Blanketrol III, Model Number 233, Cincinnati Sub-Zero Products, Inc., 12011 Mosteller Road, Cincinnati, OH 45241 Hyper-Hypothermia device used to lower or to raise a patients temperature and/or maintain a desired patient temperature through conductive heat transfer. Class II

Reason: During a label review it was discovered that both the Blanketrol III Operations (57201-N) & Operations & Technical (56201-Q) manuals had verbiage added back into the manual during ECN M111-4260 (released 12/29/2011) which was previously removed for clarification of the intended use and subsequent 501(K) approval.

Cincinnati Sub-Zero Products Inc 20130903 Terminated
Distributed: Worldwide Distribution - USA including AK, AR, AZ, CA, CO, DC, DE, FL, GA, IA, ID, IL, IN, KY, LA, MA, MD, ME, MI, MN, MT, , NC, ND, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, SC, TN, TX, UT, VA, WA, WI & WV. Internationally to ARGENTINA, AUSTRALIA, AZERBAIJAN, BELGIUM, CANADA, CHILE, CHINA, COSTA RICA, CROATIA, GERMANY, HONDURAS, HONG KONG, INDIA, INDONESIA, ISRAEL, KOREA, KUWAIT, MEXICO, NEW ZEALAND, PERU, SAUDI ARABIA, SINGAPORE, SPAIN, TAHITI, THAILAND, THE NETHERLANDS, THE PHILIPPINES, UNITED ARAB EMERITES & UNITED KINGDOM.
Osseotite Parallel Walled Certain 4mm(D)X 15mm (L) Dental Implant Product Usage: Surgical placement in the upper or lower jaw to provide a means for prosthetic attachment in single tooth restorations and in partially or fully edentulous spans with multiple single teeth utilizing delayed or immediate loading or with a terminal or intermediary abutment for fixed or removable bridgework and to retain over dentures. Class II

Reason: One of the dental implants do not have the internal hex feature. One of the dental implants do not have the internal thread feature.

Biomet 3i, LLC 20130610 Terminated
Distributed: Worldwide Distribution - USA Nationwide in the states of MA, CT, NJ, DE, NY, FL, OK, TX, PA, DC, IL, KS, LA, MD, CA, RI, TN, MN, and SC and the countries of Canada, Japan, Australia, Italy Germany, France, Switzerland, Apain, Brazil, Uruguay, Singapore, Dominican Republic, Taiwan, Austria, Portugal, and the United Kingdom.
Osseotite 2 Certain Prevail 5mm(D)X 4.1mm (P) X X 11.5mm (L) Dental Implant Product Usage: Surgical placement in the upper or lower jaw to provide a means for prosthetic attachment in single tooth restorations and in partially or fully edentulous spans with multiple single teeth utilizing delayed or immediate loading or with a terminal or intermediary abutment for fixed or removable bridgework and to retain over dentures. Class II

Reason: One of the dental implants do not have the internal hex feature. One of the dental implants do not have the internal thread feature.

Biomet 3i, LLC 20130610 Terminated
Distributed: Worldwide Distribution - USA Nationwide in the states of MA, CT, NJ, DE, NY, FL, OK, TX, PA, DC, IL, KS, LA, MD, CA, RI, TN, MN, and SC and the countries of Canada, Japan, Australia, Italy Germany, France, Switzerland, Apain, Brazil, Uruguay, Singapore, Dominican Republic, Taiwan, Austria, Portugal, and the United Kingdom.
C2244.5008 CONELOG Titanbasis CAD/CAM Conelog Titanium Base CAD/CAM Rx only ¿5.0GH 0.8 inkl HS Product Usage: Endosseous dental implant abutment CONELOG Abutments are intended to be used to fabricate prosthetic restorations in conjunction with CONELOG implants to support prostheses in the maxillary and/or mandibular arch. Class II

Reason: Product not cleared for sale/use in the United States.

Camlog Usa 20130828 Terminated
Distributed: USA Distribution in the states of NH, MD, CO, and IL.
C2244.4308 CONELOG Titanium Base CC ¿4.3GH2 0.8 inkl HS Product Usage: Endosseous dental implant abutment CONELOG Abutments are intended to be used to fabricate prosthetic restorations in conjunction with CONELOG implants to support prostheses in the maxillary and/or mandibular arch. Class II

Reason: Product not cleared for sale/use in the United States.

Camlog Usa 20130828 Terminated
Distributed: USA Distribution in the states of NH, MD, CO, and IL.
C2244.3820;, CONELOG Titanium Base CC ¿3.8GH2.0 inkl HS Product Usage: Endosseous dental implant abutment CONELOG Abutments are intended to be used to fabricate prosthetic restorations in conjunction with CONELOG implants to support prostheses in the maxillary and/or mandibular arch. Class II

Reason: Product not cleared for sale/use in the United States.

Camlog Usa 20130828 Terminated
Distributed: USA Distribution in the states of NH, MD, CO, and IL.
C2244.3808, CONELOG Titanium Base CC ¿3.8GH0.8 inkl HS Product Usage: Endosseous dental implant abutment CONELOG Abutments are intended to be used to fabricate prosthetic restorations in conjunction with CONELOG implants to support prostheses in the maxillary and/or mandibular arch. Class II

Reason: Product not cleared for sale/use in the United States.

Camlog Usa 20130828 Terminated
Distributed: USA Distribution in the states of NH, MD, CO, and IL.
Sonogage EyeScan, A-Scan Biometer, Product Usage: A-Scan Biometer diagnostic device for measurement of axial eye length. Class II

Reason: During an FDA inspection on 7/15-26/2013, it was discovered that the firm lacked registration and 510(k) clearance for the device.

Sonogage Inc 20130809 Terminated
Distributed: Worldwide Distribution - USA Nationwide in the following states: AK, AZ, CA, FL, GA, IL, KY, LA, MA, MD, MN, NE, NJ, NM, NY, OH, OR, PA, TX, VA and WA. Product was also shipped to Canada and Mexico.
GE Healthcare, CARESCAPE Monitor B850. Intended for use in multiple areas within a professional healthcare facility. Class II

Reason: GE Healthcare has recently become aware of an issue associated with the CARESCAPE Monitor B850 when used with the iCentral network. The iCentral must have the alarm management software license (L-NETM-00) and the alarm limit change option enabled to experience this issue. When changing the HR, SpO2 or Invasive pressure alarm limits from the iCentral over the network, the alarm priority on th

GE Healthcare, LLC 20110720 Terminated
Distributed: Worldwide Distribution-USA (nationwide) including the states of CA, CO, CT, HI, IL , IA, KY, LA, MA, MI, NE, NY,OH, OK, TX, and WI, and the countries of UNITED KINGDOM, UNITED ARAB EMIRATES, TURKEY, TUNISIA, THAILAND, SYRIAN ARAB REPUBLIC, SWITZERLAND, SWEDEN, SPAIN, SOUTH AFRICA, SLOVAKIA, SINGAPORE, SAUDI ARABIA, RUSSIAN FEDERATION, REPUBLIC OF SERBIA, PORTUGAL, POLAND, PARAGUAY, NORWAY, NETHERLANDS, NEPAL, MEXICO, MALAYSIA, KUWAIT, REPUBLIC OF KOREA, JORDAN, JAPAN, ITALY, ISRAEL, IRELAND, IRAQ, INDONESIA, INDIA, HONG KONG, GREECE, GERMANY, FRANCE, FINLAND, ESTONIA, EGYPT, DENMARK, CZECH REPUBLIC, COSTA RICA, COLOMBIA, CHINA, CHILE, CANADA, BOSNIA AND HERZEGOVINA, BELGIUM, BAHRAIN, AUSTRIA, and AUSTRALIA.
T8 REMOVAL HEAD, REF 8910641, QTY: 1 EA, Product Usage: The non-sterile T8 Removal Diriver is used with the Spire and Spire Z Spinal System in conjunction with the set screwdriver to remove the implant during removal surgery. The set screw driver is attached to the end of the removal head and used to disengage the remaining portion of the set screw within the device. Class II

Reason: One lot of the T8 Removal Head had incorrect hardness specification, which could result in the T8 Removal head failing during removal of the CD Horizon Spire and Spire Z Spinal System implant. There is a potential for tissue damage to occur and/or minor surgical delay to remove the remaining fragments of the tip.

Medtronic Sofamor Danek USA Inc 20130812 Terminated
Distributed: USA Nationwide Distribution in the state of GA, AK, WA, and CA.
Page 3,740 of 4,030 | Showing 10 recalls