Broncho Flexible Fiberscope, models: 11001BN1, 11002BD1, 11004BC1, and 11009BC1. " Examination and treatment of the tracheobronchial tree Contraindications for Bronchoscopy " Uncontrolled angina, uncontrolled heart failure, or serious uncontrolled ventricular arrhythmias " Uncontrolled hypertension " Bleeding disorder or use of anticoagulants. Class II

Reason: KARL STORZ Endoscopy America, Inc. is recalling the Flexible Broncho Fiberscope or a Flexible Intubation Fiberscope with a suction port because the material used in the suction channel portion was thermoplastic polyurethane (TPU), which is not listed in STERRAD systems claims for interior endoscope lumens.

Karl Storz Endoscopy America Inc 20131113 Terminated
Distributed: Nationwide Distribution.
Intubation Flexible Fiberscope, Models: 11301AA1, 11301BN1, 11302BD1, 11302BD2, 11304BC1, 11340BC1, 11301BND1, 11302BDD1, and 11302BDD2. " Transoral intubation " Transnasal intubation " Evaluation of placement of endotracheal and double lumen endotracheal tubes " Provide visual access to the larynx and tracheobronchial tree. Class II

Reason: KARL STORZ Endoscopy America, Inc. is recalling the Flexible Broncho Fiberscope or a Flexible Intubation Fiberscope with a suction port because the material used in the suction channel portion was thermoplastic polyurethane (TPU), which is not listed in STERRAD systems claims for interior endoscope lumens.

Karl Storz Endoscopy America Inc 20131113 Terminated
Distributed: Nationwide Distribution.
VITROS CA 125 II Calibrators For use in the calibration of the VITROS ECi/ECiQ Immunodiagnostic Systems, VITROS 3600 Immunodiagnostic System and VITROS 5600 Integrated System for the quantitative measurement of OC 125 defined antigen in human serum and plasma (EDTA or heparin). Class II

Reason: Ortho Clinical Diagnostics issued a Product Correction Notification for VITROS Immunodiagnostic Products CA 125 II Reagent Pack and Calibrators due to a positive shift in patient results.

Ortho-Clinical Diagnostics 20131009 Terminated
Distributed: Worldwide Distribution: US (nationwide) including states of: AK, CA, CT, CO, DC, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, ME, MI, MN, MO, MS, MT, NC, NE, NJ, NM, NY, OH, OK, OR, PA, SC, TN, TX, VA, VT, WA, WI, WV, WY, and Puerto Rico, and countries of: Australia, Brazil, Canada, Chile, China, Colombia, India, Japan, Mexico, Republic of Panama, Singapore, Venezuela, England, France, Germany, Italy & Spain.
InstaTrak¿ 3500 Plus, ENTrakTM Plus, InstaTrak¿ 3500, ENTrakTM Computer-Assisted Navigation Systems (USA), Class II

Reason: GE OEC Medical Systems Inc is recalling InstaTrak¿ 3500 Plus, ENTrakTM Plus, InstaTrak¿ 3500, ENTrakTM Computer-Assisted Navigation Systems (USA) due to several potential safety issues related to the use of these products.

GE OEC Medical Systems, Inc 20131126 Terminated
Distributed: Nationwide Distribution
outer box: Ref 906768, CurvTek Eye Needle 22 MM X-large Stainless Steel, Single Use Gamma Sterilized, Rx inner incorrect label: CurvTek 7 MM Medium, Single Pack Quantity -3, REF 906760 To pass the CurvTek needle, place the jaws of the instrument utilized to grasp the needle on the sides of the needle and orient and needle parallel to the long axis of the instrument. Position the needle to follow the path of the tunnel created by the cartridges. This aligns the needle with the radius of the transosseous tunnel. Advance the needle while rotating the instrument away from the entry point. Gently feed the needle through the length of the tunnel. Class II

Reason: Inner label of PN: 906768 Lot: 055450 CurvTek Eye Needle 22mm X-Large states incorrectly PN: 906760 7mm Medium.

Biomet, Inc. 20131126 Terminated
Distributed: Nationwide Distribution including MI, IA, IL, IN, and TN.
KayserBetten model IDA beds Product Usage: BED, AC-POWERED ADJUSTABLE HOSPITAL Class II

Reason: Some joints on the doors of a limited number of IDA beds may loosen or separate over time which can cause the doors to drop outward and cause injury.

mobility unlimited inc 20130411 Terminated
Distributed: Worldwide Distribution - USA Nationwide in the states of AL, AZ, CA, FL, GA, IL, MN, MO, NY, PA, TN, TX, VA, and Canada.
Large Needle Driver, 8MM assembly of the da Vinci S IS1200 and da Vinci Si, IS2000. Manufactured by Intuitive Surgical, Sunnyvale, CA. EndoWrist Instruments, including scissors, scalpels, forceps, needle drivers and electrocautery are intended for endoscopic manipulation of tissue, including: grasping, cutting, blunt and sharp dissection, approximation, ligation, electrocautery and suturing. Mega Needle Driver and Large Needle Driver instruments are part of the EndoWrist instrument family and are intended for use in endoscopic manipulation and suturing. Class II

Reason: Firm has become aware of the potential for detachment of the jaw insert for Large Needle Drivers and Mega Needle Drivers.

Intuitive Surgical, Inc. 20131115 Terminated
Distributed: Worldwide Distribution - USA (nationwide and Puerto Rico) and Internationally to Finland, Belgium, Netherlands, UK, Czech Republic, Sweden, Germany, Austria, Switzerland, Norway, Denmark, South Korea, France, Ireland, and Slovakia.
Mega Needle Driver, 8MM assembly of the da Vinci S IS1200 and da Vinci Si, IS2000. Manufactured by Intuitive Surgical, Sunnyvale, CA. EndoWrist Instruments, including scissors, scalpels, forceps, needle drivers and electrocautery are intended for endoscopic manipulation of tissue, including: grasping, cutting, blunt and sharp dissection, approximation, ligation, electrocautery and suturing. Mega Needle Driver and Large Needle Driver instruments are part of the EndoWrist instrument family and are intended for use in endoscopic manipulation and suturing. Class II

Reason: Firm has become aware of the potential for detachment of the jaw insert for Large Needle Drivers and Mega Needle Drivers.

Intuitive Surgical, Inc. 20131115 Terminated
Distributed: Worldwide Distribution - USA (nationwide and Puerto Rico) and Internationally to Finland, Belgium, Netherlands, UK, Czech Republic, Sweden, Germany, Austria, Switzerland, Norway, Denmark, South Korea, France, Ireland, and Slovakia.
Access Immunoassay Systems Reaction Vessels. The Access Immunoassay System is an in vitro diagnostic device used for the quantitative, semi-quantitative or qualitative determination of various analyte concentrations found in human body fluids. Class II

Reason: Beckman Coulter is initiating a recall for the Access Immunoassay Systems Reaction Vessels, including the Acess/Access 2 Immunoassay Systems, SYNCRON LXi 725 Clinical System, and UniCel DxC 600i SYNCHRON Access Clinical System because Beckman Coulter has observed an intermittent increase in Access System "Wash Carousel Motion Errors" with specific lots of Access System Reaction Vessels (RVs). I

Beckman Coulter Inc. 20131002 Terminated
Distributed: Worldwide Distribution - USA (nationwide) and Internationally to Canada, Japan, India, Hong Kong, Guatemala, China, Brazil, Brunei Darussalam, Costa Rica, El Salvador, JordanMalaysia, Mexico, Panama, Philippines, South Africa, Taiwan, Thailand, Trinidad and Tobago, United Arab Emirates, Venezuela, and Viet Nam.
Kimberly-Clark Single Shot Epidural Pain Management Trays containing the BD LOR Syringe (BD Code 405198) Product Usage: The device is designed to provide regional anesthesia. The device may contain syringes, needles and drugs Class II

Reason: Kimberly-Clark Single Shot Epidural Pain Management Trays contain the BD 7 mL Epilor plastic Luer-Lok LOR Syringes under recall by Becton Dickinson (BD) because the product may stall or stick when traveling within the barrel of the syringe.

Kimberly-Clark Corporation 20131024 Terminated
Distributed: U.S. Nationwide Distribution