Integra Mozaik Strip 10 cc Product Usage: The product is a bone graft substitute composed of collagen (Type 1) and ceramic (I3-Tricalcium phosphate) matrix in putty and strip forms. The product is intended for use as a bone void filler to fill voids or gaps of the skeletal system in the extremities, spine. and pelvis not intrinsic to the stability of the bony structure. Integra Mozaik is also indicated for use in the treatment of surgically treated osseous defects or osseous defects created from traumatic injury to the bone. Following placement in the bone void or a defect Integra Mozaik is reabsorbed and replaced with bone during the healing process. Class II

Reason: The Instructions for Use {IFU} {Putty Part Number 7700713001 and Strip Part Number 770151300l for CE marked Mozaik Strip and Putty products contains an incorrect translation of the term "bone marrow aspirate". The term is incorrectly translated in Spanish and French,

Integra LifeSciences Corp. 20111011 Terminated
Distributed: Worldwide Distribution
PlasmaJet Open Surgery Handpiece (Part of the PlasmaJet Neutral Plasma Surgery System) The PlasmaJet system is a neutral plasma surgery system that is designed for cutting and coagulation in open surgery and laparoscopic surgery. Class II

Reason: Potential failure in the glued seal between the stem and the tip of the handpiece leading to the possible leakage of a very small drop of distilled water coolant from the handpiece in use.

Plasma Surgical Inc. 20120427 Terminated
Distributed: Worldwide distribution: USA (nationwide) including states of: CA and NY; and country of: United Kingdom.
Columbus Tibia Plateau (component of the Columbus REVISION Knee System) Product Usage - The Columbus Revision Knee System is indicated for use in reconstruction of the diseased knee joint caused by osteoarthritis, rheumatoid arthritis, post-traumatic arthritis, the need to revise failed arthroplastics or osteotomies where pain, deformity or dysfunction persist, and for patients suffering from correctable valgus or varus deformity and moderate flexion contracture Class II

Reason: During an internal quality inspection, the tibia plateau was found to be out of specification. There is a potential for misalignment between the tibia plateau and attachment of the extension stem.

Aesculap Implant Systems 20120221 Terminated
Distributed: Nationwide Distribution including PA and TX.
Lipofilter 3000, model ASP-CAN-2, labeled in part ***MicroAire, 1641 Edlich Drive, Charlottesville, VA 22911, USA*** Product Usage: A fat collection canister intended for use during liposuction procedures. Class II

Reason: Medical device was not manufactured under Good Manufacturing Procedures and a user may be unable to sterilize the device according to procedures.

MicroAire Surgical Instruments, LLC 20120426 Terminated
Distributed: US Nationwide Distribution including the states of CA, OH and NY.
The Brilliance iCT and iCT SP are Computed Tomography X-Ray Systems intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data taken at different angles and planes. These devices may include signal analysis and display equipment, patient, and equipment supports, components and accessories. The Brilliance iCT and iCT SP are used to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data taken at different angles and planes. Class II

Reason: Philips is issuing a software update (version 3.2.4) to the Brilliance iCT and iCT SP systems currently running software version 3.2.0, 3.2.1 or 3.2.3, addressing a number of non-conformances. A review was conducted of all nonconformances that are fixed by the software upgrade. None of these nonconformances result in unacceptable hazards. The nonconformance documented below had the highest sc

Philips Medical Systems (Cleveland) Inc 20120507 Terminated
Distributed: Worldwide Distribution - USA including AL, CA, DC, DE, FL, GA, IA, IN, MA, MD, MN, MS, NC, NY, OH, OR, PA, TN, TX, VT, WA & WV and the countries of Argentina, Australia, Belgium, Brazil, Canada, China, Czech Republic, Denmark, EGYPT, France, Germany, Greece, Hungary, ICELAND, India, Indonesia, Israel, ISRAEL, Italy, Japan, Jordan, Lebanon, Netherlands, New Zealand, PHILIPPINES, POLAND, Portugal, Qatar, Russia, SINGAPORE, South Korea, SPAIN, SWEDEN, Switzerland, Taiwan, THAILAND, Turkey, UNITED KINGDOM & Uzbekistan.
GE Proteus XR/a Radiographic System. GE Medical Systems-Americas Milwaukee, WI. Model Number: 2259988. Product is indicated for use in generating radiographic images of human anatomy in all general-purpose diagnostic procedures. This device is not intended for mammographic applications. Class II

Reason: An incident was reported that an operator's finger was pinched between the fixed roller and the gliding tabletop of a Proteus XR/a Table while moving the tabletop with the thumb placed over its edge and fingers underneath

General Electric Med Systems LLC 20060531 Terminated
Distributed: Worldwide Distribution.
GE Healthcare Definium 8000 Digital Radiographic Systems : GE Healthcare, 3000 North Grandview , Waukesha, WI, 53188 Model Number: 5131070. The Revolution XR/d with Tomosynthesis is intended for use in generating Tomographic images of human anatomy. It is not intended for mammographic use. Class II

Reason: An incident was reported that an operator's finger was pinched between the fixed roller and the gliding tabletop of a Proteus XR/a Table while moving the tabletop with the thumb placed over its edge and fingers underneath

General Electric Med Systems LLC 20060531 Terminated
Distributed: Worldwide Distribution.
GE Healthcare Revolution XR/d; GE Healthcare, 3000 North Grandview , Waukesha, WI, 53188 Model numbers: 2259988-2 & 2351505. The Revolution XR/d Digital Radiographic Imaging System is indicated for use in generating radiographic images of human anatomy. It is intended to replace radiographic film/screen system in all general-purpose diagnostic procedures. This device is not intended for mammographic applications. Class II

Reason: An incident was reported that an operator's finger was pinched between the fixed roller and the gliding tabletop of a Proteus XR/a Table while moving the tabletop with the thumb placed over its edge and fingers underneath

General Electric Med Systems LLC 20060531 Terminated
Distributed: Worldwide Distribution.
Philips Healthcare Practix Convenio Mobile X-ray system Worldwide Distribution -- Nationwide Class II

Reason: When the system is switched off or when the tube arm is parked in the down position (X-ray generator switch off), a radiation pulse (50 kV, 2.5 pGy) is generated inadvertently. .When the system is switched off or when the tube arm is parked in the down position (X-ray generator switch off), a radiation pulse (50 kV, 2.5 pGy) is generated inadvertently. The failure happened when component supplie

Philips Healthcare Inc. 20111019 Terminated
Distributed: Nationwide Distribution
Sigma Spectrum VOLUMETRIC INFUSION PUMP, with or without Master Drug Library, Part Number 35700 -- Common/Classification Name: Infusion Pump -- Class 2 medical device -- SIGMA LLC, Medina, NY 14103 Product Usage: The Spectrum Infusion Pump is intended to be used for the controlled administration of intravenous fluids. These fluids may include blood, blood products or mixtures of pharmaceutical drugs for required patient therapy. The Spectrum is used in conjunction with legally marked intravenous administration sets and medications provided by the user. The Master Drug Library is a software package that will add additional features to the Spectrum Infusion Pump. The Master Drug Library will permit electronic communications with the Spectrum pump and other external peripheral devices. The intended use of the Spectrum pump includes common drug error prevention, through the stand alone settings features of the pump. This includes drug parameter limits and associated drug name identification. With the Master Drug Library, the intended use is to reduce user errors associated with drug selection, drug dose rates, drug dose concentrations, and patient identification associated with the prescribed drug. Class I

Reason: Recall expansion; the infusion pumps have the potential to fail causing inaccurate flow conditions during use. These conditions range from back flow to free flow, which could result in over-infusion. Sigma's evaluation of subsequent complaints involving failed bearings indicated that the loss or degradation of bearing lubricant to be a more significant contributor to bearing failure than initially believed. Based upon the results of the analysis of these additional failures, Sigma determined that any bearings prior to the latest, improved design may be at risk for failure. Sigma decided to expand the recall of the Spectrum Infusion Pumps to include units manufactured from 1/18/2005 through 11/01/2010.

Baxter Healthcare Corporation 20110706 Terminated
Distributed: Worldwide Distribution - USA Nationwide including Puerto Rico and the countries of Canada and Okinawa