ADVIA Centaur Vitamin D Assay (100 tests); Reference Number: 10491994. For use in the quantitative determination of Vit D. Class II

Reason: Negative bias in patient samples for Vitamin D

Siemens Healthcare Diagnostics, Inc 20130313 Terminated
Distributed: Nationwide Distribution.
Clinical Chemistry Phenobarbital, list number 1E08 Class II

Reason: Abbott Diagnostics is recalling Clinical Chemistry Phenobarbital reagent lots 62299UN12 and 85773UN12 due to imprecision issues which increase with the age of the reagent lot. The imprecision of the Phenobarbital assay is d 7% Total CV%; however, internal testing confirmed CVs up to approximately 12% for in-date reagent lots over five months old. Although lot 52803UN12 is now expired, it exhibit

Abbott Laboratories, Inc 20121116 Terminated
Distributed: Worldwide Distribution - USA (nationwide) and Internationally to Internationally distributed to Argentina, Australia, Austria, Bahamas, Brunei, Canada, Chile, Curacao, Denmark, France, Germany, Greece, Ireland, Italy, Lebanon, Mexico, Netherlands, Norway, Oman, Paraguay, Portugal, Puerto Rico, Russia, Saudi Arabia, Serbia, South Korea, Spain, Switzerland, Turkey, and United Kingdom.
The GemStar Pump is a small and lightweight, single channel infusion device designed for use in the home, in the hospital, or anywhere electronic infusion is required. The GemStar Pump can be powered by AC mains adaptor, rechargeable battery pack, docking station, or two disposable AA alkaline batteries. When powered by batteries, The GemStar Pump is ideal for ambulatory patients. Class II

Reason: During infusions of 2.0 mL/hr the motor assembly may rotate backwards capturing additional medication resulting in over-infusion.

Hospira Inc. 20130321 Terminated
Distributed: Worldwide Distribution - USA (nationwide) and Internationally to Australia, New Zealand, China, Hong Kong, Taiwan, Japan, Korea, Philippines, Thailand, Malaysia, Canada, Austria, Bahrain, Belgium, Croatia, Denmark, Egypt, Finland, France, Germany, Gibraltar, Greece, Hungary, Iceland, Ireland, Israel, Italy, Jordan, Kuwait, Lebanon, Libya, Luxembourg, Malta, Netherlands, Norway, Oman, Portugal, Qatar, Romania, Saudi Arabia, Slovenia, South Africa, Spain, Sweden, Switzerland, Turkey, UK, United Arab Emirates, Brazil, Chile and Colombia.
The GemStar Pump is a small and lightweight, single channel infusion device designed for use in the home, in the hospital, or anywhere electronic infusion is required. The GemStar Pump can be powered by AC mains adaptor, rechargeable battery pack, docking station, or two disposable AA alkaline batteries. When powered by batteries, The GemStar Pump is ideal for ambulatory patients. Class II

Reason: During infusions of 2.0 mL/hr the motor assembly may rotate backwards capturing additional medication resulting in over-infusion.

Hospira Inc. 20130321 Terminated
Distributed: Worldwide Distribution - USA (nationwide) and Internationally to Australia, New Zealand, China, Hong Kong, Taiwan, Japan, Korea, Philippines, Thailand, Malaysia, Canada, Austria, Bahrain, Belgium, Croatia, Denmark, Egypt, Finland, France, Germany, Gibraltar, Greece, Hungary, Iceland, Ireland, Israel, Italy, Jordan, Kuwait, Lebanon, Libya, Luxembourg, Malta, Netherlands, Norway, Oman, Portugal, Qatar, Romania, Saudi Arabia, Slovenia, South Africa, Spain, Sweden, Switzerland, Turkey, UK, United Arab Emirates, Brazil, Chile and Colombia.
Product is distributed by Zimmer (Warsaw, Indiana)/Synvasive Technology, Inc. (El Dorado Hills, CA) & manufactured under Synvasive Technology, Inc. El Dorado Hills, CA Reciprocating saw blade: A sterile surgical saw blade used to resect damaged cartilage and/or bond in an orthopaedic reconstructive procedure such as a total knee replacement. Class II

Reason: Various reciprocating and oscillating saw blades were incorrectly labeled. The labeling errors include: (1) The pouch says oscillating saw blade when it actually contains a reciprocating saw blade; (2) The box bar code date does not match the human readable expiration date; and (3) The label incorrectly states use with Zimmer/Synthes handset and not the Zimmer Universal Power System.

Synvasive Technology Inc 20130402 Terminated
Distributed: Worldwide Distribution - USA (nationwide) and countries of: Austria, Australia, Belgium, Czech Republic, Denmark, Finland, France, Germany, Italy, Japan, Netherlands, Norway, Panama, Spain, Sweden, Switzerland, Tunisia, United Kingdom and Yemen.
CoaguChek XS Pro meters (catalog number 05530199160)bar code scanner. Used with the CoagCHEK XS Pro Meters to quantitatively determines prothrombin time ("PT"). Class II

Reason: Roche investigation into complaints found Accu-Chek Inform II and CoaguChek XS Pro System barcode reading meters had the potential of intermittent erroneous decoding of patient identification if barcodes being read fail to meet size and/or quality standards. Errors have been confirmed with the EAN 13 barcodes which should not be used for patient or operator identification.

Roche Diagnostics Operations, Inc. 20130306 Terminated
Distributed: Nationwide Distribution.
ACCU-CHEK Inform II Blood Glucose Monitoring System bar code scanner. Used with the ACCU-CHEK Inform II System to quantitatively measure glucose (sugar). Class II

Reason: Roche investigation into complaints found Accu-Chek Inform II and CoaguChek XS Pro System barcode reading meters had the potential of intermittent erroneous decoding of patient identification if barcodes being read fail to meet size and/or quality standards.

Roche Diagnostics Operations, Inc. 20130306 Terminated
Distributed: Nationwide Distribution.
Alaris PC unit model 8015 with software version 9.12 Product Usage: The Alaris PC unit is the central programming, monitoring and power supply component for the Alaris System Class II

Reason: The recall was initiated because Carefusion has identified potential risk associated with bolus programming with the Alaris PC unit (model 8015) software version 9.12. Automating infusion pump parameter input from either the electronic medical record (EMR) or the Alaris Auto-ID module for a continuous infusion with a bolus dose option may result in a loss of information from the Guardrails Data s

Carefusion Corporation 20130306 Terminated
Distributed: Worldwide Distribution USA Nationwide and the countries of Australia and Canada.
Edwards Lifesciences Duraflo coated Femoral Cannula. Product Usage: The cannulae are intended for use in situations which require rapid femoral venous and arterial access for short-term (< 6 hours) cardiopulmonary bypass. Vessel access (venous or arterial) is left to the discretion of the physician. Extracorporeal circuit components with a Duraflo coating are intended for use in cardiopulmonary surgery when a heparin-coated blood path is desired. The femoral arterial cannulae are wire-reinforced, thin-wall cannulae. Class II

Reason: Two unsealed pouches were found at an Edwards' distribution location in Japan. Incorrectly sealed pouches may result in a sterility barrier breach.

Edwards Lifesciences, LLC 20130314 Terminated
Distributed: USA Nationwide Distribution including the states of: FL, MA, MI, NY, PA, TX.
IS3000 USER MANUAL for Intuitive Surgical da Vinci Si Surgical System and da Vinci Si-e Surgical System, Model SI3000. Arm drapes protect equipment from contamination. Class II

Reason: Revised instructions for users of the da Vinci S, Si and Si-e systems with additional information regarding sterile adapter engagement and instrument installation.

Intuitive Surgical, Inc. 20130315 Terminated
Distributed: Worldwide Distribution.-USA (nationwide) and countries of: Austria, Belgium, Denmark, Finland, France, Germany, Ireland, Netherlands, Norway, South Korea, Sweden, Switzerland, and United Kingdom.