MAQUET CARDIOHELP Support System: CARDIOHELP-I Software (up to version 3.3.0.0) Product Usage: The Cardiohelp System is a blood oxygenation and carbon dioxide removal system used to pump blood through the extracorporeal bypass circuit for circulatory and/or pulmonary support during procedures requiring cardiopulmonary bypass (for periods up to six hours). It is also intended to provide circulatory and/or pulmonary support during procedures not requiring cardiopulmonary bypass (for periods up to six hours). Class II

Reason: It has come to the attention of MAQUET that there have been episodes of brief unexpected shutdown, followed by an automatic device restart of the human machine interface (touch screen) of the CARDIOHELP device. The human machine interface (HMI) is the central display of the CARDIOHELP device, where measured values, settings and alarms are displayed and adjusted/confirmed, with the exception of bl

Maquet Cardiovascular, LLC 20130307 Terminated
Distributed: Worldwide Distribution - USA Nationwide and the following countries: Australia, Austria, Belarus, Belgium, Canada, Columbia, Croatia, Czech Republic, Denmark, Egypt, Finland, France, Germany, Greece, Hong Kong, Iceland, Iran, Ireland, Italy, Jordan, Kazakhstan, Korea, Kuwait, Luxembourg, The Netherlands, Norway, Oman, Poland, Portugal, Russia, Saudi Arabia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Thailand, Turkey, United Arab Emirates and the United Kingdom.
IMO-402 Mononucleosis Rapid Test Device (Whole Blood/Serum/Plasma). Catalog #IMO-402/94201 The MONO test (Whole Blood) is a rapid chromatographic immunoassay for the qualitative detection of Infectious Mononucleosis heterophile antibodies in whole blood to aid in the diagnosis of infectious Mononucleosis. Class III

Reason: Alere San Diego is recalling the Clearview Mononucleosis Cassette because the positive control included in the rapid test may produce false negative results.

Alere San Diego, Inc. 20130226 Terminated
Distributed: Nationwide Distribution including TN, WA, UT, KS, KY, MN, GA, MA, PA, NY, VT, and the Dist. of Columbia
HCS3011 AV Loop Kit For use with Transonic CO status System Only, -HCS3021 AV Loop Kit For use with Transonic CO status System Only. Product Usage: is indicated for use in patients greater than or equal to 2 years old with arterial and venous lines in the diagnostic assessment of cardiovascular status including cardiac output and associated hemodynamic parameters. Class II

Reason: Transonic Systems Inc. have received complaints of blood leakage involving the AV Loop Kit.

Transonic Systems Inc 20130328 Terminated
Distributed: Worldwide Distribution - USA Nationwide including the states of AZ, CA, FL, GA, NM, NY, TX, VA and WA and countries of Netherlands, Canada and Russia.
Pressure Monitoring sets with VAMP Plus closed blood sampling system model # T001691A, PXVP0566, PXVP2260, PXVP0826, PXVP0827, PXVP2284AT3, PXVP23X3AT3, T398807D, T100605B, T001762A, T391T00A, VP2. Blood sampling system used on patients requiring periodic withdrawal of blood samples. Class II

Reason: The device is being recalled because the firm has received multiple complaints related to incorrect assembly that may result in dilution of blood samples.

Edwards Lifesciences, LLC 20130130 Terminated
Distributed: Worldwide Distribution-USA (nationwide) including the states of AR, CA, FL, HI, ID, IN, KS, KY, LA, MA, MI, MN, MO, NC, ND, NM, NV, NY, OH, OK, PA, SD, TN, TX, VA, WA, and WI) and Europe.
ACIST Medical Systems Inc. The ACIST Angiographic Injection System is intended to be used for the controlled infusion of radiopaque contrast media for angiographic procedures. Class II

Reason: ACIST Medical Systems is issuing a voluntary recall notice for three specific lots of BT2000 Automated Manifold Kits. For these specific lots, ACIST has received field reports related to ingress of air into the manifold body. This has been detected by the Air Column Detect Sensor halting the injection, and normally replacement of the BT2000 Automated Manifold Kit is necessary in order to proceed w

Acist Medical Systems 20130315 Terminated
Distributed: Nationwide Distribution including the states of AL, AK, AZ, AR, CA, CO, CT, DC, GA, HI, ID, IL, IN, IA, KS, KY, MD,MI, MN, MS, MO, NV, NH, NJ, NM, NY,NC, OR, OK, OH, PA, TN, TX, VA and WA.
Vitrea CT Multi-Chamber Cardiac Functional Analysis application on Vitrea version 5.2 (Vitrea Enterprise Suite 1.3 and Vitrea fX 3.1) and subsequent versions of VitreaWorkstation, Vitrea fX, VitreaAdvanced and Vitrea Enterprise Suite, through version 6.4. Vitrea¿ CT Multi-Chamber Cardiac Functional Analysis provides a variety of tools for working with clinical CT images of the coronary arteries, heart, and surrounding tissue. Class II

Reason: Vital Images has made a decision to perform a field correction to the Vitrea CT Multi-Chamber Cardiac Functional Analysis application. Vital Images has identified a defect in which numerical values may be incorrect in the Results Table under the Cardiac Analysis portion of the user interface.

Vital Images, Inc. 20130315 Terminated
Distributed: Worldwide Distribution - USA including AL, AR, CA, CO, IA, MI,MN, MO, NY, NC, OH, OR, PA, TX, and UT. Internationally to AUSTRIA, BELGIUM, CANADA, CHINA, COLOMBIA, GEORGIA, JAPAN, KOREA, MACEDONIA, NETHERLANDS, POLAND, PORTUGAL, RUSSIAN FEDERATION, SINGAPORE, SWEDEN, SWITZERLAND, TURKEY, UKRAINE, UNITED KINGDOM.
Elana Catheter Sizer 2.0*** The Elana Catheter Sizer 2.0 is designed to aid the surgeon to assess the bypass graft of the ELANA technique. Class II

Reason: Distribution of an unapproved device into interstate commerce.

Elana, Inc. 20130214 Terminated
Distributed: Distributed in the states of Pennsylvania, Texas, and New York.
GE Healthcare, Innova 2100IQ, Innova 3100IQ, Innova 4100IQ, Cardiovascular and Interventional Imaging System. Class II

Reason: As part of GE Innova IQ table introduction, a label was designed to be affixed on the front Cover of the Gantry Pivot section. This safety label is missing on systems manufactured between March 2010 and July 2011. In the situation a staff member (doctor, nurse, operator, anesthetist, etc&) stands in the region between the C-arm gantry pivot and the head section of the table, there is a risk of co

GE Healthcare, LLC 20111021 Terminated
Distributed: Worldwide Distribution - Nationwide Distribution including the states of FL, IL, IN, LA, MI, MN, NJ, NM, NY, OH, PR, SC, TX and VA and the countries of Vietnam, Latvia, Italy, Israel, Indonesia, Hong Kong, United Kingdom, Finland, Spain, Czech Republic, Canada, Bahrain and Australia.
HORIBA ABX PENTRA 400 I.S.E. Module, Model Numbers: P400ISE110EN02 and P400ISE110US02, All Serial Numbers, Software Versions 5.0.7 and Below. Used to measure absorbance and ion electrodes. Class II

Reason: HORIBA Medical is informing all ABX PENTRA 400 Analyzer with I.S.E. Module customers that there are no Linearity Limit Flags for the I.S.E. assays currently programmed into the ABX PENTRA 400 Analyzer software. All other assays run on this analyzer do have Linearity Limit Flags to alert the user to values that fall below or above the validated linearity range for each assay.

Horiba Instruments, Inc dba Horiba Medical 20130320 Terminated
Distributed: Nationwide Distribution.
Duane Reade TRUEtrack monitor kit, UPC# 6 39194 01679 7 Part # A4006-81 The TRUEtrack Blood Glucose System is intended for the quantitative determination of glucose in human whole blood taken from finger or forearm. The System is intended to be used to assist the patient and healthcare professional in the management of diabetes. Class II

Reason: Meter kit outer boxes containing 10 ct. test strips state incorrect storage temperature range.

Nipro Diagnostics, Inc. 20130225 Terminated
Distributed: Nationwide Distribution only in NY.