SIGMA¿ LCS¿ HIGH PERFORMANCE" MBT KEEL PUNCH NONCEMENTED Size 1-1.5 HP M.B.T. Keel Punch Knee Instrument is used to prepare the proximal tibial canal geometry to accept the final or definitive tibial tray used during total knee arthroplasty. Class II

Reason: DePuy Orthopaedics, Inc. is issuing a voluntary device correction of the HP M.B.T. Keel Punch Knee Instrument because of the potential for tabs on the connection end to fracture and be left in the patient.

DePuy Orthopaedics, Inc. 20131118 Terminated
Distributed: Worldwide Distribution - USA (nationwide) including AR,AZ,CA,CO,FL,GA,HI,IA,IL,IN,KY,LA,MA,MD,ME,MI,MN,MT,NC,NV,NY,OH,PA, TX,VA,WA,WI and Internationally to: AUSTRALIA AUSTRIA BELGIUM BRAZIL CANADA CZECH REPUBLIC DENMARK EGYPT FRANCE GERMANY GREECE HOLLAND HUNGARY INDIA IRELAND ISRAEL ITALY JAPAN KOREA MALAYSIA NEW ZEALAND NORWAY P R CHINA PAKISTAN PANAMA POLAND PORTUGAL PUERTO RICO RUSSIA SAUDI ARABIA SINGAPORE SLOVENIA SOUTH AFRICA SPAIN SWEDEN SWITZERLAND TURKEY UAE UK
SIGMA¿ LCS¿ HIGH PERFORMANCE" MBT KEEL PUNCH CEMENTED Size 4-7 HP M.B.T. Keel Punch Knee Instrument is used to prepare the proximal tibial canal geometry to accept the final or definitive tibial tray used during total knee arthroplasty. Class II

Reason: DePuy Orthopaedics, Inc. is issuing a voluntary device correction of the HP M.B.T. Keel Punch Knee Instrument because of the potential for tabs on the connection end to fracture and be left in the patient.

DePuy Orthopaedics, Inc. 20131118 Terminated
Distributed: Worldwide Distribution - USA (nationwide) including AR,AZ,CA,CO,FL,GA,HI,IA,IL,IN,KY,LA,MA,MD,ME,MI,MN,MT,NC,NV,NY,OH,PA, TX,VA,WA,WI and Internationally to: AUSTRALIA AUSTRIA BELGIUM BRAZIL CANADA CZECH REPUBLIC DENMARK EGYPT FRANCE GERMANY GREECE HOLLAND HUNGARY INDIA IRELAND ISRAEL ITALY JAPAN KOREA MALAYSIA NEW ZEALAND NORWAY P R CHINA PAKISTAN PANAMA POLAND PORTUGAL PUERTO RICO RUSSIA SAUDI ARABIA SINGAPORE SLOVENIA SOUTH AFRICA SPAIN SWEDEN SWITZERLAND TURKEY UAE UK
SIGMA¿ LCS¿ HIGH PERFORMANCE" MBT KEEL PUNCH CEMENTED Size 2-3 HP M.B.T. Keel Punch Knee Instrument is used to prepare the proximal tibial canal geometry to accept the final or definitive tibial tray used during total knee arthroplasty Class II

Reason: DePuy Orthopaedics, Inc. is issuing a voluntary device correction of the HP M.B.T. Keel Punch Knee Instrument because of the potential for tabs on the connection end to fracture and be left in the patient.

DePuy Orthopaedics, Inc. 20131118 Terminated
Distributed: Worldwide Distribution - USA (nationwide) including AR,AZ,CA,CO,FL,GA,HI,IA,IL,IN,KY,LA,MA,MD,ME,MI,MN,MT,NC,NV,NY,OH,PA, TX,VA,WA,WI and Internationally to: AUSTRALIA AUSTRIA BELGIUM BRAZIL CANADA CZECH REPUBLIC DENMARK EGYPT FRANCE GERMANY GREECE HOLLAND HUNGARY INDIA IRELAND ISRAEL ITALY JAPAN KOREA MALAYSIA NEW ZEALAND NORWAY P R CHINA PAKISTAN PANAMA POLAND PORTUGAL PUERTO RICO RUSSIA SAUDI ARABIA SINGAPORE SLOVENIA SOUTH AFRICA SPAIN SWEDEN SWITZERLAND TURKEY UAE UK
SIGMA¿ LCS¿ HIGH PERFORMANCE" MBT KEEL PUNCH CEMENTED SIZE 1-1.5 HP M.B.T. Keel Punch Knee Instrument is used to prepare the proximal tibial canal geometry to accept the final or definitive tibial tray used during total knee arthroplasty. Class II

Reason: DePuy Orthopaedics, Inc. is issuing a voluntary device correction of the HP M.B.T. Keel Punch Knee Instrument because of the potential for tabs on the connection end to fracture and be left in the patient.

DePuy Orthopaedics, Inc. 20131118 Terminated
Distributed: Worldwide Distribution - USA (nationwide) including AR,AZ,CA,CO,FL,GA,HI,IA,IL,IN,KY,LA,MA,MD,ME,MI,MN,MT,NC,NV,NY,OH,PA, TX,VA,WA,WI and Internationally to: AUSTRALIA AUSTRIA BELGIUM BRAZIL CANADA CZECH REPUBLIC DENMARK EGYPT FRANCE GERMANY GREECE HOLLAND HUNGARY INDIA IRELAND ISRAEL ITALY JAPAN KOREA MALAYSIA NEW ZEALAND NORWAY P R CHINA PAKISTAN PANAMA POLAND PORTUGAL PUERTO RICO RUSSIA SAUDI ARABIA SINGAPORE SLOVENIA SOUTH AFRICA SPAIN SWEDEN SWITZERLAND TURKEY UAE UK
Shortened Plug Driver; Low Volume Specialty Device; Qty:1; Rx Only; Non Sterile; Biomet Spine, Parsippany, NJ 07054 The Polaris Spinal Systems is a non-cervical spinal fixation device intended for immobilization and stabilization as an adjunct to fusion as a pedical screw fixation system, a posterior hook and sacral/iliac screw fixation system, or as an anterior or anterolateral fixation system. Pedicle screw fixation is limited to skeletal mature patients for use with autograft and/or allograft. The device is indicated for all the following indications:degenerative disc disease (defined as discogeneic back pain with degeneration of the disk confirmed by history and radiographic studies), spondylolisthesis, and or lordosis) tumor, stenosis, pseudoarthrosis, or failed previous fusion Class II

Reason: Biomet received a complaint on 5/20/2013 reporting that the tip of the instrument sheared off during a removal case.

Ebi, Llc 20130924 Terminated
Distributed: US Distribution to CA
Vertebroplasty Needle Kit 10g Diamond Tip Intended to be used to inject PMMA cement into vertebral bodies during Vertebroplasty procedure. Class II

Reason: Complaints were received of inconsistencies in the expiration dates shown on the packaging of the Vertebroplasty Needle Kit 10g Diamond Tip device.

Synthes USA HQ, Inc. 20130730 Terminated
Distributed: Nationwide Distribution including IN, MT, HI, and FL.
Synthes 4.5 mm LCP Curved Condylar Plate Intended for buttressing multifragmentary distal femur fractures. Class II

Reason: The Synthes 4.5 mm LCP Curved Condylar Plate was recalled because a Straight Condylar Plate is etched with a Curved Condylar Plate part number.

Synthes USA HQ, Inc. 20130730 Terminated
Distributed: Worldwide Distribution - USA including ID, FL, TN, NY, LA and Internationally to Canada.
Slipped Capital Femoral Epiphysis System Product Usage: Intended for fracture fixation of large bones and large bone fragments and for slipped capital femoral epiphysis, pediatric femoral neck fractures, intercondylar femur fractures and sacroiliac joint disruptions. Class II

Reason: There is a possibility for the Guide Wires of the Slipped Capital Femoral Epiphysis System to stick in the cannulation of the drill bit and the drill bits to break during surgery.

Synthes USA HQ, Inc. 20130731 Terminated
Distributed: USA Nationwide Distribution
Synthes Titanium Cannulated Humeral Nail Intended to aid in the alignment and stabilization of humeral fractures Class II

Reason: The packaging incorrectly depicts the Synthes Titanium Cannulated Humeral Nail as a 3 distal hole nail instead of a 2 distal locking hole nail.

Synthes USA HQ, Inc. 20130730 Terminated
Distributed: Worldwide Distribution - USA (nationwide) including SC, IA, NC, TX, KY, VA, AL, PA, and Internationally to Canada.
Synthes 2.4 mm LC-DCP Plate 6 Holes/51mm Product Usage: Intended for fractures, osteotomies, and replantations of small bones including the foot, ankle, and hand. Class II

Reason: A complaint was reported regarding a screw being found in a package labeled for 2.4 mm LC-DCP Plate.

Synthes (USA) Products LLC 20130730 Terminated
Distributed: Worldwide Distribution - USA (nationwide) including IN, NY, WA, NE, MD, TX, AL, MI, and Internationally to Canada.