Brand: The ForeCYTE Breast Health Test, Mammary Aspiration Specimen Cytology Test (MASCT), MASCT System Kits, Clarity System Kits, MASCT Patient Sample Kits & Clarity Patient Sample Kits Instrument, Biopsy. The MASCT System Kit, Patient Sample Kits, or the Nipple Aspirate Fluid Laboratory Kit has the following codes: PRODUCT CODE: AG-MASCT; AG-FC5; DTG-MASCT, DTG-FC5, and NRLBH-5. PART NUMBER: 9002513; 9002513MD; 9002528; 9002528MD; 9002587; 9002614; and 9002717MD. THE MASCT SYSTEM KIT CONSISTS OF: 1. MASCT Breast Pump; 2. Instructions for Use (IFU); 3. Heating Pad; 4. Timer; 5. Saccomono's Fixative; 6. Nu Prep Gel; 7. Welcome and Training Materials Coversheet; 8. MASCT System Order Form; 9. ForeCYTE Training Video DVD; 10. MASCT System Instructions with pictures - pink; THE PATIENT SAMPLE KIT CONSISTS OF: 1. Two (2) flower assemblies (e.g., filter, filter retainer, and filter holder); 2. Instructions for Use (IFU); 3. Two (2) specimen collection devices; 4. Two (2) breast bags for transporting the specimen collection container; 5. Barcode labels; 6. Saccomono's Fixative; 7. Nu Prep Gel; 8. MASCT System package insert; 9. ForeCYTE Test Requisition Form; 10. Patient Information Sheet; 11. Atossa Patient Information; 12. FED EX Clinical PAK; 13. MASCT System Order Form; 14. FED EX Pouch; 15. FED EX Airbill; Product Usage: The MASCT Device is intended for use in the collection of nipple aspirate fluid for cytological testing. The collected fluid can be used in the determination and/or differentiation of normal versus premalignant versus malignant cells. Class I

Reason: Atossa Genetics is recalling ForeCYTE Breast Health Test, Mammary Aspiration Specimen Cytology Test (MASCT) due to concerns raised by the U.S. Food and Drug Administration (FDA) in a warning letter received by Atossa in February 2013. The FDA raised concerns about (1) the current Instructions for Use (IFU); (2) certain promotional claims used to market these devices; and (3) the need for FDA cl

Atossa Genetics, Inc. 20130919 Terminated
Distributed: USA Nationwide Distribution in the states of: AL, CA, FL, GA, IL, IN, KY, MD, ME, MI, MO, NC, ND, NJ, NV, NY, OK, PA, SC, TN, TX, UT, VA, WA, Washington DC, and Puerto Rico.
Xoran MiniCAT CT Scanner ID ENT00001 Class II

Reason: Failure of the gantry to rotate. The complaint alleged that the scanner would emit radiation but would not rotate.

Xoran Technologies, Inc. 20091214 Terminated
Distributed: Worldwide Distribution - US nationwide
Xoran MiniCAT00202 Class II

Reason: The MiniCAT CT Scanners subject to this notification failed to meet Xoran's specifications for dose rate of 125 ¿ 3 kVp. These 18 complaints received include dose rates both on the high and low end of Xorans specifications.

Xoran Technologies, Inc. 20100727 Terminated
Distributed: Worldwide Distribution - Nationwide
Xoran Minicat00158 Class II

Reason: The MiniCAT CT Scanners subject to this notification failed to meet Xoran's specifications for dose rate of 125 ¿ 3 kVp. These 18 complaints received include dose rates both on the high and low end of Xoran's specifications. See ATTACHMENT 1 for details. (e) A n evaluation of the hazards reasonably related to defect or the failure to comply with the Federal standard; An evaluation of the health ha

Xoran Technologies, Inc. 20100727 Terminated
Distributed: Worldwide Distribution - US nationwide
Xoran Minicat00202 Class II

Reason: The MiniCAT CT Scanners subject to this notification failed to meet Xoran's specifications for dose rate of 125 ¿ 3 kVp. These 18 complaints received include dose rates both on the high and low end of Xoran's specifications. See ATTACHMENT 1 for details. (e) A n evaluation of the hazards reasonably related to defect or the failure to comply with the Federal standard; An evaluation of the health ha

Xoran Technologies, Inc. 20100727 Terminated
Distributed: Worldwide Distribution - US nationwide
Xoran MiniCAT CT Scanner Class II

Reason: A description of the defect in the product or the manner in which the product fails to comply wvith an applicable Federal standard; The MiniCAT CT Scanners subject to this notification failed to meet Xoran's specifications for dose rate of 125 ¿ 3 kVp. These 18 complaints received include dose rates both on the high and low end of Xoran's specifications. See ATTACHMENT 1 for details. (e) A n evalu

Xoran Technologies, Inc. 20100727 Terminated
Distributed: Worldwide Distribution - Nationwide
Xoran Mini CT Scanner Class II

Reason: A description of the defect in the product or the manner in which the product fails to comply with an applicable Federal standard; The MiniCAT CT Scanners subject to this notification failed to meet Xoran's specifications for dose rate of 125 ¿¿ 3 kVp. These 18 complaints received include dose rates both on the high and low end of Xoran's specifications. See ATTACHMENT 1 for details. (e) A n evalu

Xoran Technologies, Inc. 20100727 Terminated
Distributed: Worldwide Distribution - US nationwide
GE Brivo OEC715/785/865 C-Arm, Models: 715 Prime, 785 Essential, and 865 Plus. The QEC Brivo Mobile C-Arm X-Ray Products are designed to provide digital spot film imaging and fluoroscopic image guidance for all adult and pediatric populations for orientations between patient anatomy and surgical instruments. The product is used for general surgical applications and musculoskeletal procedures to visualize, for example, implant localizations or needle positions for aspirations,injections or biopsy. Class II

Reason: GE is recalling certain lots of GE Brivo OEC715/785/865 C-Arm due to the potential for the wires in the high voltage cable bundle to fail, resulting in loss of system functionality.

GE Healthcare Beijing , Hangwei Medical Systems, Co. LTD 20130918 Terminated
Distributed: Worldwide Distribution - USA (nationwide) and Internationally to China, India, Japan, Malaysia, Australia, Ireland, France, Germany, Spain, Hungary, Belgium, and Switzerland.
VARISOURCE Breast Template System for high-dose rate Brachytherapy. ***INSTRUCTIONS FOR USE*** Breast Bridge & Template System AL13011000 ***Manufacturer: Varian Medical Systems, Inc., 3100 Hansen Way, Palo Alto, CA 94304 USA Class II

Reason: The recall was initiated after Varian Medical became aware the Lexan templates of the Breast Bridge Template System could become bent and turn opaque after steam sterilization.

Varian Medical Systems, Inc. 20130809 Terminated
Distributed: Worldwide distribution: US (nationwide) distribution in CA and in the country of India.
SmartSet¿ Endurance GMV, Gentamicin, REF 3105-040, Medium Viscosity Antibiotic Bone Cement, Sterile. SmartSet GMV Endurance Gentamicin, bone cements are self-curing, radiopaque, polymethyl methacrylate based cements, containing antibiotic, used for securing a metal or polymeric prosthesis to living bone in arthroplasty procedures where infection by gentamicin-sensitive organisms is a potential risk. Class II

Reason: The SmartSet¿ GMV Endurance Gentamicin Bone Cement, product ref: number 3105-040 is intended for use in indications that have not been cleared by the US FDA. The intended use stated in the IFU for the SmartSet¿ GMV Endurance Gentamicin Bone Cement is for the fixation of prostheses to living bone in arthroplasty procedures of joints in which infection by gentamicin-sensitive organisms is a potentia

DePuy Orthopaedics, Inc. 20130916 Terminated
Distributed: Distributed in Puerto Rico and US Virgin Islands.