Belly Bag Urine Collection Bag with Hip Belt. The product is a sterile urine collection device that collects urine by mechanical means when attached to an indwelling Foley or suprapubic catheter. Class II

Reason: Sterile packaging may be compromised.

Teleflex Medical 20130328 Terminated
Distributed: Worldwide Distribution - USA and the countries of Singapore, Canada, New Zealand, Germany, Ireland, Japan and Australia.
APLIF Implants and Instruments, Part Numbers: 184008-10 Anterior Psoas Lumbar 18 x 40 x 08mm - 10degree 184010-10 Anterior Psoas Lumbar 18 x 40 x 10mm - 10degree 184012-10 Anterior Psoas Lumbar 18 x 40 x 12mm - 10degree 184014-10 Anterior Psoas Lumbar 18 x 40 x 14mm - 10degree 184016-10 Anterior Psoas Lumbar 18 x 40 x 16mm - 10degree 184008-00 Anterior Psoas Lumbar 18 x 40 x 08mm - 00degree 184010-00 Anterior Psoas Lumbar 18 x 40 x 10mm - 00degree 184012-00 Anterior Psoas Lumbar 18 x 40 x 12mm - 00degree 184014-00 Anterior Psoas Lumbar 18 x 40 x 14mm - 00degree 184016-00 Anterior Psoas Lumbar 18 x 40 x 16mm - 00degree 184510-08 Anterior Psoas Lumbar 18 x 45 x 08mm - 10degree 184510-10 Anterior Psoas Lumbar 18 x 45 x 10mm - 10degree 184512-10 Anterior Psoas Lumbar 18 x 45 x 12mm - 10degree 184514-10 Anterior Psoas Lumbar 18 x 45 x 14mm - 10degree 184516-10 Anterior Psoas Lumbar 18 x 45 x 16mm - 10degree 184510-00 Anterior Psoas Lumbar 18 x 45 x 10mm - 00degree 184512-00 Anterior Psoas Lumbar 18 x 45 x 12mm - 00degree 184514-00 Anterior Psoas Lumbar 18 x 45 x 14mm - 00degree 184516-00 Anterior Psoas Lumbar 18 x 45 x 16mm - 00degree 185008-10 Anterior Psoas Lumbar 18 x 50 x 08mm - 10degree 185010-10 Anterior Psoas Lumbar 18 x 50 x 10mm - 10degree 185012-10 Anterior Psoas Lumbar 18 x 50 x 12mm - 10degree 185014-10 Anterior Psoas Lumbar 18 x 50 x 14mm - 10degree 185016-10 Anterior Psoas Lumbar 18 x 50 x 16mm - 10degree 185008-00 Anterior Psoas Lumbar 18 x 50 x 08mm - 00degree 185010-00 Anterior Psoas Lumbar 18 x 50 x 10mm - 00degree 185012-00 Anterior Psoas Lumbar 18 x 50 x 12mm - 00degree 185014-00 Anterior Psoas Lumbar 18 x 50 x 14mm - 00degree 185016-00 Anterior Psoas Lumbar 18 x 50 x 16mm - 00degree 185508-10 Anterior Psoas Lumbar 18 x 55 x 08mm - 10degree 185510-10 Anterior Psoas Lumbar 18 x 55 x 10mm - 10degree 185512-10 Anterior Psoas Lumbar 18 x 55 x 12mm - 10degree 185514-10 Anterior Psoas Lumbar 18 x 55 x 14mm - 10degree 185516-10 Anterior Psoas Lumbar 18 x 55 x 16mm - 10degree 185508-00 Anterior Psoas Lumbar 18 x 55 x 08mm - 00degree 185510-00 Anterior Psoas Lumbar 18 x 55 x 10mm - 00degree 185512-00 Anterior Psoas Lumbar 18 x 55 x 12mm - 00degree 185514-00 Anterior Psoas Lumbar 18 x 55 x 14mm - 00degree 185516-00 Anterior Psoas Lumbar 18 x 55 x 16mm - 00degree The Eminent Spine Interbody Fusion System is indicated for intervertebral body fusion of the lumbar spine, from L2 to S1, in skeletally mature patients who have had six months of non-operative treatment. Class II

Reason: Spinal Solutions is recalling the APLIF system because it is not supported by adequate testing and documentation to demonstrate that it meets performance or safety standards. These inadequacies might result in product performance failures that could cause patient harm due to implant breakage, movement, or inadequate sterilization.

Spinal Solutions, LLC 20130308 Terminated
Distributed: Nationwide Distribution including NV, WI, MD,and CA.
Orthopedic Alliance Spine System (alias "Blue & Gold") Part Numbers: BGC3010 Screw Cap BGR0040 Straight Rod 5.5 x 40mm BGR0050 Straight Rod 5.5 x 50mm BGR0060 Straight Rod 5.5 x 60mm BGR0070 Straight Rod 5.5 x 70mm BGR0080 Straight Rod 5.5 x 80mm BGR0090 Straight Rod 5.5 x 90mm BGR0100 Straight Rod 5.5 x 100mm BGR0110 Straight Rod 5.5 x 110mm BGR0120 Straight Rod 5.5 x 120mm BGR0130 Straight Rod 5.5 x 130mm BGR0140 Straight Rod 5.5 x 140mm BGR0150 Straight Rod 5.5 x 150mm BGR0200 Straight Rod 5.5 x 200mm BGR0300 Straight Rod 5.5 x 300mm BGR0400 Straight Rod 5.5 x 400mm BGR0500 Straight Rod 5.5 x 500mm BGS5535 5.5 x 35mm Polyaxial Screw BGS5540 5.5 x 40mm Polyaxial Screw BGS5545 5.5 x 45mm Polyaxial Screw BGS6035 6.0 x 35mm Polyaxial Screw BGS6040 6.0 x 40mm Polyaxial Screw BGS6045 6.0 x 45mm Polyaxial Screw BGS6050 6.0 x 50mm Polyaxial Screw BGS6535 6.5 x 35mm Polyaxial Screw BGS6540 6.5 x 40mm Polyaxial Screw BGS6545 6.5 x 45mm Polyaxial Screw BGS6550 6.5 x 50mm Polyaxial Screw BGS7035 7.0 x 35mm Polyaxial Screw BGS7040 7.0 x 40mm Polyaxial Screw BGS7045 7.0 x 45mm Polyaxial Screw BGS7050 7.0 x 50mm Polyaxial Screw BGS7055 7.0 x 55mm Polyaxial Screw BGS7535 7.5 x 35mm Polyaxial Screw BGS7540 7.5 x 40mm Polyaxial Screw BGS7545 7.5 x 45mm Polyaxial Screw A pedicle screw system intended to provide immobilization and stabilization of spinal segments in skeletally mature patients. Class II

Reason: Orthopedic Alliance is recalling Blue & Gold Implants and Instruments because the Blue & Gold system is not supported by adequate testing and documentation to demonstrate that it meets performance or safety standards. These inadequacies might result in product peformance failures that could cause patient harm due to implant breakage, loosening, or inadequate sterilization.

Orthopedic Alliance LLC 20130307 Terminated
Distributed: Distributed in Texas.
RoboCouch Patient Positioning System Product Usage: is intended for use in the support and positioning of a patient during radiosurgery and radiotherapy procedures and other medical procedures when precise positioning is required. Class II

Reason: A5 gearboxes responsible for pitch (head up and head down) and roll (patient left and right) motions of the RoboCouch manipulator may be defective.

Accuray Incorporated 20130327 Terminated
Distributed: Worldwide Distribution - USA Nationwide including the states of: OH, MO. OK, CA and VA and the countries of: Finland, France, Luxembourg, Poland, Russia, Turkey and Japan.
Nova StatStrip Xpress Glu-Test Strips (Glucose test strips) Catalog # 51493 (UK Only) Class II

Reason: Glucose test strips reports no result message when tested

Nova Biomedical Corporation 20130225 Terminated
Distributed: Worldwide distribution: USA (nationwide) including states of: CA, IL, KY,NY, WV and countries of: United Kingdom and Italy.
Nova StatStrip Xpress Glu-Test Strips (Glucose test strips) Catalog # 42214 Class II

Reason: Glucose test strips reports no result message when tested

Nova Biomedical Corporation 20130225 Terminated
Distributed: Worldwide distribution: USA (nationwide) including states of: CA, IL, KY,NY, WV and countries of: United Kingdom and Italy.
VITROS Chemistry Products PHBR (Phenobarbital) Slides, REF 822 1384. For the quantitative measure of phenobarbital (PHBR) concentration in serum and plasma. Class II

Reason: Ortho Clinical Diagnostics (OCD) received customer complaints of intermittent imprecision, outliers, or shifts in accuracy and control fluid values outside of expected intervals when using VITROS Chemistry Products PHBR Slides. OCD's investigation confirmed that positively biased values for patient samples could be obtained that may not be detected by quality control fluids.

Ortho-Clinical Diagnostics 20130314 Terminated
Distributed: Worldwide Distribution-USA (nationwide) and Puerto Rico and the countries of Australia, Bermuda, Brazil, Canada, Chile, China, Colombia, England, France, Germany, India, Italy, Japan, Mexico, Republic of Panama, Singapore, Spain, and Venezuela.
Harmonie software versions 5.1 uo to 6.2e : Harmonie-E Long Term Monitoring system (Models HSYS-REC-WD, HSYS-REC-WT; HSYS-REC0LT2; HSYS-REC-EAMP) Product in use with Harmonie-E Long Term Monitoring System : ICTA-S & ICTA-D Seizure Detector, Slow wave Seizure Detector Harmonie-S Sleep System: Rapid Eye Movement Detector, Spindle Detector, Heart Rate Processor, Periodic Leg Movement Detector and Oximetry/Pulse rate Threshold Notification VITA/ICU Neurological Monitoring System: Burst Suppression Detector, Cerebral Function Monitor Trend, Visual and Auditory Notifications on all trends graphs in PolyTrends Harmonie {-s}{-E} Readers , Class II

Reason: The firm is initiating this recall because the implied or stated indications in their labeling associated with Harmonie software have not yet been cleared by the FDA for sale in the USA.

Stellate Systems 20090406 Terminated
Distributed: Worldwide distribution: USA (nationwide) including states of: AL, AR, AZ, CA, CT, DE, FL, HI, IA, IN, IL, KS, KY, LA, MA, MD, ME, MI, MN, MO, NC, ND, NJ, NY, OH, OR, PA, RI, SC, TX, VA, VT, WI, WV, and WY; and the countries of: Argentina, Australia, Bahrain, Brasil, Canada, China, Denmark, Finland, France, Germany, Greece, Haiti, Hong Kong, Hungary, Indonesia, Iran, Israel, Italy, Japan, Korea, KSA, Mexico, Netherlands, Norway, Pakistan, Portugal, Rep. DE Panama, Saudi Arabia, Spain, Switzerland, Taiwan ROC, Thailand, Turkey, and United Kingdom.
Siemens Dimension Vista(R) System Vista Drug 4 Calibrator. The device is used as an in vitro diagnostic product. Class II

Reason: Siemens has confirmed complaints of low QC and patient recovery with the Dimension Vista(R) LOCI(R) Digoxin(DIGXN) Flex(R) reagent cartridge when calibrating DIGXN with either Drug 4 Cal (KC460) lot 2KD052 or Drug 4 Cal (KC460A) lot 2KD053. Internal testing has shown an average 15% low bias in the therapeutic range of 0.9 to 2.0 ng/mL.

Siemens Healthcare Diagnostics, Inc. 20130221 Terminated
Distributed: Distributed within the US to AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PR, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, and WY.
EVOTECH Endoscope Cleaner and Reprocessor System, P/N 50004. Designed to automatically clean and high-level disinfect flexible, submersible video or fiber-optic endoscopes. Class II

Reason: Advanced Sterilization Products (ASP) is recalling the EVOTECH Endoscope Cleaner and Reprocessor (ECR) because they have received reports of smoke emanating from the system due to a circuit board malfunction that may occur in the Channel Control Manifold sub-assembly.

Advanced Sterilization Products 20130328 Terminated
Distributed: Nationwide Distribution
Page 3,845 of 4,030 | Showing 10 recalls