Assembly, Patient Side Manipulator on da Vinci Surgical Systems: da Vinci S Surgical System; da Vinci SI Surgical System da Vinci Si-e Surgical System; Manufactured by Intuitive Surgical Sunnyvale, CA 94086. The Intuitive Surgical Endoscopic Instrument Control System (Intuitive Surgical da Vinci Surgical System Model IS2000/3000) is intended to assist in the accurate control of Intuitive Surgical Endoscopic Instruments including rigid endoscopes, blunt and sharp endoscopic dissectors, scissors, scalpels, ultrasonic shears, forceps/pick-ups, needle holders, endoscopic retractors, stabilizers, electro cautery and accessories for endoscopic manipulation of tissue, including grasping, cutting, blunt and sharp dissection, approximation, ligation, electro cautery, suturing, and delivery and placement of microwave and cryogenic ablation probes and accessories, during urologic surgical procedures, general laparoscopic surgical procedures, gynecologic laparoscopic surgical procedures, general cardiovascular and non-cardiovascular thoracoscopic surgical procedures and thoracoscopically-assisted cardiotomy procedures. The system is indicated for adult and pediatric use Class II

Reason: Reports of friction within certain instrument arms can interrupt smooth instrument motion. This can be felt by the surgeon as resistance in the movement of the master.In this situation, the instrument can stall momentarily and then suddenly catch-up to the master position if the surgeon pushes through the resistance.

Intuitive Surgical, Inc. 20131111 Terminated
Distributed: Worldwide Distribution: US (nationwide) and to countries of: Australia, Austria, Belgium, Brazil, Bulgaria, Canada, China, Columbia, Denmark, France, Germany, India, Israel, Italy, Japan, S. Korea, Luxembourg, Mexico, Monaco, Netherlands, Norway, Romania. Russian Federation, Singapore, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, and United Kingdom.
VersaSafe Extension Set, Model No. 21000M-007 Class II

Reason: CareFusion is recalling the VersaSafe Extension Intravascular Administration Set because it may have excessive adhesive which will prevent the male spin collar to operate as intended.

CareFusion 303, Inc. 20131108 Terminated
Distributed: Nationwide Distribution.
STERITITE Containers: STERITITE Container Kit, 1/4 size, 3" high, product code: 13902; STERITITE Mini Mid Size, 3" high, product code: 13903; STERITITE Container Kit, 1/2 size, 4" high, product code: 13904; STERITITE Container Kit, 3/4 size, 4" high, product code: 13905; STERITITE Container Kit, Full size, 4" high, product code: 13906. The ASP STERITITE sealed container system provides an efficient and cost-effective way to package and protect surgical instruments for sterilization, transportation, storage, and aseptic presentation of contents. Class III

Reason: Advanced Sterilization Products (ASP) is recalling the SteriTite containers because they do not have consistent data to show that sterility assurance level of 10(-6) is achieved during the half-cycle test runs on new, unprocessed containers.

Advanced Sterilization Products 20131003 Terminated
Distributed: Worldwide Distribution - USA (nationwide) and the countries of Australia, Vietnam, Thailand, Malaysia, China, Korea (South), Mexico, Spain, Singapore, Hong Kong and Portugal.
syngo(R) Lab Data Manager - Version VA11B and VA12A Systems Securely consolidates data from multiple lab instruments, the LIS and Siemens StreamLAB(R) Automation Solutions. It provides autoverification, integrated QC capabilities, browser-based connectivity supporting up to 25 concurrent users, remote services to help keep systems running optimally and also has audit trail capabilities. Class II

Reason: Under certain conditions the system may not perform as intended, causing the release of results to the Laboratory Information System that should have been held for manual review due to auto-verification rules or the delay/omission of result transmission to the LIS.

Siemens Healthcare Diagnostics, Inc. 20130510 Terminated
Distributed: Worldwide Distribution: USA (nationwide) and internationally to Canada.
TB-0545PC: THUNDERBEAT 5 mm , 45 em, Pistol Grip TB-0535PC: THUNDERBEAT 5 mm, 35 em, Pi stol Grip TB-0545IC: THUNDERBEAT 5 mm, 45 em, In line Grip TB-0535IC: THUNDERBEAT 5 mm, 35 em, Inline Grip TB-0520IC: THUNDERBEAT 5 mm, 20 em, In line Grip TB-051 OIC: THUNDERBEAT 5 mm, I 0 em, In line Grip TB-545FC: THUNDERBEAT 5 mm, 45 em, Front Actuated Grip TB-0535FC: THUNDERBEAT 5 mm, 35 em, Front Actuated Grip The Olympus Thunderbeat Surgical Tissue Management System ("Thunderbeat") is intended to be used for open, laparoscopic (including single-site surgery), and endoscopic surgery to cut (dissect) or coagulate soft tissue or to ligate (seal and cut) vessels. Class II

Reason: Complaints of damage to various models of Thunderbeat Hand Instruments during surgical procedures, including cracking, breaking, or deformation of the components of the probe tip or jaw.

Olympus America Inc. 20130927 Terminated
Distributed: Worldwide Distribution - USA (nationwide) including Puerto Rico and Internationally to Canada.
Easy Diagnost Eleva Universal Fluoroscopic applications Class II

Reason: The delivered application software was an outdated version to the model of equipment.

Philips Medical Systems, Inc. 20131030 Terminated
Distributed: Nationwide Distribution including MN, NH, DE, OR, MA, TX, OH, FL, CA, WA, and SC.
Dimension Vista(R) 500 and Dimension Vista(R) 1500 - Reloaded Vista Vials - limited to systems using software versions 3.4 and 3.5.1. Diagnostic device intended to determine a variety of analytes in human body fluids. Class II

Reason: There is the potential for the vial volume of "in use" Vista calibrators, controls, or sample diluent to incorrectly revert to full volume when removed and reloaded. Dimension Vista(R) 500 and Dimension Vista(R) 1500 - Reloaded Vista Vials - limited to systems using software versions 3.4 and 3.5.1

Siemens Healthcare Diagnostics, Inc. 20130520 Terminated
Distributed: Worldwide Distribution: USA (nationwide) and iinternationally to Canada.
Synthes Soft Tissue Retractor Small Extendible. Offset blade facilitates easy preparation of the epipereosteal cavity for percutaneous plate insertion. Class II

Reason: The small extendible Soft Tissue Retractor was recalled due to locking nut (clamping sleeve) malfunction.

Synthes USA HQ, Inc. 20130801 Terminated
Distributed: Worldwide Distribution - USA (nationwide) and Internationally to Canada and Mexico.
3M ESPE, Unitek Stainless Steel Permanent Molar Crowns Stainless Steel Crowns, are designed to fill the need for long term provisional coverage on adult molars and bicuspids as well as pediatric posterior and anterior teeth. Class III

Reason: 3M is recalling specific lots of 3M ESPE Unitek Primary Stainless Steel Crowns - Series UR4 contain LL3 crowns which are incorrectly labeled as UR4 crowns.

3M Company - Health Care Business 20130605 Terminated
Distributed: Worldwide Distribution - USA including IN, TX, PA, FL, NV, SC, and TN and Internationally to Canada and Saudi Arabia.
Custom Ultrasonics System 83 Plus 2 and Plus 9 Endoscope Washer/Disinfector. Class II

Reason: Certain System 83 Plus 2 and System 83 Plus 9 units running the Windows 7 operating system were not fully validated prior to distribution.

Custom Ultrasonics, Inc. 20130312 Terminated
Distributed: Worldwide Distribution: USA (nationwide) and Internationally to: Canada, Slovakia Republic, Sydney and Peru.