LifeScan brand OneTouch¿ Ultra Control Solution, Part Numbers/Model Numbers: 01045808, 02141603. Testing of Blood Glucose Levels. Class II

Reason: When OneTouch¿ Ultra and OneTouch¿ Select test strips are tested with Ultra and Select control solution the meter may display a result outside the control solution test range printed on the strip vial label even though the system is working properly.

Lifescan Inc 20130225 Terminated
Distributed: Worldwide Distribution-USA (nationwide) including DC and Puerto Rico and the states of AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, 10, IL, IN, KS, KY, LA, MA, MD, ME, Ml, MN, MO, MS, MT, NC, NO, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SO, TN, TX, UT, VA, VI, VT, WA, WI,WV, and WY, and the countries of Russia and Ukraine.
SR Instruments FG3105 Patient Scales Used on various Joerns Healthcare Patient Lifters. Model# FG3105 Patient Scales. Product Usage: The FG3105 Scale is designed for Joerns Healthcare for use in portable patient weighing systems. The scale is designed to be connected between the patient hoist mechanism and patient sling apparatus. The scale is not intended to be used for patient transport. Class II

Reason: SR Instruments Inc. is implementing a field correction for the Joerns Healthcare Patient Lifts after they received two incident reports regarding two field failures of a tensile type load cell involving the FG3105 Patient Scale. The failures resulted in patient falls.

Sr Instruments Inc 20121213 Terminated
Distributed: USA Nationwide Distribution in the state of Texas.
The Oxford Partial Knee; Oxford Knee Resection Procedure 3 Pack CEMENTED For use with Stryker System 5, System 4, System 2000 and EDH. Product Usage: For use with Stryker System 5, System 4, System 2000 and EDH. A set of three surgical saw blades is used to resect damaged cartilage and/or bone in an orthopedic reconstructive procedure such as a partial knee replacement. The set contains both oscillating blades for tissue resection along a small arc in a left/right or up/down direction, and reciprocating blades for resection back and forth in a single plane. Class II

Reason: Biomet part # 506076, lot 928182 was received from Synvasive Technology containing the incorrect blade.

Synvasive Technology Inc 20130314 Terminated
Distributed: USA Nationwide Distributor including the state of IN
Accuray CyberKnife System Medical Physicist, Robotic Radiosurgery System, Medical charged-particle radiation therapy system, Model: Iris Variable Aperture Collimator P/Ns 028986 and 031777. The CyberKnife System is indicated for radiation treatment. Class II

Reason: A defect on the master-side connector block of the pneumatic tool changer could introduce a tilt in the collimator housings, resulting in a potential shift to the central axis of the radiation beam. The issue could pose significant clinical impact to plans created using the smallest collimators.

Accuray Incorporated 20130319 Terminated
Distributed: Worldwide distribution, including Nationwide in the US and the countries of Korea, Myanmar, Canada, Belgium, England, Czech Republic, Finland, France, Germany, Greece, India, Italy, Poland, Russia, Spain, Switzerland, Turkey, Columbia, Mexico, Japan and Saudi Arabia.
Instrumentation Industries BE183-SU and BE 183 SU-12 Manual Jet Ventilators; BE 183-SUR and BE 183 SUR-12 Manual Jet Ventilators with Regulator and Gauge; BE 183-2 4 ft. Small Bore Tubing Assembly w/Inline Filter; and BE 183-R Regulator The BE 183-SUR and BE 183-SU Manual Jet Ventilators are designed to provide transtracheal ventilation in specific emergency situations of upper airway obstructions. Class II

Reason: Firm initiated an update to Directions for Use necessitating replacement of previously distributed Directions for Use.

Instrumentation Industries Inc 20120907 Terminated
Distributed: Worldwide Distribution - USA including AK, AL, AZ, CA, CO, CT, DC, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, ND, NH, NJ, NV, NY, OH, OK, OR, PA, PR, SC, SD, TN, TX, UT, VA, VT, WA, WI and WV. Shipped internationally to Canada, Ireland, Hong Kong, the Netherlands, New Zealand, and Thailand.
REF 32-422760 Oxford Uni Knee System,Toffee Hammer, non-sterile, found in REF 32-422763 Oxford Mlcroplasty Knee System Tibial instrument Tray small surgical hammer supplied with the Oxford Partial Knee Microplasty Instrument set, used for impacting implants. Class II

Reason: Investigation of complaints found excessive use of the instrument could result in cracking on the weld between the head and the handle, which could lead to part of the head becoming detached from the handle.

Biomet U.K., Ltd. 20130320 Terminated
Distributed: Worldwide Distribution - USA (nationwide), Europe, Japan, Australia, Canada, Singapore, South Africa, and Costa Rica.
EVOTECH Endoscope Cleaner & Reprocessor System, P/N 50004 Designed to automatically clean and high-level disinfect flexible, submersible video or fiber-optic endoscopes. Class II

Reason: Advanced Sterilization Products (ASP) has determined that a limited number of EVOTECH ECRs manufactured between 2011 and 2012 did not receive sufficient High Potential (HIPOT) DC voltage testing.

Advanced Sterilization Products 20130322 Terminated
Distributed: Nationwide Distribution-USA including DC and the states of AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, ME, MD, MI, MN, MO, MS, NC, ND, NE, NH, NJ, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, and WI.
ADVIA Centaur Vitamin D Calibrator 6-pack; Reference Number:10630911. For use in the quantitative determination of Vit D. Class II

Reason: Negative bias in patient samples for Vitamin D

Siemens Healthcare Diagnostics, Inc 20130313 Terminated
Distributed: Nationwide Distribution.
ADVIA Centaur Vitamin D Calibrator 2-pack; Reference Number:10493589. For use in the quantitative determination of Vit D. Class II

Reason: Negative bias in patient samples for Vitamin D

Siemens Healthcare Diagnostics, Inc 20130313 Terminated
Distributed: Nationwide Distribution.
ADVIA Centaur Vitamin D Assay (500 tests); Reference Number:10631021. For use in the quantitative determination of Vit D. Class II

Reason: Negative bias in patient samples for Vitamin D

Siemens Healthcare Diagnostics, Inc 20130313 Terminated
Distributed: Nationwide Distribution.