Stryker Instruments 0400-710-000 T4 Pullover Toga (S/M) Sterile, EO; single use; Product Usage: Protection System and are intended to protect the patient, healthcare personnel and operating room personnel against contamination, exposure of infectious bodily fluids, and the transfer of microorganismsand particulate material. Class II

Reason: Based on an accelerated aging test for this product packaging, there is an indicator of potential for a breach in the sterile barrier in the header bag and potentially the blue sterilization wrap for togas within the second and third year of shelf life. The test method used involved high heat to accelerate the aging of the packaging. Stryker is conducting additional testing and assessing this pr

Stryker Instruments Div. of Stryker Corporation 20130523 Terminated
Distributed: USA Nationwide Distribution
Hand Controls for Operon D750, D752, D760, D820 and D850 The hand controls are used to operate OPERON surgical tables Class II

Reason: The firm received complaints for unintended movements.

Berchtold Corp. 20130312 Terminated
Distributed: Nationwide Distribution
AGFA Digital Radiography X-Ray System DX-D 100 Agfa's DX-D 100 is indicated for use in providing diagnostic quality images to aid the physician with diagnosis. Systems can be used with MUSICA2 image processing to create radiographic images of the skeleton including skull, spinal column and extremities: chest, abdomen, and other body parts. Class II

Reason: Customers could potentially experience intermittent, unintended and illogical movement when using the product.

AGFA Corp. 20130515 Terminated
Distributed: Worldwide Distribution - US Nationwide including the states of AL, FL, IL, IN, IA, KY, MA, MI, MN, MS, MO, NE, NJ, NY, NC, OH, OR, SC, TN, VA, WV, WY and the country of Canada
Digitex Absorbable Polydiaxanone Suture, Size 2-0. Catalog Number / Code of product: 52031, 5203101400. The Digitex Delivery Device is a sterile, disposable device intended to deliver a suture to the operative site. The device assists in suturing by passing a needle through the tissue to capture the suture, and suture ligation and know-typing by holding the suture. Class II

Reason: Coloplast is conducting a recall on three lots of Digitex Absorbable Polydiaxanone Suture, Size 2-0 because it is mislabeled with a three year expiration date.

Coloplast Manufacturing US, LLC 20130513 Terminated
Distributed: US Distribution to: KY, MI, OH, PR, TX, and VA.
Zimmer Air Dermatome II, REF 00-8851-001-01 handpiece, w/o hose, Rx, non-sterile grafting instruments intended to provide variable graft thickness and width capabilities. Class II

Reason: This removal is in response to complaints alleging that the Zimmer Air Dermatomes and Zimmer Air Dermatome II products would either not operate or operate intermittently. Zimmer's investigation determined that the planetary gear teeth were broken. This may cause a delay or postponement of the surgery or result in a suboptimal graft retrieval with potential of requiring an additional donor site.

Zimmer Surgical Inc 20130507 Terminated
Distributed: Worldwide distribution: USA Nationwide including Puerto Rico; UAE, TAIWAN, SWEDEN, THAILAND, KOREA, SINGAPORE, HONG KONG CHINA, CHILE, CANADA, AUSTRALIA, JAPAN, ENGLAND, GERMANY, and EGYPT.
Zimmer Air Dermatome II, REF 00-8851-001-00 handpiece Rx, non-sterile grafting instruments intended to provide variable graft thickness and width capabilities Class II

Reason: This removal is in response to complaints alleging that the Zimmer Air Dermatomes and Zimmer Air Dermatome II products would either not operate or operate intermittently. Zimmer's investigation determined that the planetary gear teeth were broken. This may cause a delay or postponement of the surgery or result in a suboptimal graft retrieval with potential of requiring an additional donor site.

Zimmer Surgical Inc 20130507 Terminated
Distributed: Worldwide distribution: USA Nationwide including Puerto Rico; UAE, TAIWAN, SWEDEN, THAILAND, KOREA, SINGAPORE, HONG KONG CHINA, CHILE, CANADA, AUSTRALIA, JAPAN, ENGLAND, GERMANY, and EGYPT.
The Zimmer Air Dermatome, REF 00- 8801- 001- 00 , Rx, non-sterile grafting instruments intended to provide variable graft thickness and width capabilities Class II

Reason: This removal is in response to complaints alleging that the Zimmer Air Dermatomes and Zimmer Air Dermatome II products would either not operate or operate intermittently. Zimmer's investigation determined that the planetary gear teeth were broken. This may cause a delay or postponement of the surgery or result in a suboptimal graft retrieval with potential of requiring an additional donor site.

Zimmer Surgical Inc 20130507 Terminated
Distributed: Worldwide distribution: USA Nationwide including Puerto Rico; UAE, TAIWAN, SWEDEN, THAILAND, KOREA, SINGAPORE, HONG KONG CHINA, CHILE, CANADA, AUSTRALIA, JAPAN, ENGLAND, GERMANY, and EGYPT.
Straight Impactor - Dorado IBC, Part Number: SI02140. The device is used to position implants after they have been inserted into the disc space during implantation of intervertebral body fusion devices. Class II

Reason: During two separate occasions an impactor plate on the handle broke off while being struck by a mallet.

SpineFrontier, Inc. 20080510 Terminated
Distributed: Worldwide distribution: US (nationwide) to states of: CO, MA, and TX; and country of: Jamaica.
BrightView SPECT gamma camera for Emission Computed Tomography, Philips Medical Systems, Cleveland, OH 44143 BrightView SPECT is a gamma camera designed for single or dual detector nuclear imaging accommodating a broad range of emission computed tomography Emission Computed Tomography (ECT) studies. Class II

Reason: Philips has received one report from the field that there was an unexpected motion of the detector due to a failure of the mechanical assembly.

Philips Medical Systems (Cleveland) Inc 20130425 Terminated
Distributed: Worldwide Distribution - USA including AK, AR, AZ, CA, CO, CT, DC, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NJ, NM, NV, NY, OH, OK, OR, PA, PR, SC, SD, TN, TX, VA, WA, WI, WV & WY. Internationally to Australia, Austria, Bahrain, Belgium, Brazil, Canada, China, Czech Republic, Denmark, Dominican Republic, Egypt, France, Germany, India, Indonesia, Iran, Italy, Japan, Japan, Korea, Lebanon, Lithuania, Malaysia, Mexico, Portugal, Russian Federation, Saudi Arabia, Singapore, Slovenia, South Africa, Spain, Syria, Thailand, Turkey, Taiwan, Ukraine, United Arab Emirates, United Kingdom, Venezuela & Viet Nam.
E-Z WIDER BARIATRIC CHAIR BED; Single Motion Expandable Width Technology Class II

Reason: An EZ Wider bed (capacity of 1000 lb. safe working load collapsed during patient use, leaving the patient in a forward tilt position.

Kreg Medical Inc. 20120802 Terminated
Distributed: Nationwide Distribution including IL and IN.