Presource PBDS, Knee Arthroscopy, Kit, Circulator Class I

Reason: Various Presource Kits contain a pre-assembled Filter and Anesthesia Circuit that may contain outer plastic packaging on one or more components. If the packaging is removed without disassembling the components, remnants of the plastic from the packaging material may become lodged in the filter potentially causing an obstruction of airflow.

Cardinal Health, Medical Products & Services 20130326 Terminated
Distributed: Distributed in the states of AR, MS, SC, and TN.
Presource PBDS, Major Abdominal, Kit, Circulator Class I

Reason: Various Presource Kits contain a pre-assembled Filter and Anesthesia Circuit that may contain outer plastic packaging on one or more components. If the packaging is removed without disassembling the components, remnants of the plastic from the packaging material may become lodged in the filter potentially causing an obstruction of airflow.

Cardinal Health, Medical Products & Services 20130326 Terminated
Distributed: Distributed in the states of AR, MS, SC, and TN.
Presource PBDS, General Laparoscopy, Kit, Circulator Class I

Reason: Various Presource Kits contain a pre-assembled Filter and Anesthesia Circuit that may contain outer plastic packaging on one or more components. If the packaging is removed without disassembling the components, remnants of the plastic from the packaging material may become lodged in the filter potentially causing an obstruction of airflow.

Cardinal Health, Medical Products & Services 20130326 Terminated
Distributed: Distributed in the states of AR, MS, SC, and TN.
A) Presource PBDS, Lap Chole, Kit, Clean Up; B) Presource PBDS, Greenwood LeFlore Hosp, Lap Chole, Kit Class I

Reason: Various Presource Kits contain a pre-assembled Filter and Anesthesia Circuit that may contain outer plastic packaging on one or more components. If the packaging is removed without disassembling the components, remnants of the plastic from the packaging material may become lodged in the filter potentially causing an obstruction of airflow.

Cardinal Health, Medical Products & Services 20130326 Terminated
Distributed: Distributed in the states of AR, MS, SC, and TN.
Presource PBDS, Gyn Laparoscopy Kit, Circulator Class I

Reason: Various Presource Kits contain a pre-assembled Filter and Anesthesia Circuit that may contain outer plastic packaging on one or more components. If the packaging is removed without disassembling the components, remnants of the plastic from the packaging material may become lodged in the filter potentially causing an obstruction of airflow.

Cardinal Health, Medical Products & Services 20130326 Terminated
Distributed: Distributed in the states of AR, MS, SC, and TN.
NxStage PureFlow-B Solution, Model/Catalog Number: RFP-401, packaged in a case The NxStage PureFLow-B Solution is indicated for use with renal replacement therapy systems that utilize sterile premixed dialysate during hemodialysis. Class II

Reason: Product may be mislabeled.

NxStage Medical, Inc. 20130318 Terminated
Distributed: Nationwide Distribution including IL, SC, TX, VA, FL, AR, and NY.
3DKnee e+ Tibial Insert. The insert is used along with the other components that comprise the 3DKnee System. Class II

Reason: U.S. Agent contacted Director Commercial Logistics, notifying him of incorrect color coded labels on two packages of e+tibial inserts. There were 8 e+ tibial inserts of the same lot number in Finished Goods as the two from the agency. These devices were inspected and all were incorrectly labeled.

Encore Medical, Lp 20130401 Terminated
Distributed: US Distribution to the state of California and International Distribution to the country of Venezuela.
Optima MR430s MRI Scanner; 5000-0002 is indicated to use as a magnetic resonance imaging device of the leg (excluding the thigh), knee, ankle, foot, elbow, forearm, wrist, and hand. Class II

Reason: Coolant supply lines may leak liquid onto the floor resulting in a slippery surface and fall hazard.

GE Healthcare, LLC 20130328 Terminated
Distributed: Worldwide Distribution - USA including AL, CA, CO, FL, ID, IL, MA, NH, NJ, NY, OH, PA, SC, SD, TN, TX, UT, VA, WA, WI; Worldwide: Argentina, Australia, Austria, Brazil, Canada, Chile, Denmark, Egypt, Finland, Germany, India, Italy, Japan, Korea, Malaysia, Netherlands, Norway, Peru, Poland, Russia, Singapore, South Africa, Spain, Sweeden, Switzerland, UK, Ukraine, and UAE.
MSK EXTREME MR SCANNER, MODEL AA5000 The intended for use as a diagnostic imaging device to produce axial, sagittal, coronal and oblique images of the internal structures of the leg (excluding the thigh), knee, ankle, foot, forearm, elbow, wrist and hand. Class II

Reason: Coolant supply lines may leak liquid onto the floor resulting in a slippery surface and fall hazard.

GE Healthcare, LLC 20130328 Terminated
Distributed: Worldwide Distribution - USA including AL, CA, CO, FL, ID, IL, MA, NH, NJ, NY, OH, PA, SC, SD, TN, TX, UT, VA, WA, WI; Worldwide: Argentina, Australia, Austria, Brazil, Canada, Chile, Denmark, Egypt, Finland, Germany, India, Italy, Japan, Korea, Malaysia, Netherlands, Norway, Peru, Poland, Russia, Singapore, South Africa, Spain, Sweeden, Switzerland, UK, Ukraine, and UAE.
Stryker Howmedica Osteonics Corp. AxSOS System Instrument. The AxSOS Locking Plate System is designed to treat periarticular or intraarticular fractures of the Distal Femur, Proximal Humerus, Proximal Tibia, and the Distal Tibia. Per the operative technique, " A 4.3 mm Drill (REF 702743) is then used to drill the core hole for the locking screw. Using a fluoroscopy, check the correct depth of the drill, and measure the length of the screw. The Drill Sleeve should now be removed, and the correct length 5.0 mm Locking Screw is inserted using the Screwdriver T20 and Screw Holding Sleeve." Class II

Reason: Stryker Orthopaedics has received a report indicating that the scale of the AxSOS Calibrated Drill Bit 4.3,,X263mm, flat , manufactured by Stryker Osteosynthesis, is incorrect. The incorrect scale could result in the surgeon selecting a screw that is 10mm longer than intended.

Stryker Howmedica Osteonics Corp. 20130328 Terminated
Distributed: Nationwide Distribution including the states of AL, AZ, IL, KY, LA, NJ, NC, FL, CA, NM, MO, TN, VA, CO, MA, TX, IN and MT.