CareFusion Gravity Set, Model #44000-07 The CareFusion Gravity Sets are used to administer fluid and medication through a needle or catheter inserted into the patient's artery or vein. The CareFusion Gravity Set is comprised of components commonly found on intravascular administration sets and extension sets. It includes a check valve, injection ports, 4-way stopcock, anti-siphon valve, roller clamp and tubing. Class II

Reason: CareFusion is recalling the Gravity Set (Model 44000-07) because of an incorrect expiration date. The affected lot numbers of the Gravity Set are labeled with a 5 year expiration instead of three (3) years.

CareFusion 303, Inc. 20130830 Terminated
Distributed: IL
ABX PENTRA Calcium CP Reagent, Model # A11A01633 Calcium plays an essential role in many cell functions: intracellularly in muscle contraction and glycogen metabolism, extracellularly, in bone mineralization, in blood coagulation and in transmission of nerve impulses. Calcium is present in plasma in three forms: free, bound to proteins or complexed with anions as phosphate, citrate and bicarbonate. Under physiological conditions, calcium balance is determined by the relationship between calcium intake and calcium absorption and excretion. Urinary excretion is an important determinant of calcium retention in the body. Decreased total calcium levels can be associated with diseases of the bone apparatus (especially osteoporosis), kidney diseases (especially under dialysis), defective intestinal absorption and hypoparathyroidism. Increased total calcium can be measured in hyperparathyroidism, malignant diseases with metastases and sarcoidosis. Calcium measurements also help in monitoring of calcium supplementation mainly in the prevention of osteoporosis. Class III

Reason: HORIBA Medical is informing all PENTRA 200 and PENTRA 400 customers of a potential for reporting incorrect calcium results when using the ABX PENTRA CALCIUM CP reagent.

Horiba Instruments, Inc dba Horiba Medical 20130906 Terminated
Distributed: Worldwide Distribution - USA including the country of Canada.
SmartSite Low Sorbing Infusion Set, Model#72313E The SmartSite Low Sorbing Infusion Sets are used to administer fluid and medications through a needle or catheter inserted into the patient's artery or vein. The SmartSite Low Sorbing set is comprised of components commonly found on intravascular administration sets and extension sets. It includes a 1.2 micron filter, needleless SmartSite connector, low sorbing tubing segment, pinch clamps, and luer connectors. Class II

Reason: CareFusion is recalling the SmartSite Low Sorbing Infusion Set, Model 72313E, Lot Number 13016314, because of an incorrect filter was assembled on this lot. The correct filter is a 1.2 micron filter (blue color) but a 0.2 micron filter (clear color) may have been added instead.

CareFusion 303, Inc. 20130830 Terminated
Distributed: CA, FL, AZ, and TN
Synthes 3.5 MM LCP(R) Distal Humerus System The Synthes 3.5 MM LCP Distal Humerus System is intended for fixation of fractures of the distal humerus, olecranon, and ulna in adults and adolescents (12-21) in which the growth plates have fused. Specifically, distal humerus plates are indicated for intra-articular fractures, comminuted supracondylar fractures, osteotomies, malunions and non-unions of the distal humerus. Class II

Reason: Recall is being initiated due to the part being mislabeled (Part Number 241.267 was mislabeled with Part Number 241.276).

Synthes USA HQ, Inc. 20130308 Terminated
Distributed: Nationwide Distribution including CA,and VA. .
Monaco RTP System Radiation Therapy Treatment Planning System. Radiation treatment planning. Class II

Reason: Patient needs to be positioned as indicated by the treatment plan.

Elekta, Inc. 20130522 Terminated
Distributed: Distributed in Puerto Rico and the states of AK, AZ, CA, CO, CT, FL, IL, IN, MA, MD, MI, MN, MO, NC, NJ, NM, NY, OH, OK, OR, PA, RI, SC, TN, TX, VA, and WA.
Monaco RTP System Radiation treatment planning Class II

Reason: Upgrade installation script is designed to overwrite the same pouch files in the Treatment Couch Library with the original default values.

Elekta, Inc. 20130702 Terminated
Distributed: Nationwide Distribution including AK, CA, CO, CT, FL, IL, IN, MI, MN, MO, NC, NJ, NY, OH, OK, OR, PA, SC, TN, TX, WA and Puerto Rico.
TEGRIS system Four labels of the Tegris System include: Emergency button (1703.72) MAQUET GmbH & Co. KG Kehler-Strabe 31 Baden-Wuerttemberg 76437 Rastatt, Germany Touch screen (elo) Elo TouchSystems 301 Constitution Drive Menlo Park, CA 94025 www.elotouch.com Manufacturer: Tyco Electronics Brand Name: Elo Touch Systems Made in China Tegris system (1703.01) MAQUET GmbH & Co. KG Kehler Strabe 31 76437 Rastatt, Germany OR Table Module (1703.71) MAQUET GmbH & Co. KG Kehler Strabe 31 Baden-Wuerttemberg 76437 Rastatt, Germany Designed to be used as the central operating system in an operating room. The integration system has two main functions. -recording and distribution of images and videos -interaction with medical and non-medical devices. The system has a modular design and comprises the basic units as well as s various optional modules,which may be used or integrated in conjunction with the basic unit. TEGRIS is operated using a touchscreen. Medical devices that can be operated by TEGRIS include OR tables an OR lights. In the case of OR tables and OR lights, the main functions of these devices may be separated, e.g. adjustment of OR tables or switching the OR lights on.off. TEGRIS has no effected on the primary operating functions of the original control device. TEGRIS does not provide any other functions in addition to these primary operating functions. The purpose of TEGRIS namely the central operation of these various devices, is dependant on the individual devices. Please refer to the operating instructions of the individual devices for instructions. The integrated surgical safety checklist is a raft and must be modified to meet local requirements and approved by the User. TEGRIS is intended for short-term use pursuant to Appendix IX of the Directive 93/42/EEC of medical products in medical rooms in accordance with Group 1 and 2 VDR 0100-710, providing, among other things, protection against electric shock and connections for potential equalization lines. Intended for the following users: medical experts Class II

Reason: MAQUET has identified a potential issue that might occur because of the lack of an emergency stop button on a secondary touch screen display on the TEGRIS System.

Maquet Cardiovascular Us Sales, Llc 20130819 Terminated
Distributed: Worldwide Distribution - USA including NY and PA and internationally to Australia, Brazil, Canada, Germany, Hong Kong, Japan, Sweden.
BD Epilor Syringe BD Franklin Lakes, NJ 07417 BD EDC Laagstraat 57, B-9140 Temse-Belgium The Epilor syringe is used in conjunction with an epidural needle with the purpose of finding the epidural space. Class II

Reason: BD has received several reports indicating that the plunger of the 7 mL BD Epilor plastic BD Luer-Lok Loss of Resistance Syringe may stall or stick within the syringe preventing the user from detecting a loss of resistance upon entering the epidural space.

Becton Dickinson & Company 20130823 Terminated
Distributed: Worldwide Distribution - USA including FL, VA, TX, NY, IL, AZ, MI, RI, NJ, SC, MO, TN, CA, MN, UT, GA, NE, OH, MD, PA, MT and internationally to Brazil, Australia, Canada, and Puerto Rico..
Stryker Vision Mounting Arm, for use with other Stryker Endoscopy medical device products for human patients. Model number 0240-095-200, manufactured by Stryker Endoscopy San Jose The intended use of the Stryker Vision Mount is to hold and position an LCD monitor. The Stryker Vision Mounting Arm can be used with any Stryker flat panel display and is It is compatible with Stryker Carts (0240099001, 0240099011 or 0240099020). It is indicated for use in a surgical operating room environment and is exposed to environmental conditions associated with the storage and shipping of the product. It is not intended to enter the sterile field. The Stryker Vision Mount does not deliver to and/or extract from the patient any energy or material. It does not process any biological materials, as it is not intended to be sterilized by the user. The Stryker Vision Mount is only used to display surgical images and video and does not sustain or support life. Class II

Reason: When positioning the display located on top of the video cart, it is possible for the user to apply excessive force and push the display up and out of the mounting arm. This can result in the display falling upon the user/patient.

Stryker Endoscopy 20130830 Terminated
Distributed: Worldwide Distribution - USA including AL, AK, AZ, AR, CA, CO, CT, DE, FL, GA, HI, ID, IL, IN, IA, KS, KY, LA, ME, MD, MA, MI, MN, MS, MO, MT, NE, NV, NH, NJ, NM, NY, NC, ND, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VT, VA, WA, WV, WI, WY and internationally to Australia, Canada, Chile, China, Argentina, Guam, Taiwan, India, Japan, Korea, Latin America, Philippines, Mexico, New Zealand, Hong Kong, Singapore, and Switzerland.
Invacare Power Wheelchair, one per box To provide mobility to a person restricted to a sitting position. Class II

Reason: Unintended slowing or deceleration, and on rare occasion, unanticipated and unattended acceleration of power wheelchairs equipped with SPJ and MPJ type joysticks.

Invacare Corporation 20130604 Terminated
Distributed: Worldwide Distribution - USA including AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, and WY. Internationally to Canada, Australia, Bermuda, Brazil, China, Costa Rica, Germany, France, United Kingdom, Guam, Hong Kong, Israel, India, Kuwait, New Zealand, Puerto rico, Qatar, Saudi Arabia, Singapore, Thailand, Trinidad, Tobago, and South Africa.
Page 3,733 of 4,030 | Showing 10 recalls