Fluoroscan Mini C-arm InSight, software versions 1.x, 2.x, 3.x The device is a Mini C-arm Fluroscopic Imaging System designed to provide physicians with general fluoroscopic visualization of a patient, including but no limited to surgical orthopedic and podiatry use, critical and emergency procedures, and light anatomy imaging situation. Class II

Reason: The audible alarm to alert the operator when cumulative irradiation time reaches five minutes is not accurate and the total cumulative irradiation time is not accurate.

Hologic, Inc. 20130425 Terminated
Distributed: Worldwide distribution: USA (nationwide) and countries of: DE, RU, AU, IT, GB, ES, FR, KR, ZA, CA, AR, LT, TW, CN, BE, AT, JP, QA, SA, SG, HK, SI, NL, and SE.
Varian High Energy Clinacs, High Energy Accelerator, Radiation Treatment System, Model Numbers: H14, H27, H29, HCX. Product Usage: The Varian High Energy Accelerator is intended to provide stereotactic radiosurgery and precision radiotherapy for lesions, tumors and conditions anywhere in the body when radiation treatment is indicated. Class II

Reason: Under certain conditions, photon beams in High Energy Clinacs may experience a gradual change in beam symmetry, potentially reaching an asymmetry of up to about 7% before interlock occurs. At maximum asymmetry, this may result in no more than about 3.5% dose deviation from expected at any point in the beam, and may result in minor injury to the patient or slightly decreased local disease control.

Varian Medical Systems, Inc. Oncology Systems 20110531 Terminated
Distributed: Worldwide distribution - USA (nationwide) including Puerto Rico, and countries of: Algeria, Angola, Australia, Austria, Bahrain, Bangladesh, Belarus, Belguim, Brazil, Bosnia and Hezregovina, Bulgaria, Canada, Chile, China, Columbia, Cyprus, Denmark, Dominican Republic, Ecuador, Egypt, Estonia, Finland, Frances, French Polynesia, Georgia, Germany, Greece, Gaudaloupe, Gaum, Hong Kong, India, Indonesia, Ireland, Italy, Israel, Latvia, Lithuania, Luxemburg, Japan, Kazarhstan, Kuwait, Lebanon, Macau, Macedonia, Malaysia, Martinique, Mauritania, Mexico, Moldova, Monaco, Morroco, Nepal, Netherlands, New Zealand, Norway, Oman, Pakistan, Panama, Peru, Phillipines, Poland, Portugal, Qatar, Reunion, Romania, Russia, Saudia Arabia, Serbia, Singapore, Slovakia, Spain, South Africa, South Korea, Sri Lanka, Syria, Sweden, Switzerland, Taiwan, Thailand, Trinidad and Tobago, Tunisia, Turkey, Ukraine, United Arab Emirates, United Kingdom, Uruguay, Venezula, Vietnam, and Zimbabwe.
Segmented Cylinder Applicator Set, Model # GM11004150, when used for Pulsed Dose Rate (PDR) Product Usage: Usage: Used during pulsed dose rate brachytherapy to treat vaginal and rectal cancer. Class II

Reason: Segmented cylinder applicator set may slip during treatment, causing the delivery of radiation to areas outside the target.

Varian Medical Systems, Inc. 20130411 Terminated
Distributed: Worldwide distribution: USA (nationwide) and in the countries of: Belgium, France, Martinique, Slovenia, Sweden, and the United Kingdom.
Heartware HVAD Pump Implant Kit (Heartware Ventricular Assist Device) Class II

Reason: HeartWare, Inc. of Miami Lakes, FL is recalling their Ventricular Assist System due to the loosening of the HVAD Pump driveline connector.

Heartware Inc 20121227 Terminated
Distributed: Worldwide Distribution-USA (nationwide) including the states of MN, IL, MI, FL, WI, TX, MO, CA, MA, TN, OH, NC, GA, VA, UT, KY, MD, IN, PA, NY, NJ, LA, OR, WA, AL and CO, and the countries of Poland, Australia, Austria, Italy, Germany, New Zealand, South Africa, Czech Republic, France, Brazil, Switzerland, Singapore, United Kingdom, Saudi Arabia, Netherlands, China, Chile, Belgium, Canada, Norway, Lebanon, Israel, Malaysia, Turkey, India, Argentina, Greece, Japan and Lithuania.
Proplege Coronary Sinus Catheter model PR9, 11 French Introducer, kitted with PR9. Product Usage: The catheter introducer sheath kit provides access to the central venous circulation to facilitate catheter and guidewire insertion. The Proplege coronary sinus catheter provides retrograde cardioplegia during cardiopulmonary by-pass surgery. Class II

Reason: Edwards Lifesciences is recalling certain Proplege Coronary Sinus Catheter, model PR9 due to confirmed reports of blood leakage through the hemostasis valve in the introducer that is kitted with the catheter.

Edwards Lifesciences, LLC 20130125 Terminated
Distributed: Worldwide Distribution - USA Nationwide including the states of: MD, NY, PA, CA, FL, AZ, AL, TX, VA, OH, MO, NC, GA, TN, CO, OR, SC, ME, IL, HI, NJ, WA, WI, CT, KY, LA and Canada.
QIAGEN artus CMV RG PCR ASR (24) (Catalog number 4503223) Product Usage: The artus CMV RG PCR ASR is designed to detect and to quantify nucleic acid specific to human cytomegalovirus (CMV) using real-time PCR technology (PCR) in the Rotor-Gene 2000/3000 (RG). Class II

Reason: CMV RG PCR kits may contain a mixture of incorrect vials among the correct vials.

QIAGEN Gaithersburg, Inc. 20130130 Terminated
Distributed: USA Nationwide Distribution including the states of: CA, DC, FL, GA, MA, MD, MI, and IL.
QIAGEN artus CMV RG PCR ASR (96) (Catalog number 4503225) Product Usage: Test is intended for use in the detection and quantitation of human cytomegalovirus (CMV) specific DNA, as part of an in house validated real-time PCR assay. One service package containing the handbook. Class II

Reason: CMV RG PCR kits may contain a mixture of incorrect vials among the correct vials.

QIAGEN Gaithersburg, Inc. 20130130 Terminated
Distributed: USA Nationwide Distribution including the states of: CA, DC, FL, GA, MA, MD, MI, and IL.
Summit Medical, BL-2006, Myringotomy Lance (Juvenile) Blade, Sterile P, Summit Medical, Inc., 815 Northwest Parkway, Suite 100, St. Paul, MN 55121 Use to make an incision in the tympanic membrane. Class II

Reason: Summit Medical has initiated a recall of Myringotomy Lance Blade (Juvenile) because the box contains the wrong blade. The box contains Myringotomy Spear Blade (Juvenile).

Summit Medical, Inc. 20130327 Terminated
Distributed: Distributed in MS, NY, PA and TX.
GE Healthcare Innova 2100 IQ, Innova 3100 IQ, Innova 4100 IQ, Innova 2121 IQ, Innova 3131 IQ , Innova 3100, Innova 4100. For use in generating fluoroscopic images of human anatomy for diagnostic and interventional procedures. Class II

Reason: While performing a fluoroscopic examination on the GE Innova systems, there is a potential of nonrecoverable loss of imaging modes with no advance warning, due to a relay failure in the power distribution box (PDB). Although no injuries have been reported due to this issue to date, there is risk of complications during the procedure as a result of imaging functionality loss.

GE Healthcare, LLC 20111220 Terminated
Distributed: Worldwide Distribution -- Nationwide including the states of AL, AZ, AR, CA, CO, CT, DC, FL, GA, GUAM, HI, ID, IL , IN, KS, KY, LA, ME, MD, MA, MI, MN, MS, MO, MT, NE, NV, NH, NH, NM, NY, NC, ND, OH, OK, OR, PA, PR, SC, TN, TX, UT, VT, VA, WA, WY and WI., and the countries of Philippines, Peru, Paraguay, Panama, Mexico, Malaysia, Korea, Japan, Guatemala, Ecuador, Dominican Republic , Columbia, China, Chile, Canada, Brazil.,Argentina and Venezuela
PadPro(R) Adult Radiotransparent Multifunction Electrodes, REF/Catalog No. 2001M-C. The PadPro(R) Adult Radiotransparent Multifunction Electrodes with Carbon Leadwires and Medtronic(R) Connector are intended to be used in external pacing, defibrillation and monitoring applications Class II

Reason: ConMed was notified by the contract manufacturer that their wire/connector supplier reported damage to the connector tooling which could pinch or compress the lead wire filaments and potentially render the device incapable of delivering the appropriate energy or shock to the patient and/or could cause potential injury to user or patient.

ConMed Corporation 20130503 Terminated
Distributed: Worldwide distribution: USA (nationwide) and the country of: Saudi Arabia.
Page 3,828 of 4,030 | Showing 10 recalls