GE Healthcare, Innova 3100 and Innova 4100 systems. The Digital Fluoroscopic Imaging System is indicated for use in diagnostic and interventional angiographic procedures of human anatomy. It is intended to replace image intensifier fluoroscopic systems in all diagnostic or interventional procedures. Class II

Reason: GE Healthcare reported a potential safety issue due to temporary loss of X-Ray Imaging Modes during Collimator Auto-Calibration which affects Innova 3100 and Innova 4100 Systems.

GE Healthcare, LLC 20110516 Terminated
Distributed: Worldwide distribution - USA including AZ, CA, CO, FL, ID, IL, ME, NV,NY, NC, OK, TN, UT, and WI; and Internationally to Canada, Chile, France, Germany, Italy, Japan, Korea, Norway and Spain.
VITROS Chemistry Products CRP Slides (Reagent), REF/Catalog Number 192 6740 and 809 7990, Product Usage: For in vitro diagnostic use only. VITROS Chemistry Products CRP Slides quantitatively measure C-reactive protein (CRP) concentration in serum and plasma using VITROS 250/350/950/5,1 FS and 4600 Chemistry Systems and the VITROS 5600 Integrated System. Class III

Reason: While performing interval testing Ortho Clinical Diagnostics (OCD) determined that an excess amount of surfactant was unintentionally added to VITROS Chemistry Products CRP Slides during the manufacturing process.

Ortho-Clinical Diagnostics 20130509 Terminated
Distributed: Worldwide Disribution - USA Nationwide including Puerto Rico, and the countries of Australia, Bermuda, Brazil, Canada, Chile, China, Colombia, England, France, Germany, India, Italy, Japan, Mexico, Republic of Panama, Singapore, Spain and Venezuela.
VITROS Chemistry Products K+ (Potassium) Slides (Reagent), REF/Catalog Number 815 7596, Product Usage: For in vitro diagnostic use only. VITROS Chemistry Products K+ Slides quantitatively measure potassium (K+) concentration in serum, plasma, and urine using VITROS 250/350/950/5,1 FS and 4600 Chemistry Systems and the VITROS 5600 Integrated System. Class III

Reason: While performing interval testing, Ortho Clinical Diagnostics (OCD) determined that a surfactant was unintentionally added to VITROS Chemistry Products K+ Slides during the manufacturing process.

Ortho-Clinical Diagnostics 20130509 Terminated
Distributed: Worldwide Disribution - USA Nationwide including Puerto Rico, and the countries of Australia, Bermuda, Brazil, Canada, Chile, China, Colombia, England, France, Germany, India, Italy, Japan, Mexico, Republic of Panama, Singapore, Spain and Venezuela.
SYNCHRON System(s) Lactate (LACT) Reagent. Product Usage: Lactate (LACT) reagent, when used in conjunction with Synchron LX¿ System(s), UniCel¿ DxC 600/800 System(s) and Synchron¿ Systems Multi Calibrator, is intended for the quantitative determination of lactate concentration in human plasma and cerebrospinal fluid (CSF). Class II

Reason: Beckman Coulter has confirmed customer reports of LACT Lot M210077 failing calibration with error condition OCR LOW. The calibration failures appear to be related to improper shipping or storage conditions (frozen reagent). Failed calibration of LACT may cause a delay in reporting results.

Beckman Coulter Inc. 20130528 Terminated
Distributed: Worldwide Distribution - USA Nationwide and country of Canada.
Terumo Glidesheath (tm), 6 French, Nitinol wire, product code 15-1060. Used to facilitate placing a catheter through the skin into a vein or artery. Class II

Reason: Medical devices were incorrectly labeled with extended expiration dates.

Terumo Medical Corporation 20130530 Terminated
Distributed: Nationwide Distribution.
Pinnacle¿ Destination¿ Peripheral Guiding Sheath, 8 French, 90cm, Straight Tip, Cross Cut Valve. Acts as a conduit to facilitate the introduction of interventional and diagnostic devices into the human vasculature. Class II

Reason: Medical devices were incorrectly labeled with extended expiration dates.

Terumo Medical Corporation 20130530 Terminated
Distributed: Nationwide Distribution.
Brand Name: Multiday lnfusor, Seven Day lnfusor and the Basal/Bolus lnfusor: Single day Infusor 2 ml/h, Single Day INFUSOR 2 ml/h System, Half Day INFUSOR SV 5 ml/h System, Two Day INFUSOR 2 ml/h System, Two day Infusor 2 ml/h, Multiday Infusor 0.5 ml/h, Multiday INFUSOR 0.5ml/h System, Seven Day Infusor 0.5 ml/h, Seven Day INFUSOR 0.5ml/h System, Basal-Bolus Infusor-15min., Basal/Bolus INFUSOR 0.5 x 2 ml/h System with 15 Minute Lockout, Basal-Bolus Infusor Device Listing: D018045 The Multiday lnfusor. Seven Day lnfusor and the BasaVBolus lnfusor are indicated for patients requiring slow, continuous intravenous, intra-arterial, subcutaneous or epidural administration of medications. Class II

Reason: Due to an increase in complaints for leaks at the distal male Luer and Luer cap. Baxter implemented enhancements to the blue winged cap to improve the seal between the cap and the Luer.

Baxter Healthcare Corp. 20130607 Terminated
Distributed: Worldwide Distribution - USA (nationwide), Puerto Rico, and Internationally to Brazil, Colombia, Panama, Mexico, Singapore Japan, Australia, New Zealand, Hong Kong, India, Thailand, Taiwan, and Korea.
Brand Name: Coiled Tube Infusors (Portable Elastomeric Infusion Systems). Baxter's Infusor SV Elastomeric Infusion Device Device Listing: D020871 Baxter's lnfusors SV Elastomeric Infusion Device is a single-use, disposable elastomeric infusion pump indicated tor patients requiring slow, continuous intravenous, intra-arterial, subcutaneous, or epidural administration of medications at a constant flow rate. Class II

Reason: Due to an increase in complaints for leaks at the distal male Luer and Luer cap. Baxter implemented enhancements to the blue winged cap to improve the seal between the cap and the Luer.

Baxter Healthcare Corp. 20130607 Terminated
Distributed: Worldwide Distribution - USA (nationwide), Puerto Rico, and Internationally to Brazil, Colombia, Panama, Mexico, Singapore Japan, Australia, New Zealand, Hong Kong, India, Thailand, Taiwan, and Korea.
Cordis a Johnson & Johnson Company, RadialSource Spring Wire Transradial Access Kit. Manufacturer Greatbatch Medical 2300 Berkshire Lane North, Minneapolis, MN 55441, Distributor Cordis Corporation 14201 North West 60th Avenue, Miami Lakes, Florida 33014. Baxter's lnfusors SV Elastomeric Infusion Device is a single-use, disposable elastomeric infusion pump indicated tor patients requiring slow, continuous intravenous, intra-arterial, subcutaneous, or epidural administration of medications at a constant flow rate. The device is also indicated for the administration of bolus doses of medication upon patient demand when used in conjunction with the Patient Control Module. lt is suitable for use in the hospital or home setting. Class II

Reason: Cordis Corporation, Inc and Greatbatch Medical are recalling one lot of the cordis RADIALSOURCE Spring Wire Transradial Access Kit product. Some units were manufactured with an undersized vessel dilator tip, too small to slide over the guidewire.

Greatbatch Medical 20130531 Terminated
Distributed: Worldwide Distribution - AUSTRIA, BELGIUM, SWITZERLAND, GERMANY, SPAIN, GREECE, HUNGARY, IRELAND, ISRAEL and ITALY.
Medtronic, Consulta¿ CRT-P, Model C4TR01, C3TR01 (not approved for US), and Syncra¿ CRT-P, Model C2TR01, Sterile EO. Dual chamber and atrial tracking modes are indicated for patients who may benefit from maintenance of AV synchrony. Class II

Reason: Medtronic recently identified an issue with a subset of Consulta¿ CRT-P and Syncra¿ CRT-P devices during production. As of May 30, 2013, there have been no reported or confirmed device failures. However, because of the potential for malfunction, Medtronic is requiring the return of non-implanted devices manufactured between April 1 and May 13, 2013 for re-inspection. The issue is unique to specifi

Medtronic Inc. Cardiac Rhythm Disease Management 20130607 Terminated
Distributed: Worldwide Distribution - USA including CA, FL, GA, KY, LA, LMN, MO, NE, NY, OK, PA, TN, TX, WI, and PR and Internationally to Australia, Austria, Belgium, Brazil, Bulgaria, Canada, Chile, China, Colombia, Denmark, Ecuador, Egypt, Finland, France, Germany, Hungary, Ireland, Italy, Japan, Latvia, Lebanon, Netherlands, Norway, Poland, Singapore, South Africa, Spain, Sweden, Switzerland, Taiwan, Tunisia ,United Arab Emirates, United Kingdom, Uruguay , and Yemen. UPDATED Physicians location in USA: AL, AR, FL, GA, IL, IN, KS. LA, ME, MI, MN, MO, NY, OH, PA, SC, TN, TX and VA. OUS: Luxembourg
Page 3,799 of 4,030 | Showing 10 recalls
Data provided by openFDA