GENESIS(R) II, PS HIGH FLEXION ARTICULAR INSERT, SIZE 3-4, 9 MM QTY: (1), REF 71421508, STERILE EO, Smith & Nephew, Inc. Memphis, TN 38116 USA Orthopedic Class II

Reason: The Tyvek(TM) lid on one, or both, of the two trays that form a double barrier around the insert may not have been adequately sealed.

Smith & Nephew Inc 20130827 Terminated
Distributed: Worldwide Distribution - USA including Nevada and Internationally to United Arab Emirates, Italy, Spain, Belgium, Bulgaria, Poland, France, Japan, Netherlands, Australia, Turkey, Russian Federation, Germany and United Kingdom.
GENESIS(R) II, PS HIGH FLEXION ARTICULAR INSERT, SIZE 3-4, 13 MM QTY: (1), REF 71420820, STERILE EO, Smith & Nephew, Inc. Memphis, TN 38116 USA Orthopedic Class II

Reason: The Tyvek(TM) lid on one, or both, of the two trays that form a double barrier around the insert may not have been adequately sealed.

Smith & Nephew Inc 20130827 Terminated
Distributed: Worldwide Distribution - USA including Nevada and Internationally to United Arab Emirates, Italy, Spain, Belgium, Bulgaria, Poland, France, Japan, Netherlands, Australia, Turkey, Russian Federation, Germany and United Kingdom.
GENESIS(R) II, POSTERIOR STABILIZED ARTICULAR INSERT, SIZE 1-2, 13 MM QTY: (1), REF 71420806, STERILE EO, Smith & Nephew, Inc. Memphis, TN 38116 USA Orthopedic Class II

Reason: The Tyvek(TM) lid on one, or both, of the two trays that form a double barrier around the insert may not have been adequately sealed.

Smith & Nephew Inc 20130827 Terminated
Distributed: Worldwide Distribution - USA including Nevada and Internationally to United Arab Emirates, Italy, Spain, Belgium, Bulgaria, Poland, France, Japan, Netherlands, Australia, Turkey, Russian Federation, Germany and United Kingdom.
Blanketrol III Data Export Software Version 1.10 The Blanketrol III Data Export Software is intended for use in the electronic transfer of medical device data from the Blanketrol III to a computer, without controlling or altering the function or parameters of the Blanketrol III. Class II

Reason: During a customer inquiry regarding a beta version of the Bill Data Export Software, Model USB-127, it was discovered that the wrong software version was released to a customer.

Cincinnati Sub-Zero Products Inc 20130515 Terminated
Distributed: Worldwide Distribution - US Distribution including the following states AK, CO, FL, MA, MI, and NC. and Internationally to Canada, Germany and United Kingdom.
Navigator HD l3F/15F x 46cm Ureteral Access Sheath Set Product No M0062502280, RDF Catalog M0062502290, Sterile. Indicated for use in endoscopic procedures to facilitate the passage of endoscopes, urological instruments and for the injection of fluids into the urinary tract. Class II

Reason: The 13F/15F Navigator HD Ureteral Access SheathSet are being recalled after receipt of 5 reports where the tip of the dilator has separated fro the body of the dilator.

Boston Scientific Corporation 20131105 Terminated
Distributed: Worldwide Distribution-USA (nationwide) including Puerto Rico and the countries of Australia, Ireland, Denmark, Italy, France, Canada, Great Britain, Switzerland, South Africa, Germany, and the Netherlands.
Navigator HD l3F/15F x 36cm Ureteral Access Sheath Set Product No M0062502280, RDF Catalog No 250-228, Sterile. Indicated for use in endoscopic procedures to facilitate the passage of endoscopes, urological instruments and for the injection of fluids into the urinary tract. Class II

Reason: The 13F/15F Navigator HD Ureteral Access SheathSet are being recalled after receipt of 5 reports where the tip of the dilator has separated fro the body of the dilator.

Boston Scientific Corporation 20131105 Terminated
Distributed: Worldwide Distribution-USA (nationwide) including Puerto Rico and the countries of Australia, Ireland, Denmark, Italy, France, Canada, Great Britain, Switzerland, South Africa, Germany, and the Netherlands.
Navigator HD l3F/15F x 28 cm Ureteral Access Sheath Set Product No M0062502270, RDF Catalog No 250-227, Sterile. Indicated for use in endoscopic procedures to facilitate the passage of endoscopes, urological instruments and for the injection of fluids into the urinary tract. Class II

Reason: The 13F/15F Navigator HD Ureteral Access SheathSet are being recalled after receipt of 5 reports where the tip of the dilator has separated fro the body of the dilator.

Boston Scientific Corporation 20131105 Terminated
Distributed: Worldwide Distribution-USA (nationwide) including Puerto Rico and the countries of Australia, Ireland, Denmark, Italy, France, Canada, Great Britain, Switzerland, South Africa, Germany, and the Netherlands.
C-Series: Clinac, Trilogy, Trilogy Tx. , Novalis high energy linear accelerators and UNIQUE single energy linear accelerator; Versions 7, 8 and 9. The UNIQUE is not sold in the US. Product Usage: The Varian High Energy Linear Accelerator is indicated for stereotactic radiosurgery and precision radiotherapy for lesions, tumors and conditions anywhere in the body when radiation treatment is indicated. The UNIQUE is a single energy medical linear accelerator. UNIQUE is indicated for precision radiotherapy for lesions, tumors and conditions anywhere in the body where radiation treatment is indicated. Class II

Reason: This correction is to notify users that a solution to a previous correction has been developed and Varians reps will be contacting locations to schedule installation.

Varian Medical Systems, Inc. 20131024 Terminated
Distributed: Worldwide Distribution - USA Nationwide and in the countries of Algeria, Angola, Argentina, Australia, Austria, Azerbaijan, Bangladesh, Belgium, Belorussia, Bosnia and Herzegovina, Brazil, Bulgaria, Canada, Chile, China, Colombia, Costa Rica, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, Estonia, Finland, France, French Polynesia, Georgia, Germany, Greece, Guadeloupe, Guam, Guatemala, Hong Kong, Hungary, India, Indonesia, Iraq, Ireland, Israel, Italy, Japan, Kazakhstan, Kenya, Kuwait, Latvia, Lebanon, Uthuania, luxembourg, Macau, Macedonia, Malaysia, Martinique, Mauritania, Mexico, Moldova, Monaco, Morocco, Netherlands, New Zealand, Norway, Oman, Pakistan, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Puerto Rico, Qatar, Reunion, Romania, Russia, Saudi Arabia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, SriLanka, Sweden, Switzerland, Taiwan, Thailand, Trinidad and Tobago, Tunisia, Turkey, Ukraine, United Arab Emirates, United Kingdom, Venezuela, Vietnam, and Zimbabwe.
SIEMENS brand ARTISTE Linear Accelerator Linac systems with component: any photon unflat beam option (Multiple X feature) in combination with the IMRT or mARC option. A family of linear accelerator systems is to deliver X-ray photon and electron radiation for the therapeutic treatment of cancer. The ARTISTE/ONCOR systems are manufactured by: SIEMENS AG, MEDICAL SOLUTIONS, GERMANY and distributed SIEMENS Healthcare IM CR RO, CR Radiation Oncology, Martinez, CA Class II

Reason: A safety risk regarding a potentially existing dark current radiation phenomenon on the Linear Accelerator (LINAC) in combination with IMRT or mARC (rotational IMRT) treatments using unflat (flattening filter free) beams. In case there are many long periods of field shape changes in a treatment plan, the dose due to dark current radiation may sum up to a significant value.

Siemens Medical Solutions USA, Inc 20131023 Terminated
Distributed: Worldwide Distribution: USA (nationwide) including states of: FL, NC, NY, PA, TN, UT, and WI; and countries of: Angola, Brazil, Canada, P.R. China, Germany, India, Japan, Republic Korea, Poland, and Thailand.
SIEMENS brand ONCOR series Linear Accelerator Linac systems with component: any photon unflat beam option(Multiple X feature) in combination with the IMRT or mARC option. A family of linear accelerator systems is to deliver X-ray photon and electron radiation for the therapeutic treatment of cancer. The ARTISTE/ONCOR systems are manufactured by: SIEMENS AG, MEDICAL SOLUTIONS, GERMANY and distributed SIEMENS Healthcare IM CR RO, CR  Radiation Oncology, Martinez, CA. Class II

Reason: A safety risk regarding a potentially existing dark current radiation phenomenon on the Linear Accelerator (LINAC) in combination with IMRT or mARC (rotational IMRT) treatments using unflat (flattening filter free) beams. In case there are many long periods of field shape changes in a treatment plan, the dose due to dark current radiation may sum up to a significant value.

Siemens Medical Solutions USA, Inc 20131023 Terminated
Distributed: Worldwide Distribution: USA (nationwide) including states of: FL, NC, NY, PA, TN, UT, and WI; and countries of: Angola, Brazil, Canada, P.R. China, Germany, India, Japan, Republic Korea, Poland, and Thailand.