Reason: Reports of friction within certain instrument arms can interrupt smooth instrument motion. This can be felt by the surgeon as resistance in the movement of the master.In this situation, the instrument can stall momentarily and then suddenly catch-up to the master position if the surgeon pushes through the resistance.
Distributed: Worldwide Distribution: US (nationwide) and to countries of: Australia, Austria, Belgium, Brazil, Bulgaria, Canada, China, Columbia, Denmark, France, Germany, India, Israel, Italy, Japan, S. Korea, Luxembourg, Mexico, Monaco, Netherlands, Norway, Romania. Russian Federation, Singapore, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, and United Kingdom.Reason: CareFusion is recalling the VersaSafe Extension Intravascular Administration Set because it may have excessive adhesive which will prevent the male spin collar to operate as intended.
Distributed: Nationwide Distribution.Reason: Advanced Sterilization Products (ASP) is recalling the SteriTite containers because they do not have consistent data to show that sterility assurance level of 10(-6) is achieved during the half-cycle test runs on new, unprocessed containers.
Distributed: Worldwide Distribution - USA (nationwide) and the countries of Australia, Vietnam, Thailand, Malaysia, China, Korea (South), Mexico, Spain, Singapore, Hong Kong and Portugal.Reason: Under certain conditions the system may not perform as intended, causing the release of results to the Laboratory Information System that should have been held for manual review due to auto-verification rules or the delay/omission of result transmission to the LIS.
Distributed: Worldwide Distribution: USA (nationwide) and internationally to Canada.Reason: Complaints of damage to various models of Thunderbeat Hand Instruments during surgical procedures, including cracking, breaking, or deformation of the components of the probe tip or jaw.
Distributed: Worldwide Distribution - USA (nationwide) including Puerto Rico and Internationally to Canada.Reason: The delivered application software was an outdated version to the model of equipment.
Distributed: Nationwide Distribution including MN, NH, DE, OR, MA, TX, OH, FL, CA, WA, and SC.Reason: There is the potential for the vial volume of "in use" Vista calibrators, controls, or sample diluent to incorrectly revert to full volume when removed and reloaded. Dimension Vista(R) 500 and Dimension Vista(R) 1500 - Reloaded Vista Vials - limited to systems using software versions 3.4 and 3.5.1
Distributed: Worldwide Distribution: USA (nationwide) and iinternationally to Canada.Reason: The small extendible Soft Tissue Retractor was recalled due to locking nut (clamping sleeve) malfunction.
Distributed: Worldwide Distribution - USA (nationwide) and Internationally to Canada and Mexico.Reason: 3M is recalling specific lots of 3M ESPE Unitek Primary Stainless Steel Crowns - Series UR4 contain LL3 crowns which are incorrectly labeled as UR4 crowns.
Distributed: Worldwide Distribution - USA including IN, TX, PA, FL, NV, SC, and TN and Internationally to Canada and Saudi Arabia.Reason: Certain System 83 Plus 2 and System 83 Plus 9 units running the Windows 7 operating system were not fully validated prior to distribution.
Distributed: Worldwide Distribution: USA (nationwide) and Internationally to: Canada, Slovakia Republic, Sydney and Peru.