Presource Packs containing Medtronic Microstream Components: Intubation Kit Catalog Numbers AN23366084 AN23366085 AN2336608H AN2336608J Intubating Bag Kit Catalog Number OMSVM4166A Intubation UH Kit Catalog Number AN2336608G Class II

Reason: Kits contain Microstream components. Medtronic issued an update to the Instructions for Use of these components due reports of difficulty in disconnecting the adapter from the patient's endotracheal tube to perform a procedure. The issue can result in a delay to treatment and unintended extubation. The potential exists for respiratory failure, hypoxia, aspiration, or low oxygen saturation.

Cardinal Health 200, LLC 20260819 Ongoing
Distributed: Pending
DxC 500i Clinical Analyzer Modules: 1. DxC 500 AU Module w/ISE, DxC 500i, Part No. C63522 2. DxC 500 AU Module, DxC 500i, Part No. C63521 3. Access 2 Module, DxC 500i, Part No. C13252 Class II

Reason: Beckman Coulter has identified a software issue affecting DxC 500i Clinical Analyzers when testing plasma samples. Following user configuration of a Clinical Chemistry test to use units different from the system default, a subsequent configuration change and activation of the draft configuration in the System Configuration page may cause the analyzer to revert the test units to the system default setting without notifying the user. The system does Pre-Review Reminders Include OII and CDRHRecallGroup@fda.hhs.gov in any discussions on change/elevation of recalls to Class I classification and situations involving additional public communications (FDA and/or firm issued). Post-Review Reminders Final Check: Ensure your documentation accurately reflects the final recommendation prior to signature. Complete CTS: Recall Product tab, Additional Recall Event Info, Regulation, and Act. Upload PDFs of Consult, Review Memo, and Classification Letter for Class I Recall in DocMan. [Click to enter extra reminder]To add reminders, type the reminder in the text field to the left then press the Tab key. Doc ID 03534.05.06 2 of 8 not provide a warning or notification to alert the user that the configured units have reverted to the default units. This issue can result in falsely elevated reported results.

Beckman Coulter Inc. 20260902 Ongoing
Distributed: Worldwide - US Nationwide distribution.
SEDIPLAST; Model/Catalog Numbers: S-100 and S-1000; Class II

Reason: During use of an affected tube, the operator may encounter increased resistance when inserting the pipette. The resulting pressure buildup may cause the tube cap to detach or be forcibly ejected, potentially resulting in the release of blood and exposure of the operator or surrounding work area to a biological hazard (blood).

POLYMEDCO, INC. / POLYMEDCO CANCER DIAGNOSTIC PRODUCTS, LLC 20260825 Ongoing
Distributed: Worldwide - US Nationwide distribution in the states and territory of AK, AR, AZ, CA, CO, FL, GA, HI, IA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, ND, NH, NJ, NM, NY, OH, OK, OR, PA, PR, SC, SD, TN, TX, UT, VA, WA, WI, WV and the country of United Kingdom.
DAVINCI LAP/GYN; Model/Item Number (Kit): WRNM037-16; Class II

Reason: Product may be potentially contaminated with a lubricant used during the manufacturing process.

AVID Medical, Inc. 20260810 Ongoing
Distributed: US Nationwide distribution in the states of AZ, CA, FL, MD, NE, TX.
BeneVision Series Telemetry Monitor. Operator's Manuals: TM70/TM80 Class III

Reason: Certain monitor operator's manuals contain descriptions that may be insufficient or may require clarification regarding selected physiological alarm functions

Mindray DS USA, Inc. 20260831 Ongoing
Distributed: Worldwide - Nationwide and the country of Canada.
ENDOVASCULAR PACK; Model/Item Number (Kit): VAPX014-04; Class II

Reason: Product may be potentially contaminated with a lubricant used during the manufacturing process.

AVID Medical, Inc. 20260810 Ongoing
Distributed: US Nationwide distribution in the states of AZ, CA, FL, MD, NE, TX.
HALYARD Laparoscopy Comprehensive Pack II REF: 49332 Class II

Reason: Laparoscopy packs labeled as "latex free" were found to contain a latex urethral catheter

GRI Medical & Electronic Technology Co., Ltd. 20260826 Ongoing
Distributed: US Nationwide distribution in the state of MS.
Alphenix INFX-8000C/B5 Class II

Reason: Loosening of the bolts securing the longitudinal movement mechanism may cause the tabletop to shift, resulting in [positional/imaging/patient] misalignment and a longitudinal motion error

Canon Medical System, USA, INC. 20260826 Ongoing
Distributed: US Nationwide distribution in the state of ID.
DxC 500 AU Clinical Chemistry Analyzers: 1. AU w/out ISE, Part No. C63519 2. AU w/ISE, Part No. C63520 Class II

Reason: Beckman Coulter has identified a software issue affecting DxC 500 AU Clinical Analyzers when testing plasma samples. Following user configuration of a Clinical Chemistry test to use units different from the system default, a subsequent configuration change and activation of the draft configuration in the System Configuration page may cause the analyzer to revert the test units to the system default setting without notifying the user. The system does Pre-Review Reminders Include OII and CDRHRecallGroup@fda.hhs.gov in any discussions on change/elevation of recalls to Class I classification and situations involving additional public communications (FDA and/or firm issued). Post-Review Reminders Final Check: Ensure your documentation accurately reflects the final recommendation prior to signature. Complete CTS: Recall Product tab, Additional Recall Event Info, Regulation, and Act. Upload PDFs of Consult, Review Memo, and Classification Letter for Class I Recall in DocMan. [Click to enter extra reminder]To add reminders, type the reminder in the text field to the left then press the Tab key. Doc ID 03534.05.06 2 of 8 not provide a warning or notification to alert the user that the configured units have reverted to the default units. This issue can result in falsely elevated reported results.

Beckman Coulter Inc. 20260902 Ongoing
Distributed: Worldwide - US Nationwide distribution.
cPM Series Patient Monitor. Operator's Manual: cPM 8/cPM 12 Class III

Reason: Certain monitor operator's manuals contain descriptions that may be insufficient or may require clarification regarding selected physiological alarm functions

Mindray DS USA, Inc. 20260831 Ongoing
Distributed: Worldwide - Nationwide and the country of Canada.
Page 4 of 4030 | Showing 10 recalls
Data provided by openFDA