Synthes Implant Holder for Synfix (TM)-LR the Synthes Implant Holder for Synfix (TM)-LR is a non-powered hand-held device intended for medical purposes to manipulate tissue, or for use with other devices in orthopedic surgery Class II

Reason: Complaints were received which described the SynFix LR implant holder breaking at the interface between the implant and the holder.

Synthes USA HQ, Inc. 20131008 Terminated
Distributed: Worldwide Distribution - USA (nationwide) and Internationally to Canada .
Synthes 14.5 mm Outer Protection Sleeve for Suprapatellar Synthes 14.5 mm Outer Protection Sleeve for Suprapatellar is used as part of the Synthes Tibial Nail System EX to stabilize fractures of the proximal and distal tibia and the tibia shaft, open and closed tibia shaft fractures, certain pre- and post- isthmic fractures, and tibial malunions and non-unions. Class II

Reason: Complaints were received reporting that a product packaged as a 14.5 mm Outer Protection Sleeve for Suprapatellar contained a 12.0 mm Outer Protection Sleeve for Suprapatellar.

Synthes (USA) Products LLC 20131014 Terminated
Distributed: Nationwide Distribution
STERRAD NX, Product Code: 10033 The STERRAD Steri lization System is a low temperature general purpose sterilizer used to sterilize heat and moisture sensitive reusable medical devices. Class II

Reason: Advanced Sterilization Products (ASP) is recalling the STERRAD NX and STERRAD 100NX sterilization systems because ASP has determined that the bronchoscopes may contain a polyurethane material in the suction channel location that is not listed in the current claims for interior channels (lumens) for processing in STERRAD systems.

Advanced Sterilization Products 20131111 Terminated
Distributed: Nationwide Distribution
STERRAD 100NX, Product Code: 10104 The STERRAD Steri lization System is a low temperature general purpose sterilizer used to sterilize heat and moisture sensitive reusable medical devices. Class II

Reason: Advanced Sterilization Products (ASP) is recalling the STERRAD NX and STERRAD 100NX sterilization systems because ASP has determined that the bronchoscopes may contain a polyurethane material in the suction channel location that is not listed in the current claims for interior channels (lumens) for processing in STERRAD systems.

Advanced Sterilization Products 20131111 Terminated
Distributed: Nationwide Distribution
Synthes 4.5MM TI VA-LCP CRVD Condylar Plate/6H/159MM/RT-STER Synthes 4.5MM TI VA-LCP CRVD Condylar Plate/6H/159MM/RT-STER is used for buttressing multi fragmentary diatal femur fractures including: supra-condylar, intra-articular and extra articular condylar fractures, periprosthetic fractures, fractures in normal or osteopenic bone, nonunions and malunions. Class II

Reason: A plate was inadvertently released to a sales consultant that was restricted for sale.

Synthes USA HQ, Inc. 20130529 Terminated
Distributed: US Distribution in Colorado.
LANX Timberline MPF Lateral Modular Plate Fixation System Straight Split Tip Driver - Short ("Driver"). Model Number 8630-0107. The Timberline MPF device is an intervertebral body fusion device consisting of a PEEK intervertebral spacer, titanium plate, and screws. The PEEK spacer has a generally rounded shape with various heights and footprints and has a hollowed out central area to accommodate autogenous bone graft. The upper and lower surfaces have a series of transverse grooves formed to improve stability and fixation once the device is inserted. The titanium plate has holes for receiving bone screws and a central hole for receiving a cover plate to prevent screw back-out. The Timberline MPF system is available in a variety of sizes and configurations to approximate anatomical variation in different vertebral levels and/or patient anatomy. The Timberline MPF system is provided nonsterile. The Driver is used with the Timberline MPF system to drive titanium screws through the plate and into the vertebrae thus locking plate to the vertebrae. Class II

Reason: Lanx, Inc. is recalling various lots of Timberline MPF Straight Split Tip Driver - Short ("Driver") because the width of the most distal end from the Driver handle (the tip of the hex portion of the Driver) is oversized.

Lanx, Inc. 20131104 Terminated
Distributed: Nationwide Distribution including AZ, PA, WI, NY, TX, PR, and TN.
Medtronic Strata II Shunt Assembly Kit, Small, Catalog No. 27848. Strata II Shunt Assemblies are designed to provide continued cerebrospinal fluid (CSF) flow from the ventricles of the brain into the right atrium of the heart of the peritoneal cavity. Class II

Reason: Medtronic Neurosurgery is recalling the CSF-Unitized Shunts, Delta Shunt Assembly Kit, and the Strata II Shunt Assembly Kit because they were found to have label discrepancies.

Medtronic Neurosurgery 20131118 Terminated
Distributed: Worldwide Distribution - USA (nationwide) and Internationally to US and worldwide: Austria, Belgium, Bosnia, Herzegovina, Canada, Croatia, Germany, Japan, Poland, Portugal, Taiwan, United Arab Emirates, United Kingdom.
Medtronic Delta Shunt Assembly Kit with Bioglide Small, Performance Level 1.0, Catalog No. 276201. Delta Shunt Assemblies are designed for use in shunting CSF from the lateral ventricle of the brain into the peritoneal cavity. Class II

Reason: Medtronic Neurosurgery is recalling the CSF-Unitized Shunts, Delta Shunt Assembly Kit, and the Strata II Shunt Assembly Kit because they were found to have label discrepancies.

Medtronic Neurosurgery 20131118 Terminated
Distributed: Worldwide Distribution - USA (nationwide) and Internationally to US and worldwide: Austria, Belgium, Bosnia, Herzegovina, Canada, Croatia, Germany, Japan, Poland, Portugal, Taiwan, United Arab Emirates, United Kingdom.
CSF-Unitized Shunt, Contoured Regular, Low Pressure, Catalog No. 46022; CSF-Unitized Shunt, Contoured Regular, Medium Pressure, Catalog No. 46024; CSF-Unitized Shunt, Contoured Regular, High Pressure, Catalog No. 46026. CSF-Unitized Shunt Kits are designed as integrated CSF-Flow Control Shunts for use in shunting cerebrospinal fluid from the lateral ventricle of the brain into the peritoneal cavity. Class II

Reason: Medtronic Neurosurgery is recalling the CSF-Unitized Shunts, Delta Shunt Assembly Kit, and the Strata II Shunt Assembly Kit because they were found to have label discrepancies.

Medtronic Neurosurgery 20131118 Terminated
Distributed: Worldwide Distribution - USA (nationwide) and Internationally to US and worldwide: Austria, Belgium, Bosnia, Herzegovina, Canada, Croatia, Germany, Japan, Poland, Portugal, Taiwan, United Arab Emirates, United Kingdom.
Synthes Titanium Minimally Invasive Reduction Screw System Locking Cap, Part Number 04.631.502 Non-cervical spinal fixation devices intended for posterior pedicle screw fixation, posterior hook fixation, or anterolateral fixation. Class II

Reason: This action is being initiated following a detailed investigation in response to reported complaints of post-operative locking cap loosening. Synthes has concluded that if the construct is mis-aligned during final tightening, the potential exists for locking caps to loosen post-operatively in certain situations.

Synthes USA HQ, Inc. 20120907 Terminated
Distributed: Nationwide Distribution including CA, CO, CT, FL, GA, ID, IL, IN, LA, MA, MD, MI, MN, MT, NC, NJ, NY, OH, OR, PA, SC, TN, TX, UT, VA, WA, and WI.