Duane Reade (DR) Wound Closure Adhesive Surgical Tape Strips 30 strips 1/4 in. x 4 in. Sterile latex free UPC # 639194043885 Used to secure, close and support small cuts and wounds. Class II

Reason: This recall has been initiated by Aspen Surgical Products, Inc., due to the inability to confirm sterility.

ASO, LLC 20130726 Terminated
Distributed: Nationwide Distribution including FL, IA, IL, MI, NC, NY, OH, OR, PA, RI, TX, WA and WI.
Discount Drug Mart Wound Closure Adhesive Surgical Tape Strips 30 strips 1/4 in. x 4 in. Sterile latex free UPC # 093351120374 Used to secure, close and support small cuts and wounds. Class II

Reason: This recall has been initiated by Aspen Surgical Products, Inc., due to the inability to confirm sterility.

ASO, LLC 20130726 Terminated
Distributed: Nationwide Distribution including FL, IA, IL, MI, NC, NY, OH, OR, PA, RI, TX, WA and WI.
CVS Pharmacy Wound Closure Adhesive Surgical Tape Strips 30 strips 1/4 in. x 4 in. Sterile latex free UPC # 050428102909 item # 409170 Used to secure, close and support small cuts and wounds. Class II

Reason: This recall has been initiated by Aspen Surgical Products, Inc., due to the inability to confirm sterility.

ASO, LLC 20130726 Terminated
Distributed: Nationwide Distribution including FL, IA, IL, MI, NC, NY, OH, OR, PA, RI, TX, WA and WI.
Cordis OPTEASE Retrieval Inferior Vena Cava (IVC) Filter***REF Cat. No.***466F220A and 466F220B***Vena Cava Filter and Introduction Kit***(BRITE TIP Catheter Sheath Introducer and Angiographic Vessel Dilator)***Cordis***a Johnson and Johnson Company***ENDOVASCULAR*** Label No. SB 10247519.2/***LB466F220AO.5***omoraoro/2012.050794***Do not resterilize***. Product Usage: The OPTEASE Vena Cava Filter is indicated for use in the prevention of recurrent pulmonary embolism (PE) via percutaneous placement in the vena cava in several situations: Pulmonary thromboembolism when anticoagulants are contraindicated; Failure of anticoagulant therapy for thromboembolic disease; Emergency treatment following massive pulmonary embolism where anticipated benefits of conventional therapy are reduced and Chronic, recurrent pulmonary embolism where anticoagulant therapy has failed, or is contraindicated. Class I

Reason: Correction to provide clarification and additional labeling to minimize likelihood on implanting filter backwards.

Cordis Corporation 20130329 Terminated
Distributed: Worldwide Distribution - USA Nationwide in the states of AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, Ny, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY, including Hawaii and Puerto Rico and the countries of Canada & Panama
IMMULITE Systems Rapid TSH (REF LKRT1 - 100T, LKRT5 - 500 T, L2KRT2 - 200T, L2KRT6 - 600T). Intended to quantitatively measure human thyroid stimulating hormone (TSH, thyrotropin). Class II

Reason: A rare variant of TSH, identified in a small cluster of patients, is not detected.

Siemens Healthcare Diagnostics, Inc 20130510 Terminated
Distributed: Worldwide Distribution-USA (nationwide) and the countries of Canada, AE, AR, AT, AU, BA, BD, BE, BG, BO, BR, BY, CH, CL, CM, CN, CO, CR, CY, CZ, DE, DK, DO, DZ, EC, EE, EG, ES, FI, FR, GB, GR, GT, HK, HN, HR, HU, ID, IE, IL, IN, IR, IS, IT, JO, JP, KA, KR, KW, KZ, LB, LK, LT, LV, LY, MA, MK, MT, MX, MY, NI, NL, NO, NP, OM, PA, PE, PH, PK, PL, PT, PY, QA, RO, RU, SA, SE, SG, SK, SV, SY, TH, TM, TN, TR, TW, UA, UY, UZ, VE, VN, XS, and ZA.
IMMULITE Systems Third Generation TSH (REF LKTS1 - 100T, LKTS5 - 500T, L2KTS2 - 200T, L2KTS6 - 600T). Intended to quantitatively measure human thyroid stimulating hormone (TSH, thyrotropin). Class II

Reason: A rare variant of TSH, identified in a small cluster of patients, is not detected.

Siemens Healthcare Diagnostics, Inc 20130510 Terminated
Distributed: Worldwide Distribution-USA (nationwide) and the countries of Canada, AE, AR, AT, AU, BA, BD, BE, BG, BO, BR, BY, CH, CL, CM, CN, CO, CR, CY, CZ, DE, DK, DO, DZ, EC, EE, EG, ES, FI, FR, GB, GR, GT, HK, HN, HR, HU, ID, IE, IL, IN, IR, IS, IT, JO, JP, KA, KR, KW, KZ, LB, LK, LT, LV, LY, MA, MK, MT, MX, MY, NI, NL, NO, NP, OM, PA, PE, PH, PK, PL, PT, PY, QA, RO, RU, SA, SE, SG, SK, SV, SY, TH, TM, TN, TR, TW, UA, UY, UZ, VE, VN, XS, and ZA.
Dimension Vista System TSH Flex Reagent Cartridge (REF K6412). Intended to quantitatively measure human thyroid stimulating hormone (TSH, thyrotropin). Class II

Reason: A rare variant of TSH, identified in a small cluster of patients, is not detected.

Siemens Healthcare Diagnostics, Inc 20130510 Terminated
Distributed: Worldwide Distribution-USA (nationwide) and the countries of Canada, AE, AR, AT, AU, BA, BD, BE, BG, BO, BR, BY, CH, CL, CM, CN, CO, CR, CY, CZ, DE, DK, DO, DZ, EC, EE, EG, ES, FI, FR, GB, GR, GT, HK, HN, HR, HU, ID, IE, IL, IN, IR, IS, IT, JO, JP, KA, KR, KW, KZ, LB, LK, LT, LV, LY, MA, MK, MT, MX, MY, NI, NL, NO, NP, OM, PA, PE, PH, PK, PL, PT, PY, QA, RO, RU, SA, SE, SG, SK, SV, SY, TH, TM, TN, TR, TW, UA, UY, UZ, VE, VN, XS, and ZA.
Dimension EXL LOCI Module TSHL Flex Reagent Cartridge (REF RF612). Intended to quantitatively measure human thyroid stimulating hormone (TSH, thyrotropin). Class II

Reason: A rare variant of TSH, identified in a small cluster of patients, is not detected.

Siemens Healthcare Diagnostics, Inc 20130510 Terminated
Distributed: Worldwide Distribution-USA (nationwide) and the countries of Canada, AE, AR, AT, AU, BA, BD, BE, BG, BO, BR, BY, CH, CL, CM, CN, CO, CR, CY, CZ, DE, DK, DO, DZ, EC, EE, EG, ES, FI, FR, GB, GR, GT, HK, HN, HR, HU, ID, IE, IL, IN, IR, IS, IT, JO, JP, KA, KR, KW, KZ, LB, LK, LT, LV, LY, MA, MK, MT, MX, MY, NI, NL, NO, NP, OM, PA, PE, PH, PK, PL, PT, PY, QA, RO, RU, SA, SE, SG, SK, SV, SY, TH, TM, TN, TR, TW, UA, UY, UZ, VE, VN, XS, and ZA.
Dimension TSH Flex Reagent Cartridge (REF RF412). Intended to quantitatively measure human thyroid stimulating hormone (TSH, thyrotropin). Class II

Reason: A rare variant of TSH, identified in a small cluster of patients, is not detected.

Siemens Healthcare Diagnostics, Inc 20130510 Terminated
Distributed: Worldwide Distribution-USA (nationwide) and the countries of Canada, AE, AR, AT, AU, BA, BD, BE, BG, BO, BR, BY, CH, CL, CM, CN, CO, CR, CY, CZ, DE, DK, DO, DZ, EC, EE, EG, ES, FI, FR, GB, GR, GT, HK, HN, HR, HU, ID, IE, IL, IN, IR, IS, IT, JO, JP, KA, KR, KW, KZ, LB, LK, LT, LV, LY, MA, MK, MT, MX, MY, NI, NL, NO, NP, OM, PA, PE, PH, PK, PL, PT, PY, QA, RO, RU, SA, SE, SG, SK, SV, SY, TH, TM, TN, TR, TW, UA, UY, UZ, VE, VN, XS, and ZA.
ADVIA Centaur Systems TSH3 Ultra ReadyPack ((REF 06491072 - 100 Tests, REF 06491080 - 500 Tests, REF 04862625 - 2500T). Intended to quantitatively measure human thyroid stimulating hormone (TSH, thyrotropin). Class II

Reason: A rare variant of TSH, identified in a small cluster of patients, is not detected.

Siemens Healthcare Diagnostics, Inc 20130510 Terminated
Distributed: Worldwide Distribution-USA (nationwide) and the countries of Canada, AE, AR, AT, AU, BA, BD, BE, BG, BO, BR, BY, CH, CL, CM, CN, CO, CR, CY, CZ, DE, DK, DO, DZ, EC, EE, EG, ES, FI, FR, GB, GR, GT, HK, HN, HR, HU, ID, IE, IL, IN, IR, IS, IT, JO, JP, KA, KR, KW, KZ, LB, LK, LT, LV, LY, MA, MK, MT, MX, MY, NI, NL, NO, NP, OM, PA, PE, PH, PK, PL, PT, PY, QA, RO, RU, SA, SE, SG, SK, SV, SY, TH, TM, TN, TR, TW, UA, UY, UZ, VE, VN, XS, and ZA.
Page 3,772 of 4,030 | Showing 10 recalls
Data provided by openFDA