TX1 Tissue Removal System disposable handpieces. Product Usage: The TX1 Tissue Removal System is indicated for use in surgical procedures where fragmentation, emulsification and aspiration of soft tissue are desirable, including General Surgery, Orthopedic Surgery, Laparoscopic Surgery and Plastic and Reconstructive Surgery. Class II

Reason: American Optisurgical, Inc. is recalling certain lots of TX1 Tissue Removal System disposable handpiece due to inventory being released into distribution without a cleared FDA 510(k) to cover recent product changes. Use of this product may lead to an underperformance of the device.

American Optisurgical Inc 20130430 Terminated
Distributed: Worldwide Disitribution - USA Nationwide in the states of Alabama, Arkansas, Arizona, California, Colorado, Delaware, Florida, Georgia, Iowa, Illinois, Indiana, Kansas, Kentucky, Massachusetts, Maryland, Michigan, Minnesota, Missouri, North Carolina, Nebraska, New Hampshire, New Jersey, New Mexico, Nevada, New York, Ohio, Oklahoma, Oregon, Pennsylvania, South Carolina, Tennessee, Texas, Utah, Virginia, Vermont, Washington, Wisconsin, West Virginia, Wyoming and the countries of Quezon City, Philippines, Singapore, and Panama City, Panama.
Brand Name: AFX" Introducer System, Model Number S17-45. Lot Numbers: 1079840, 1079843, 1079844, 1079845. Product Usage: The AFX Introducer System is intended for use to facilitate the introduction of catheters and other medical devices into the vasculature and to minimize blood loss associated with such introduction. Class I

Reason: Endologix, Inc. initiated this voluntary recall of certain lots of Endologix, Inc. due to reports of dilator breakage during clinical procedures.

Endologix Inc 20130513 Terminated
Distributed: Worldwide Distribution - USA Nationwide in the states of: NY, NH, IN, MI, FL and NJ and in the countries of: Europe and Latin America.
lnvatec SpA brand Diver C.E., Model/Catalog Numbers: DVR010014145, DVR011014145, Product is manufactured by lnvatec SpA, Roncadelle, Italy, and distributed by Invatec, Inc., Bethlehem, PA and by Medtronic Endovascular Therapies, Santa Rosa, CA The Diver C.E. product is an aspiration catheter indicated for the removal of fresh, soft emboli and thrombi from vessels in the arterial system. Class II

Reason: Devices labeled for an intended use was not included in the existing 510(k) or PMA, in that, the Indications for use was modified.

Invatec Llc 20090304 Terminated
Distributed: Nationwide Distribution
American Optisurgical's TX1 Tissue Removal System Console. The console is a software unit that is indicated for use in surgical procedures. Class II

Reason: The recall has been initiated due to an overheating protection voltage suppressor on the redundant safety protection circuit board. Secondary to the recall, American Optisurgical, Inc. will be making a correction to add spacers between the valve and cassette to correct a small irrigation leak that can occur after the priming cycle. Use of this product may cause the device to lose power and stop wo

American Optisurgical Inc 20130330 Terminated
Distributed: Worldwide distribution, US, including the states of United States: Alabama, Arkansas, Arizona, California, Colorado, Florida, Georgia, Iowa, Illinois, Indiana, Kansas, Kentucky, Massachusetts, Maryland, Michigan, Minnesota, Missouri, North Carolina, Nebraska, New Jersey, New York, Ohio, Pennsylvania, South Carolina, Tennessee, Texas, Utah, Virginia, Washington Wisconsin, West Virginia and the countries of Singapore and the Philippines.
ADVIA Centaur¿ iPTH Calibrator 6 pk; REF: 10492387, 10630628. For in vitro diagnostic use in calibrating Intact PTH assays. Class II

Reason: iPTH Assay- reagent and calibrator lots may have demonstrated higher than anticipated variability

Siemens Healthcare Diagnostics, Inc 20130514 Terminated
Distributed: Nationwide Distribution.
ADVIA Centaur iPTH Calibrator 2 pk; REF 10492394. For in vitro diagnostic use in calibrating Intact PTH assays. Class II

Reason: iPTH Assay- reagent and calibrator lots may have demonstrated higher than anticipated variability

Siemens Healthcare Diagnostics, Inc 20130514 Terminated
Distributed: Nationwide Distribution.
ADVIA Centaur¿ iPTH ReadyPack Assay 500 Tests; REF 10492369, 10493615, 1049361. For in vitro diagnostic use in the quantitative determination of intact parathyroid hormone (iPTH). Class II

Reason: iPTH Assay- reagent and calibrator lots may have demonstrated higher than anticipated variability

Siemens Healthcare Diagnostics, Inc 20130514 Terminated
Distributed: Nationwide Distribution.
ADVIA Centaur¿ iPTH ReadyPack Assay 100 Tests; REF 10492368. For in vitro diagnostic use in the quantitative determination of intact parathyroid hormone (iPTH). Class II

Reason: iPTH Assay- reagent and calibrator lots may have demonstrated higher than anticipated variability

Siemens Healthcare Diagnostics, Inc 20130514 Terminated
Distributed: Nationwide Distribution.
FLEXSTENT Femoropopliteal Self-Expanding Stent System. Product Usage: The FSS FLEXSTENT¿ Femoropopliteal SES clinical product is indicated for treatment of atherosclerotic de novo native superficial femoral artery lesions and proximal popliteal lesions. Class II

Reason: Holes and/or tears can be introduced in the FLEXSTENT¿ inner pouch under some conditions, and to a lesser degree, in the outer pouch.

Cordis Corporation 20130423 Terminated
Distributed: Worldwide Distribution - USA (nationwide) and the states of: AZ, CA, CT, FL, GA, IA, IL, LA, MD, MI, MS, NC, NJ, NY, OH, PA, RI, SC, TX, WA, WI, WV, and the countries of: Czech Republic, Italy, Netherlands, and New Zealand
FLEXSTENT Femoropopliteal Self-Expanding Stent System. Product Usage: The FSS FLEXSTENT¿ Femoropopliteal SES clinical product is indicated for treatment of atherosclerotic de novo native superficial femoral artery lesions and proximal popliteal lesions. Class II

Reason: Holes and/or tears can be introduced in the FLEXSTENT¿ inner pouch under some conditions, and to a lesser degree, in the outer pouch.

Cordis Corporation 20130423 Terminated
Distributed: Worldwide Distribution - USA (nationwide) and the states of: AZ, CA, CT, FL, GA, IA, IL, LA, MD, MI, MS, NC, NJ, NY, OH, PA, RI, SC, TX, WA, WI, WV, and the countries of: Czech Republic, Italy, Netherlands, and New Zealand