: KLTD214 airways found inside KLTSD414 packages. The KLTSD414 airway has a lumen for use of a gastric suction tube; the KLTD214 airway does not include a lumen for a gastric suction tube.

King Systems Corp. 20131213 Terminated
Distributed: Distributed in the states of CA, NY
Multiples Anspach Micro Curved Attachments (MCA) and Curved Burr Support S leves (Cutters and Attachments) MCA - MICRO CURVED ATTACHMENT, MCA5-05SD - 0.5MM DIAMOND BALL, 5.6CM, MCA5-06SD - 0.6MM DIAMOND BALL, 5.6CM, MCA5-07SD - 0.7MM DIAMOND BALL, 5.6CM, MCA5-08SD - 0.8MM DIAMOND BALL, 5.6CM, MCA5-15SB - 1.5MM FLUTED BALL, 5.6CM, MCA5-15SD - 1.5MM DIAMOND BALL, 5.6CM, MCA5-1SB - lMM FLUTED BALL, 5.6CM, MCA5-1SD - lMM DIAMOND BALL, 5.6CM, MCA5-2SB - 2MM FLUTED BALL, 5.6CM, MCA5-2SD - 2MM DIAMOND BALL, 5.6CM, MCA7-05SD - 0.5MM DIAMOND BALL, 7.3CM, MCA7-06SD - 0.6MM DIAMOND BALL, 7 .3CM, MCA7-07SD - 0.7MM DIAMOND BALL, 7.3CM, MCA7-08SD - 0.8MM DIAMOND BALL, 7.3CM, MCA7-15SB - 1.5MM FLUTED BALL, 7.3CM, MCA7-15SD  1.5MM DIAMOND BALL, 7.3CM, MCA7-1SB - lMM FLUTED BALL, 7.3CM, MCA7-1SD - 1 MM DIAMOND BALL, 7 .3CM, MCA7-2SB - 2MM FLUTED BALL, 7.3CM and MCA7-2SD - 2MM DIAMOND BALL, 7.3CM. The MCA attachment is a reusable device used with dissection tools. Dissection tools are single use, disposable cutter designed for cutting and shaping delicate bone, primary in otology procedures for delicate bone cutting. Class II

Reason: During a recent review of documentation Anspach determined that the current directions for use for the MCA cutting burrs and MCA Micro Curved Attachment do not provide clear details regarding the intended use specifically for the MCA burrs. The MCA burrs are intended for use in delicate bone in otologic procedures such as cochleostomy. The use of the device in dense bone removal may result in cu

The Anspach Effort, Inc. 20121026 Terminated
Distributed: Worldwide Distribution - USA (nationwide) and Internationally to Canada, Netherlands, Great Britain, France, Sweeden, Italy, Spain, Japan, South Africa, Australia, Switzerland,
17 cm (7") Smallbore Trifuse Ext Set w/3 NanoClave¿ ( Light Blue, Lavender Rings), 3 Anti-Syphon Valves, Rotating Luer, Item #011-A1096, Lot # 2796668. The ICU Needleless Connector is a single use, sterile, non-pyrogenic device intended for use as an accessory to intravascular administration set for the administration of fluids to a patient through a cannula placed in the vein or artery. Class II

Reason: ICU Medical is recalling certain lots of the NanoClave Needlefree Connector product line because of a small percentage of devices that may leak.

ICU Medical, Inc. 20131224 Terminated
Distributed: Worldwide Distribution - USA including UT, TX, GA, WI, VT, NH, AL, OH, TN, IL, CA, KY,HI, NC, and Internationally to CA, BE, IN, KW, and AU.