ACCU-CHEK Aviva Combo Blood Glucose Meter (as a part of the ACCU-CHEK Aviva Combo System)\ The ACCU-CHEK Combo System is indicated for the treatment of insulin-requiring diabetes and for the quantitative measurement of glucose in fresh capillary whole blood from the finger. Class II

Reason: There is a software synchronization issue with the ACCU-CHEK Combo system. In rare cases, when the Manual Pump option is chosen on the ACCU-CHEK Aviva Combo meter, there is a possibility of receiving an incorrect bolus advice recommendation that may cause a temporary under delivery of insulin.

Roche Diagnostics Operations, Inc. 20131025 Terminated
Distributed: Nationwide Distribution.
Synthes Matrix Threaded Persuader Deformity is a comprehensive thoracolumbar pedical screw system designed to provide flexibility biomechanical performance and a solution to complex posterior pathological challenges. Class II

Reason: Synthes became aware that during a procedure using the Matrix Threaded Persuader, the Reduction Insert malfunctioned causing pieces to break off. Once there is a breakage to the Reduction Insert it will not allow the instrument to function as required.

Synthes USA HQ, Inc. 20130731 Terminated
Distributed: Worldwide distribution: US (nationwide) and countries of Canada and Switzerland.
MAS Urinalysis Control Level 1, Multi-analyte control (Assayed and Unassayed), Catalog/Part Numbers: 024222; Product is manufactured and distributed by Microgenics Corporation, a subsidiary of Thermo Fisher Scientific, Clinical Diagnostic Division, Fremont, CA MAS Urinalysis is intended for use in the clinical laboratory as a control for qualitative and semi-quantitative procedures used in routine urinalysis testing. Assay values are provided for specific systems listed. The user can compare observations with expected ranges as a means of assuring consistent performance of reagent and equipment. The product is suitable for use as a control material for physiochemical, chemical, and microscopic methods of routine urinalysis. Urinalysis may be used in conjunction with commercially available standardized systems for urine microscopic analysis. Class III

Reason: Via one (1) consumer complaint and in-house testing has confirmed that the product does not meet the 30 day open vial stability specifications for Lot: UA1514011.

Thermofisher Scientive 20131029 Terminated
Distributed: Worldwide Distribution - USA (nationwide) including Puerto Rico and internationally to Hong Kong.
Tri-Nex Healing Abutment, 0.50 mm x 5 mm height, Catalog No. HA-L-50-5. The NSI Implant System is intended to be implanted in the upper and lower jaw arches to provide support for fixed or removable dental prostheses in a single tooth, partially edentulous prostheses, or full arch prostheses. It further adds the option for immediate loading on single and splinted multiple unit restorations when good primary stability is achieved and with appropriate occlusal loading, to restore chewing function. Class II

Reason: Southern Implants is recalling the Tri-Nex healing abutment because the package was incorrectly labeled and packaged as the Tri-Nex healing abutment, 05.0 mm x 5 mm height (HA-L-50-5). The package contains the Tri-Nex healing abutment 05.0 mm x 6.0 mm Flare x 5 mm Height (HA-L-50W-5).

Southern Implants, Inc 20130913 Terminated
Distributed: Worldwide Distribution - USA (nationwide) and Internationally to Canada, Guatemala, Taiwan.
REF 200068 Security (TM) Enclosed Carpal Tunnel System Blade ( Single) ABS /Stainless Steel Sterile Device, Single Use Only. Designed to release the transverse carpal ligament. Class II

Reason: Security Blades were manufactured oversized and potentially won't advance or may be difficult to advance through the mating Security Clip.

Biomet, Inc. 20131025 Terminated
Distributed: Worldwide Distribution: USA (nationwide) and Internationally to Japan.
2.4 and 2.7 MM Locking Screw SLF-TPNG with T8 Stardrive Recess 12 MM Fixation of fractures, osteotomies, nonunions, replantations, and fusions of small bones and small bone fragments, particularly in osteopenic bone. Class II

Reason: Specific lots of the SS 2.4mm and SS 2.7mm Locking Screw (self-tapping) due to a mislabeling issue. The SS 2.7mm Screw was labeled as SS 2.4 mm Screw.

Synthes USA HQ, Inc. 20130731 Terminated
Distributed: Nationwide Distribution including PA, CA, FL, MO, KS, IN, IL, WV, KY, LA, NM, SC, IA, TX, AZ, GA, NC, OK, NY, TN, MI, CO, and MT.
Product Brand Name(s): Alere Cholestech LDX¿ Lipid Profile"GLU Test Cassette, Model 10-991. Product Generic Name: Alere Cholestech LDX¿ Lipid Profile"GLU Test Cassette. Description of the product: Lipid. Glucose panel Test cassettes. Product packaging: 10 individually pouched cassettes in a labeled kit box. Intended for the quantitative determination of total cholesterol, HDL (high-density lipoprotein) cholesterol, triglycerides, and/or glucose in whole blood, serum, and plasma. Class III

Reason: Alere initiated a recall of Cholestech LDX Lipid Profile-GLU test cassettes, Catalog Number 10-991, Lot # 320167, becuase an incorrect lot number was printed on the outer 10 pack kit box label. The incorrect Lot # number printed was 321067.

Alere San Diego, Inc. 20131023 Terminated
Distributed: Worldwide Distribution: USA (nationwide) and Internationally to Canada.
Mizuho Sugita Titanium Standard Aneurysm Clip II , Temporary Mini Type No 58 Temporary and permanent occlusion of cervical segment of cerebral aneurysms and peripheral blood vessels. They are designed for Single Use. Class II

Reason: Specific item numbers form Lots 27-J and 29-J of Sugita T2 Aneurysm Clips. Mini Temporary Type, are discolored.

Mizuho America, Inc. 20131010 Terminated
Distributed: Nationwide Distribution
Mizuho Sugita Titanium Standard Aneurysm Clip II , Temporary and permanent occlusion of cervical segment of cerebral aneurysms and peripheral blood vessels. They are designed for Single Use. Class II

Reason: Specific item numbers form Lots 27-J and 29-J of Sugita T2 Aneurysm Clips. Mini Temporary Type, are discolored.

Mizuho America, Inc. 20131010 Terminated
Distributed: Nationwide Distribution
(Siemens Rad Fluoro Uro Systems) Ysio, Luminos Agile, Luminos dRF, and Uroskop Omnia systems with software version VC10 Image intensified fluoroscopic x-ray system Class II

Reason: A potential malfunction and hazard to patients exists when using the Ysio, Luminos Agile, Luminos dRF, and Uroskop Omnia systems with software version VC10, which may result in radiation continuing longer than required and overexposed images that are not of diagnostic quality.

Siemens Medical Solutions USA, Inc 20130903 Terminated
Distributed: Worldwide Distribution - USA (nationwide) and internationally to Canada.