Brand Names: Integra¿ Camino 110-4 Intracranial Pressure (ICP) Monitoring Catheters. Product / Catalogue No.s: 110-4B, 110-4BC, 110-4G, 110-4H, 110-4HC, 110-4L, 110-4LC. The fully packaged 110-4B, 110-4H, and 110-4L catheters are also provided as a component in kits that contain other cranial access devices. These kits are Catalogue No. 110-4BC, 110- 4HC, and 110-4LC respectively. Common Name: Intracranial Pressure and Pressure-Temperature Monitoring Kits Classification Name: Device, Monitoring, Intracranial Pressure 510(k) Number: K102875 The Camino 110-4 Intracranial Pressure monitoring catheters are sterile transducer-tipped pressure monitoring catheters that are used by neurosurgeons to rapidly determine and continuously monitor intracranial pressure or pressure/temperature.  The Camino 110-4B catheter is used for Intracranial Pressure Monitoring in the parenchyma or subarachnoid space.  The Camino 110-4G catheter is used for Post Craniotomy Subdural Pressure Monitoring.  The Camino 110-4H catheter is intended for Intracranial Pressure Monitoring and Cerebrospinal fluid drainage.  The Camino 110-4L catheter is intended for use with the Licox¿ Brain Oxygen Monitoring System to measure intracranial pressure in the parenchyma.  The fully packaged 110-4B, 110-4H, and 110-4L catheters are also provided as a component in kits that contain other cranial access devices. These kits are Catalogue No. 110-4BC, 110- 4HC, and 110-4LC respectively. Class II

Reason: Integra LifeSciences has identified through an internal evaluation that in order to comply with US FDA requirements, certain Camino¿ Intracranial Pressure Monitoring Kits require additional labelling to indicate they are UNSAFE in an MR environment. Only the following Camino¿ Intracranial Pressure Monitoring Kits are affected: 110-4B, 110-4BC, 110-4G, 110-4H, 110-4HC, 110-4L, 110-4LC.

Integra Limited 20130910 Terminated
Distributed: Nationwide Distribution
The Centricity PACS Workstation is intended for use as a primary diagnostic and analysis tool for diagnostic images by trained healthcare professionals, including radiologists, physicians, technologists, clinicians and nurses. It is also intended for use as a clinical review workstation throughout the healthcare facility. The workstation interface provides the user with a means to display, manipulate, archive, print and export images when connected with the Centricity PACS infrastructure. Class II

Reason: GE has recently become aware of potential issues due to out of context issues associated with specific workflows of the Centricity PACS RA1000 workstation. These issues are as follows: 1) In Centricity RIS-IC, and a RIS-Driven integration, the "close" button closes the exam in RIS-IC, but does not close the exam in Centricity PACS RA1000 Workstation. 2) Interrupted workflow could break the

GE Healthcare It 20130611 Terminated
Distributed: Worldwide Distribution: US (nationwide) including states of: AK, AL, AR, AZ, CA, CO, CT, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI and WV) and the countries of the UK, Taiwan, Switerzland, Sweden, Spain, Singapore, Saudi Arabia, Netherlands, Malaysia, Korea, Japan, Italy, Israel, India, Hong Kong, Germany, France, Denmark, China, Canada, Bermuda, Beligum and Australia.
Breeze C Scooter is a motorized electric scooter for elderly people. Brand name Breeze, model C. It is intended to provide increased mobility for elderly and/or disabled person and it is also an aid to independent living. Products can be identified by their color; the recalled products are either metallic blue or metallic silver. The products have two rear wheels and have either one or two wheels in the front. The scooters have an adjustable swivel seat and a front storage basket; some scooters may be fitted with an optional larger rear basket. Class II

Reason: Breeze C Scooter, motorized 3-wheeled electric scooter for elderly people, is recalled because it may become unresponsive during operation due to an electrical contact which may fault due to extensive wear, which potentially could result in the driver not being able to stop the scooter during operation except by turning off the ignition key.

S.A.E. AFIKIM 20130927 Terminated
Distributed: Nationwide Distribution: US including the states of California, Florida, Georgia, Illinois, Maine, Maryland, Massachusetts, Michigan, Minnesota, New Jersey, New York, Ohio, Oklahoma, Pennsylvania, Texas, Virginia, and Wisconsin.
Synthes Application Instrument Sternal Zip Flex System intended for use in primary or secondary closure/repair of the sternum following sternotomy or fracture of the sternum to stabilize the sternum and promote fusion Class II

Reason: When using the 1st generation Zip Fix Application Instrument it is possible to cut the implant while the tensioning trigger is being squeezed which could lead to loosening of the implant.

Synthes USA HQ, Inc. 20130813 Terminated
Distributed: Worldwide Distribution - US (nationwide) and country of Mexico.
COULTER LH 750 and LH 780 Analyzers. The COULTER LH 750 Hematology Analyzer is a quantitative, automated hematology analyzer and leukocyte differential counter For In Vitro Diagnostic Use in clinical laboratories. The COULTER LH 750 Hematology Analyzer provides automated Reticulocyte analysis and enumeration of nucleated red blood cells (NRBCs) as well as an automated method for enumeration of RBCs and WBCs in body fluids. The COULTER LH 780 Hematology Analyzer is a quantitative, automated hematology analyzer and leukocyte differential counter For In Vitro Diagnostic Use in clinical laboratories. The COULTER LH 780 Hematology Analyzer also provides automated Reticulocyte analysis and enumeration of nucleated red blood cells (NRBCs) as well as an automated method for enumeration of RBCs and WBCs in body fluids. Class II

Reason: The check valves may fail on the Coulter LH 750 Hematology Analyzer (PN 6605632 and A85570) and Coulter LH 780 Hematology Analyzer (PN 723585 and A90728). A failure may result in an air or liquid leak at the connection to the Vacuum Overflow Tank waste line or at the Backwash Tank drain line.

Beckman Coulter Inc. 20130924 Terminated
Distributed: Worldwide Distribution - Nationwide Distribution and the country of Canada..
Fujinon/Fujifilm Ultrasonic Endoscope Intended to provide ultrasonic images of submucosal and peripheral organs of the upper gastrointestinal tract for observation, diagnosis and endoscopic treatment. Class II

Reason: A limited number of endoscopes were shipped to FMSU-ESD with a manufacturing defect which included missing set screws used to more securely attach the Forceps Inlet (biopsy inlet port) to the endoscope.

Fujinon Inc 20130731 Terminated
Distributed: Worldwide distribution: US (nationwide) including states of: NJ, PA and VA; and country of Canada.
Single Site Curved Needed Driver Intuitive Surgical, Sunnyvale, CA. 94086 The Intuitive Surgical da Vinci Single-Site Instruments and Accessories used with the da Vinci Si Surgical System (IS3000) are indicated for use by trained physicians in an operating room environment for endoscopic manipulation of tissue, grasping, cutting, blunt and sharp dissection, approximation, clip-ligation, electrocautery and suturing during single incision laparoscopic cholecystectomy, benign hysterectomy and salpingo-oophorectomy with the da Vinci Single-Site Instruments and Accessories, including graspers, dissectors, needle drivers, scissors, suction irrigators, monopolar cautery, bipolar cautery, 5 mm curved cannulae, 5 mm and 10 mm straight cannulae, flexible blunt obturators, and the 5mm Single-Site Port. Class II

Reason: A limited number of FDA-cleared instruments for da Vinci Single Site use, Bipolar Maryland and curved needle Driver, were shipped without the correct revision of the user manual which provided added instruction for use of these additional Single Site instruments including indication, warnings and precautions.

Intuitive Surgical, Inc. 20131002 Terminated
Distributed: Worldwide distribution: US (nationwide) and countries of:Australia, Belgium, France, Italy and United Kingdom. .
Single Site Bipolar Maryland Forceps Intuitive Surgical, Sunnyvale, CA. 94086 The Intuitive Surgical da Vinci Single-Site Instruments and Accessories used with the da Vinci Si Surgical System (IS3000) are indicated for use by trained physicians in an operating room environment for endoscopic manipulation of tissue, grasping, cutting, blunt and sharp dissection, approximation, clip-ligation, electrocautery and suturing during single incision laparoscopic cholecystectomy, benign hysterectomy and salpingo-oophorectomy with the da Vinci Single-Site Instruments and Accessories, including graspers, dissectors, needle drivers, scissors, suction irrigators, monopolar cautery, bipolar cautery, 5 mm curved cannulae, 5 mm and 10 mm straight cannulae, flexible blunt obturators, and the 5mm Single-Site Port. 2. Class II

Reason: A limited number of FDA-cleared instruments for da Vinci Single Site use, Bipolar Maryland and curved needle Driver, were shipped without the correct revision of the user manual which provided added instruction for use of these additional Single Site instruments including indication, warnings and precautions.

Intuitive Surgical, Inc. 20131002 Terminated
Distributed: Worldwide distribution: US (nationwide) and countries of:Australia, Belgium, France, Italy and United Kingdom. .
Vented Spike Adapter Product Code 2C0471, Sterile, nopyrogenic. Manufactured by an affiliate of Baxter Healthcare Corporation, Clintec Nutrition Division, Deerfield IL 60015 USA. An IV transfer device, used to transfer fluids in pharmacy compounding system. Class II

Reason: Potential tears in the pouch of six lots of Vented Spike Adapter Product Code 2C0471

Baxter Healthcare Corp. 20130830 Terminated
Distributed: Nationwide distribution.
Siemens RAPIDLab¿ 1265 system blood gas analyzer Product Usage: RAPIDLab 1200 Systems: The Rapidlab 1200 systems are intended for in vitro diagnostic use by healthcare professionals in the quantitative testing of human whole blood. The systems can determine the following parameters: RAPIDLab 1240 pH, pC02, p02 RAPIDLab 1245 pH, pC02, p02, tHb, nBili, F02Hb, FCOHb, FMetHb, FHHb RAPIDLab 1260 pH, pC02, p02, Na+, K+, Ca++, Cl-, glucose, lactate RAPIDLab 1265 pH, pC02, p02, Na+, K+, Ca++, Cl-, glucose, lactate, tHb, nBili, F02Hb, FCOHb, FMetHb, FHHb Class II

Reason: RAPIDPoint¿ 400 Series, RAPIDPoint¿ 500, or RAPIDLab¿ 1200 Series blood gas analyzer, users can inadvertently enter a wrong patient ID on the demographics screen, which ultimately could lead to a misidentification of the sample

Siemens Healthcare Diagnostics Inc 20120917 Terminated
Distributed: Worldwide Distribution - USA Nationwide and in the countries of Algeria, Argentina, Australia, Austria, Bahrain, Bangladesh, Belgium, Bosnia Herzeg., Botswana, Brazil, Brunei, Bulgaria, Chile, Colombia, Croatia, Czech Republic, Denmark, Egypt, Estonia, Finland, France, Fren.Polynesia, Germany, Greece, Hong Kong, Hungary, Iceland, India, Indonesia, Iraq, Ireland, Israel, Italy, Ivory Coast, Japan, Kazakhstan, Kenya, Kuwait, Latvia, Lithuania, Luxembourg, Macedonia, Malaysia, Mayotte, Netherlands, New Caledonia, New Zealand, Norway, P.R. China, Philippines, Poland, Portugal, Qatar, Rep. of Yemen, Republic Korea, Romania, Russian Fed., Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, Turkmenistan, U.A.E., United Kingdom, Vatikancity, Venezuela, Vietnam,