Roche Molecular Systems cobas x 480 System 50 mL reagent reservoir, Catalog number 05232732001. Used with the cobas¿ CT/NG 4800 System, for in vitro diagnostics. Class II

Reason: Certain lots of Reagent reservoirs, may not have a separation in the lower part of the reservoir. As a consequence, pipetting errors can occur due to uneven distribution of reagent. These pipetting errors may lead to delayed results and wasted reagent.

Roche Diagnostics Operations, Inc. 20130301 Terminated
Distributed: Nationwide Distribution including Puerto Rico and the states of MO, NE, CO, MT, NJ, FL, WA, TX, GA, NY, KS, NH, VA, HI, PA, MI, IA, CA, NC, AZ, UT, LA, IL, TN, ME, MN, KY, CT, OR, NV, AL and NM.
STERRAD 200 Sterilization System, P/N 10201 Product Usage: The STERRAD 200 Sterilization System is a general purpose, low temperature sterilizer which uses the STERRAD 200 process to inactivate microorganisms on a board range of medical devices and surgical instruments. Class II

Reason: The recall was initiated because Advanced Sterilization Products (ASP) has determined that the STERRAD 100NX and STERRAD 200 Sterilization Systems may emit an odor or smell into the surrounding environment that is not typical to normal operating conditions.

Advanced Sterilization Products 20130322 Terminated
Distributed: US Nationwide Distribution
STERRAD 100NX Sterilization System, P/N 10104 Product Usage: The STERRAD 100NX Sterilization System is a general purpose, low temperature sterilizer which uses the STERRAD 100NX process to inactivate microorganisms on a board range of medical devices and surgical instruments. Class II

Reason: The recall was initiated because Advanced Sterilization Products (ASP) has determined that the STERRAD 100NX and STERRAD 200 Sterilization Systems may emit an odor or smell into the surrounding environment that is not typical to normal operating conditions.

Advanced Sterilization Products 20130322 Terminated
Distributed: US Nationwide Distribution
Dual Tip Marker, Sterile Product Usage: A writing instrument intended to mark paper and plastic items around the sterile field during surgery. It has two tips. One is black for miscellaneous use. The other is for surgical skin marking. Class III

Reason: Dual Tip Marker has inconsistent ink flow on the black permanent ink side.

Viscot Medical LLC 20120822 Terminated
Distributed: USA Nationwide Distribution
Z-800 Volumetric Infusion System. For intravenous infusion of parenteral fluids, blood and blood products to a human patient. Class II

Reason: Update Instruction For Use for Preventative Maintenance (PM) Schedule and addition of Odometer software to monitor accumulative fluid volume infused by the pump

Zyno Medical LLC 20110613 Terminated
Distributed: Nationwide Distribution.
Smiths Medical, BCI¿ Remote Alarm Cable (Normally Open & Normally Closed), Reorder Numbers: WW3398NO & WW3398NC. For use with respiration carbon dioxide measurement equipment. Product Usage: The BCI¿ Remote Alarm Cable is designed for use with the 9004 Capnocheck¿ Capnography (Plus or Sleep) system to transfer alarm signals from the Capnocheck¿ Capnography system to a remote alarm system. The BCI¿ Remote Alarm Cable and the 9004 Capnocheck¿ Capnography systems are intended for use in all critical environments including ventilatory applications, patient transport, and anesthesia. Class I

Reason: In rare instances, the BCI¿ Remote Alarm Cables (BCI¿ Cable) are not transferring alarms when used with some remote nurse alarm systems.

Smiths Medical ASD, Inc. 20130206 Terminated
Distributed: USA Nationwide Distribution - including the states of: CO, IL, IN, MA, SC and WI.
Animas(R) 2020 Insulin Infusion Pump Product Usage: This product is indicated for continuous subcutaneous infusion of insulin for the treatment of insulin-requiring diabetes. Class I

Reason: Animas has identified a component issue affecting a small supply of the Animas(R) 2020 insulin pumps. The component issue may trigger the pumps to sound a false alarm or warning related to one of the following: - Loss of prime; -Occlusion; or -No cartridge detected.

Animas Corporation 20130103 Terminated
Distributed: Worldwide Distribution - USA (nationwide) including the states of: AL, AR, AZ, CA, CO, CT, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV and Puerto Rico and the countries of: Australia, Austria, Bermuda, Canada, Czech Republic, Finland, France, Germany, Ireland, Israel, Italy, Mexico, New Zealand, Norway, Spain and the United Kingdom.
20-Pole Eco Cable, Catalog #EM-5050-060 Product Usage: The 20-Pole Eco Cable is intended to be used as an interface cable between the LASSO NAV Eco Catheter and CARTO 3 System. Class II

Reason: Biosense is recalling the 20-Pole Eco Cable because it was incorrectly calibrated. When a catheter is connected to an affected to an affected 20-Pole Eco Cable, the catheter will not be recognized by the impacted 54 cable serial numbers within the field.

Biosense Webster, Inc. 20130305 Terminated
Distributed: Worldwide Distribution - USA (Nationwide) including the states of: DE, NL, CH, FR, GB, DK.
Gen-Probe LIFECODES PF4 Enhanced assay, Catalog No. X-HAT45 GTI PF4 Enhanced is designed as a solid phase enzyme-linked immunosorbent assay (ELISA). Class II

Reason: Lifecodes PF4 Enhanced assay, lot number 3000389, is being recalled because of observed inconsistent replicate values when testing the kit Positive Control and patient samples. The variation in reactivity can lead to invalid assays. If an invalid assay is obtained, those results should not be reported and the test should be re-run. An invalid assay would not result in significant harm to a pati

Genetic Testing Institute,inc 20130204 Terminated
Distributed: Worldwide Distribution - USA (nationwide) and Internationally to Canada, China, Slovakia, and Spain.
Microbiologics KWIK-QC SLIDES ( KWIK-QC Acid Fast Stain Slides and KWIK-QC Mycobacterium Slides). Slide preparations serving as quality control challenges to demonstrate satisfactory performance of staining reagents. Class II

Reason: Microbiologics Inc is initiating a voluntary recall on products KWIK-QC Acid Fast Stain Slide, SL42-10 and KWIK-QC Mycobacterium, SL43-10. These products are being recalled because the control organisms are not properly fixated on the slide, which may cause the user to unintentionally contaminate the patient area of the slide by flooding the control well with stain. This contamination could pote

Microbiologics Inc 20130211 Terminated
Distributed: Worldwide Distribution-USA including the states of CA, IN, MS, NY, OH, PA, and SD, and the countries of AUSTRALIA, AUSTRIA, UNITED ARAB EMIRATES, MALAYSIA, TAIWAN, JAPAN, TURKEY, BELGIUM, COLOMBIA, CANADA, SOUTH AFRICA, and VENEZUELA.